Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

  • Loading metrics

Noninvasive Measurement of Carbon Dioxide during One-Lung Ventilation with Low Tidal Volume for Two Hours: End-Tidal versus Transcutaneous Techniques

  • Hong Zhang,

    Affiliation Department of Anesthesiology and Critical Care Medicine, Peking University First Hospital, Beijing, China

  • Dong-Xin Wang

    wangdongxin@vip.sina.com

    Affiliation Department of Anesthesiology and Critical Care Medicine, Peking University First Hospital, Beijing, China

Abstract

Background

There may be significant difference between measurement of end-tidal carbon dioxide partial pressure (PetCO2) and arterial carbon dioxide partial pressure (PaCO2) during one-lung ventilation with low tidal volume for thoracic surgeries. Transcutaneous carbon dioxide partial pressure (PtcCO2) monitoring can be used continuously to evaluate PaCO2 in a noninvasive fashion. In this study, we compared the accuracy between PetCO2 and PtcCO2 in predicting PaCO2 during prolonged one-lung ventilation with low tidal volume for thoracic surgeries.

Methods

Eighteen adult patients who underwent thoracic surgeries with one-lung ventilation longer than two hours were included in this study. Their PetCO2, PtcCO2, and PaCO2 values were collected at five time points before and during one-lung ventilation. Agreement among measures was evaluated by Bland-Altman analysis.

Results

Ninety sample sets were obtained. The bias and precision when PtcCO2 and PaCO2 were compared were 4.1 ± 6.5 mmHg during two-lung ventilation and 2.9 ± 6.1 mmHg during one-lung ventilation. Those when PetCO2 and PaCO2 were compared were -11.8 ± 6.4 mmHg during two-lung ventilation and -11.8 ± 4.9 mmHg during one-lung ventilation. The differences between PtcCO2 and PaCO2 were significantly lower than those between PetCO2 and PaCO2 at all five time-points (p < 0.05).

Conclusions

PtcCO2 monitoring was more accurate for predicting PaCO2 levels during prolonged one-lung ventilation with low tidal volume for patients undergoing thoracic surgeries.

Introduction

Arterial carbon dioxide partial pressure (PaCO2) is the gold standard in monitoring ventilation during general anesthesia. End-tidal carbon dioxide partial pressure (PetCO2) reflects PaCO2 and becomes a standard monitoring during surgery. However, various pathologic processes of the cardio-respiratory system such as ventilation-perfusion mismatch or shunt as well as changes in patient positioning have been shown to influence the correlation between PaCO2 and PetCO2 [1]. One-lung ventilation (OLV) and lateral decubitus position during thoracic surgery impair ventilation-perfusion matching and, as a result, the difference between PaCO2 and PetCO2.

Transcutaneous carbon oxide partial pressure (PtcCO2) monitoring provides a noninvasive and continuous estimation of PaCO2 by sampling from arterialized capillary blood and is not influenced by ventilation-perfusion mismatch [2]. Previous studies [34] found that during short time OLV (≤ 1 hour) for thoracic surgery, the value of PtcCO2 is closer to PaCO2 than PetCO2.

With the wide-spread use of mini-invasive thoracic surgery and the introduction of lung-protective ventilation strategy during thoracic anesthesia, prolonged hypercapnia originated from low tidal volume OLV in these patients are not uncommon [56]. We designed this study to evaluate the accuracy of PtcCO2 in predicting PaCO2 values during prolonged OLV and permissive hypercapnia during mini-invasive thoracic surgery.

Materials and Methods

The study protocol was approved by the Clinical Research Ethics Committee of Peking University First Hospital (2012[504]). Written informed consent was obtained from each patient.

Eighteen adult patients of ASA physical status I or II who were scheduled to undergo mini-invasive thoracic surgery with an expected OLV duration of two hours or more were recruited for this study. Patients with diagnosed cardiovascular disease were excluded.

No premedication was administrated. Before the induction of general anesthesia, an epidural puncture was performed between the fifth and eighth thoracic interspace and an epidural catheter was inserted. A test dose of 3 ml 1% lidocaine was administered and no other epidural medication was used during anesthesia.

