Evaluation of Elevated Mean Pulmonary Arterial Pressure Based on Magnetic Resonance 4D Velocity Mapping: Comparison of Visualization Techniques

  • Ursula Reiter ,

    Contributed equally to this work with: Ursula Reiter, Gert Reiter

    ursula.reiter@klinikum-graz.at

    Affiliation Division of General Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

  • Gert Reiter ,

    Contributed equally to this work with: Ursula Reiter, Gert Reiter

    Affiliation Siemens AG, Healthcare Sector, Graz, Austria

  • Gabor Kovacs,

    Affiliation Division of Pulmology, Department of Internal Medicine, Medical University of Graz & LBI for Lung Vascular Research, Graz, Austria

  • Aurelien F. Stalder,

    Affiliation Siemens AG, Healthcare Sector, Erlangen, Germany

  • Mehmet A. Gulsun,

    Affiliation Siemens Corporate Research, Baltimore, United States

  • Andreas Greiser,

    Affiliation Siemens AG, Healthcare Sector, Erlangen, Germany

  • Horst Olschewski,

    Affiliation Division of Pulmology, Department of Internal Medicine, Medical University of Graz & LBI for Lung Vascular Research, Graz, Austria

  • Michael Fuchsjäger

    Affiliation Division of General Radiology, Department of Radiology, Medical University of Graz, Graz, Austria

Abstract

Purpose

Three-dimensional (3D) magnetic resonance phase contrast imaging (PC-MRI) allows non-invasive diagnosis of pulmonary hypertension (PH) and estimation of elevated mean pulmonary arterial pressure (mPAP) based on vortical motion of blood in the main pulmonary artery. The purpose of the present study was to compare the presence and duration of PH-associated vortices derived from different flow visualization techniques with special respect to their performance for non-invasive assessment of elevated mPAP and diagnosis of PH.

Methods

Fifty patients with suspected PH (23 patients with and 27 without PH) were investigated by right heart catheterization and time-resolved PC-MRI of the main pulmonary artery. PC-MRI data were visualized with dedicated prototype software, providing 3D vector, multi-planar reformatted (MPR) 2D vector, streamline, and particle trace representation of flow patterns. Persistence of PH-associated vortical blood flow (tvortex) was evaluated with all visualization techniques. Dependencies of tvortex on visualization techniques were analyzed by means of correlation and receiver operating characteristic (ROC) curve analysis.

Results

tvortex values from 3D vector visualization correlated strongly with those from other visualization techniques (r = 0.98, 0.98 and 0.97 for MPR, streamline and particle trace visualization, respectively). Areas under ROC curves for diagnosis of PH based on tvortex did not differ significantly and were 0.998 for 3D vector, MPR vector and particle trace visualization and 0.999 for streamline visualization. Correlations between elevated mPAP and tvortex in patients with PH were r = 0.96, 0.93, 0.95 and 0.92 for 3D vector, MPR vector, streamline and particle trace visualization, respectively. Corresponding standard deviations from the linear regression lines ranged between 3 and 4 mmHg.

Conclusion

3D vector, MPR vector, streamline as well as particle trace visualization of time-resolved 3D PC-MRI data of the main pulmonary artery can be employed for accurate vortex-based diagnosis of PH and estimation of elevated mPAP.

Introduction

Pulmonary hypertension (PH) is a life-threatening complex pathophysiological condition characterized by mean pulmonary arterial pressure (mPAP) equal to or exceeding 25 mmHg at rest [1], [2]. It is associated with vortical motion of blood in the main pulmonary artery [3], which can be evaluated from time-resolved 3-dimensional (3D) phase contrast magnetic resonance imaging (PC-MRI). The duration of PH-associated vortical blood motion correlates well with invasively obtained measurements of elevated mPAP. Thus, 3D PC-MRI of the main pulmonary artery is a reliable, non-invasive, non-ionizing screening and longitudinal tool for follow-up of patients with PH [3], [4].

