  • Sona Ghadimi,

    Affiliations Faculty of Electrical Engineering, K.N. Toosi University of Technology, Tehran, Iran, Inserm UMR 1105, Faculté de Médecine, Université de Picardie Jules Verne, Amiens, France

  • Mehrana Mohtasebi,

    Affiliation Faculty of Electrical Engineering, K.N. Toosi University of Technology, Tehran, Iran

  • Hamid Abrishami Moghaddam ,

    hamid.abrishami@u-picardie.fr

    Affiliations Faculty of Electrical Engineering, K.N. Toosi University of Technology, Tehran, Iran, Inserm UMR 1105, Faculté de Médecine, Université de Picardie Jules Verne, Amiens, France

  • Reinhard Grebe,

    Affiliation Inserm UMR 1105, Faculté de Médecine, Université de Picardie Jules Verne, Amiens, France

  • Masoumeh Gity,

    Affiliation Tehran University of Medical Sciences, Tehran, Iran

  • Fabrice Wallois

    Affiliations Inserm UMR 1105, Faculté de Médecine, Université de Picardie Jules Verne, Amiens, France, Inserm UMR 1105, Centre Hospitalier Universitaire d'Amiens, Amiens, France

Abstract

Neonatal MR templates are appropriate for brain structural analysis and spatial normalization. However, they do not provide the essential accurate details of cranial bones and fontanels-sutures. Distinctly, CT images provide the best contrast for bone definition and fontanels-sutures. In this paper, we present, for the first time, an approach to create a fully registered bimodal MR-CT head template for neonates with a gestational age of 39 to 42 weeks. Such a template is essential for structural and functional brain studies, which require precise geometry of the head including cranial bones and fontanels-sutures. Due to the special characteristics of the problem (which requires inter-subject inter-modality registration), a two-step intensity-based registration method is proposed to globally and locally align CT images with an available MR template. By applying groupwise registration, the new neonatal CT template is then created in full alignment with the MR template to build a bimodal MR-CT template. The mutual information value between the CT and the MR template is 1.17 which shows their perfect correspondence in the bimodal template. Moreover, the average mutual information value between normalized images and the CT template proposed in this study is 1.24±0.07. Comparing this value with the one reported in a previously published approach (0.63±0.07) demonstrates the better generalization properties of the new created template and the superiority of the proposed method for the creation of CT template in the standard space provided by MR neonatal head template. The neonatal bimodal MR-CT head template is freely downloadable from https://www.u-picardie.fr/labo/GRAMFC.

Introduction

Neonatal brain atlases play a crucial role to study normal growth and developmental changes in brain tissues, diagnosis of abnormal anatomical variants, surgical planning, and neuroimaging analysis [17]. These atlases are constructed from a population of neurological healthy neonates and represent the typical head structure of healthy neonates.

The majority of published neonatal brain atlases are based on T1- and T2-weighted MR images [715]. They are appropriate for brain structural analysis and spatial normalization. Brain atlases include a gray scale anatomical average image (known as template), a set of tissue probability maps (TPMs) and in some cases a set of anatomical parcellation maps These neonatal templates and their associated tissue probability map are used as a priori reference to improve segmentation methods [1619]. However, due to the low intensity of magnetic resonance signals derived from bone, MR templates do not provide the essential accurate details of cranial bones and fontanels-sutures, which are mandatory for the diagnosis of abnormal fontanels [20, 21], a better understanding of the wide variation of normal fontanels and sutures [22] and, more specifically, for functional studies taking into account the impact of fontanels and sutures on neuronal source localization in infants by either high-density (HD) electroencephalography [2325] or HD functional optical imaging [26]. No appropriate tool is therefore available to segment and create a realistic neonatal skull model from MR images. In a previous study [19], we presented a probabilistic map of the skull based on neonatal MR images. However, this model is not sufficiently accurate and does not provide essential information about the precise location of the fontanels and sutures.

Computed tomography (CT) is the reference modality for skull evaluation, as it provides the best contrast for bone definition. Fontanels and sutures can be distinguished as discontinuities between skull bones. CT images can therefore be used to develop a more realistic model of the neonatal skull. However, because of X-ray exposure, CT acquisition is not recommended for research purposes especially in neonates, but it constitutes the reference modality for clinical assessment of the skull, allowing retrospective research on neonatal skull modeling from normal CT data acquired for clinical purposes in newborns. To the best of our knowledge, only one neonatal CT template is currently available [27], constructed using CT images from 5 subjects using only one image as reference. This template was developed in the framework of a model-based algorithm for skull and fontanel segmentation from CT images.

Since CT images lack sufficient contrast and do not provide useful information for soft tissues, a template based exclusively on CT data only provides information about cranial bones and fontanels-sutures. The CT template, providing information about hard tissues, must therefore be coregistered with the MR template, providing information about soft tissues in the same stereotaxic space. This type of bimodal structural template and associated probability maps for soft tissues and hard tissues would be suitable, after spatial normalization, for the analysis of all structures of the head allowing electrical and optical properties to then be attributed to each precisely segmented structure to facilitate direct and inverse problem in source analysis using EEG or NIRS. Moreover, integration of this type of realistic model of cranial bones based on CT images with the available MR probabilistic maps would provide more accurate results for neonatal MR skull segmentation. In other words, a bimodal MR-CT template may help to construct probabilistic map of cranial bones in the same standard stereotaxic space as introduced by MR template. By using such novel probabilistic map as a priori hard tissue information obtained from the CT template and by developing specific algorithms for processing of hard tissue information from MR images, the need for CT acquisition in neonates could be alleviated.

