1H/31P Polarization Transfer at 9.4 Tesla for Improved Specificity of Detecting Phosphomonoesters and Phosphodiesters in Breast Tumor Models

  • Jannie P. Wijnen ,

    Contributed equally to this work with: Jannie P. Wijnen, Lu Jiang

    Affiliations Johns Hopkins University In vivo Cellular and Molecular Imaging Center, The Russell H. Morgan Department of Radiology and Radiological Science, Division of Cancer Imaging Research, The Johns Hopkins University School of Medicine, Baltimore, Maryland, United States of America, Department of Radiology, University Medical Centre Utrecht, Utrecht, Netherlands

  • Lu Jiang ,

    Contributed equally to this work with: Jannie P. Wijnen, Lu Jiang

    Affiliation Johns Hopkins University In vivo Cellular and Molecular Imaging Center, The Russell H. Morgan Department of Radiology and Radiological Science, Division of Cancer Imaging Research, The Johns Hopkins University School of Medicine, Baltimore, Maryland, United States of America

  • Tiffany R. Greenwood,

    Affiliation Johns Hopkins University In vivo Cellular and Molecular Imaging Center, The Russell H. Morgan Department of Radiology and Radiological Science, Division of Cancer Imaging Research, The Johns Hopkins University School of Medicine, Baltimore, Maryland, United States of America

  • Wybe J. M. van der Kemp,

    Affiliation Department of Radiology, University Medical Centre Utrecht, Utrecht, Netherlands

  • Dennis W. J. Klomp,

    Affiliation Department of Radiology, University Medical Centre Utrecht, Utrecht, Netherlands

  • Kristine Glunde

    kglunde@mri.jhu.edu

    Affiliations Johns Hopkins University In vivo Cellular and Molecular Imaging Center, The Russell H. Morgan Department of Radiology and Radiological Science, Division of Cancer Imaging Research, The Johns Hopkins University School of Medicine, Baltimore, Maryland, United States of America, Sidney Kimmel Comprehensive Cancer, Johns Hopkins University School of Medicine, Baltimore, Maryland, United States of America

Abstract

Purpose

To assess the ability of a polarization transfer (PT) magnetic resonance spectroscopy (MRS) technique to improve the detection of the individual phospholipid metabolites phosphocholine (PC), phosphoethanolamine (PE), glycerophosphocholine (GPC), and glycerophosphoethanolamine (GPE) in vivo in breast tumor xenografts.

Materials and Methods

The adiabatic version of refocused insensitive nuclei enhanced by polarization transfer (BINEPT) MRS was tested at 9.4 Tesla in phantoms and animal models. BINEPT and pulse-acquire (PA) 31P MRS was acquired consecutively from the same orthotopic MCF-7 (n = 10) and MDA-MB-231 (n = 10) breast tumor xenografts. After in vivo MRS measurements, animals were euthanized, tumors were extracted and high resolution (HR)-MRS was performed. Signal to noise ratios (SNRs) and metabolite ratios were compared for BINEPT and PA MRS, and were also measured and compared with that from HR-MRS.

Results

BINEPT exclusively detected metabolites with 1H-31P coupling such as PC, PE, GPC, and GPE, thereby creating a significantly improved, flat baseline because overlapping resonances from immobile and partly mobile phospholipids were removed without loss of sensitivity. GPE and GPC were more accurately detected by BINEPT in vivo, which enabled a reliable quantification of metabolite ratios such as PE/GPE and PC/GPC, which are important markers of tumor aggressiveness and treatment response.

Conclusion

BINEPT is advantageous over PA for detecting and quantifying the individual phospholipid metabolites PC, PE, GPC, and GPE in vivo at high magnetic field strength. As BINEPT can be used clinically, alterations in these phospholipid metabolites can be assessed in vivo for cancer diagnosis and treatment monitoring.

Introduction

Elevated concentrations of water-soluble phospholipid metabolites such as the phosphomonoesters (PMEs) phosphocholine (PC) and phosphoethanolamine (PE) and the phosphodiesters (PDEs) glycerophosphocholine (GPC) and glycerophosphoethanolamine (GPE) are a metabolic hallmark of cancer [1], [2]. 1H and 31P magnetic resonance spectroscopy (MRS) are able to detect this activated phospholipid metabolism in cancers in vivo, which can be used clinically for cancer diagnosis and treatment monitoring [2]. However, the total choline (tCho) signal detected by 1H MRS cannot be spectrally resolved into individual metabolites, mainly free choline (Cho), PC and GPC, due to the low spectral resolution in vivo at clinical field strengths of 1.5 and 3 Tesla (T), nor can it be resolved at high field strengths of 4 and 7 T [3], [4], [5]. Non-invasive detection of these individual phospholipid metabolites is of high interest as it has been shown that, for example, the PC/GPC ratio can increase with increasing aggressiveness of breast cancer cells [6]. With 31P MRS, individual PME and PDE signals can be detected, however, it suffers from relatively low signal to noise ratio (SNR) due to the intrinsically low sensitivity of the 31P nucleus [7], [8]. Even at high field strength, the detection of individual PE, PC, GPE and GPC is difficult, particularly in heterogeneous cancer tissues in which the homogeneity of the magnetic field is poor [9]. Another significant problem for quantitative in vivo detection with 31P MRS is that these signals typically overlap with signals of other molecules with 31P nuclei, such as sugar phosphates and immobile membrane phospholipids (e.g. phosphatidylcholine, phosphatidylethanolamine), which cause a broad, uneven baseline, which in turn severely hampers the accuracy of metabolite measurement and quantification [10]. Accurate measurement and quantification of these phospholipid metabolites is particularly important in the clinic as consistent changes in phospholipid metabolite levels can aid in cancer diagnosis, prognosis, and treatment response monitoring.