Visualization of 3D velocity fields is crucial for detecting flow patterns in general and vortices in particular, but it remains challenging. Densely scattered in volume, 3D velocity vectors overlap and obscure each other, hampering analysis of regional flow patterns. To improve the visualization of 3D velocity fields, various approaches have been introduced that selectively reduce overall 3D information [5][7]. Post-processing techniques commonly applied to visualize 3D PC-MRI data represent velocities as 3D vectors or multi-planar reformatted (MPR) 2-dimensional (2D) vectors restricted to anatomic slices [8][11], or show calculated 3D time-varying integral curves such as streamlines or particle traces. Streamlines, defined as tangent curves to velocity vectors at a particular time point, describe instantaneous velocity directions in the 3D volume [8], [11], [12]; particle traces show the trajectories of particles moving in the 3D velocity field, providing a time-integrated picture of flow [11], [13].

The use of 3D vector representation of PC-MRI data to assess vortical blood motion in the main pulmonary artery and thereby diagnose PH and estimate elevated mPAP has been investigated previously [3]. Other flow visualization techniques, however, might be advantageous for vortex detection. It remains unknown whether PH-associated vortical blood motion can be evaluated from 2D MPR vector, streamline, or particle trace representation and how findings from 3D vector visualization might relate to findings from these commonly used PC-MRI visualization techniques.

The purpose of the present study was to compare the presence and duration of PH-associated vortical motion of blood in the main pulmonary artery on MPR vector, streamline, and particle trace visualization with results obtained from 3D vector visualization with special respect to their performance for non-invasive assessment of elevated mPAP and diagnosis of PH.

Materials and Methods

Study Population and Right Heart Catheterization

The prospective study was approved by the local ethical review board of the Medical University of Graz, Austria, and all subjects gave written informed consent. Subjects with known contraindications to MR were not enrolled. Fifty-three patients with known or suspected PH underwent PC-MRI of the main pulmonary artery after successful right heart catheterization (RHC). Three patients were excluded from evaluation due to inadequate PC-MRI data (one patient with severe arrhythmia, two patients with claustrophobia). Time delay between the two investigations of the remaining fifty patients was 10±14 days. No clinically relevant changes in drug treatment or disease state occurred between the two examinations.

RHC was performed with a 7F quadruple-lumen, balloon-tipped, flow directed Swan-Ganz catheter (Baxter Healthcare Corp, Irvine, CA, USA) using transjugular approach. Measurements were obtained in free breathing with the patient in the supine position. Invasively obtained mPAP was used as the reference standard for the diagnosis of PH [2]. The RHC-based classification and demographic characteristics of the study population are summarized in Table 1.

thumbnail
Table 1. Demographic characteristics of the study population.

https://doi.org/10.1371/journal.pone.0082212.t001

MR Imaging

MR imaging was performed at 1.5 T (MAGNETOM Sonata, Siemens, Erlangen, Germany) using a 6-channel cardiac array coil, with the patient in the supine position. PC-MRI data were acquired in free breathing in right ventricular outflow tract (RVOT) orientation; the main pulmonary artery was covered in 5–10 gapless slices of a retrospectively ECG-gated, segmented, 2D spoiled gradient-echo-based phase contrast sequence with three-directional velocity encoding by a simple four-point velocity encoding scheme [14]. Velocity encoding (VENC) was set to 90 cm/s in all directions and adapted if necessary to prevent aliasing in the main pulmonary artery. Further protocol parameters were as follows: Field of view, 234–276×340 mm2; matrix, 96–114×192; slice thickness, 6 mm; bandwidth, 451 Hz/pixel; GRAPPA (generalized auto-calibrating partially parallel acquisition) factor, 2; number of reference lines, 21–26; flip angle, 15°; echo time, 4.1 ms; repetition time, 7.5 ms. 3 acquired phase encoding steps per segment resulted in a temporal resolution of 89 ms, which was interpolated to 20 cardiac phases per cardiac cycle. Three-fold averaging to suppress breathing artifacts yielded a measurement time of 66–72 heart beats per slice.