Intraoperative monitoring included a non-invasive blood pressure, pulse oxygen saturation, an electrocardiogram, nasopharyngeal temperature, urine output, peak airway pressure, and direct arterial blood pressure measurement through a radial artery catheter. General anesthesia was induced with propofol (1–2 mg/kg) and remifentanil (effect site target control infusion at a target of 4–6 ng/ml). And rocuronium (0.6mg/kg) was administered to facilitate endotracheal intubation with a double-lumen tube by direct laryngoscopy. Patients were mechanically ventilated in a volume-controlled manner both in the supine and the lateral decubitus position. During two-lung ventilation (TLV), the fresh gas flow, tidal volume, respiratory rate and inspiratory/expiratory ratio were set at 1 L/min oxygen and 1 L/min air, 6–8 ml/kg, 10–12 breath/min, and 1:2, respectively. Anesthesia was maintained with sevoflurane inhalation (end-tidal anesthetic concentration of 0.8 MAC or above) and remifentanil infusion until the end of the surgery. Sufentanil was administered as a bolus when deemed necessary during surgery and before the end of surgery.

For all patients, the position of the double-lumen endobronchial tube was confirmed under direct vision with a fiberoptic bronchoscope (FOB). The patients were then turned to the lateral decubitus position. The tube position was then checked again with the FOB just before OLV, and the effectiveness of lung collapse during OLV was confirmed by direct observation in the operative field. During OLV, the fresh gas flow, tidal volume, respiratory rate and inspiratory/expiratory ratio were set at 1 L/min oxygen, 4–6 ml/kg, 10–16 breath/min, and 1:1.5, respectively, to maintain a SpO2 of 90% or higher and a peak airway pressure lower than 25 cmH2O. Intravenous ephedrine or phenylephrine or nicardipine was administrated to maintain blood pressure fluctuation within 30% from baseline. Additional doses of rocuronium were administered to maintain muscle relaxation.

PtcCO2 was measured with a TCM3 transcutaneous CO2/oxygen device (Radiometer, Copenhagen, Denmark). The monitoring technique was standardized by applying the probe to the upper part of the patient’s dependent arm in the lateral decubitus position. Before each study, the device was calibrated by using a two-point self-calibration (5% and 10% CO2) and the working temperature of the electrode was maintained at 42°C to “arterialize” the skin capillary blood flow according to the manufacturer’s recommendations. The monitor used an internal adjustment to compensate for the effects of the heated probe on CO2 tension. It took appropriately 20 minutes for initial stabilization after probe attachment. The end-tidal concentrations of the anesthetics and CO2 were measured with a AS/5 monitor (Datex-Ohmeda, Helsinki, Finland) which was calibrated in 5% CO2 and 20.9% oxygen gas before the study. Continuous sampling was obtained from a connector attached to the proximal end of heat moisture exchanger in the respiratory circuit. Arterial blood samples were obtained during TLV, just before the initiation of OLV, and every 30 minutes during OLV until 120 minutes. Arterial blood gas analysis was performed using a GEM premier 3000 analyzer (Instrumentation Laboratory, USA). Data of PtcCO2, PetCO2, and PaCO2 monitoring results were collected simultaneously. The heart rate, mean arterial pressure, pulse oxygen saturation, and nasopharyngeal temperature were also recorded at same time-points.

Quantitative data were presented as means ± standard deviation (SD). Bland-Altman method was used to analyze the agreement between PaCO2 and PetCO2 or between PaCO2 and PtcCO2. The bias (the mean difference between the values) and the precision (the SD of the bias) were calculated. Student’s unpaired t-tests were also used to compare the differences between PaCO2 and PetCO2 or between PaCO2 and PtcCO2. A p value of less than 0.05 was regarded as statistically significant. Statistical analysis was conducted using SPSS version 14.0 (Chicago, IL, USA).

Results

All eighteen patients completed the study protocol. The demographic data were shown in Table 1. Surgical procedures included lobectomy or pneumonectomy for lung cancer, and thymectomy for thymoma. A total of 90 data sets consisting of the simultaneous measurements of PtcCO2, PetCO2 and PaCO2 at five time points were obtained (Table 2). The heart rate and mean arterial pressure did not significantly change from the preoperative values during the study period. The body temperature remained constant between 35.5 and 36.5°C.