Although a combined MR-CT template is not available for newborns, such a template has been developed in adults stroke patients [28] and has been implemented in SPM toolbox (http://www.mricro.com/clinical-toolbox/spm8-scripts). MR and CT templates are both constructed separately to match the MNI template [29]. Since these adult templates were not directly coregistered with each other, the MR template only approximately matches the shape of the CT template and the accurate shapes of the templates are somewhat different.

Registration of CT images to the neonatal MR template is a crucial step to create a bimodal MR-CT template. This task requires inter-subject inter-modality registration, which is the most challenging aspect of registration, usually used to compare the patient’s data with anatomical atlases. Most of the available methods were developed for adults and intra-subject inter-modality category [3037]. Typically, pairwise registration is used for generating a template in which a subject image is considered as reference and other images are transferred separately to the space provided by it. This method inevitably causes some bias in the resulted template due to the image selected as reference. Furthermore, since the spatial transformation used for registering the subjects’ image to selected template is estimated independently, it’s not the method of choice in analyzing the group similarity and variation within a population [3840]. Contrastingly, groupwise registration can simultaneously estimate the transformation fields for all subjects without explicitly specifying an individual subject as reference which provides more accurate and consistent registration among the population [4145]. By using groupwise registration, the new CT template was then created in full alignment with the MR template. Various published MR templates are available for this age-group [8,1115], but only two of them [8,13] concern the whole head and contain the skull layer, which is the feature of special interest for the creation of an MR-CT bimodal template. The other MR templates are skull stripped that are not suitable for the creation of the bimodal template. Moreover, these two templates were created nonlinearly, which constitutes an advantage over linearly constructed templates by providing a higher level of detail of anatomical structures of the brain [14].

First, we aim to develop a fully aligned MR-CT bimodal template for neonates with a gestational age (GA) of 39 to 42 weeks. Then, to determine the best approach for inter-subject and inter-modal registration between CT images and MR templates, we develop a two-step intensity-based registration method by globally and locally aligning CT images to the MR template. Finally, we assess the full alignment of our CT template with the corresponding MR one and evaluate it against existing CT templates.

Materials and Methods

Subjects and Data Acquisition

In order to create the neonatal brain template, 7 participants (4 girls, 3 boys) were selected from MR images acquired as part of our previously published study [8]. Newborns were imaged with a General Electric 1.5 T MR scanner. The structural 3D volumetric T1-weighted imaging sequence was acquired with the following parameters: TR = 10.1ms, TE = 2.208ms and TI = 500ms. Each 3D volume was constructed from a sequence of slices of 512×512 pixels with 220 mm field of view resulting in a voxel size of 0.47×0.47×0.7mm3. Non-axial images were reoriented to the axial plane and all images were re-sliced to 0.47×0.47×0.47mm3 isotropic voxels. The neonatal MR template proposed in [8] was used as stereotaxic space to create the CT template.

26 newborn CT images (39–42 weeks GA) were selected from the large database acquired over recent years at Amiens University Hospital, France, many of which have been described in [46]. 16 (12 boys, 4 girls) of these were used to create the CT template and the remaining 10 (5 boys, 5 girls) were used for the cross-validation study. Images were acquired using a LightSpeed 16, GE Medical Systems, with a matrix size of 512×512 pixels and voxel size of [0.26–0.49]×[0.26–0.49]×[0.6–1.25] mm3, median 0.32×0.32×0.63 mm3. Similar to MR data, all non-axial CT images were reoriented to the axial plane and re-sliced to 0.47×0.47×0.47mm3 isotropic voxels. It is worthwhile noting that there was no overlap between the subjects used to create the CT and MR templates.

CT Template Creation

Development of a neonatal bimodal MR-CT template is comprised of two crucial steps: 1) development of an effective two-step intensity-based registration method to align neonatal CT data with the MR template, 2) construction of a neonatal CT template using an advanced groupwise registration and template building method. The selected MR template for this study is the one presented in [8]. Because the only alternative template constructed in [13] presents excessive variations, especially in extracranial parts (http://cmrm.med.jhmi.edu/). The selected template contains the whole neonatal head and matches with the MNI stereotaxic space. An overview of the proposed methodology is shown in Fig 1.

thumbnail
Fig 1. Flow diagram of the proposed pipeline to construct the CT template.

https://doi.org/10.1371/journal.pone.0166112.g001

Data preprocessing.

In order to obtain accurate registration results, CT images were manually rotated and translated in order to be almost aligned with the ‘GRAMFC_T39-42’ neonatal MR brain template [8] using 3D Slicer software [47]. During this rigid transformation by 3D Slicer, some voxels with zero intensity may appear in the corners of the image. As typical CT scans were calibrated using Hounsfield units (approximate intensities are: bone: +1000, water: ~0, CSF: ~15, WM: ~20–30, GM: ~37–45, and air: ~-1000), these newly appeared voxels which correspond to background (air voxels) could be wrongly considered as water (CSF) due to their zero intensity. In addition to the aforementioned problem, there are noisy patterns around the head (especially posteriorly due to the pillow and mattress), inappropriate objects such as a pacifier, anesthesia mask or a nasogastric or orogastric feeding tube, and an artificial rim around the air with an intensity of -3000 Hounsfield units (HU) in CT images. These irrelevant structures must be eliminated from CT scans due to their unwanted effect on registration. For this purpose, the head mask of each CT image was extracted using histogram thresholding by Otsu’s method [48] and morphology operators such as opening, closing, filling and identifying connected components. All voxels outside the head mask of each image were then considered to be air and were assigned an intensity value of -1000 HU.