Image Processing and Analysis

For calculation and visualization of 3D velocity fields, PC-MRI data were imported into dedicated prototype software (4D Flow, Siemens, Erlangen, Germany) [15], [16]. After background phase correction [13], [17], [18] and semi-automatic segmentation of the RVOT and main pulmonary artery, blood flow patterns were analyzed in 3D vector, MPR vector, streamline, and particle trace representations by two experienced readers who were blinded to mPAP measurements derived from RHC. First, the readers evaluated blood flow patterns independently so that interobserver variability could be assessed. Subsequently, they evaluated blood flow patterns in consensus.

With 3D vector visualization, the magnitudes (with color encoding) and directions of the measured velocities are projected onto opaque anatomical images (Figure 1A). Through-plane components of velocities can be assessed by adjusting the opacity of anatomical images as well as the spatial rotation of image planes joined with 3D vectors. MPR vector visualization enables reconstruction of arbitrary cross-sectional planes in volume, showing interpolated in-plane velocity vectors projected onto multi-planar reformatted, anatomical images (Figure 1B). In our study, the magnitude and direction of in-plane velocities were displayed as the magnitude (enhanced with color coding) and direction of in-plane velocity vectors. For visualization of streamlines (Figure 1C) and particle traces (Figure 1D), starting points (“seeding” points) were chosen that were uniformly distributed in the segmented volume in each cardiac phase. Particle trace length was adapted to the seeding period to obtain continuous particle traces of adequate length. Magnitudes of velocities along streamlines and particle traces were further enhanced by color encoding. Projection into opaque 3D phase contrast angiographic reconstructions of PC-MRI data provided anatomical and perspective context.

thumbnail
Figure 1. PH-associated vortical blood flow in the main pulmonary artery (MPA) in 3D vector (A), MPR vector (B), streamline (C) and particle trace (D) visualization together with schematic drawings of identification criteria of this flow pattern in 3D vector (E), MPR vector (F), streamline (G) and particle trace (H) visualization.

PV: pulmonary valve, RV: right ventricle.

https://doi.org/10.1371/journal.pone.0082212.g001

Blood flow patterns were analyzed visually with respect to the presence and duration of PH-associated vortical motion of blood in the main pulmonary artery. A PH-associated vortex was defined as non-valvular rotational blood flow in the RVOT orientation. In 3D vector and MPR vector visualization, rotational blood flow was identified from the existence of notional, smooth, closed concentric tangent curves in the velocity vector field (Figure 1E and 1F). With streamline visualization, rotational blood flow was detected when closed streamlines were observed (Figure 1G), whereas with particle trace visualization, rotational blood flow was identified as a closed ring of particle traces with particle velocities not vanishing along the entire paths (Figure 1H).

If a PH-associated vortex was identified in a patient, its duration (tvortex), its time of onset (tstart) and its time of termination (tend) were specified. tvortex was defined as the duration of PH-associated vortical blood flow (divided by the cardiac interval and given as a percentage); tstart was defined as the time interval from pulmonary valve opening (derived from 3D vector visualization) to the onset of PH-associated vortical blood flow (divided by the cardiac interval and given as a percentage); and tend was defined as the time interval from pulmonary valve opening to termination of PH-associated vortical blood flow (divided by the cardiac interval and given as a percentage).

Statistical Analysis

Mean values are given together with standard deviations. Statistical analysis was performed using NCSS (Hintze, J. (2008) NCSS, LLC. Kaysville, UT, USA). For statistical tests a significance level of 0.05 was employed.

Cohen’s Kappa coefficient (κ) was calculated to specify interobserver agreement with respect to the detection of PH-associated vortices. Interobserver variability levels in the determination of tvortex, tstart and tend were specified as within-subject standard deviations in variance component analysis together with the intraclass correlation coefficients rIC.

