SUV-quantification of physiological lung tissue in an integrated PET/MR-system: Impact of lung density and bone tissue

  • Ferdinand Seith,

    Affiliation Diagnostic and Interventional Radiology, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

  • Holger Schmidt ,

    holger.schmidt@uni-tuebingen.de

    Affiliation Diagnostic and Interventional Radiology, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

  • Sergios Gatidis,

    Affiliation Diagnostic and Interventional Radiology, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

  • Ilja Bezrukov,

    Affiliations Preclinical Imaging and Radiopharmacy, Laboratory for Preclinical Imaging and Imaging Technology of the Werner Siemens Foundation, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany, Max Planck Institute for Intelligent Systems, Spemannstr. 38, Tuebingen, Germany

  • Christina Schraml,

    Affiliation Diagnostic and Interventional Radiology, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

  • Christina Pfannenberg,

    Affiliation Diagnostic and Interventional Radiology, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

  • Christian la Fougère,

    Affiliation Nuclear Medicine, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

  • Konstantin Nikolaou,

    Affiliation Diagnostic and Interventional Radiology, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

  • Nina Schwenzer

    Affiliation Diagnostic and Interventional Radiology, Department of Radiology, Eberhard Karls University, Hoppe-Seyler-Straße 3, Tuebingen, Germany

Abstract

Purpose

The aim of the study was to investigate the influence of lung density changes as well as bone proximity on the attenuation correction of lung standardized uptake values (SUVs).

Methods and materials

15 patients with mostly oncologic diseases were examined in 18F-FDG-PET/CT and subsequently in a fully integrated PET/MR scanner. From each PET dataset acquired in PET/MR, four different PET reconstructions were computed using different attenuation maps (μ-maps): i) CT-based μ-map (gold standard); ii) CT-based μ-map in which the linear attenuation coefficients (LAC) of the lung tissue was replaced by the lung LAC from the MR-based segmentation method; iii) based on reconstruction ii), the LAC of bone structures was additionally replaced with the LAC from the MR-based segmentation method; iv) the vendor-provided MR-based μ-map (segmentation-based method). Those steps were performed using MATLAB. CT Hounsfield units (HU) and SUVmean was acquired in different levels and regions of the lung. Relative differences between the differently corrected PETs were computed.

Results

Compared to the gold standard, reconstruction ii), iii) and iv) led to a relative underestimation of SUV in the posterior regions of -9.0%, -13.4% and -14.0%, respectively. Anterior and middle regions were less affected with an overestimation of about 6–8% in reconstructions ii)–iv).

Conclusion

It could be shown that both, differences in lung density and the vicinity of bone tissue in the μ-map may have an influence on SUV, mostly affecting the posterior lung regions.

Introduction

In positron emission tomography (PET), the correction of the attenuation (attenuation correction, AC) of annihilated photons on their way through the patient's body plays a key role in quantifying the local tracer-uptake, expressed as standardized uptake value (SUV) [1]. In positron emission tomography/computed tomography (PET/CT), a direct transformation of Hounsfield Units (HU) into the attenuation of 511keV-photons is possible, leading to precise linear attenuation coefficients (LACs) of the tissue which is used to compute an attenuation map (μ-map). Since the signal intensity in magnetic resonance (MR) imaging cannot be directly related to photon attenuation, an analogous procedure is not possible in integrated PET/MR-systems. To date, several approaches have been proposed to overcome this drawback. The most commonly used one is the segmentation-method [2] as it is used e.g. in the Siemens Biograph mMR. In principle, it separates the body into 3–4 different tissue types based on a T1-weighted MR sequence (air, lung, soft tissue and fat) and corresponding LACs are assigned. As the segmentation-method is based on standard MR sequences, bone is routinely not recognized due to low signal intensity of osseous tissue and therefore not considered in the μ-map. However, bone causes the highest attenuation of photons in the human body which can lead to significant underestimation of SUV in regions with a high amount of bone tissue [3]. Another limitation lies in the fact that the segmentation-method relies on the attribution of four predefined tissue classes to all body parts, however, only one single LAC is attributed to the whole lung. This LAC for lung is patient-independent and does not account for regional lung density changes. Thus, both inter-subject as well as regional differences in lung tissue may not be considered for AC using the segmentation-method in the Biograph mMR.

PET/MRI can be used as an alternative to PET/CT in staging oncologic patients [4], providing lower radiation exposure for patients and functional imaging techniques such as diffusion weighted imaging. It is known that—compared to CT—MR is inferior in detecting small lung nodules [5] and therefore, the PET information in PET/MR might even be more important for the detection and characterization of pulmonary nodules than it is in PET/CT [6]. Previously conducted studies could show that the segmentation-method might not be sufficient for reliable LAC of lung tissue [715]. Keereman et al. [9, 10] showed in simulations, that misclassifications in lung and cortical bone tissue can have a significant impact on PET quantification. However, the potential clinical impact of these effects has not yet been evaluated in systematic studies.