Another way to improve the SNR, baseline flatness, and detection accuracy of PMEs and PDEs is to apply polarization transfer (PT) methods [11] such as refocused insensitive nuclei enhanced by polarization transfer (RINEPT) [12], [13]. PT techniques transfer the polarization of the excited 1H spins through J-coupling to the 31P spins during the echo time period TE1H as shown in Fig. 1, which increases the SNR and removes all resonances without 1H-31P coupling, thereby flattening the baseline. Previous studies at lower field strengths have focused on the ability of 1H-31P PT to increase the SNR. At 1.5T the SNR of PC obtained by using INEPT was comparable to that obtained by pulse-acquire (PA) acquisition [12]. However, for PDEs the SNR gain was 2 when using RINEPT [11]. At 3 T an SNR gain of 2.6, 1.6, 2.2, and 2.4 for PE, PC, GPE, and GPC, respectively, was reported when comparing a selective-RINEPT to PA acquisition in human brain [10]. At 7 T the concept of PT was expanded by combining it with a direct 31P detection sequence in one repetition time, which increases the SNR per unit of time as compared to a PT sequence alone [9].

thumbnail
Figure 1. Schematic overview of the adiabatic version of the refocused insensitive nuclei enhanced by polarization transfer (BINEPT) sequence.

Segmented BIR4 pulses (400 µs per segment, 35 ppm band width) and a full BIR4 180° pulse (400 µs, 35 ppm band width) were used instead of conventional 90 and 180 degree pulses. The echo time of 1/4 J was set to 34 ms, which gave optimal signals for the acquisition of PE and PC. Abbreviations: BIR4, B1 insensitive refocusing; JPH, J-coupling constant; TE1H, effective echo time for proton refocusing; TE31P, effective echo time for phosphorus refocusing.

https://doi.org/10.1371/journal.pone.0102256.g001

As shown in the schematic overview of the adiabatic version of the refocused insensitive nuclei enhanced by polarization transfer (BINEPT) sequence in Fig. 1, the optimal duration for TE1H to achieve polarization transfer is relatively long due to the small J-coupling constants between 31P and 1H spins, which causes substantial signal loss to molecules that have short T2 values for transverse relaxation. Consequently, apart from a higher SNR for PE, PC, GPE, and GPC, PT will simultaneously decrease the SNR of larger 31P-containing molecules (with H-P coupling) that typically exhibit short T2 values and may overlap with these phospholipid metabolites [14]. This results in a flat baseline, which improves the detection of PMEs and PDEs.

To test these considerations, we performed PT at the high field strength of 9.4 T on a preclinical scanner. We demonstrate that an optimized PT technique at 9.4T results in a flat and clean baseline without loss of sensitivity compared to Ernst-angle pulse-acquire acquisition, and in an improved in vivo detection and quantification of PC, PE, GPE, and GPC in two breast cancer xenograft models.

Materials and Methods

Breast cancer xenograft models

Highly aggressive MDA-MB-231 and weakly aggressive MCF-7 breast cancer cells were purchased from the American Type Culture Collection and cultured as previously described [15]. 2×106 MDA-MB-231 or MCF-7 cells in 50 µL of Hank's balanced salt solution (HBSS, Mediatech) were injected into the upper right thoracic mammary fat pad of anesthetized female athymic nude mice with an average weight of 26 g. Estrogen-dependent MCF-7 tumor growth was supported by a 0.18-mg 17β-estradiol 60-day release pellet (Innovative Research of America, Sarasota, FL) implanted subcutaneously into the back of mice 1 week before tumor inoculation. We studied 10 MCF-7 tumors of 1.67±0.59 cc that developed within nine weeks and 10 MDA-MB-231 tumors of 0.94±0.09 cc that developed within seven weeks. The maximum tumor size threshold for animals to be sacrificed was 2.5 cc. All surgical procedures and animal handling were approved by the Johns Hopkins University Institutional Animal Care and Use Committee, and conformed to the NIH Guide for the Care and Use of Laboratory Animals.

Non-invasive in vivo 31P MRS studies

In vivo 31P MRS was performed on a 9.4T Bruker Horizontal Bore Small Animal Scanner (Bruker BioSpin Corp.). A 1H/31P double-tuned solenoid coil with an inner diameter of 12 mm was used (MRcoils BV, Drunen, The Netherlands). The tumor was hanging into the coil while the animal lay on a platform above the coil with an opening for the tumor [16]. Mice were anesthetized by breathing isoflurane (2% in air) through a nose cone. Body temperature was maintained by using a blanket with circulating warm water. Breathing rate was monitored with a movement sensor attached to the abdomen of each mouse. 3D RARE images were acquired using an echo time (TE) of 7.2 ms, repetition time (TR) of 500 ms, RARE factor of 4, flip angle of 90°, field of view (FOV) of 1 cmx1 cmx1 cm, 64 phase encode steps (64×64×64), and number of averages (NA) of 4. 1st and 2nd order B0 shimming was performed manually by minimizing the water line width of the entire tumor. Non-localized PA 31P MR spectra were acquired with adiabatic excitation (BIR4 45°, 200 µs, 120 ppm bandwidth), repetition time of 1 s, and 2000 averages, using a saturation slab to eliminate signals from adjacent muscles [17]. The combination of a short TR and 45° flip angle approximates PA acquisition with Ernst-angle excitation for PE and PC. Subsequently, an MR spectrum was acquired using an adiabatic version of the refocused insensitive nuclei enhanced polarization transfer technique BINEPT with a repetition time of 1 s and 2000 averages, which had an equal acquisition time of 33 min 20 s as PA. Segmented BIR4 pulses and a full BIR4 180° pulse were used as shown in Fig. 1. Both TE1H and TE31P were set to 34 ms (Fig. 1), which was the optimal echo time for the detection of PC and PE [18].