Further analyses were performed using tvortex, tstart and tend values derived by the two readers in consensus. Agreement of 3D vector visualization with MPR vector, streamline and particle trace visualization with respect to the detection of PH-associated vortices was calculated as Kendall’s τB with corrections for ties. Symmetry was analyzed by using McNemar’s test. Comparison of tvortex, tstart and tend measurements derived from 3D vector, MPR-vector, streamline, and particle trace visualization was performed by means of Pearson’s correlation coefficient r and Bland-Altman analysis.

The diagnostic performance of tvortex in predicting PH from the various visualization techniques was investigated by means of receiver operating characteristic (ROC) curve analysis. Empirical areas under ROC curves (AUCs) describing the diagnostic performance of tvortex as detected by MPR vector, streamline and particle trace visualization were compared to the empirical AUC describing the diagnostic performance of tvortex detected by 3D vector visualization by z test. The relationship of tvortex to mPAP was analyzed by means of correlation and linear regression analysis. Comparison of correlation coefficients between mPAP and tvortex determined from 3D vector, MPR vector, streamline and particle trace visualization was performed by Williams-Hotelling test.

Results

Interobserver Variability of Vortex Detection

Interobserver variability levels in the measurement of tvortex from 3D vector, MPR-vector, streamline and particle trace visualization were 2% (rIC = 0.99), 4% (rIC = 0.97), 5% (rIC = 0.96) and 4% (rIC = 0.97) of RR-interval, respectively. Interobserver agreement in the identification of patients with PH-associated vortices was high for 3D vector (κ = 0.88) and particle trace (κ = 0.87) visualization and medium for MPR vector (κ = 0.48) and streamline (κ = 0.59) visualization.

Both observers found PH-associated vortices in all patients with PH with all visualization techniques, so tstart and tend of PH-associated vortical blood flow could be compared. Interobserver variability levels in the determination of tvortex, tstart and tend in patients with PH are presented in Table 2.

thumbnail
Table 2. Interobserver variability in determination of tvortex, tstart and tend from different flow visualizations in patients with PH (n = 23).

https://doi.org/10.1371/journal.pone.0082212.t002

Comparison of Flow Visualization Techniques

tvortex determined by MPR vector, streamline and particle trace visualization correlated strongly with tvortex derived from 3D vector representation (r = 0.98, 0.98 and 0.97, respectively). tvortex determined from MPR vectors, streamlines and particle traces were, however, slightly larger than tvortex determined from 3D vectors, with a standard deviation of measurement differences of 5–6% of the RR-interval (Figure 2).

thumbnail
Figure 2. Bland-Altman plots of vortex duration (tvortex) determined from MPR vector (A), streamline (B) and particle trace (C) visualization compared to tvortex determined by 3D vector visualization for all patients (n = 50).

SD denotes standard deviation of measurement differences.

https://doi.org/10.1371/journal.pone.0082212.g002

Compared to the agreement on tvortex observed between 3D vector visualization and the various other visualization techniques, agreement on the identification of patients with PH-associated vortical blood flow between 3D vector visualization and MPR vectors, streamlines and particle traces was weaker (τB = 0.71, 0.62 and 0.77, respectively). PH-associated vortices were identified in more subjects with analysis of MPR vector, streamline and particle trace visualization than with 3D vector visualization (p = 0.005, 0.011 and 0.102, respectively).

Mean values of tvortex, tstart and tend in patients with PH are summarized in Table 3. As presented in Table 4, correlations between tvortex measurements determined by different visualization techniques were lower in patients with PH than in all patients. Moreover, tstart and tend of PH-associated vortical blood flow showed considerable dependence on visualization technique.

thumbnail
Table 3. Mean values and standard deviations of tvortex, tstart and tend derived from different flow visualizations in patients with PH (n = 23).

https://doi.org/10.1371/journal.pone.0082212.t003

thumbnail
Table 4. Comparison of tvortex, tstart and tend determined from MPR-vector, streamline and particle trace visualizations with the same parameters determined from 3D vector visualization in patients with PH (n = 23).