Therefore, the aim of this study was to assess the accuracy of the regional SUV quantification based on the segmentation-method compared to the CT-based attenuation correction, with focus on the impact of regional lung density changes as well as the anatomical vicinity of bone.

Material and methods

Patient cohort

In a prospective study conducted between March 2011 and October 2013, 200 patients underwent a clinically indicated PET/CT and subsequent PET/MR examination in a fully integrated PET/MR system. The study was designed as a feasibility study to compare PET/MR with PET/CT including dedicated examinations of the brain, the upper and lower abdomen as well as whole body examinations. The study was approved by the German Federal Institute for Drugs and Medical Devices as well as the local ethics committee. Written informed consent was obtained from all patients concerning both examination and the scientific evaluation of their data. The first 60 patients had to be excluded due to a vendor-provided software update that could influence PET quantification. Furthermore, 19 examinations had to be excluded due to technical problems such as strong artifacts or registration problems. From the remaining 121 cases, 21 PET/MR examinations had a diagnostic lung imaging protocol from whom 15 were investigated with 18F-FDG and were thus included for evaluation (8 female, mean age 56.7 ± 13.7 years). Lung regions with consolidations or malignant lesions (e.g. infiltrates, dystelectasis, atelectasis, metastases, pleural carcinomatosis) were excluded. Most of the patients had known oncologic diseases: oesophageal cancer (n = 2), head-and-neck-cancer (n = 1), melanoma (n = 1), lymphoma (n = 1), ovarian cancer (n = 2), pulmonary cancer (n = 2), breast cancer (n = 2), thyroid cancer (n = 1). The other participants were suffering from fever of unknown origin (n = 2) and vasculitis (n = 1).

In order to identify patients with severe lung emphysema as a potential influencing factor on the SUV measurement, an additional CT-based analysis of lung density was performed in all included patients using the imaging software Pulmo3D (syngo.via, Siemens Healthcare GmbH, Erlangen, Germany). Each lung was separated into three segments (upper part, middle part and lower part) and the mean lung density (MLD) was measured in each segment. MLD below -950 HU was considered as pathologic and suspicious of severe emphysema [16, 17].

Hybrid examinations

All PET/CT scans were performed on an integrated scanner with 64-slice CT technology (Biograph mCT, Siemens Healthcare GmbH, Erlangen, Germany). Depending on the clinical indication, a diagnostic contrast-enhanced CT (contrast media: Ultravist 370, Bayer Healthcare, n = 8) or a low-dose unenhanced CT (n = 7) was performed. While the whole-body CT examination was performed in end-expiratory breath-hold in order to acquire a μ-map, an additional CT scan of the lung in inspiratory breath-hold was performed (120 kv, 2 mm axial reconstruction slice thickness, convolution kernel B31f or B30f). Before intravenous injection of 18F-FDG, patients fasted for at least 6 h. Tracer injection took place before the PET/CT examination. Average injected dose of 18F-FDG was 351.8 ± 23.7 MBq, average uptake time for PET/MRI was 130 ± 15.6 min. The injected activity was sufficient for subsequent PET/MR imaging which was performed on average 68.7 ± 13.7 minutes after the PET/CT examination had started. Thus, no additional tracer injection had to be performed. All combined PET/MR examinations were performed on a fully integrated whole-body PET/MR system (Biograph mMR, Siemens Healthcare GmbH, Erlangen, Germany) which is able to acquire PET and MR simultaneously. To create a segmentation-based μ-map, a 3D T1-weighted spoiled gradient-echo sequence (VIBE) was performed in end-expiratory breath-hold, with Dixon-based fat-water separation and with the following sequence parameters: time of acquisition (TA), 19 s; matrix size, 79 x 192; repetition time (TR), 3.6 ms; excitation angle, 10°; echo times (TE), TE1 = 1.23 ms, TE2 = 2.46 ms; pixel size, 2.6 x 2.6 x 2.6mm; bandwidth, 965Hz/pixel; 128 slices per slab; parallel imaging acceleration factor 2. The μ-map is calculated by the system using the compartments air (0cm-1), lung (0.0224cm-1), fat (0.0854cm-1) and soft tissue (0.1cm-1). In all PET/MR exams, a coronal short time inversion recovery (STIR) sequence was acquired in free breathing during PET acquisition with TR/TE, 3000 ms/81 ms; inversion time (TI), 220 ms; excitation angle, 120°; bandwidth, 383 Hz/pixel; matrix size, 145 x 256; resolution, 2.0 x 1.8 x 5.0mm; parallel imaging acceleration factor 3; averages 2; TA, 2 min 13 s. Patients’ arms were positioned above the head in PET/CT and alongside the abdomen in PET/MR.