When PtcCO2 and PaCO2 were compared, the bias and precision were 4.1 ± 6.5 mmHg during TLV and 2.9 ± 6.1 mmHg during OLV, respectively (Fig 1). When PetCO2 and PaCO2 were compared, the bias and precision were -11.8 ± 6.4 mmHg during TLV and -11.8 ± 4.9 mmHg during OLV, respectively (Fig 2). The values of bias and precision were stable and the difference between PtcCO2 and PaCO2 was significantly lower than that between PetCO2 and PaCO2 throughout the 2-hour period of OLV (Table 3).

thumbnail
Fig 1. Agreement between transcutaneous CO2 (PtcCO2) and arterial CO2 (PaCO2).

Bland-Altman analysis of PtcCO2 versus PaCO2 during two-lung ventilation (TLV) and one-lung ventilation (OLV). Bias was labeled. The 95% limits of agreement of the average PtcCO2 –PaCO2 difference during TLV and OLV were 4.1 ± 6.5 mmHg and 2.9 ± 6.1 mmHg (mean ± 1.96 standard deviation), respectively.

https://doi.org/10.1371/journal.pone.0138912.g001

thumbnail
Fig 2. Agreement between end-tidal CO2 (PetCO2) and arterial CO2 (PaCO2).

Bland-Altman analysis of PetCO2 versus PaCO2 during two-lung ventilation (TLV) and one-lung ventilation (OLV). Bias was labeled. The 95% limits of agreement of the average PetCO2 –PaCO2 difference during TLV and OLV were -11.8 ± 6.4 mmHg and -11.8 ± 4.9 mmHg (mean ± 1.96 standard deviation), respectively.

https://doi.org/10.1371/journal.pone.0138912.g002

Discussion

This study demonstrates that PtcCO2 monitoring is a more accurate estimation of PaCO2 than PetCO2 during OLV of two hours or more with low tidal volume for thoracic surgery. Our results are consistent with previous ones on the accuracy of PtcCO2 monitoring during thoracic surgery of shorter duration [34]. Moreover, we find that the PtcCO2 is accurate of in estimating the PaCO2 in the clinical condition of permissive hypercapnia which is not observed in previous studies [34].

PtcCO2 monitoring used in the present study is a non-invasive method for continuous measurement of transcutaneous CO2 partial pressure. It uses skin electrodes to quantify the amount of CO2 that diffuses to the electrode on the surface of the skin. Local heating is required in order to increase the local blood circulation in the capillary bed below the sensor. The accuracy of PtcCO2 monitoring is influenced by several factors including methodological limitations (e.g., stabilization time, reaction time, periodic repositioning of the sensor, the need for membrane restoration, and baseline calibration), technical mistakes (e.g., improper application of the sensor, trapped air bubbles in the electrolyte solution, damage to the sensor membrane, improper calibration) and hypoperfusion (e.g., vasoconstriction, hypothermia, shock, low cardiac output or local edema) [7].

Since its introduction into clinical practice, PtcCO2 monitoring has received great attention in neonates [89]. Later on, with the improvement of technology and increasing knowledge of this monitoring modality, its use has been increasing in pediatric patient [1011], in adult patients undergoing thoracic [34] and laparoscopic surgery [1215] and in patients after surgery [1617]. The accuracy of PtcCO2 monitoring has been confirmed by these studies. However, some others reported that PtcCO2 monitoring cannot precisely predict PaCO2 in preterm infants during in their first 24 hours [18] and in some patients receiving artificial ventilation during general anesthesia [19], possibly because of the aforementioned reasons. Considering the relative high price and the failure possibility, appropriate indication should be considered in chosing this monitoring method. One suitable clinical condition for PtcCO2 monitoring is OLV during thoracic surgery.

OLV is essential for the success of mini-invasive thoracic surgery. Advances in this technique have enabled more complex intrathoracic procedures being performed. Protective ventilation strategy [56] is now widely used during OLV and consists of small tidal volumes, low inspired oxygen fraction, low airway pressures, permissive hypercapnia, and positive end expiratory pressure. Our results demonstrated the applicability of PtcCO2 during a 2-hour period of OLV with hypercapnia for thoracic surgery. The study of Oshibuchi et al. [3] found that PtcCO2 monitoring provides a more accurate estimation of PaCO2 than PetCO2 during an 1-hour period of OLV with normocapnia (PaCO2 in the range of 30–50 mmHg) for thoracic surgery. In another study, Kelly et al. [20] reported that the agreement between PtcCO2 and PaCO2 deteriorated at high PaCO2 levels (>60 mmHg). Our results extrapolate these ranges, i.e., a 2-hour period of OLV with hypercapnia (PaCO2 in the range of 35–70 mmHg) during thoracic surgery. Long time PtcCO2 monitoring (longer than 2 hours) was only observed in nonsurgical patients in previous studies [2122].