https://doi.org/10.1371/journal.pone.0082212.t004

Non-invasive Diagnosis of PH based on Vortex Duration

Areas under ROC curves for non-invasive diagnosis of PH from tvortex values from the various visualization techniques did not differ significantly. The AUCs were 0.998 (95% confidence interval [CI], 0.983–1.000) for 3D vector visualization, MPR vector visualization, and particle trace visualization and 0.999 (95% CI, 0.987–1.000) for streamline visualization (Figure 3). Optimal cut-off values of tvortex for maximizing the sum of sensitivity and specificity are given in Table 5.

thumbnail
Figure 3. ROC curves for the diagnosis of manifest PH employing vortex duration (tvortex) determined from 3D vector, MPR vector, streamline and particle trace visualization.

https://doi.org/10.1371/journal.pone.0082212.g003

thumbnail
Table 5. Cut-off values of tvortex determined from 3D vector, MPR-vector, streamline and particle trace visualizations for non-invasive diagnosis of PH.

https://doi.org/10.1371/journal.pone.0082212.t005

tvortex correlated strongly with mPAP for patients with PH (r = 0.96 in case of 3D vector, r = 0.93 in case of MPR vector, r = 0.95 in case of streamlines, and r = 0.92 in case of particle trace visualization) and the correlations did not differ significantly. Linear regression lines between mPAP and tvortex values derived from 3D vector, MPR vector streamline and particle trace visualization are shown in Figure 4.

thumbnail
Figure 4. Scatter plot and linear regression lines of mPAP and vortex duration (tvortex) determined from 3D vector (A), MPR vector (B), streamline (C) and particle trace (D) visualization for patients with PH (n = 23).

Regression equations are to be understood in mmHg. SD denotes standard deviation from regression line.

https://doi.org/10.1371/journal.pone.0082212.g004

Discussion

The present study showed that 3D vector, MPR vector, streamline and particle trace visualization techniques for analyzing PH-associated vortical blood flow in the main pulmonary artery on 3D PC-MRI all allow highly accurate diagnosis of PH and estimation of elevated mPAP. However, interobserver variability in vortex identification was smallest with 3D vector visualization.

A previous study employing 3D vector field representation of PC-MRI data established a linear relationship between PH-associated vortical blood flow in the main pulmonary artery and elevated mPAP [3]. In agreement with the findings of the earlier study, the present study found low inter-observer variability in vortex detection, high sensitivity and specificity of PH diagnosis based on tvortex, and strong linear correlation between tvortex and elevated mPAP with 3D vector visualization. Additionally, ROC analysis allowed the specification of an optimal cut-off value of tvortex = 15% for non-invasive diagnosis of PH (mPAP≥25 mmHg).

tvortex values determined from MPR vector, streamline and particle trace visualization correlated strongly with those derived from 3D vector representation, and correlations between elevated mPAP and tvortex values derived from 3D vector, MPR vector, streamline and particle trace visualization did not differ significantly. Cut-off values, as well as linear regression equations and standard deviations for prediction of elevated mPAP were rather similar in all visualization techniques (Figure 4), indicating that all of the techniques tested are appropriate for non-invasive diagnosis of PH.

Numbers of patients with PH-associated vortices, as well as tvortex values, were larger with the use of MPR vector, streamline and particle trace visualization techniques than when 3D vector visualization was used. Furthermore, interobserver variability of tvortex was greater with the former three visualization techniques than with 3D vector analysis. These differences can be explained by the differing abilities of the particular visualization techniques to display and separate complex dynamic blood flow patterns such as valvular-vortical, bifurcation-related, helical, and PH-associated vortical blood flow [18][21].

Interobserver agreement on the presence of PH-associated vortical blood flow was lowest with MPR vector visualization. MPR vectors can be reformatted on arbitrary image orientations in volume, which is helpful to optimize the RVOT view but causes higher variability in visualized patterns because of the observer-dependent angulation of reconstruction. Additionally - and even more importantly - MRP vector visualization, representing 2D in-plane velocities projected onto anatomical images, neglects through-plane components of 3D velocities, impeding differentiation of helical from vortical blood flow. Helical blood flow in the main pulmonary artery, frequently observed in healthy volunteers as well as patients, has been related to anatomy, curvature and contraction characteristics of the right ventricle and outflow tract [20], [22], [23]. As shown in Figure 5, these flow patterns might lead to misinterpretation of helical as PH-associated vortical blood flow in MPR vector visualization. Moreover, beginning in late systole, helical motion of blood does not severely affect definition of tstart but hinders identification of tend and therefore limits the validity of tvortex from MPR vector visualization (Table 2 and 4).

thumbnail
Figure 5. Helical blood flow in the main pulmonary artery mimicking PH-associated vortical blood flow in MPR vector representation (A).