Other sequences were chosen depending on the clinical question. Based on AC templates, the AC for the spine array coil and the patient table was performed automatically. For MR signal detection, a standard phased-array body coil was used, optimized for minimal photon attenuation. This flexible coil was not considered in the μ-maps.

Reconstruction of PET images from PET/MR using different attenuation correction modes

The CT-based μ-map was derived from the CT dataset of the PET/CT examination, the CT table was removed manually using MATLAB. The CT image was non-rigidly registered to the T1-weighted in-phase MR images which are also used to calculate the MR-based attenuation μ-map and resampled to the same spatial resolution. HU values were converted to LAC values using a piecewise bilinear transformation also used in the PET/CT scanners [18]. This was performed with the software tool elastix [19]. Since the CT and MR were acquired with different arm positions, the arm regions in the CT-derived attenuation maps were replaced using the original attenuation map created by the system [20].

From each PET dataset acquired in PET/MR, four different PET reconstructions were computed using different μ-maps: i) a CT-based μ-map (PETCTAC, gold standard); ii) a CT-based μ-map in which the LAC of the lung tissue was replaced by the lung LAC from the MR-based segmentation method (PETCTAC_MRLUNG); iii) based on reconstruction ii), the LAC of bone structures was additionally replaced by the LAC of soft tissue from the MR-based segmentation method (PETCTAC_MRLUNG_NOBONE); iv) the vendor-provided MR-based μ-map (PETMRAC, segmentation-based method). The detailed workflow is depicted in Fig 1.

thumbnail
Fig 1. Overview of the workflow of reconstruction of PET images using different attenuation correction modes.

https://doi.org/10.1371/journal.pone.0177856.g001

To correct for different arm positions in PET/CT (elevated arms) and PET/MR, arms were removed in the CT datasets and replaced with arm regions of the corresponding MR-based μ-map. These steps were performed using MATLAB (The MathWorks Inc., Natick, MA, USA), results were controlled visually.

Masks of the lung volumes from these CT-based μ-maps were created using a region growing algorithm implemented in Imagine (Version 2.0; http://www.mathworks.com/matlabcentral/fileexchange/40440-imagine). The LACs of all voxels within this mask were set to the lung LAC of the MR-based μ-map (0.022 cm-1) using MATLAB. Finally, voxels of bony structures (LAC > 0.10 cm-1) were set to soft-tissue LAC of the MR-based μ-map (0.1 cm-1) using MATLAB.

To assess the precision of the registration of CT images to MR images, we performed a voxelwise subtraction of the CT-based μ-maps in which the LACs of lung and bone were replaced (μ-maps iii)) from the MR-based μ-map. A histogram analysis of the voxels within each patient was performed after segmentation of the patient from the background. Results are given in Fig 2.

thumbnail
Fig 2.

Example of a subtraction image of the MR-based μ-map and the registered CT-based μ-map of a patient in which the LACs from the lung and the bone tissue were replaced by the LACs from the MR- μ-map in axial view (A), sagittal view (B) and coronal view (C). The histogram analysis of all patients shows that most of the voxels within the patient from the subtraction images provide HU of around 0 (>90% of the voxels show a deviation of more than 10% from the maximum value (0.1cm-1)) which stands for the high precision of the performed registration.

https://doi.org/10.1371/journal.pone.0177856.g002

PET data were reconstructed with the different μ-maps using the PET reconstruction software provided by the vendor (e7-tools, Siemens Healthcare GmbH, Knoxville, TN, USA). A 3D ordered-subset expectation-maximization algorithm with 2 iterations, 21 subsets, matrix size 256 x 256, Gaussian filtering of 4 mm was used.

Image analysis

Image analysis was performed using the imaging software PMOD (PMOD Technologies Ltd., Zurich, Switzerland). The STIR sequences as well as the MR and CT μ-maps from each patient were rigidly registered to the corresponding PET data from the PET/MR system. 12 ROIs with a diameter of 1 cm were placed in physiological lung tissue of each patient. For each lung, three were set at the level of the hilum (anterior, middle, posterior) and three were set in the basal lung (anterior, middle, posterior). Fig 3 shows an example for the ROI placement.

thumbnail
Fig 3. Example of ROI placements in the basal lung parts.

