The successive projection algorithm as an initialization method for brain tumor segmentation using non-negative matrix factorization

  • Nicolas Sauwen,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Software, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliations KU Leuven, Department of Electrical Engineering (ESAT), STADIUS Centre for Dynamical Systems, Signal Processing and Data Analytics, Leuven, Belgium, imec, Leuven, Belgium

  • Marjan Acou,

    Roles Data curation, Investigation, Writing – original draft, Writing – review & editing

    Affiliation Ghent University Hospital, Department of Radiology, Ghent, Belgium

  • Halandur N. Bharath,

    Roles Conceptualization

    Affiliations KU Leuven, Department of Electrical Engineering (ESAT), STADIUS Centre for Dynamical Systems, Signal Processing and Data Analytics, Leuven, Belgium, imec, Leuven, Belgium

  • Diana M. Sima,

    Roles Conceptualization, Funding acquisition, Methodology, Project administration, Software, Writing – original draft, Writing – review & editing

    Affiliations KU Leuven, Department of Electrical Engineering (ESAT), STADIUS Centre for Dynamical Systems, Signal Processing and Data Analytics, Leuven, Belgium, imec, Leuven, Belgium, Icometrix, R&D Department, Leuven, Belgium

  • Jelle Veraart,

    Roles Formal analysis, Investigation, Software

    Affiliation University of Antwerp, iMinds Vision Lab, Department of Physics, Antwerp, Belgium

  • Frederik Maes,

    Roles Project administration, Writing – original draft, Writing – review & editing

    Affiliation KU Leuven, Department of Electrical Engineering (ESAT), PSI Centre for Processing Speech and Images, Leuven, Belgium

  • Uwe Himmelreich,

    Roles Funding acquisition, Project administration, Supervision, Writing – original draft, Writing – review & editing

    Affiliation KU Leuven, Department of Imaging and Pathology, Biomedical MRI/MoSAIC, Leuven, Belgium

  • Eric Achten,

    Roles Data curation, Funding acquisition, Project administration, Resources, Supervision

    Affiliation Ghent University Hospital, Department of Radiology, Ghent, Belgium

  • Sabine Van Huffel

    Roles Conceptualization, Funding acquisition, Project administration, Resources, Supervision, Writing – original draft, Writing – review & editing

    sabine.vanhuffel@kuleuven.be

    Affiliations KU Leuven, Department of Electrical Engineering (ESAT), STADIUS Centre for Dynamical Systems, Signal Processing and Data Analytics, Leuven, Belgium, imec, Leuven, Belgium

Abstract

Non-negative matrix factorization (NMF) has become a widely used tool for additive parts-based analysis in a wide range of applications. As NMF is a non-convex problem, the quality of the solution will depend on the initialization of the factor matrices. In this study, the successive projection algorithm (SPA) is proposed as an initialization method for NMF. SPA builds on convex geometry and allocates endmembers based on successive orthogonal subspace projections of the input data. SPA is a fast and reproducible method, and it aligns well with the assumptions made in near-separable NMF analyses. SPA was applied to multi-parametric magnetic resonance imaging (MRI) datasets for brain tumor segmentation using different NMF algorithms. Comparison with common initialization methods shows that SPA achieves similar segmentation quality and it is competitive in terms of convergence rate. Whereas SPA was previously applied as a direct endmember extraction tool, we have shown improved segmentation results when using SPA as an initialization method, as it allows further enhancement of the sources during the NMF iterative procedure.

Introduction

Non-negative matrix factorization (NMF) has become a widely used tool for multivariate data analysis in the fields of blind source separation and pattern recognition. NMF decomposes a non-negative input matrix X into the product of 2 non-negative factor matrices W and H, providing a low-rank (rank r) approximation: (1) NMF provides an additive parts-based representation of the input data. As such, it reveals the basic components which are present in the data, the so-called sources, and models each input signal (i.e. column of the input matrix) as a weighted sum of the sources. The columns of W represent the sources and each column of H contains the weights, or so-called abundances, of the sources for one particular column of X. The most commonly used similarity measure for X and its factorization is the Frobenius norm. It will also serve as the cost function for solving the optimization problem: (2)

As NMF is a non-convex optimization problem, the obtained factorization will be a local rather than the global minimum of the cost function. The final result therefore depends on the initialization of the factor matrices, W0 and H0. Random initialization is the benchmark and is used in the vast majority of NMF studies. However, the quality and reproducibility of the NMF result is rarely questioned when using random initialization. As was pointed out in [1], one would have to run NMF with random initialization a sufficient number of times, then select the ‘best’ run based on some criterion (e.g. lowest Frobenius residual error), to warrant robust and reproducible NMF results. This would of course increase computation time dramatically.