A difference of 5 mmHg or less between PaCO2 and other carbon dioxide measurement was a clinically acceptable discrepancy [4,13]. Our results found that the bias between PtcCO2 and PaCO2 was less than 5 mmHg during either TLV or OLV, whereas the bias between PetCO2 and PaCO2 was lower than -11 mmHg during either TLV or OLV. In the scatter diagram of PetCO2 and PaCO2, 3 of 90 points were outside of the limits of agreement during OLV and all 3 points were beyond 22 mmHg. Whereas in the scatter diagram of PtcCO2 and PaCO2, 4 of 90 points were outside of the limits but only 1 point was beyond 22 mmHg. These also indicated the superiority of PtcCO2 in predicting PaCO2.

PetCO2 monitoring is still the most convenient method in CO2 measurement and has a unique role in judging the position of artificial airway and ventilation status. The difference between PaCO2 and PetCO2 increases with age, pulmonary disorders, pulmonary embolism, reduced cardiac output, hypervolemia, anesthesia, and other conditions that increase the ventilation-perfusion mismatch. In the present study, the mean difference between PetCO2 and PaCO2 during either TLV or OLV was higher than the previously reported ones in similar patients [34]. This is perhaps because the tidal volume settings during TLV (6–8 ml/kg) and OLV (4–6ml/kg) were lower in our study than in previous one (10 ml/kg throughout the operation) [3]. According to respiratory physiology, the ratio of dead space to tidal volume determines the gradient between PaCO2 and PetCO2 assuming that the PaCO2 remains constant. The higher difference between PetCO2 and PaCO2 in our study supports the use of PtcCO2 monitoring during prolonged OLV with lung-protective strategy of low tidal volume ventilation.

Conclusions

In conclusion, our study demonstrated that PtcCO2 is more accurate than PetCO2 in estimating PaCO2 during prolonged OLV with low tidal volume ventilation for thoracic surgery.

Supporting Information

Acknowledgments

We thank Dr. Chen-Xiao Zhu (Department of Anesthesiology, Peking University First Hospital) for her help in conducting this study. We thank Ms. Sai-Nan Zhu (Department of Biostatistics, Peking University First Hospital) for her help in statistical analysis. We also thank Dr. Hong Liu (Department of Anesthesiology and Pain Medicine, University of California Davis Health System) for his critical comments during the manuscript preparation stage.

Author Contributions

Conceived and designed the experiments: HZ DXW. Performed the experiments: HZ. Analyzed the data: HZ. Contributed reagents/materials/analysis tools: HZ. Wrote the paper: HZ DXW.