Left hand side: ROIs superposed with coregistered μ-map from CT. Right hand side: ROIs superposed with the correlating PET.

https://doi.org/10.1371/journal.pone.0177856.g003

μ-maps from MR as well as the computed μ-maps from CT were rigidly overlaid to the PET to avoid positioning of ROIs in artifacts. The ROIs exported from PMOD were copied to the different corresponding PETs and to the CTs from PET/CT. The relative difference of SUVmean in percent in the different lung parts of the different PET images X to the reference PETCTAC was defined as: (SUVmean (X)—SUVmean (PETCTAC)) / SUVmean (PETCTAC) * 100. Moreover, Hounsfield units (HU) were measured in each ROI in the CT. Since CT contrast agents can have an influence on CT HU in the lung [21, 22], we also provided a separate analysis of HU differences in lung regions for unenhanced and enhanced CT data.

CT HUs and the SUVmean between the differently reconstructed PETs and the relative differences of HUs and SUVmean between the different regions were statistically compared using analysis of variance with Bonferroni post-hoc correction. P-values < 0.05 were considered significant (SPSS Statistics 19, IBM, Armonk, NY).

Results

In the MLD evaluation, lung density was higher than -950 HU in all evaluated segments so that per definition, no emphysematous lung areas were present in the evaluated patient group. HU values are given in Table 1.

thumbnail
Table 1. Overview of mean values and SD of measured Hounsfield units (HU) of CTs from PET/CT in different lung regions.

A gradient from anterior to posterior can be seen in both contrast-enhanced and non-contrast-enhanced CTs, pronounced between the middle and posterior regions. Values of contrast-enhanced CTs and non-contrast-enhanced CTs as well as the differences between the anterior and posterior regions did not differ significantly.

https://doi.org/10.1371/journal.pone.0177856.t001

HUs of contrast-enhanced and non-contrast-enhanced CTs and the differences between the anterior and posterior regions did not differ significantly between these two groups (p = 0.17 and p = 0.92, respcectively). A gradient in CT HUs from the anterior to the posterior region is visible with higher HUs in the posterior regions. The SUVmean values are given in Table 2.

thumbnail
Table 2. SUVmean including standard deviations in different regions and different PET images.

P-values describe the significance of differences between the anterior and posterior lung regions (p-value a.-p.) in the respective PET.

https://doi.org/10.1371/journal.pone.0177856.t002

A gradient of SUV from anterior to posterior is clearly visible with decreasing intensity from PETCTAC to PETMRAC. The relative differences between the differently corrected PETs are shown in Table 3. An example is given in Fig 4.

thumbnail
Fig 4. Typical example of an axial view of the right lung of a patient to demonstrate the effect of the replacement of the CT-based LACs of lung tissue (μ-map CT) with the LAC from the MR-based segmentation method from the Biograph mMR (μ-map CT_MRLUNG).

To show the differences of LACs, a subtraction map is given in the upper row on the right-hand side (μ-map CT_MRLUNG—CT); as one can see, the subtraction of μ-map CT from μ-map CT_MRLUNG leads to an underestimation of LACs in the posterior regions and an overestimation in the anterior regions (given in LACs). This results in higher SUVs in PET_CTAC as compared to PET_CTAC_MRLUNG in the posterior regions, or, in other words, an underestimation of SUVs in the posterior regions of PET_CTAC_MRLUNG as shown in the subtraction image PET CTAC_MRLUNG–CTAC.

https://doi.org/10.1371/journal.pone.0177856.g004

thumbnail
Table 3. Overview of relative differences in % between the differently reconstructed PETs and regions.

Note the differences of SUV-underestimation in the posterior lung parts while the overestimations in the anterior regions do not change significantly. P-Values describe the significance of differences between the particular lung regions in the different PET images. The SUVs of PETCTAC_MRLUNG_NOBONE and PETMRAC images did not differ significantly in any region.

https://doi.org/10.1371/journal.pone.0177856.t003

The maximum relative percentage of overestimation of PETMRAC compared to PETCTAC was +35.4% (anterior part of the lung); the maximum relative percentage of underestimation of PETMRAC to PETCTAC was -48.9% (posterior part of the lung). In all reconstructed PETs, there was no significant difference between the measured SUVmean of the right and left lung.

The SUVmean for the different lung regions of the individual patients are given in S1S4 Tables for PET_CTAC, PET_MRLUNG, PET_MRLUNG_NOBONE, PET_MRAC, respectively. The analysis of the lung densities and the respective Hounsfield units of the individual patients are given in S5 and S6 Tables, respectively.

Discussion

It is well known that the density of lung tissue differs between individuals. Previously conducted studies could show that the interpatient variability and the differences between the right and left lung of a patient can influence the quantification in PET [9, 10, 15]. However, there are additional effects on lung tissue density, such as gravitational dependency, breathing position and underlying pathology in each patient [2327].