CT and MR images are very different in terms of intensity range. CT images typically have an intensity range from -1024 to 2048 and MR images have an intensity range from 0 to 1023 (255 for 8-bit MR images). Ideally, for registration purposes, the intensity value of CT image voxels should be mapped so that the intensity range is similar to the MR intensity range. In particular, it is important to map negative CT values to the non-negative range and expand the dynamic range of soft tissue voxels inside the cranial bones (as in MR images) [28, 36, 4950]. This type of mapping would allow more accurate differentiation of intracranial soft tissues that present low contrast on raw brain CT images. According to [28], CT image intensities are transformed as follows: values -1000 to -100 are transformed to 0–900, values from -99 to 100 are linearly scaled to the range of 901 to 3100 and intensities higher than 100 (I > 100) are converted to I + 3000 (Fig 2). This CT intensity transformation has a significant impact on CT-MR multimodal registration.

thumbnail
Fig 2. Original and preprocessed CT images and related histogram representation.

https://doi.org/10.1371/journal.pone.0166112.g002

The first row shows the original CT image and corresponding histogram. The second row presents the preprocessed CT image in intensity-transformed scale and its related histogram.

Multimodal registration.

Registration of CT neonatal head images to the MR neonatal template is a challenging problem not only due its inter-modality nature, but also due to specificities related to neonatal head anatomy and imaging issues. In neonatal head CT/MR images, extracranial tissues are highly variable due to the very fine soft tissues of the scalp, skull discontinuities and fontanels-sutures. In contrast, intracranial soft tissues are physically more constrained and consequently less variable in terms of morphology and appearance. The above considerations led us to choose a two-step intensity-based approach for registration of CT images to the MR neonatal template, as illustrated in the multimodal registration part (dashed box) of Fig 1. At the first step of multimodal registration, the use of intracranial image intensity is expected to provide better results than the use of intensity information from the whole head. This strategy eliminates the effect of extracranial variations in multimodal registration. In the second step, the extracranial parts of the head are used for affine registration between the whole CT image and the MR template. Due to the unavailability of the GRAMFC_T39-42 MR intracranial template, this template had to be constructed in the framework of this research. All required steps of co-registration and segmentation, as shown in Fig 1, will be described with more details in the following subsections.

MR Intracranial Template Creation: The same images and the same GRAMFC_T39-42 template building procedure described in [8] were used to construct the MR intracranial template. After aligning all MR images to the same stereotaxic space, brain and CSF were extracted from each MR image by the method proposed by Ashburner and Friston [51] using SPM8 software in conjunction with neonatal brain and CSF probabilistic models [19]. The extracted tissues were then binarized by histogram thresholding using the method described by Otsu [48] and were then added and integrated as a brain-CSF mask by applying various morphology operators such as “or”, “closing” and “filling” functions. Intracranial can be extracted by detecting outer surface of dura mater [52, 53]. As no automatic tools for extracting dura mater from neonatal T1 images are available, an expert radiologist performed manual correction on the extracted brain and CSF to add dura mater. After manual correction and revision of the surfaces by the expert and after obtaining the intracranial mask in all images, the MR intracranial images were obtained, averaged and smoothed with a 2 mm Full Width at Half Maximum (FWHM) Gaussian kernel as used in [8, 10] to construct the MR intracranial template. Fig 3 shows the GRAMFC_T39-42 template and our proposed intracranial template.

thumbnail
Fig 3. GRAMFC_T39-42 template and the proposed intracranial template.

(a) GRAMFC_T39-42 template [8], (b) Proposed MR intracranial template.

https://doi.org/10.1371/journal.pone.0166112.g003

Extraction of Intracranial CT Data: The coupled level set approach for skull segmentation and fontanel reconstruction in neonatal CT scans was used to extract intracranial CT data [46]. This approach uses hard tissue contrast in CT images, prior information concerning head shape integrated in level set initialization, and a predefined constraint to impose surface reconstruction properties. It applies a pair of interior/exterior surfaces as geodesic active regions propagating towards and interacting with each other. Level set evolution is stopped when they touch each other or encounter the outer (convex) and inner (concave) surfaces of cranial bones using edge information. In the coupled level set method, each moving surface (e.g. interior surface) is influenced not only by its own internal and external forces, but also by a mutual term which is controlled by another (exterior) moving surface. In locations corresponding to fontanels-sutures, these moving surfaces touch each other without crossing over. The resulting contour (inner surface of cranial bones and fontanels-sutures) is assumed to be the intracranial outer surface (Fig 4B). After expert revision of the surface, a mask of the surface was created by the filling operator (Fig 4C) and was used to extract the intracranial CT image (Fig 4D).

thumbnail
Fig 4. CT intracranial extraction using coupled level sets.