References

  1. 1. Whitesell R, Asiddao C, Gollman D, Jablonski J. Relationship between arterial and peak expired carbon dioxide pressure during anesthesia and factors influencing the difference. Anesth Analg 1981;60:508–12. pmid:6787952
  2. 2. Huttmann SE, Windisch W, Storre JH. Techniques for the measurement and monitoring of carbon dioxide in the blood. Ann Am Thorac Soc 2014; 11:645–52. pmid:24701974
  3. 3. Oshibuchi M, Cho S, Hara T, Tomiyasu S, Makita T, Sumikawa K. A comparative evaluation of transcutaneous and end-tidal measurements of CO2 in thoracic anesthesia. Anesth Analg 2003; 97:776–9. pmid:12933401
  4. 4. Tobias JD. Noninvasive carbon dioxide monitoring during one-lung ventilation: End-Tidal versus transcutaneous techniques. J Cardiothorac Vasc Anesth 2003; 17:306–8. pmid:12827576
  5. 5. Kilpatrick B, Slinger P. Lung protective strategies in anaesthesia. Br J Anaesth 2010; 105 Suppl 1:i108–16. pmid:21148650
  6. 6. Brassard CL, Lohser J, Donati F, Bussieres JS. Step-by-step clinical management of one-lung ventilation: continuing professional development. Can J Anesth 2014; 61:1103–21. pmid:25389025
  7. 7. Eberhard P. The design, use, and results of transcutaneous carbon dioxide analysis: Current and future directions. Anesth Analg 2007; 105(6 Suppl):S48–52. pmid:18048898
  8. 8. Binder N, Atherton H, Thorkelsson T, Hoath SB. Measurement of transcutaneous carbon dioxide in low birthweight infants during the first two weeks of life. Am J Perinatol 1994; 11:237–41. pmid:8048993
  9. 9. Bernet V, Döll C, Cannizzaro V, Ersch J, Frey B, Weiss M. Longtime performance and reliability of two different ptcCO2 and spOO2 sensors in neonates. Paediatr Anaesth 2008; 18:872–7. pmid:18768047
  10. 10. Dullenkopf A, Bernardo SD, Berger F, Fasnacht M, Gerber AC, Weiss M. Evaluation of a new combined spOO2/ptcCO2 sensor in anaesthetized paediatic patients. Paediatr Anaesth 2003; 13:777–84. pmid:14617118
  11. 11. Tobias JD. Transcutaneous carbon dioxide monitoring in infants and children. Paediatr Anaesth 2009; 19:434–44. pmid:19236597
  12. 12. Bhavani-Shankar K, Steinbrook RA, Mushlin PS, Freiberger D. Transcutaneous PCO2 monitoring during laparoscopic cholecystectomy in pregnancy. Canadian Journal of Anaesthesia 1998; 45:164–9 pmid:9512853
  13. 13. Xue Q, Wu X, Jin J, Yu B, Zheng M. Transcutaneous carbon dioxide monitoring accurately predicts arterial carbon dioxide partial pressure in patients undergoing prolonged laparoscopic surgery. Anesth Analg 2010; 111:417–20 pmid:20584872
  14. 14. Liu S, Sun J, Chen X, Yu Y, Liu X, Liu C. The application of transcutaneous CO2 pressure monitoring in the anesthesia of obese patients undergoing laparoscopic bariatric surgery. PLoS One 2014; 9(4):e91563. pmid:24699267
  15. 15. De Oliveira GS, Ahmad S, Fitzgerald PC, McCarthy RJ. Detection of hypoventilation during deep sedation in patients undergoing ambulatory gynaecological hysteroscopy: A comparison between transcutaneous and nasal end-tidal carbon dioxide measurements. Br J Anaesth 2010; 104:774–8. pmid:20418266
  16. 16. Hirabayashi M, Fujiwara C, Ohtani N, Kagawa S, Kamide M. Transcutaneous PCO2 monitors are more accurate than end-tidal PCO2 monitors. J Anesth 2009; 23:198–202. pmid:19444557
  17. 17. Roediger R, Beck-Schimmer B, Theusinger OM, Rusch D, Seifert B, Spahn DR, et al. The revised digital transcutaneous PCO2/spOO2 ear sensor is a reliable noninvasive monitoring tool in patients after cardiac surgery. J Cardiothorac Vasc Anesth 2011; 25:243–9. pmid:20851636
  18. 18. Aliwalas LL, Noble L, Nesbitt K, Fallah S, Shah V, Shah PS. Agreement of carbon dioxide levels measured by arterial, transcutaneous and end tidal methods in preterm infants < or = 28 weeks gestation. J Perinatol 2005; 25:26–9. pmid:15496874
  19. 19. Nishiyama T, Nakamura S, Yamashita K. Effects of the electrode temperature of a new monitor, TCM4, on the measurement of transcutaneous oxygen and carbon dioxide tension. J Anesth 2006; 20:331–4. pmid:17072703
  20. 20. Kelly AM, Klim S. Agreement between arterial and transcutaneous PCO2 in patients undergoing non-invasive ventilation. Respir Med 2011; 105:226–9. pmid:21131188
  21. 21. Storre JH, Steurer B, Kabitz HJ, Dreher M, Windisch W. Transcutaneous PCO2 monitoring during initiation of noninvasive ventilation. Chest 2007; 132:1810–6. pmid:18079217
  22. 22. Janssens JP, Laszlo A, Uldry C, Titelion V, Picaud C, Michel JP. Non-invasive (transcutaneous) monitoring of PCO2 (tcpCO2) in older adults. Gerontology 2005; 51:174–8. pmid:15832044