To create a μ-map for PET/CT, the CT part is usually acquired in end-expiratory breath-hold and in supine position, which seems to provide best results in AC [28]. This offers different HU values, LAC and therefore a gradient of SUVmean for anterior, middle and posterior parts of the lung (Table 1 and Fig 4). In contrast, the segmentation-method for MR-based AC defines one single lung LAC for the entire lung and does not account for potential variances in patients’ individual lung density. This would imply that the lung tissue was very homogeneous. With the replacement of LACs of the CT-derived lung tissue with LAC, as provided in the segmentation–method (PETCTAC_MRLUNG), an overestimation of SUV in the anterior parts and a SUV underestimation in the posterior parts of the lungs in the resulting PET occurs, most probably because the gravitational dependency in the posterior parts of the lung is ignored (Fig 4). By an additional replacement of the LACs of bony structures with LAC of soft tissue of the MR-based μ-map (PETCTAC_MRLUNG_NOBONE), a further increase of SUV-underestimation compared to PETCTAC in the posterior regions was observed. This suggests that the underestimation of SUV as it is seen in the posterior lung parts in the original PETMRAC is not only caused by a gradient of lung tissue density from anterior to posterior, but also from the surrounding bone tissue, e.g., the spine and the costovertebral joints. In the anterior parts of the lung, the replacement of LACs of bony structures in the CT μ-map did not to have additional significant influence on the SUV in our evaluation, probably caused by the lower amount of surrounding bony tissue in those regions. There was no significant difference observed between SUVs of the PETMRAC and PETCTAC_MRLUNG_NOBONE images. This let suggest that differences in lung density and surrounding bone tissue are the most influencing factors on the SUV in the lung. Corresponding to the observations acquired in the relative differences, the gradient of SUVmean from anterior to posterior which can be seen in PETCTAC is continuously decreasing with the replacement of lung-LAC and bone-LAC to PETMRAC. However, there is still a small gradient of tracer uptake from anterior to posterior seen in the PETMRAC. As the MRAC does not take differences of lung density into account this finding represents a slightly higher accumulation of 18F-FDG-PET tracer in the posterior regions. As we focussed our work on the influence of different attenuation maps on the quantification of tracer uptake in the lung, the evaluation of regional distribution of different PET tracers in physiological lung tissue was beyond the scope of our work.

Several studies demonstrated that PET/MR is reliable in staging lung cancer with no significant differences compared to PET/CT regarding the clinical impact [2932]. However, modern oncology will ask for quantifiable and reproducible results of imaging technologies in the initial assessment of the disease, therapy monitoring and for evaluation of tumor response to treatment. As a consequence, PET information does not only serve for lesion detection but also for quantifying tracer uptake compared to reference points (e.g., the liver) as it is common, e.g., in lymphoma [33] but also in the recently proposed PERCIST criteria [34]. Atlas-based methods are expected to be implemented into vendor software for AC in fully integrated PET/MR-systems [20, 35]. This will lead to an implementation of bone tissue into the μ-map for attenuation correction of PET in PET/MR. Based on our results it can reasonably be assumed that this will lead to a reduction of SUV underestimation in the posterior lung regions which is partially caused by bone tissue. However, the deviation of SUV based on differences in lung density compared to PET/CT might still persist [14]. A gradient-adjusted μ-map for the lung as well as the correction of patient-specific differences [9, 10] might further reduce the system-based error of PET-quantification in the lung in fully integrated PET/MR-systems. Patient-based, individual differences might be corrected by measuring the lung density as proposed by Marshall et al. [11] or specific MR sequences like ultra-short echo time MR imaging [7, 12]. Another possibility to improve PET quantification in lung tissue has recently been proposed by Mehranian et al. by using continuous and patient-specific lung LACs, derived from time-of-flight (TOF)-PET [13].

Our study has several limitations: As the scope of our work was to investigate the influence of different methods of AC on the SUVmean in physiological lung tissue, our patient cohort did not contain a substantial number of patients with PET-avid lesions in the lung. A specific investigation of lung lesions and a potential clinical impact (e.g. lesion detection and PET quantification of lesions) will be the scope of future studies. However, it should be noticed that it is generally difficult to differentiate between respiratory motion effects, non-recognition of pulmonary lesions in the μ-map leading itself to SUV underestimation, misalignment between PET and the μ-map and the erroneous recognition of lung tissue and bone as it was elaborated in the present study. Moreover, one has to consider that PET/CT as gold standard may also be hampered in the evaluation of small lung lesions by these effects, which limits the significance of comparing PET/CT and PET/MR regarding lesion assessment. We only included lung regions which not obviously affected by diseases (in CT, MRI or PET) and, moreover, performed a measurement of lung density in CT in the different regions to make sure that we did not analyze patients with e.g. emphysema. However, this approach does not exclude all kinds of diseases which might influence the distribution of PET tracers. A more profound evaluation of lung tissue would include histological analysis or an investigation of healthy volunteers in PET which is ethically not feasible.