Quantification of in vivo 31P MRS data

Lorentzian lines were fitted to the PA and BINEPT 31P MR spectra using JMRUI 4.0 software [19] and the AMARES algorithm [20]. For PA spectra, the phosphocreatine (PCr) resonance was set to 0 ppm, and line widths of PE, PC, GPE, and GPC were constrained to that of PCr during fitting (Fig. 2). In the BINEPT spectra, which only contained PE, PC, GPE, and GPC, line width was constrained to that of PE, and the chemical shifts of PE and GPC were estimated by JMRUI using a starting value of 6.8 ppm for PE and 2.9 ppm for GPC (Fig. 2). Frequency differences between the resonance positions of PC and PE, and GPC and GPE were fixed to 100 Hz in PA and BINEPT spectra. To be able to compare MR spectra of different mice with different coil loads and gain settings, metabolite levels were quantified as ratios to the noise, which was measured from the standard deviation of the last 200 points in the time domain signal. PE, PC, GPE, and GPC levels from the PA spectra were corrected for differences in T1 relaxation. 31P T1 values of PE, PC, GPE, and GPC were measured in vivo by progressive saturation series in four MDA-MB-231 tumors as shown in Table 1. PE, PC, GPE, and GPC levels from the BINEPT spectra were corrected for polarization transfer efficiency (about 50%) and for 1H T1, 1H T2, and 31P T2 relaxation with estimates based on literature values. 1H relaxation times of tCho were based on values obtained at 7 and 9.4 T [21], [22], and 31P T2 relaxation times were based on values obtained at 3 and 7 T [10], [23], [24] (Table 1).

thumbnail
Figure 2. Raw RARE images and segmentation of corresponding slices.

Necrotic tissue is depicted in red. Segmentation of (a) representative MCF-7 tumor and (b) representative MDA-MB-231 tumor. (c) Correlation between line width of PCr and necrotic fraction for all MCF-7 and MDA-MB-231 tumors measured.

https://doi.org/10.1371/journal.pone.0102256.g002

Dual-phase extraction of tumors

Directly after in vivo MRS measurements, mice were deeply euthanized as verified by toe pinch, and sacrificed by cervical dislocation. The entire tumor was immediately removed, freeze-clamped, carefully weighed for quantification, pulverized in liquid N2, and homogenized in 4 mL of ice-cold methanol with a tissue tearor homogenizer. Both lipid and water-soluble tumor extract fractions were obtained using a dual-phase extraction method based on methanol/chloroform/water (1∶1∶1; v/v/v) as previously described [25], [26].

31P high-resolution (HR)-MRS studies

The lyophilized, water-soluble extract fractions were dissolved in D2O containing 0.24×10−6 mol 3-(trimethylsilyl)propionic-2,2,3,3-d4 acid (TSP; Sigma-Aldrich) and 6.0×10−6 mol phenylphosphonic acid (PPA; Sigma-Aldrich) as internal reference concentration and chemical shift standards. 31P HR-MRS was performed on a Bruker Avance 500 (11.7 T) spectrometer (Bruker BioSpin Corp.) using a 10-mm 31P probe. The MR spectra were acquired using the following acquisition parameters: 60° flip angle, 10162 Hz sweep width, 8 K time domain size, repetition time of 3 s and 2500 averages. The spectra were processed using the MestReC 4.9.9.6 software (Mestrelab Research). Lorentzian lines were fitted to the signals of PPA, PE, PC, GPE, and GPC. The areas under the curves were corrected for differences in T1 relaxation time and potential saturation effects. To this end, we measured the T1 relaxation times of PPA, PE, PC, and GPC with a progressive saturation series in a phantom solution in D2O. The T1-corrected PE, PC, GPE and GPC levels were normalized to tumor weight.

MR data analysis

To assess the quality of our in vivo MR data, we measured the line width of PCr in the in vivo PA data of each tumor. In addition, we calculated the necrotic fraction of each tumor by using the dark areas in the corresponding 3D RARE T1-weighted images of each tumor [27], [28]. The tumor – air boundary was segmented in 3D using a threshold, set to the highest 10% in the histogram. Necrotic dark spots inside the tumor were segmented by manually adjusting a threshold to approximately the lowest 10% in the histogram. Necrotic regions were quantified as necrotic fraction per tumor by counting the number of necrotic voxels in the dark spots inside the tumor divided by the total number of voxels inside the tumor. The correlation between PCr line width and necrotic fraction was measured for each tumor as shown in Figure 2. As the fitting of MR spectra becomes unreliable in cases where the line width is broad, the line width of PCr in the in vivo MR spectra was used as a criterion to exclude tumors with poor spectral resolution and large necrotic regions from further analysis. Data from tumors in which the line width of PCr in the in vivo PA MR spectrum was larger than 140 Hz were excluded from further analysis (vertical line in Fig. 2c), although when comparing the group results based on the data from all tumors, the conclusions do not change. This cutoff value in line width was a tradeoff between including as much data as possible and excluding data with poor quality.