The same perspective in 3D vector visualization is not interpreted as showing a PH-associated vortex (B). Rotation of perspective (C) indicates the helical character of the blood flow. RV: right ventricle, PV: pulmonary valve, MPA: main pulmonary artery.

https://doi.org/10.1371/journal.pone.0082212.g005

In the current study interobserver variability in streamline visualization was moderate; furthermore, in determining which patients had PH-associated vortices, streamline visualization agreed with 3D vector visualization less often than any other visualization technique investigated. Both results can be explained by the fact that notional curves of velocity vectors, synonymously notional streamlines, are subject to some kind of visual filtering: A single closed streamline, or a few concentric calculated streamlines, which might already be assigned as PH-associated vortical blood flow, will often not be noticed in vector representation. This also explains the larger tvortex values and later tends values found in streamline compared to vector visualization. The fact that tstart was generally detected later in streamline than vector visualization relates to the typical behavior of streamlines: that is, uniformly distributed seeding points do not result in evenly spaced streamlines but in clusters in regions with high velocities, and appear low-density in low-velocity regions [5], [24]. As the onset of PH-associated vortical blood flow typically starts in systole (Table 3) during high-velocity blood flow in the main pulmonary artery, it may easily happen that small regions of closed streamlines are occluded or underrepresented and therefore missed. An example of tstart being detected later in streamline than in 3D vector visualization is shown in Figure 6.

thumbnail
Figure 6. Blood flow patterns in the main pulmonary artery of a patient with PH in streamline (A, B) and 3D vector visualization (C, D) in two consecutive systolic cardiac phases.

Onset of vertical blood flow tstart is defined later in streamline (B) than in 3D vector visualization (C). RV: right ventricle, PV: pulmonary valve, MPA: main pulmonary artery.

https://doi.org/10.1371/journal.pone.0082212.g006

As particle traces observed in a specific cardiac phase are trajectories of particles seeded in previous times, they do not necessarily coincide with streamlines in time-varying velocity fields [12], [25], [26]. The strong correlation between tvortex in particle trace and 3D vector visualization first of all confirms that RVOT-oriented, ring-shaped blood circulation in the main pulmonary artery (the definition of a PH-associated vortex in particle trace visualization) corresponds to a PH-associated vortex observed in vector and streamline visualization. The striking difference between PH-associated vortices detected in particle trace and those detected in 3D vector visualization was that onset and termination of vortical blood flow was identified substantially later in particle trace visualization. This result can be explained by the fact that 1) particles were defined to have zero-length in the first time-frame, 2) seeded particles need some time to cover distances long enough to be identified as a ring-shaped particle trace (in particular when velocity is low), and 3) long-living vortices observed at very high mPAP might last until the next cardiac cycle, which was not considered in the current particle trace implementation. The high tvortex values shown in Figure 4D might therefore be slightly underestimated.

In the present study, there were limitations with respect to data acquisition and visualization that will need to be addressed. Underestimation of high mPAP values from particle trace analysis due to underrating of tvortex could be prevented by cyclic particle seeding, which was not performed. In general the results for streamlines and particle traces are limited to standard 3D integral curve visualizations provided by the employed 4D Flow prototype software, and additional techniques such as opacity, illumination or topology passed seeding strategies might further improve non-invasive mPAP estimation [5][7], [24]. In order to cover the main pulmonary artery in reasonable imaging time, PC-MRI data were acquired with 6-mm slice thickness and measured time resolution of 89 ms. Although spatially interpolated, calculation of streamlines and particle traces might be limited due to the anisotropic 3D velocity field acquired. The moderate time resolution might have furthermore influenced the precision of calculated particle traces. Finally, performance of 3D vector visualization was privileged by the coincidence of the chosen imaging and the rotational plane of PH-associated vortices.