The patients`arms in PETCTAC were replaced by the corresponding MR-segmented arms, which were not containing bone. This leads to a potential slight misinterpretation of PET in the shoulder girdle and the apex of the lungs which currently cannot be avoided. However, since all PET images of a patient were reconstructed with the same arms, this should not have influence on the relative differences. We used flexible MR coils for the examinations which are not attenuation corrected and which might therefore also have an impact on the anterior-posterior gradient found [3638].

Conclusions

In conclusion, it could be shown that both surrounding bone tissue and differences in lung density in the μ-map may have noticeable influence on SUV in physiological lung tissue, mostly affecting the posterior lung regions. This should be taken into account for PET/MR lung reading and for the comparison of PET/CT and PET/MR examinations.

Supporting information

S1 Table. SUVmean in different lung regions in PET_CTAC.

This table shows the SUVmean from ROI analysis in the different lung regions in the PET image reconstructed with the attenuation map from PET/CT (PET_CTAC).

https://doi.org/10.1371/journal.pone.0177856.s001

(PDF)

S2 Table. SUVmean in different lung regions in PET_MRLUNG.

This table shows the SUVmean from ROI analysis in the different lung regions in the PET image reconstructed with the CT-based attenuation map in which the LACs of lung tissue were replaced by the respective LAC from the MR-based segmentation method (PET_MRLUNG).

https://doi.org/10.1371/journal.pone.0177856.s002

(PDF)

S3 Table. SUVmean in different lung regions in PET_MRLUNG_NOBONE.

This table shows the SUVmean from ROI analysis in the different lung regions in the PET images reconstructed with the CT-based attenuation map in which the LACs of lung tissue and bone were replaced by the respective LACs from the MR-based segmentation method (PET_MRLUNG_NOBONE).

https://doi.org/10.1371/journal.pone.0177856.s003

(PDF)

S4 Table. SUVmean in different lung regions in PET_MRAC.

This table shows the SUVmean from ROI analysis in the different lung regions in the PET image reconstructed with the MR attenuation map (PET_MRAC).

https://doi.org/10.1371/journal.pone.0177856.s004

(PDF)

S5 Table. Analysis of lung density.

This table shows the results of the analysis of the mean lung density in different lung regions. This analysis takes a whole lung region into account. MLD = Mean lung density. SD = Standard deviation. AL/R = Apex left/right. ML/R = Middle left /right. BL/R = Basal left/right.

https://doi.org/10.1371/journal.pone.0177856.s005

(PDF)

S6 Table. Hounsfield units.

This table shows the Hounsfield Units from ROI analysis in the CT images in the different lung regions, corresponding to the regions analysed in the PET images.

https://doi.org/10.1371/journal.pone.0177856.s006

(PDF)

Author Contributions

  1. Conceptualization: HS FS NS SG.
  2. Data curation: HS.
  3. Formal analysis: HS FS.
  4. Investigation: FS HS.
  5. Methodology: FS HS IB.
  6. Project administration: HS NS.
  7. Software: HS IB.
  8. Supervision: NS HS CP CF KN.
  9. Validation: NS HS CP CF KN.
  10. Visualization: FS HS SG.
  11. Writing – original draft: FS.
  12. Writing – review & editing: HS NS CP CS KN.