(a) The initial coupled level sets. (b) The extracted inner and outer surfaces of cranial bones and fontanels-sutures using the method described in [46]. (c) Corresponding mask of the resulting interior contour. (c) The extracted intracranial CT image obtained by multiplying the mask to the preprocessed CT image. The interior and exterior contours are represented by the red and green colour, respectively.

https://doi.org/10.1371/journal.pone.0166112.g004

CT-MR Intracranial Co-registration: Intracranial registration consists of two steps. The first step consists of a 12-parameter affine transformation that is used for translation and linear deformation (rotation, scaling and shearing) of the input images to the stereotaxic space as defined by the constructed MR intracranial template. In the second step, a nonlinear deformation with symmetric normalization transformation (SyN) [54] is used to regionally match the subject to the template. The SyN method was shown to have the best performance among 14 different nonlinear registration methods in a comparative study by Klein [55]. It uses symmetric parameterization for geodesic connection of two neuroimages (I and J) in the diffeomorphic space and is invertible in the discrete domain [54]. Regardless of whether it is fixed or moving, both images are deformed in time along the geodesic path. Fig 5 presents the moving intracranial CT image and the GRAMFC_T39-42 intracranial space as a fixed image. The registered intracranial CT data shown in Fig 5C was achieved by applying both affine and non-linear SyN transformations.

thumbnail
Fig 5. CT-MR intracranial co-registration, first step: Intracranial intensity-based registration.

(a) The input moving CT intracranial space. (b) The reference MR intracranial space. (c) The CT intracranial registered to GRAMFC_T39-42 intracranial space using affine and non-linear SyN approach [54].

https://doi.org/10.1371/journal.pone.0166112.g005

CT'-MR Co-registration: After registering the CT intracranial to the MR intracranial template, the resulting deformation field, which is similar to the elastic model, was applied to the preprocessed CT data to warp the image based on intracranial information and to provide CT' image (Fig 6A). The CT' image was then registered to the MR template by affine transformation using mutual information (MI) similarity measure (Fig 6B). All transformations were applied using the open source toolkit Advanced Normalization Tools (ANTs) [56] built on an Insight ToolKit (ITK) framework [57]. The outputs of this stage (CT'') are CT images that are almost fully aligned with each other and with the MR template. These images are the input of the last part of the proposed method to create the CT template.

thumbnail
Fig 6. CT'-MR co-registration, second step: whole head intensity-based registration.

(a) CT' image obtained by applying the resulting deformation field from CT-MR intracranial co-registration sub-section to the preprocessed CT image. (b) CT'' image achieved by linearly registering CT' to GRAMFC_T39-42.

https://doi.org/10.1371/journal.pone.0166112.g006

Template building.

The major challenge when constructing a template is to eliminate the bias of the constructed template with respect to the specific subject(s). Groupwise registration is one of the powerful methods used for template building. This method not only incorporates all image information in the registration process (by simultaneously registering all images to the mean of the population), but also eliminates the bias related to a specific reference image in data analysis and computational anatomy. The groupwise registration method proposed by Avants et al. [58] has been integrated into the template construction function by ANTs software. It provides an optimal unbiased template with respect to the input data by using a symmetric diffeomorphic approach and by proposing an unbiased optimization strategy that accounts for appearance and shape variation. This symmetric groupwise normalization (SyGN) method applies SyN energy terms to measure image similarity and minimize diffeomorphism lengths. SyGN has been shown to be an effective tool for template building [58].

To construct the template, a script named “antsMultivariateTemplateConstruction.sh” using cross-correlations [59] as a similarity measure was run on the CT'' images. The number of iterations was set to 4 by default, however as the images were in good alignment with each other, setting the iteration limit to 2 was sufficient for the convergence and creation of a reliable template. Our experiments with different number of iterations from 2 to 5 revealed that the created templates are similar in terms of mean and standard deviation of the MI similarity index. To maintain the connection between CT data and the MR template, the MR template was chosen as the initial template guess. Firstly, the diffeomorphisms between the fixed MR template and individual input images are optimized by minimizing the computed total length between the two. The template appearance with fixed shape and the mappings resulting from the previous step are then optimized by maximizing the similarity of intensity pattern between the template and the set of currently deformed CT images using cross-correlation measure. Finally, by computing a small diffeomorphism that maps the image toward the average shape, the template shape is updated. The above procedure is then reiterated until convergence.

Validation.

Mutual information (MI) is one of the best similarity measures to compare co-registered inter-modality images. It was used to evaluate the similarity between MR and CT templates. The mutual information I(X, Y) between image X (MR template) and image Y (CT template) is given by [60]: (1) where p(x,y) is the joint probability distribution of the two MR and CT templates, and p(x) and p(y) are the marginal probability distribution functions of MR and CT templates, respectively. As derived from (5), the MI value is zero when the images share nothing in common and is equal to the entropy of one image when the images are completely registered.

To quantitatively evaluate the constructed neonatal CT template, CT test images were registered to the template in order to determine the similarity between registered images and the template. The affine registration using cross-correlation (CC) measure was initially applied to globally align the test images to the template. The nonlinear SyN registration method and CC measure were then used to locally map the test images to the template. Since CC was used for intra-modality registration to normalize the test images to the CT template, the registered images were evaluated using MI with the proposed CT template, which was smoothed by a Gaussian kernel with 2 mm FWHM as used in [8,10] for fetuses and newborns. For evaluation, two different scenarios were considered. In the first scenario, the leave-one-out cross-validation was applied for 16 images used during template creation and in the second scenario, cross-validation was performed with 10 test CT images.