Statistical analysis

A two-sided t-test assuming unequal variances between metabolite levels or ratios was performed to test for statistically significant differences between MDA-MB-231 and MCF-7 tumors, or PA and BINEPT methods.

Results

Our initial quality check of MR data demonstrated a strong positive correlation (Pearson correlation, r = 0.7053, P = 0.0011) between necrotic fraction and spectral line width as shown in Figure 2. This could be explained by the fact that in tumors with larger necrotic fractions more tissue-tissue transitions between viable and necrotic areas were present which degraded the B0 homogeneity, which in turn led to larger spectral line widths (Fig. 2).

PA 31P MRS, as shown in Figure 3, detected the PMEs PE and PC and the PDEs GPE and GPC, in addition to inorganic phosphate (Pi), PCr, and α-, β-, and γ-nucleoside triphosphates (NTP). Data from mice that did not pass the initial spectral quality check (line width of PCr<140 Hz) were removed from further quantification, which resulted in n = 6 for MDA-MB-231 and n = 7 for MCF-7 tumors. BINEPT 31P MRS only detected PE, PC, GPE, and GPC. The PDE signals of GPE and GPC were better visible in the BINEPT spectra compared to the corresponding PA spectra as no overlapping signals from sugar phosphates, large membrane phospholipids, and other phosphorylated compounds were detected in the 0–10 ppm region, leading to a flat baseline throughout the spectrum (Fig. 2). Also, the PME in BINEPT spectra were not overlapping with other phospholipid signals.

thumbnail
Figure 3. Example of in vivo pulse-acquire (PA, top) and BINEPT (bottom) 31P MR spectra of a representative MCF-7 (left) and MDA-MB-231 (right) tumor.

Lorentzian lines as fitted by JMRUI are shown below each MR spectrum. All phosphorylated metabolites are visible in the PA spectrum, whereas the BINEPT spectrum only contains signals from phospholipid metabolites with H-P-coupling such as PE, PC, GPE, and GPC. Note the broad, uneven baseline in the 0–5 ppm region of the PA spectra, where signals from mobile membrane phospholipids are resonating [35], [45], [46]. The signal of β-NTP is formed by β-NTP only. The signal labeled α-NTP is an overlapping signal from α-NTP, α-NDP, NAD, and DPDE. The signal labeled γ-NTP is an overlapping signal from γ-NTP and β-NDP. Typically, β-NTP is the smallest peak of the three NTP signals, however, here, γ-NTP overlaps with a broad baseline signal that makes it appear smaller than β-NTP.

https://doi.org/10.1371/journal.pone.0102256.g003

Figure 4 shows the quantification of the in vivo levels (average ± standard error) of PE, PC, GPE, and GPC obtained with PA and BINEPT 31P MRS. We did not detect any significant differences in PE, PC, GPE, or GPC levels in MDA-MB-231 (n = 6) versus MCF-7 (n = 7) tumors. However, BINEPT 31P MRS detected higher levels of GPE and GPC as compared to the detection by PA 31P MRS within the same tumor model. This was the case for both MBA-MB-231 as well as MCF-7 tumors. The pH of these tumors as determined from the chemical shift difference between PCr and Pi was 6.92±0.10 for MCF-7 tumors and 7.00±0.11 for MDA-MB-231 tumors, which was not significantly different from each other.

thumbnail
Figure 4. In vivo phospholipid metabolite levels as measured by (a, c) PA and (b, d) BINEPT 31P MRS.

Average metabolite levels ± standard error are shown for (a, b) MCF-7 (n = 7) and (c, d) MDA-MB-231 (n = 6) breast tumor xenografts. Statistical results: PA vs BINEPT MCF-7: GPE p = 0.004 *, GPC p = 0.03 **; PA vs BINEPT MDA-MB-231: GPE p = 0.02 #, GPC p = 0.06 ##; MCF-7 vs MDA-MB-231: PA, for all metabolites p>0.20; MCF-7 vs MDA-MB-231: BINEPT, for all metabolites p>0.21.

https://doi.org/10.1371/journal.pone.0102256.g004

Figure 5 shows a representative 31P HR-MR spectrum of the aqueous phase of an MCF-7 breast tumor extract. The PME signals PE and PC as well as the PDE signals GPE and GPC were clearly separated in the 31P HR-MR spectra of tumor extracts (Fig. 5). Figure 6 shows the quantification of the PE, PC, GPE, and GPC concentration in MCF-7 and MDA-MB-231 tumors obtained from the 31P HR-MR spectra. When normalized to tumor weight and the concentration reference PPA, no statistically significant differences in PE, PC, GPE, and GPC concentrations were detected between MBA-MB-231 (n = 6) and MCF-7 (n = 7) tumors (Fig. 6).

thumbnail
Figure 5. 31P High-Resolution MR spectrum of the same MCF-7 tumor as shown in Fig. 2.

The full spectral range is displayed in the spectrum on the bottom. Signals of NTPs are visible in the 0 to -15 ppm region. An expanded region of 0 to 8 ppm is displayed on the top, as well as the region with the PPA reference signal at 13 ppm. The signals of PE, PC, GPE, and GPC are clearly separated.

https://doi.org/10.1371/journal.pone.0102256.g005

thumbnail
Figure 6. Quantification of PE, PC, GPE, and GPC levels from 31P HR-MRS of (a) MCF-7 and (b) MDA-MB-231 breast tumor extracts.