In conclusion, 3D vector, MPR vector, streamline and particle trace visualization of time-resolved 3D PC-MRI data of the main pulmonary artery can be employed for accurate vortex-based diagnosis of PH and estimation of elevated mPAP. As all visualizations have advantages and drawbacks, their combined usage might contribute to a refined understanding and an establishment of automatic extraction of topological features of blood flow patterns in the main pulmonary artery.

Acknowledgments

The authors thank Ada Muellner, MS, for editing the manuscript.

Author Contributions

Conceived and designed the experiments: UR GR GK HO MF. Performed the experiments: UR GR GK. Analyzed the data: UR GR. Contributed reagents/materials/analysis tools: UR GR AFS MAG AG. Wrote the paper: UR GR GK AFS MAG AG HO MF.

References

  1. 1. McLaughlin VV, Archer SL, Badesch DB, Barst RJ, Farber HW, et al. (2011) American College of Cardiology Foundation Task Force on Expert Consensus Documents; American Heart Association; American College of Chest Physicians; American Thoracic Society, Inc; Pulmonary Hypertension Association. ACCF/AHA 2009 expert consensus document on pulmonary hypertension a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents and the American Heart Association developed in collaboration with the American College of Chest Physicians; American Thoracic Society, Inc.; and the Pulmonary Hypertension Association. J Am Coll Cardiol 53(17): 1573–1619.
  2. 2. Galiè N, Torbicki A, Barst R, Dartevelle P, Haworth S, et al. (2004) Guidelines on diagnosis and treatment of pulmonary arterial hypertension. The Task Force on Diagnosis and Treatment of Pulmonary Arterial Hypertension of the European Society of Cardiology. Eur Heart J 25: 2243–2278.
  3. 3. Reiter G, Reiter U, Kovacs G, Kainz B, Schmidt K, et al. (2008) Magnetic resonance-derived 3-dimensional blood flow patterns in the main pulmonary artery as a marker of pulmonary hypertension and a measure of elevated mean pulmonary arterial pressure. Circ Cardiovasc Imaging 1(1): 23–30.
  4. 4. Bradlow WM, Gibbs JS, Mohiaddin RH (2012) Cardiovascular magnetic resonance in pulmonary hypertension. J Cardiovasc Magn Reson 14: 6.
  5. 5. Post FH, Vrolijk B, Hauser H, Laramee RS, Doleisch H. (2002) Feature Extraction and Visualization of Flow Fields. In Eurographics 2002 State-of-the-Art Reports; The Eurographics Association: 69–100.
  6. 6. McLoughlin T, Laramee RS, Peikert R, Post FH, Chen M (2013) Over Two Decades of Integration-Based, Geometric Flow Visualization. Computer Graphics Forum 29(6): 1807–1829.
  7. 7. Krishnan H, Garth C, Gühring J, Gülsün MA, Greiser A, et al. (2012) Analysis of time-dependent flow-sensitive PC-MRI data. IEEE Trans Vis Comput Graph 18(6): 966–977.
  8. 8. Napel S, Lee DH, Frayne R, Rutt BK (1992) Visualizing three-dimensional flow with simulated streamlines and three-dimensional phase contrast MR imaging. J Magn Reson Imaging 2: 143–153.
  9. 9. Mohiaddin RH, Yang GZ, Kilner PJ (1994) Visualization of flow by vector analysis of multidirectional cine magnetic resonance velocity mapping: technique and application. J Comput Assist Tomogr 18: 383–392.
  10. 10. Buonocore MH (1998) Visualizing blood flow patterns using streamlines, arrows, and particle paths. Magn Reson Med 40: 210–226.