Experimental Results and Discussion

Bimodal MR-CT Head Template Evaluation

Fig 7 shows the constructed bimodal MR-CT template for neonates with 39–42 weeks GA. The new GRAMFC_CT39-42 template in Hounsfield units and the intensity rescaled using intensity transformation, as described in the preprocessing subsection, is shown in the first two rows of Fig 7. Multi-slice visualization of the bimodal template qualitatively demonstrates the accuracy of the proposed method and the full alignment of CT and MR templates.

thumbnail
Fig 7. Multi-slice visualization of the proposed neonatal MR-CT head template.

The first and second row show some axial slices in Hounsfield units and the intensity-transformed scale of GRAMFC_CT39-42, respectively, and the third row shows the corresponding slices from the GRAMFC_T39-42 MR template.

https://doi.org/10.1371/journal.pone.0166112.g007

The MI value obtained for the GRAMFC_CT39-42 template in Hounsfield units is 1.02 and for the intensity-transformed CT template is 1.17. The resulting value for intensity-transformed CT is greater than the Hounsfield unit CT atlas, which shows that intensity changes of the CT template with an invertible transformation, increases the similarity of these MR and CT templates. Compared to the values of MR template entropy (1.94) and CT template entropy (1.59), the value of 1.17 indicates a good alignment of the two templates.

The similarity of the MR-CT template proposed by Rorden et al. [28] for older adults was computed in order to provide a clearer interpretation of the MI value. The MI value for their high-resolution bimodal MR-CT template was 0.84, when the CT template is expressed in Hounsfield units and 0.88 for the intensity-transformed CT template, which demonstrates that our MR and CT templates are more similar to each other than the template proposed by Rorden et al. and confirms the advantages of the proposed method to create a bimodal neonatal brain template.

CT Template Evaluation

The CT template was presented in both the intensity-transformed scale and Hounsfield units (Fig 7). To assess the performance of the CT template with the two different intensities and to determine which representation provides better results, all evaluations of image normalization will hereafter be applied to the proposed CT template with both scaling modalities. Table 1 shows the MI values for the two scenarios and for the proposed CT template in Hounsfield units and the intensity-transformed scales. The second and third columns of Table 1 clearly illustrate the out-performance of the proposed CT template using the intensity-transformed scale.

thumbnail
Table 1. Mutual information (Mean±SD) between normalized images and neonatal CT templates.

https://doi.org/10.1371/journal.pone.0166112.t001

To compare our template with the only available CT template for newborns proposed by Jafarian et al. [27], 10 test images were first registered to their template using the same registration procedure and parameters as described above. The similarity of the normalized images with the template was then evaluated by MI similarity measure. The average MI values for cross-validation using 10 test images is reported in Table 1. Comparison of the MI values in the first and second columns of Table 1 demonstrates the superiority of the proposed method used for template creation and its capacity to be used as a template. To study the performance of the templates in normalized images, the similarity between normalized images was also examined. For this purpose, forty five different pairs were created from ten test images and each image of each pair was registered to the proposed CT template (GRAMFC_CT39-42) and to the CT template proposed by Jafarian et al. (CTJ) using the same registration parameters. By assuming that Ii and Ij are a pair of images (Ii,Ij), following registration of these images to the GRAMFC_CT39-42 and CTJ templates to obtain and , respectively, cross-correlation was then used to measure the similarity between the two data sets, i.e. and . The mean and standard deviation of similarity measures of all pairs for each CT template are reported in Table 2, which shows a higher level of similarity between normalized images using our CT template.

thumbnail
Table 2. Mean and standard deviation of cross-correlation between 45 image pairs after registration to GRAMFC_CT39-42 and CTJ.

https://doi.org/10.1371/journal.pone.0166112.t002

Fig 8 qualitatively shows the difference between GRAMFC_CT39-42 and CTJ. The template proposed here provides greater anatomic definition compared to CTJ.

thumbnail
Fig 8. Qualitative comparison of the GRAMFC_CT39-42 template (left) and the template proposed by Jafarian et al. [27] (right) for neonates with 39–42 weeks GA.

https://doi.org/10.1371/journal.pone.0166112.g008

Our bimodal MR-CT neonatal head template was constructed based on the only available and eligible T1 weighted (T1w) MR template which can be considered as a limitation. T1 weighting is mostly used during the preterm and perinatal periods, but the contrast becomes poorer with age until it recovers during the second post-natal year. T2 weighting transitorily enables a better contrast between term and 4–6 months post-term [61]. Another issue is the differing voxel sizes and non-isotropic voxels in the images we used for creating the template. Obviously, high resolution CT images with the same and isotropic voxel sizes are preferred for being used in template construction. However, these images are rarely available in practice due to the difficulties in imaging neonates and high risk imposed by X-ray exposure. Finally, our methodology applied some interpolation/smoothing operations to the images before and during the registration procedures. Although these operations are almost mandatory, further considerations are required to optimize the procedure to improve the accuracy of the created template.

Conclusion

We present an approach for construction of a bimodal MR-CT head template for newborns with 39–42 weeks GA. Regardless of the neonatal MR template chosen, a two-step intensity-based method was proposed for inter-subject inter-modality registration and a groupwise registration approach was applied to create a CT template. The constructed CT template is the first neonatal template in a standard stereotaxic space similar to MNI which could be used for spatial normalization in CT modality. The CT template was built using intensity-transformed CT data and presented as an intensity-transformed CT template. All evaluations showed that the intensity-transformed CT template is more reliable than the Hounsfield unit template for bimodal MR-CT template and image normalization using popular open source methods. It is noteworthy that the proposed method is not dependent on a specific MR template and can be used to create a CT template in full alignment with any MR template satisfying the previously defined eligibility criteria.