Average metabolite levels ± standard errors are shown and were expressed as concentration in μmol per gram of tumor tissue. Statistical results: MCF-7 vs MDA-MB-231: PE p =  0.22 +, PC p = 0.77 x, GPE p = 0.31 *, GPC p = 0.16 #.

https://doi.org/10.1371/journal.pone.0102256.g006

Figure 7 shows a direct comparison between the in vivo PA, in vivo BINEPT, and tumor extract 31P HR-MRS measurements, which were taken sequentially from the same tumors. The PE/GPE ratio could not be assessed accurately in the PA spectra because GPE was barely detectable, resulting in very small GPE values, leading to high ratios of PE/GPE and large standard deviations, particularly in MDA-MB-231 tumors. There was a difference between PC/GPC ratios obtained from in vivo BINEPT spectra and PA spectra. The value from BINEPT spectra more closely resembled the PC/GPC ratio that was obtained by HR-MRS of tumor extracts, for MCF-7 as well as MDA-MB-231 tumors. The PE/PC ratios obtained from in vivo PA and BINEPT measurements were similar. No significant differences were detected between any of the phospholipid metabolite ratios of MCF-7 versus MDA-MB-231 tumors. However there was a trend towards a higher PE/PC ratio in MCF-7 tumors as compared to MDA-MB-231 tumors.

thumbnail
Figure 7. Comparison of phospholipid metabolite ratios between in vivo PA, in vivo BINEPT, and tumor extract HR 31P MRS measurements of the same (a) MCF-7 and (b) MDA-MB-231 tumors.

Values are given as average ± standard error. The PE/GPE ratios are not shown for in vivo PA measurements because the GPE resonance could not be reliably detected in several PA spectra due to low SNR. Statistical results: PA vs BINEPT: MCF-7 PC/GPC p = 0.02 *, all other ratios in MCF-7 and MDA-MB-231 p>0.12; PA vs tumor extract: MCF-7 PE/PC p = 0.04 #, PC/GPC p = 0.05 ##, all other ratios in MCF-7 and MDA-MB-231 p>0.21; BINEPT vs tumor extract: MCF-7 PE/PC p = 0.05 +, all other ratios in MCF-7 and MDA-MB-231 p>0.24.

https://doi.org/10.1371/journal.pone.0102256.g007

Discussion

The polarization transfer technique BINEPT significantly improved the in vivo detection of the phospholipid metabolites PE, PC, GPE and GPC at high magnetic field strength in human breast tumor xenografts. The BINEPT sequence removed all signals originating from 31P compounds without P-H coupling because its detection is based on P-H polarization transfer. As a consequence, PCr, Pi, and NTP signals, as well as signals from sugar phosphates and large membrane phospholipids with broad line widths in the spectral region between 0 ppm and 10 ppm, were not detected by BINEPT, leading to a flat baseline, which significantly improved the quantification of PE, PC, GPE and GPC by enabling the use of simple line-fitting algorithms.

An alternative way to obtain a flat baseline could be the use of an off resonance saturation pulse. The effect of such a pulse would be that magnetization transfer and chemical exchange would eliminate broad resonances from very large molecules [7]. However, the signal intensities of PE, PC, GPE, and GPC would be attenuated by such a pulse due to magnetization transfer, whereas BINEPT enhances these signal intensities. Proton decoupling applied during acquisition could improve the spectral resolution and SNR. However, at 9.4 Tesla the chemical shift anisotropy is the dominant relaxation mechanism, which will only result in a marginal increase of SNR, as the water line width is in the order of 80 Hz and the J-coupling in PMEs and PDEs is ∼7 Hz. In addition, the possible signal enhancement by nuclear overhauser enhancement (NOE) effect is different for each metabolite and spatially heterogeneous due to the use of a surface coil. This will introduce more variation into the data, which is not desired, and therefore proton decoupling and NOE were not applied in this study. Although the polarization transfer efficiency can also be different between compounds, this efficiency is easily calculated using spin operator calculations, and can therefore be taken into consideration during quantification. Only differences in T2 relaxation will attribute to an unknown variance in polarization transfer efficiency. However, recent innovations in pulse sequence design may even recover this missing information in T2 [23].

Considering absolute quantification, it is important to note that BINEPT, unlike PA, does not depend on the 31P T1 relaxation, but rather on the T1 relaxation of 1H and the T2 relaxation of 31P and 1H spins. We measured the 31P T1 relaxation times at 9.4 T in vivo and estimated the 31P T2 relaxation time to be 150 ms for PE and PC and 80 ms for GPE and GPC based on values reported in the brain at 3 T and the breast at 7 T [24], [29] (Table 1). The 1H T1 relaxation time of the tCho resonance in the brain at 9.4 T is approximately 1400 ms [21] and its 1H T2 relaxation time at 7 T has been reported to be between 150 and 200 ms [22], [30]. Therefore, we assumed the 1H T1 relaxation time of PE, PC, GPE and GPC in the tumor at 9.4 T to be 1400 ms, we estimated their T2 relaxation time to be 120 ms (slightly lower than at 7 T). The fact that we used such estimates made it questionable to directly compare the quantified phospholipid metabolite levels from in vivo BINEPT, in vivo PA, and tumor extract HR 31P MRS measurements.