  11. 11. Markl M, Frydrychowicz A, Kozerke S, Hope M, Wieben O (2012) 4D flow MRI. J Magn Reson Imaging 36(5): 1015–1036.
  12. 12. Oertel H (2004) Dynamics of fluid flow. In: New York, NY: Springer Oertel H (editor). Prandtl’s Essentials of Fluid Mechanics 2nd, ed. 2004: 50.
  13. 13. Wigström L, Ebbers T, Fyrenius A, Karlsson M, Engvall J, et al. (1999) Particle trace visualization of intracardiac flow using time-resolved 3D phase contrast MRI. Magn Reson Med 41(4): 793–9.
  14. 14. Pelc NJ, Herfkens RJ, Shimakawa A, Enzmann DR (1991) Phase Contrast Cine Magnetic Resonance Imaging. Magn Reson Q. 7: 229–254.
  15. 15. Reiter G, Reiter U, Kainz B, Greiser A, Bischof H, et al.. (2007) MR vector field measurement and visualization of normal and pathological time-resolved three-dimensional cardiovascular blood flow patterns. J Cardiovasc Magn Reson 9: 237–238. Abstract.
  16. 16. Gulsun MA, Jolly M-P, Guehring J, Guetter C, Littmann A, et al.. (2012) A Novel 4D Flow Tool for Comprehensive Blood Flow Analysis. ISMRM 2012: 1176. Abstract.
  17. 17. Bernstein MA, Zhou XJ, Polzin JA, King KF, Ganin A, et al. (1998) Concomitant gradient terms in phase contrast MR: analysis and correction. Magn Reson Med 39(2): 300–308.
  18. 18. Markl M, Draney MT, Hope MD, Levin JM, Chan FP, et al. (2004) Time-resolved 3-dimensional velocity mapping in the thoracic aorta: visualization of 3-directional blood flow patterns in healthy volunteers and patients. J Comput Assist Tomogr 28(4): 459–468.
  19. 19. Frydrychowicz A, Markl M, Hirtler D, Harloff A, Schlensak C, et al. (2011) Aortic hemodynamics in patients with and without repair of aortic coarctation: in vivo analysis by 4D flow-sensitive magnetic resonance imaging. Invest Radiol 46(5): 317–325.
  20. 20. Bächler P, Pinochet N, Sotelo J, Crelier G, Irarrazaval P, et al. (2013) Assessment of normal flow patterns in the pulmonary circulation by using 4D magnetic resonance velocity mapping. Magn Reson Imaging 31(2): 178–188.
  21. 21. Frydrychowicz A, Berger A, Munoz Del Rio A, Russe MF, Bock J, et al. (2012) Interdependencies of aortic arch secondary flow patterns, geometry, and age analysed by 4-dimensional phase contrast magnetic resonance imaging at 3 Tesla. Eur Radiol 22(5): 1122–1130.
  22. 22. François CJ, Srinivasan S, Schiebler ML, Reeder SB, Niespodzany E, et al. (2012) 4D cardiovascular magnetic resonance velocity mapping of alterations of right heart flow patterns and main pulmonary artery hemodynamics in tetralogy of Fallot. J Cardiovasc Magn Reson 14: 16.
  23. 23. Sundareswaran KS, Haggerty CM, de Zélicourt D, Dasi LP, Pekkan K, et al. (2012) Visualization of flow structures in Fontan patients using 3-dimensional phase contrast magnetic resonance imaging. J Thorac Cardiovasc Surg 143(5): 1108–1116.
  24. 24. Ye X., Kao D., Pang A (2005) Strategy for seeding 3D streamlines. In: Proceedings of IEEE Visualization: 471–476.
  25. 25. Hope MD, Hope TA, Meadows AK, Ordovas KG, Urbania TH, et al. (2010) Bicuspid aortic valve: four-dimensional MR evaluation of ascending aortic systolic flow patterns. Radiology 255(1): 53–61.
  26. 26. Bogren HG, Buonocore MH (2010) Helical-shaped streamlines do not represent helical flow. Radiology 257(3): 895–896.