Future directions of this work include: 1) constructing a bimodal MR-CT head atlas consisting of probabilistic maps of cranial bones, fontanels and the scalp in order to enhance the skull and scalp segmentation in MR images, We are developing the methodology of such application and the primary results are being evaluated, 2) application of the created bimodal atlas to neuronal source localization in infants by either high-density electroencephalography or functional optical imaging to demonstrate its superiority versus its monomodal counterparts in terms of localization accuracy.

Ethical Statement

Institutional Review Board approval was obtained for this study. Ethical permission for this study was provided by the local ethics committee (Commission d’Evaluation Ethique de Recherches Non Interventionelles, CEERNI), CHU Amiens, approval No. 66, 2011. Written informed consent was obtained from each subject’s caregivers.

Author Contributions

  1. Conceptualization: FW HAM RG.
  2. Methodology: SG HAM MM.
  3. Project administration: FW HAM.
  4. Supervision: HAM RG.
  5. Validation: SG HAM MG MM.
  6. Visualization: SG MM.
  7. Writing – original draft: SG.
  8. Writing – review & editing: HAM FW.

References

  1. 1. Hüppi PS, Warfield S, Kikinis R, Barnes PD, Zientara GP, Jolesz FA, Tsuji MK, Volpe JJ. Quantitative magnetic resonance imaging of brain development in premature and mature newborns. Annals of neurology. 1998 Feb 1;43(2):224–35. pmid:9485064
  2. 2. Thompson PM, Giedd JN, Woods RP, MacDonald D, Evans AC, Toga AW. Growth patterns in the developing brain detected by using continuum mechanical tensor maps. Nature. 2000 Mar 9;404(6774):190–3. pmid:10724172
  3. 3. Rutherford MA. MRI of the Neonatal Brain. WB Saunders Co; 2002.
  4. 4. Evans AC, Brain Development Cooperative Group. The NIH MRI study of normal brain development. Neuroimage. 2006 Mar 31;30(1):184–202. pmid:16376577
  5. 5. Boardman JP, Counsell SJ, Rueckert D, Hajnal JV, Bhatia KK, Srinivasan L, et al. Early growth in brain volume is preserved in the majority of preterm infants. Annals of neurology. 2007 Aug 1;62(2):185–92. pmid:17696128
  6. 6. Kikinis R, Shenton ME, Iosifescu DV, McCarley RW, Saiviroonporn P, Hokama HH, et al. A digital brain atlas for surgical planning, model-driven segmentation, and teaching. Visualization and Computer Graphics, IEEE Transactions on. 1996 Sep;2(3):232–41.
  7. 7. Aljabar P, Bhatia KK, Murgasova M, Hajnal JV, Boardman JP, Srinivasan L, et al. Assessment of brain growth in early childhood using deformation-based morphometry. Neuroimage. 2008 Jan 1;39(1):348–58. pmid:17919930
  8. 8. Kazemi K, Moghaddam HA, Grebe R, Gondry-Jouet C, Wallois F. A neonatal atlas template for spatial normalization of whole-brain magnetic resonance images of newborns: preliminary results. Neuroimage. 2007 Aug 15;37(2):463–73. pmid:17560795
  9. 9. Altaye M, Holland SK, Wilke M, Gaser C. Infant brain probability templates for MRI segmentation and normalization. Neuroimage. 2008 Dec 31;43(4):721–30. pmid:18761410
  10. 10. Habas PA, Kim K, Corbett-Detig JM, Rousseau F, Glenn OA, Barkovich AJ, Studholme C. A spatiotemporal atlas of MR intensity, tissue probability and shape of the fetal brain with application to segmentation. Neuroimage. 2010 Nov 1;53(2):460–70. pmid:20600970
  11. 11. Kuklisova-Murgasova M, Aljabar P, Srinivasan L, Counsell SJ, Doria V, Serag A, et al. A dynamic 4D probabilistic atlas of the developing brain. NeuroImage. 2011 Feb 14;54(4):2750–63. pmid:20969966
  12. 12. Shi F, Yap PT, Wu G, Jia H, Gilmore JH, Lin W, Shen D. Infant brain atlases from neonates to 1-and 2-year-olds. PloS one. 2011 Apr 14;6(4):e18746. pmid:21533194
  13. 13. Oishi K, Mori S, Donohue PK, Ernst T, Anderson L, Buchthal S, et al. Multi-contrast human neonatal brain atlas: application to normal neonate development analysis. Neuroimage. 2011 May 1;56(1):8–20. pmid:21276861
  14. 14. Serag A, Aljabar P, Ball G, Counsell SJ, Boardman JP, Rutherford MA, et al. Construction of a consistent high-definition spatio-temporal atlas of the developing brain using adaptive kernel regression. Neuroimage. 2012 Feb 1;59(3):2255–65. pmid:21985910
  15. 15. Momeni M, Abrishami Moghaddam H, Grebe R, Gondry-Jouet C, Wallois F. Neonatal Atlas Templates for the Study of Brain Development Using Magnetic Resonance Images. Current Medical Imaging Reviews. 2015 Feb 1;11(1):38–48.
  16. 16. Prastawa M, Gilmore JH, Lin W, Gerig G. Automatic segmentation of MR images of the developing newborn brain. Medical image analysis. 2005 Oct 31;9(5):457–66. pmid:16019252