BINEPT 31P MRS detected higher levels of GPE and GPC as compared to the detection by PA 31P MRS within the same tumor model. This underestimation of GPE and GPC levels by PA is most likely due to limitations of fitting with the JMRUI algorithm, which typically underestimates the fitting of signals with narrow line width and low SNR, such as GPE and GPC, which are overlapping with other broad signals. The 1H and 31P T2 relaxation times used in the quantification are influencing the final metabolite levels measured from the BINEPT spectra displayed in Figure 4. The relaxation times of GPE and GPC are typically less accurately measured than the relaxation times of PE and PC, which have a much higher concentration in most tissues and hence appear with higher SNR. Therefore, the PE/GPE and PC/GPC ratios were measured more reliably by BINEPT than by PA 31P MRS in vivo. BINEPT 31P MRS was best suited for the detection of the PE/GPE ratio since in contrast to the spectra obtained with PA, both PE and GPE appeared with sufficient SNR in the BINEPT spectrum. Accurate detection of PE/GPE and PC/GPC ratios is of great importance in studying cancer tissues, as these ratios are strongly associated with cancer aggressiveness [2], [6].

Comparing ratios of metabolites in tumor extracts and in vivo measurements showed that the PC/GPC ratio obtained with BINEPT better resembled the tumor extract PC/GPC ratio than did the PC/GPC ratio obtained with PA. The quantification of BINEPT MR spectra using estimated values for relaxation times could be used in longitudinal studies on the effects of a given anticancer treatment on choline and ethanolamine metabolism.

The PC/GPC ratios detected in this study are in good agreement with a recent study of Morse et al [31], in which MCF-7 tumors also displayed a slightly higher PC/GPC ratio compared to that of MDA-MB-231 tumors. These findings do not match the PC/GPC levels reported in the MCF-7 and MBA-MB-231 breast cancer cell lines grown in cell culture [6]. This is not surprising in light of the various differences that exist between cell lines in culture versus tumor xenografts. The PE/PC ratios in MCF-7 and MDA-MB-231 breast tumor xenografts in our study were comparable at around 1.5, which points towards a similar activation in choline and ethanolamine phospholipid metabolism in both tumor models, with a higher level of PE than PC. This is in good agreement with the finding that choline kinase alpha, which is overexpressed and activated in human breast cancers [32], has the ability to use ethanolamine as well as choline as substrates to produce PE and PC, respectively [33].

In early attempts with in vivo 31P MRS of human cancers, a combined peak for PME and PDE was detected [34], [35]. Some early studies demonstrated the potential of 31P MRS for monitoring the effects of anticancer treatments [36], [37] by evaluating the ratio of PME to PDE, and for assessing proliferation by evaluating the ratio of PME to γ-NTP [38]. However, at that time, the spectral resolution was insufficient for resolving the individual metabolites of PE, PC, GPE and GPC. With recent technological developments of improved human high-field MR scanners, coils, and pulse sequences for 31P MRS, resolving these individual metabolites has become feasible in the clinical setting [9], [13]. Furthermore, using human tissue biopsies and magic angle spinning 31P HR-MRS, it has been possible to use phospholipid metabolites as markers to discriminate between tumor types and grades [39], [40], evaluate response to treatment [36], predict response to treatment [41], and assess resection margins during breast cancer surgery [42].

As 1H MRS is more sensitive than 31P MRS, it has been more widely used to study tCho in breast cancer. tCho has been used as a marker for malignancy and treatment response, however, the data from different groups show large variability and discrepancy between the findings [3]. Moreover, the 1H MRS detection of tCho in shrinking tumors often fails due to large signals from lipids [43] and poor magnetic field homogeneity (B0), both of which are hampering accurate detection of tCho levels in breast lesions. This inaccuracy in tCho detection is particularly limiting for the assessment of tCho endpoints in longitudinal studies of response to treatment [3]. As 31P MRS does not pose the problem of large overlapping lipid signals and provides detailed information on anabolic and catabolic metabolites in phospholipid metabolism such as PE, PC, GPE, and GPC, it may be a better-suited technique to evaluate the role of choline- as well as ethanolamine-related metabolites in breast cancer. Even though the sensitivity of 31P MRS is relatively low, the PME levels in tumors can be very high, which allows for the possibility of assessing phospholipid metabolism even in small tumors. Furthermore, in the near future, better coils and improved pulse sequences will become available for clinical use, which will increase the sensitivity of 31P MRS. Finally, the ability to detect the individual phospholipid metabolites PE, PC, GPE, and GPC would significantly improve the diagnosis, treatment monitoring, and assessment of tumor response to therapy of breast tumors.

Supporting Information

Acknowledgments

We thank Drs. Artemov and Chacko for help with the experimental set-up of the MR scanners.

This work originates from The Johns Hopkins University School of Medicine, Radiology Department.

Author Contributions

Conceived and designed the experiments: JPW LJ KG. Performed the experiments: JPW LJ TRG. Analyzed the data: JPW JL. Contributed reagents/materials/analysis tools: JPW LJ KG WJMK DJWK. Wrote the paper: JPW LJ KG. Critically revised the manuscript and approved the final manuscript: JPW LJ KG WJMK DWJK TRG.