References

  1. 1. Gallamini A, Zwarthoed C, Borra A. Positron Emission Tomography (PET) in Oncology. Cancers. 2014;6(4):1821–89. Epub 2014/10/01. PubMed Central PMCID: PMC4276948. pmid:25268160
  2. 2. Keereman V, Mollet P, Berker Y, Schulz V, Vandenberghe S. Challenges and current methods for attenuation correction in PET/MR. Magma (New York, NY). 2013;26(1):81–98. Epub 2012/08/10.
  3. 3. Samarin A, Burger C, Wollenweber SD, Crook DW, Burger IA, Schmid DT, et al. PET/MR imaging of bone lesions—implications for PET quantification from imperfect attenuation correction. European journal of nuclear medicine and molecular imaging. 2012;39(7):1154–60. Epub 2012/04/25. pmid:22526955
  4. 4. Spick C, Herrmann K, Czernin J. 18F-FDG PET/CT and PET/MRI Perform Equally Well in Cancer: Evidence from Studies on More Than 2,300 Patients. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2016;57(3):420–30. Epub 2016/01/09. PubMed Central PMCID: PMC5003572.
  5. 5. Bruegel M, Gaa J, Woertler K, Ganter C, Waldt S, Hillerer C, et al. MRI of the lung: value of different turbo spin-echo, single-shot turbo spin-echo, and 3D gradient-echo pulse sequences for the detection of pulmonary metastases. Journal of magnetic resonance imaging: JMRI. 2007;25(1):73–81. Epub 2006/12/13. pmid:17154370
  6. 6. Chandarana H, Heacock L, Rakheja R, DeMello LR, Bonavita J, Block TK, et al. Pulmonary nodules in patients with primary malignancy: comparison of hybrid PET/MR and PET/CT imaging. Radiology. 2013;268(3):874–81. Epub 2013/06/06. pmid:23737537
  7. 7. Keereman V, Fierens Y, Vanhove C, Lahoutte T, Vandenberghe S. Magnetic resonace-based attenuation correction for micro-single-photon emission computed tomography. Molecular imaging. 2012;11(2):155–65. Epub 2012/04/04. pmid:22469243
  8. 8. Wagenknecht G, Kaiser H-J, Mottaghy FM, Herzog H. MRI for attenuation correction in PET: methods and challenges. Magma (New York, NY). 2013;26(1):99–113.
  9. 9. Keereman V, Holen RV, Mollet P, Vandenberghe S. The effect of errors in segmented attenuation maps on PET quantification. Medical physics. 2011;38(11):6010–9. Epub 2011/11/04. pmid:22047365
  10. 10. Izquierdo-Garcia D, Sawiak SJ, Knesaurek K, Narula J, Fuster V, Machac J, et al. Comparison of MR-based attenuation correction and CT-based attenuation correction of whole-body PET/MR imaging. European journal of nuclear medicine and molecular imaging. 2014;41(8):1574–84. Epub 2014/03/22. PubMed Central PMCID: PMC4090355. pmid:24652234
  11. 11. Marshall HR, Prato FS, Deans L, Theberge J, Thompson RT, Stodilka RZ. Variable lung density consideration in attenuation correction of whole-body PET/MRI. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2012;53(6):977–84. Epub 2012/05/09.
  12. 12. Zeimpekis K, Delso G, Wiesinger F, Veit-Haibach P, von Schulthess G, Grimm R. Investigation of 3D UTE MRI for lung PET attenuation correction. Journal of Nuclear Medicine. 2014;55(supplement 1):2103.
  13. 13. Mehranian A, Zaidi H. Emission-based estimation of lung attenuation coefficients for attenuation correction in time-of-flight PET/MR. Physics in medicine and biology. 2015;60(12):4813–33. Epub 2015/06/06. pmid:26047036
  14. 14. Hofmann M, Bezrukov I, Mantlik F, Aschoff P, Steinke F, Beyer T, et al. MRI-based attenuation correction for whole-body PET/MRI: quantitative evaluation of segmentation- and atlas-based methods. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2011;52(9):1392–9. Epub 2011/08/11.
  15. 15. Keereman V, D'Asseler Y, Van Holen R, Mollet P, Vandenberghe S. The effect of inter-patient attenuation coefficient variability on segmented attenuation correction for PET. J NUCL MED MEETING ABSTRACTS. 2010;51(2_MeetingAbstracts):1378.
  16. 16. Park KJ, Bergin CJ, Clausen JL. Quantitation of emphysema with three-dimensional CT densitometry: comparison with two-dimensional analysis, visual emphysema scores, and pulmonary function test results. Radiology. 1999;211(2):541–7. Epub 1999/05/06. pmid:10228540
  17. 17. Watanuki Y, Suzuki S, Nishikawa M, Miyashita A, Okubo T. Correlation of quantitative CT with selective alveolobronchogram and pulmonary function tests in emphysema. Chest. 1994;106(3):806–13. Epub 1994/09/01. pmid:8082363
  18. 18. Carney JP, Townsend DW, Rappoport V, Bendriem B. Method for transforming CT images for attenuation correction in PET/CT imaging. Medical physics. 2006;33(4):976–83. Epub 2006/05/16. pmid:16696474
  19. 19. Klein S, Staring M, Murphy K, Viergever MA, Pluim JP. elastix: a toolbox for intensity-based medical image registration. IEEE transactions on medical imaging. 2010;29(1):196–205. Epub 2009/11/20. pmid:19923044