  17. 17. Shi F, Fan Y, Tang S, Gilmore JH, Lin W, Shen D. Neonatal brain image segmentation in longitudinal MRI studies. Neuroimage. 2010 Jan 1;49(1):391–400. pmid:19660558
  18. 18. Shi F, Yap PT, Fan Y, Gilmore JH, Lin W, Shen D. Construction of multi-region-multi-reference atlases for neonatal brain MRI segmentation. Neuroimage. 2010 Jun 30;51(2):684–93. pmid:20171290
  19. 19. Kazemi K, Ghadimi S, Moghaddam HA, Grebe R, Gondry-Jouet C, Wallois F. Neonatal probabilistic models for brain, CSF and skull using T1-MRI data: preliminary results. In Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE 2008 Aug 20: 3892–3895.
  20. 20. Kiesler J, Ricer R. The abnormal fontanel. American family physician. 2003 Jun;67(12):2547–52. pmid:12825844
  21. 21. Cunningham ML, Heike CL. Evaluation of the infant with an abnormal skull shape. Current opinion in pediatrics. 2007 Dec 1;19(6):645–51. pmid:18025930
  22. 22. Gallagher ER, Hing AV, Cunningham ML. Evaluating fontanels in the newborn skull. Contemporary Pediatrics. 2013 30(11), 12.
  23. 23. Lew S, Sliva DD, Choe MS, Grant PE, Okada Y, Wolters CH, et al. Effects of sutures and fontanels on MEG and EEG source analysis in a realistic infant head model. Neuroimage. 2013 Aug 1;76:282–93. pmid:23531680
  24. 24. Roche-Labarbe N, Aarabi A, Kongolo G, Gondry-Jouet C, Dümpelmann M, Grebe R, et al. High-resolution electroencephalography and source localization in neonates. Human brain mapping. 2008 Feb 1;29(2):167–76. pmid:17390314
  25. 25. Gargiulo P, Belfiore P, Friðgeirsson EA, Vanhatalo S, Ramon C. The effect of fontanel on scalp EEG potentials in the neonate. Clinical Neurophysiology. 2015 Sep 30;126(9):1703–10. pmid:25553851
  26. 26. Dehaes M, Kazemi K, Pélégrini-Issac M, Grebe R, Benali H, Wallois F. Quantitative effect of the neonatal fontanel on synthetic near infrared spectroscopy measurements. Human brain mapping. 2013 Apr 1;34(4):878–89. pmid:22109808
  27. 27. Jafarian N, Kazemi K, Moghaddam HA, Grebe R, Fournier M, Helfroush MS, et al. Automatic segmentation of newborns’ skull and fontanel from CT data using model-based variational level set. Signal, Image and Video Processing. 2014 Feb 1;8(2):377–87.
  28. 28. Rorden C, Bonilha L, Fridriksson J, Bender B, Karnath HO. Age-specific CT and MRI templates for spatial normalization. Neuroimage. 2012 Jul 16;61(4):957–65. pmid:22440645
  29. 29. Collins DL, Neelin P, Peters TM, Evans AC. Automatic 3D intersubject registration of MR volumetric data in standardized Talairach space. Journal of computer assisted tomography. 1994 Mar 1;18(2):192–205. pmid:8126267
  30. 30. Hsu LY, Loew MH. Fully automatic 3D feature-based registration of multi-modality medical images. Image and Vision Computing. 2001 Jan 31;19(1):75–85.
  31. 31. Hartkens T, Hill D, Castellano-Smith A, Hawkes D, Maurer C, Martin A, et al. Using points and surfaces to improve voxel-based non-rigid registration. Medical Image Computing and Computer-Assisted Intervention—MICCAI 2002. 2002:565–72.
  32. 32. Law W, Feng DD, Wong KP, Wang X. An automated elastic image registration method for CT and MR brain images. In Intelligent Multimedia, Video and Speech Processing, IEEE 2004. Proceedings of 2004 International Symposium on 2004 Oct 20: 350–353.
  33. 33. Xie Z, Farin GE. Image registration using hierarchical B-splines. Visualization and Computer Graphics, IEEE Transactions on. 2004 Jan;10(1):85–94.
  34. 34. Loeckx D, Slagmolen P, Maes F, Vandermeulen D, Suetens P. Nonrigid image registration using conditional mutual information. Medical Imaging, IEEE Transactions on. 2010 Jan;29(1):19–29.
  35. 35. De Nigris D, Collins DL, Arbel T. Multi-modal image registration based on gradient orientations of minimal uncertainty. Medical Imaging, IEEE Transactions on. 2012 Dec;31(12):2343–54.
  36. 36. Kuijf HJ, Biesbroek JM, Viergever MA, Biessels GJ, Vincken KL. Registration of brain CT images to an MRI template for the purpose of lesion-symptom mapping. In Multimodal Brain Image Analysis 2013 Sep 22 (pp. 119–128). Springer International Publishing.
  37. 37. Heinrich MP, Jenkinson M, Bhushan M, Matin T, Gleeson FV, Brady M, Schnabel JA. MIND: Modality independent neighbourhood descriptor for multi-modal deformable registration. Medical Image Analysis. 2012 Oct 31;16(7):1423–35. pmid:22722056
  38. 38. Christensen GE, Johnson HJ. Consistent image registration. IEEE transactions on medical imaging. 2001 Jul;20(7):568–82. pmid:11465464