References

  1. 1. Negendank W (1992) Studies of human tumors by MRS: a review. NMR Biomed 5: 303–324.
  2. 2. Glunde K, Bhujwalla ZM, Ronen SM (2011) Choline metabolism in malignant transformation. Nature reviews Cancer 11: 835–848.
  3. 3. Haddadin IS, McIntosh A, Meisamy S, Corum C, Styczynski Snyder AL, et al. (2009) Metabolite quantification and high-field MRS in breast cancer. NMR Biomed 22: 65–76.
  4. 4. Meisamy S, Bolan PJ, Baker EH, Pollema MG, Le CT, et al. (2005) Adding in vivo quantitative 1H MR spectroscopy to improve diagnostic accuracy of breast MR imaging: preliminary results of observer performance study at 4.0 T. Radiology 236: 465–475.
  5. 5. Klomp DW, van de Bank BL, Raaijmakers A, Korteweg MA, Possanzini C, et al. (2011) 31P MRSI and 1H MRS at 7 T: initial results in human breast cancer. Nmr in Biomedicine 24: 1337–1342.
  6. 6. Aboagye EO, Bhujwalla ZM (1999) Malignant transformation alters membrane choline phospholipid metabolism of human mammary epithelial cells. Cancer Res 59: 80–84.
  7. 7. de Graaf R (2008) in vivo NMR Spectroscopy; Principles and Techniques. Second edtion: John Wiley & Sons Ltd. 570 p.
  8. 8. Arias-Mendoza F, Payne GS, Zakian KL, Schwarz AJ, Stubbs M, et al. (2006) In vivo 31P MR spectral patterns and reproducibility in cancer patients studied in a multi-institutional trial. NMR Biomed 19: 504–512.
  9. 9. van der Kemp WJ, Boer VO, Luijten PR, Wijnen JP, Klomp DW (2012) Increase in SNR for 31P MR spectroscopy by combining polarization transfer with a direct detection sequence. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 68: 353–357.
  10. 10. Klomp DW, Wijnen JP, Scheenen TW, Heerschap A (2008) Efficient 1H to 31P polarization transfer on a clinical 3T MR system. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 60: 1298–1305.
  11. 11. Gonen O, Mohebbi A, Stoyanova R, Brown TR (1997) In vivo phosphorus polarization transfer and decoupling from protons in three-dimensional localized nuclear magnetic resonance spectroscopy of human brain. Magn Reson Med 37: 301–306.
  12. 12. Mancini L, Payne GS, Leach MO (2003) Comparison of polarization transfer sequences for enhancement of signals in clinical 31P MRS studies. Magn Reson Med 50: 578–588.
  13. 13. Wijnen JP, Scheenen TW, Klomp DW, Heerschap A (2010) 31P magnetic resonance spectroscopic imaging with polarisation transfer of phosphomono- and diesters at 3 T in the human brain: relation with age and spatial differences. Nmr in Biomedicine 23: 968–976.
  14. 14. de Haan JH, Klomp DW, Tack CJ, Heerschap A (2003) Optimized detection of changes in glucose-6-phosphate levels in human skeletal muscle by 31P MR spectroscopy. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 50: 1302–1306.
  15. 15. Glunde K, Guggino SE, Solaiyappan M, Pathak AP, Ichikawa Y, et al. (2003) Extracellular acidification alters lysosomal trafficking in human breast cancer cells. Neoplasia 5: 533–545.
  16. 16. Jiang L, Greenwood TR, Artemov D, Raman V, Winnard PT Jr, et al. (2012) Localized hypoxia results in spatially heterogeneous metabolic signatures in breast tumor models. Neoplasia 14: 732–741.
  17. 17. Choi IY, Tkac I, Gruetter R (2000) Single-shot, three-dimensional "non-echo" localization method for in vivo NMR spectroscopy. Magn Reson Med 44: 387–394.
  18. 18. Mancini L, Payne GS, Leach MO (2005) Implementation and evaluation of CSI-localized J cross-polarization for detection of 31P magnetic resonance spectra in vivo. Magn Reson Med 54: 1065–1071.
  19. 19. Naressi A, Couturier C, Devos JM, Janssen M, Mangeat C, et al. (2001) Java-based graphical user interface for the MRUI quantitation package. Magn Reson Mater Phys 12: 141–152.
  20. 20. Vanhamme L, van den Boogaart A, van Huffel S (1997) Improved method for accurate and efficient quantification of MRS data with use of prior knowledge. J Magn Reson 129: 35–43.
  21. 21. Cudalbu C, Mlynarik V, Xin L, Gruetter R (2009) Comparison of T1 relaxation times of the neurochemical profile in rat brain at 9.4 tesla and 14.1 tesla. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 62: 862–867.
  22. 22. Xin L, Schaller B, Mlynarik V, Lu H, Gruetter R (2012) Proton T(1) relaxation times of metabolites in human occipital white and gray matter at 7 T. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine
  23. 23. van der Kemp WJ, Boer VO, Luijten PR, Stehouwer BL, Veldhuis WB, et al. (2013) Adiabatic multi-echo (3)(1)P spectroscopic imaging (AMESING) at 7 T for the measurement of transverse relaxation times and regaining of sensitivity in tissues with short T(2) values. Nmr in Biomedicine 26: 1299–1307.
  24. 24. Stehouwer BL, van der Kemp WJ, Luijten PR, van den Bosch MA, Veldhuis WB, et al. (2014) (31)P magnetic resonance spectroscopy of the breast and the influence of the menstrual cycle. Breast Cancer Res Treat 144: 583–589.