  20. 20. Bezrukov I, Schmidt H, Mantlik F, Schwenzer N, Brendle C, Scholkopf B, et al. MR-based attenuation correction methods for improved PET quantification in lesions within bone and susceptibility artifact regions. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2013;54(10):1768–74. Epub 2013/09/07.
  21. 21. Kinahan PE, Hasegawa BH, Beyer T. X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Seminars in nuclear medicine. 2003;33(3):166–79. Epub 2003/08/22. pmid:12931319
  22. 22. Antoch G, Freudenberg LS, Egelhof T, Stattaus J, Jentzen W, Debatin JF, et al. Focal tracer uptake: a potential artifact in contrast-enhanced dual-modality PET/CT scans. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2002;43(10):1339–42. Epub 2002/10/09.
  23. 23. Heremans A, Verschakelen JA, Van fraeyenhoven L, Demedts M. Measurement of lung density by means of quantitative CT scanning. A study of correlations with pulmonary function tests. Chest. 1992;102(3):805–11. Epub 1992/09/01. pmid:1516407
  24. 24. Kalender WA, Rienmuller R, Seissler W, Behr J, Welke M, Fichte H. Measurement of pulmonary parenchymal attenuation: use of spirometric gating with quantitative CT. Radiology. 1990;175(1):265–8. Epub 1990/04/01. pmid:2315492
  25. 25. Gattinoni L, Pelosi P, Vitale G, Pesenti A, D'Andrea L, Mascheroni D. Body position changes redistribute lung computed-tomographic density in patients with acute respiratory failure. Anesthesiology. 1991;74(1):15–23. Epub 1991/01/01. pmid:1986640
  26. 26. Verschakelen JA, Van fraeyenhoven L, Laureys G, Demedts M, Baert AL. Differences in CT density between dependent and nondependent portions of the lung: influence of lung volume. AJR American journal of roentgenology. 1993;161(4):713–7. Epub 1993/10/01. pmid:8372744
  27. 27. Hartley PG, Galvin JR, Hunninghake GW, Merchant JA, Yagla SJ, Speakman SB, et al. High-resolution CT-derived measures of lung density are valid indexes of interstitial lung disease. Journal of applied physiology (Bethesda, Md: 1985). 1994;76(1):271–7. Epub 1994/01/01.
  28. 28. Vogel MN, Brechtel K, Klein MD, Aschoff P, Horger M, Eschmann S, et al. [Evaluation of different breathing and contrast-protocols concerning quality and alignment in 18F-FDG PET/CT]. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2007;179(1):72–9. Epub 2006/12/06. pmid:17146750
  29. 29. Huellner MW, de Galiza Barbosa F, Husmann L, Pietsch CM, Mader CE, Burger IA, et al. TNM Staging of Non-Small Cell Lung Cancer: Comparison of PET/MR and PET/CT. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2016;57(1):21–6. Epub 2015/10/17.
  30. 30. Fraioli F, Screaton NJ, Janes SM, Win T, Menezes L, Kayani I, et al. Non-small-cell lung cancer resectability: diagnostic value of PET/MR. European journal of nuclear medicine and molecular imaging. 2015;42(1):49–55. Epub 2014/08/15. PubMed Central PMCID: PMC4244545. pmid:25120040
  31. 31. Schaarschmidt BM, Grueneisen J, Metzenmacher M, Gomez B, Gauler T, Roesel C, et al. Thoracic staging with 18F-FDG PET/MR in non-small cell lung cancer—does it change therapeutic decisions in comparison to 18F-FDG PET/CT? European radiology. 2016. Epub 2016/05/18.
  32. 32. Heusch P, Buchbender C, Kohler J, Nensa F, Gauler T, Gomez B, et al. Thoracic staging in lung cancer: prospective comparison of 18F-FDG PET/MR imaging and 18F-FDG PET/CT. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2014;55(3):373–8. Epub 2014/02/08.
  33. 33. Barrington SF, Qian W, Somer EJ, Franceschetto A, Bagni B, Brun E, et al. Concordance between four European centres of PET reporting criteria designed for use in multicentre trials in Hodgkin lymphoma. European journal of nuclear medicine and molecular imaging. 2010;37(10):1824–33. Epub 2010/05/28. pmid:20505930
  34. 34. Wahl RL, Jacene H, Kasamon Y, Lodge MA. From RECIST to PERCIST: Evolving Considerations for PET response criteria in solid tumors. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2009;50 Suppl 1:122S–50S. Epub 2009/06/24. PubMed Central PMCID: PMC2755245.
  35. 35. Paulus DH, Quick HH, Geppert C, Fenchel M, Zhan Y, Hermosillo G, et al. Whole-Body PET/MR Imaging: Quantitative Evaluation of a Novel Model-Based MR Attenuation Correction Method Including Bone. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2015;56(7):1061–6. Epub 2015/05/31.
  36. 36. Paulus DH, Braun H, Aklan B, Quick HH. Simultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coils. Medical physics. 2012;39(7):4306–15. Epub 2012/07/27. pmid:22830764
  37. 37. Paulus DH, Tellmann L, Quick HH. Towards improved hardware component attenuation correction in PET/MR hybrid imaging. Physics in medicine and biology. 2013;58(22):8021–40. Epub 2013/10/31. pmid:24168832
  38. 38. Kartmann R, Paulus DH, Braun H, Aklan B, Ziegler S, Navalpakkam BK, et al. Integrated PET/MR imaging: automatic attenuation correction of flexible RF coils. Medical physics. 2013;40(8):082301. Epub 2013/08/10. pmid:23927344