  39. 39. Shen D, Davatzikos C. HAMMER: hierarchical attribute matching mechanism for elastic registration. IEEE transactions on medical imaging. 2002 Nov;21(11):1421–39. pmid:12575879
  40. 40. Vercauteren T, Pennec X, Perchant A, Ayache N. Diffeomorphic demons: Efficient non-parametric image registration. NeuroImage. 2009 Mar 31;45(1):S61–72.
  41. 41. Holden M, Hill DL, Denton ER, Jarosz JM, Cox TC, Rohlfing T, Goodey J, Hawkes DJ. Voxel similarity measures for 3-D serial MR brain image registration. IEEE transactions on medical imaging. 2000 Feb;19(2):94–102. pmid:10784281
  42. 42. Sabuncu MR, Balci SK, Shenton ME, Golland P. Image-driven population analysis through mixture modeling. IEEE transactions on medical imaging. 2009 Sep;28(9):1473–87. pmid:19336293
  43. 43. Jia H, Wu G, Wang Q, Shen D. ABSORB: Atlas building by self-organized registration and bundling. NeuroImage. 2010 Jul 1;51(3):1057–70. pmid:20226255
  44. 44. Wang Q, Chen L, Shen D. Group-wise registration of large image dataset by hierarchical clustering and alignment. InSPIE Medical Imaging 2009 Feb 26 (pp. 72590N-72590N). International Society for Optics and Photonics.
  45. 45. Joshi S, Davis B, Jomier M, Gerig G. Unbiased diffeomorphic atlas construction for computational anatomy. NeuroImage. 2004 Dec 31;23:S151–60. pmid:15501084
  46. 46. Ghadimi S, Abrishami-Moghaddam H, Grebe R, Wallois F. Skull segmentation and reconstruction from newborn CT images using coupled level sets. IEEE J Biomed Health Inform. 20(2), 563–573. pmid:25667361
  47. 47. Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin JC, Pujol S, et al. 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magnetic resonance imaging. 2012 Nov 30;30(9):1323–41. pmid:22770690
  48. 48. Otsu N. A threshold selection method from gray-level histograms. Automatica. 1975 Jun;11(285–296):23–7.
  49. 49. Van Den Elsen PA, Pol EJ, Sumanaweera TS, Hemler PF, Napel S, Adler JR. Grey value correlation techniques used for automatic matching of CT and MR brain and spine images. In Visualization in Biomedical Computing 1994 1994 Sep 9 (pp. 227–237). International Society for Optics and Photonics.
  50. 50. Meyer, J., 2007. Histogram transformation for inter-modality image registration. Conf Proc IEEE Bioinformatics and Bioengineering, 2007. 1118–1123.
  51. 51. Ashburner J, Friston KJ. Unified segmentation. Neuroimage. 2005 Jul 1;26(3):839–51. pmid:15955494
  52. 52. Yasuda M, Mori E, Kitagaki H, Yamashita H, Hirono N, Shimada K, et al. Apolipoprotein E epsilon 4 allele and whole brain atrophy in late-onset Alzheimer's disease. Am J Psychiatry, 1998; 155(6), 779–84. pmid:9619150
  53. 53. Whitwell JL, Crum WR, Watt HC, Fox NC. Normalization of cerebral volumes by use of intracranial volume: implications for longitudinal quantitative MR imaging. American Journal of Neuroradiology. 2001 Sep 1;22(8):1483–9. pmid:11559495
  54. 54. Avants BB, Epstein CL, Grossman M, Gee JC. Symmetric diffeomorphic image registration with cross-correlation: evaluating automated labeling of elderly and neurodegenerative brain. Medical image analysis. 2008 Feb 29;12(1):26–41. pmid:17659998
  55. 55. Klein A, Andersson J, Ardekani BA, Ashburner J, Avants B, Chiang MC, et al. Evaluation of 14 nonlinear deformation algorithms applied to human brain MRI registration. Neuroimage. 2009 Jul 1;46(3):786–802. pmid:19195496
  56. 56. Avants BB, Tustison NJ, Song G, Cook PA, Klein A, Gee JC. A reproducible evaluation of ANTs similarity metric performance in brain image registration. Neuroimage. 2011 Feb 1;54(3):2033–44. pmid:20851191
  57. 57. Avants BB, Tustison NJ, Stauffer M, Song G, Wu B, Gee JC. The Insight ToolKit image registration framework. Frontiers in neuroinformatics. 2014 Apr 28;8:44. pmid:24817849
  58. 58. Avants BB, Yushkevich P, Pluta J, Minkoff D, Korczykowski M, Detre J, Gee JC. The optimal template effect in hippocampus studies of diseased populations. Neuroimage. 2010 Feb 1;49(3):2457–66. pmid:19818860
  59. 59. Lemieux L, Wieshmann UC, Moran NF, Fish DR, Shorvon SD. The detection and significance of subtle changes in mixed-signal brain lesions by serial MRI scan matching and spatial normalization. Medical Image Analysis. 1998 Sep 30;2(3):227–42. pmid:9873901
  60. 60. Shannon CE. A mathematical theory of communication. ACM SIGMOBILE Mobile Computing and Communications Review. 2001 Jan 1;5(1):3–55.
  61. 61. Leroy F, Mangin JF, Rousseau F, Glasel H, Hertz-Pannier L, Dubois J, Dehaene-Lambertz G. Atlas-free surface reconstruction of the cortical grey-white interface in infants. PloS one. 2011 Nov 16;6(11):e27128. pmid:22110604