  25. 25. Glunde K, Raman V, Mori N, Bhujwalla ZM (2005) RNA interference-mediated choline kinase suppression in breast cancer cells induces differentiation and reduces proliferation. Cancer Res 65: 11034–11043.
  26. 26. Tyagi RK, Azrad A, Degani H, Salomon Y (1996) Simultaneous extraction of cellular lipids and water-soluble metabolites: evaluation by NMR spectroscopy. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 35: 194–200.
  27. 27. Osbakken MD, Kreider JW, Taczanowsky P (1986) Nuclear magnetic resonance imaging characterization of a rat mammary tumor. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 3: 1–9.
  28. 28. Dodd NJ, Moore JV, Poppitt DG, Wood B (1989) In vivo magnetic resonance imaging of the effects of photodynamic therapy. British journal of cancer 60: 164–167.
  29. 29. Wijnen JP, van der Kemp WJ, Luttje MP, Korteweg MA, Luijten PR, et al. (2012) Quantitative 31P magnetic resonance spectroscopy of the human breast at 7 T. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 68: 339–348.
  30. 30. Marjanska M, Auerbach EJ, Valabregue R, Van de Moortele PF, Adriany G, et al. (2012) Localized 1H NMR spectroscopy in different regions of human brain in vivo at 7 T: T2 relaxation times and concentrations of cerebral metabolites. Nmr in Biomedicine 25: 332–339.
  31. 31. Morse DL, Carroll D, Day S, Gray H, Sadarangani P, et al. (2009) Characterization of breast cancers and therapy response by MRS and quantitative gene expression profiling in the choline pathway. Nmr in Biomedicine 22: 114–127.
  32. 32. Ramirez de Molina A, Gutierrez R, Ramos MA, Silva JM, Silva J, et al. (2002) Increased choline kinase activity in human breast carcinomas: clinical evidence for a potential novel antitumor strategy. Oncogene 21: 4317–4322.
  33. 33. Gallego-Ortega D, Ramirez de Molina A, Ramos MA, Valdes-Mora F, Barderas MG, et al. (2009) Differential role of human choline kinase alpha and beta enzymes in lipid metabolism: implications in cancer onset and treatment. PloS one 4: e7819.
  34. 34. Leach MO, Verrill M, Glaholm J, Smith TA, Collins DJ, et al. (1998) Measurements of human breast cancer using magnetic resonance spectroscopy: a review of clinical measurements and a report of localized 31P measurements of response to treatment. NMR Biomed 11: 314–340.
  35. 35. Smith TA, Glaholm J, Leach MO, Machin L, Collins DJ, et al. (1991) A comparison of in vivo and in vitro 31P NMR spectra from human breast tumours: variations in phospholipid metabolism. Br J Cancer 63: 514–516.
  36. 36. Redmond OM, Stack JP, O'Connor NG, Carney DN, Dervan PA, et al. (1992) 31P MRS as an early prognostic indicator of patient response to chemotherapy. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 25: 30–44.
  37. 37. Bryant DJ, Bydder GM, Case HA, Collins AG, Cox IJ, et al. (1988) Use of phosphorus-31 MR spectroscopy to monitor response to chemotherapy in non-Hodgkin lymphoma. J Comput Assist Tomogr 12: 770–774.
  38. 38. Kalra R, Wade KE, Hands L, Styles P, Camplejohn R, et al. (1993) Phosphomonoester is associated with proliferation in human breast cancer: a 31P MRS study. British journal of cancer 67: 1145–1153.
  39. 39. Solivera J, Cerdan S, Pascual JM, Barrios L, Roda JM (2009) Assessment of 31P-NMR analysis of phospholipid profiles for potential differential diagnosis of human cerebral tumors. NMR Biomed 22: 663–674.
  40. 40. Smith TA, Bush C, Jameson C, Titley JC, Leach MO, et al. (1993) Phospholipid metabolites, prognosis and proliferation in human breast carcinoma. Nmr in Biomedicine 6: 318–323.
  41. 41. Cao MD, Sitter B, Bathen TF, Bofin A, Lonning PE, et al. (2012) Predicting long-term survival and treatment response in breast cancer patients receiving neoadjuvant chemotherapy by MR metabolic profiling. Nmr in Biomedicine 25: 369–378.
  42. 42. Bathen TF, Geurts B, Sitter B, Fjosne HE, Lundgren S, et al. (2013) Feasibility of MR metabolomics for immediate analysis of resection margins during breast cancer surgery. PloS one 8: e61578.
  43. 43. Bolan PJ, DelaBarre L, Baker EH, Merkle H, Everson LI, et al. (2002) Eliminating spurious lipid sidebands in 1H MRS of breast lesions. Magn Reson Med 48: 215–222.
  44. 44. Lagemaat MW, Maas MC, Vos EK, Kobus T, Bitz AK, et al. (2013) Quantification of T1 relaxation times and nuclear Overhauser effect of 31P metabolites in the human prostate at 7T. International Society for Magnetic Resonance in Medicine, conference proceedings 21th Scientitic meeting and exhibition
  45. 45. Cerdan S, Subramanian VH, Hilberman M, Cone J, Egan J, et al. (1986) 31P NMR detection of mobile dog brain phospholipids. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 3: 432–439.
  46. 46. Schiller J, Arnold K (2002) Application of high resolution 31P NMR spectroscopy to the characterization of the phospholipid composition of tissues and body fluids - a methodological review. Medical science monitor : international medical journal of experimental and clinical research 8: MT205–222.