Alternatively, more advanced initialization strategies have been suggested for NMF. Besides random initialization, other randomization based methods have been suggested. These methods obtain the columns of W0 by averaging a number of random columns of X or a subset of X [2]. Despite having a low computational cost and providing a more realistic first estimate of the sources compared to random initialization, these methods also suffer from a lack of reproducibility. Another group of initialization schemes is based on low-rank approximation methods of the input data matrix, such as the singular value decomposition [1, 3, 4] or independent component analysis [5]. These methods rely on the most significant low-rank components and their corresponding source vectors to initialize W0 and H0. As these methods impose constraints such as orthogonality or statistical independence to the source vectors, the non-negative structure of the input data is lost, introducing negative values into W0 and H0. Some straightforward ways of dealing with these negative values have been proposed, such as setting them to zero [1], replacing them by a mean value from the input matrix [1] or taking their absolute values [3]. Clustering algorithms have also been suggested for NMF initialization, such as spherical k-means clustering [6], fuzzy C-means clustering [7, 8] and subtractive clustering [9]. Clustering-based initialization schemes will provide more realistic source estimates compared to low-rank approximation methods, but they can become computationally expensive. Furthermore, clustering methods usually require some initialization themselves. Most of the proposed initialization methods have been compared with random initialization in terms of convergence rate and/or quality of the solution. However, different random initializations will lead to different NMF results, making it a questionnable reference. It is unclear how previous studies have coped with this lack of reproducibility.

NMF has been used in several biomedical applications, such as ECG-EMG signal unmixing [10], prostate tumor detection [11] and brain tumor segmentation [12, 13]. Brain tumor segmentation is a crucial task for planning surgical resection, for radiotherapy planning and to monitor tumor growth or shrinkage during follow-up [14]. It aims at outlining the total tumor volume as well as its main constituting tissue compartments, i.e. active tumor, necrosis and edema. NMF has been applied for tumor segmentation of multi-parametric MRI data [12, 13]. In this context, the columns of the input matrix X correspond to the different voxels and the rows represent the MRI parameters. The sources in W are interpreted as tissue-specific signatures and the abundances in H as the proportions of the different tissue types in each voxel. As such, NMF models each voxel’s MRI feature set as a weighted sum of the tissue-specific signatures. The abundances associated with each source can be visualised as a segmentation of the image in various tissue components and evaluated by comparison against manual expert segmentation.

The Successive Projection Algorithm (SPA) is a forward selection method which minimizes collinearity of the selected variables in vector space. It was introduced by Araújo et al. [15] and has been commonly used for feature selection. Several studies have considered SPA as an endmember extraction tool for hyperspectral unmixing [16, 17]. SPA is fast and reproducible, and it takes advantage of the geometrical convexity that is seen in a wide range of NMF problems [16]. To the authors’ knowledge, this is the first study in which SPA is being proposed as an initialization method for NMF. We illustrate the use of SPA initialization on 2 multi-parametric MRI datasets, applying NMF for brain tumor segmentation. The performance of SPA in terms of segmentation quality and convergence rate is compared with other common initialization methods, i.e. random initialization, non-negative double singular value decomposition (NNDSVD) and fuzzy c-means clustering (FCM). Non-deterministic methods such as random initialization are run repetitively, making sure the reported results are reproducible. Several NMF algorithms are discussed and their sensitivity to the initialization methods is investigated. The paper is further organized as follows: Section 2 describes the multi-parametric MRI datasets. Section 3 introduces the NMF methods, the SPA algorithm and the other initialization methods. NMF segmentation results comparing the different initialization methods are given in Section 4, and an in-depth discussion of the results follows in Section 5.

Materials and methods

Multi-parametric MRI datasets

The NMF methods are applied for brain tumor segmentation on 2 multi-parametric MRI datasets, acquired at the Ghent University Hospital (UZ Ghent) and the University Hospital of Leuven (UZ Leuven). Both datasets consisted of structural MRI, perfusion-weighted MRI, diffusion-weighted MRI and MR spectroscopic imaging, but they were acquired using a different scanning protocol. The MRI modalities provide us with complementary information about the structural, haemodynamic and biochemical properties of the brain and tumor tissue. It was shown previously that combining these MRI modalities leads to improved brain tumor segmentation [12].

The UZ Ghent dataset consisted of 21 patients who were diagnosed with a high-grade glioma. The UZ Ghent local ethics committee allowed a retrospective analysis of the data. The MR examinations were performed on a 3T Siemens Trio Tim scanner (Erlangen, Germany), using a standard 12-channel phased array head coil. A detailed description of the UZ Ghent acquisition protocol and image processing methods can be found in Appendix 1. All MRI modalities were rigidly coregistered using SPM8 (Wellcome Department of Imaging Neuroscience, University College London, UK). Skull-stripping was applied to all modalities prior to coregistration. The normalized mutual information criterion was used for coregistration [18]. All images were brought to the same spatial resolution of 1 × 1 × 3mm3, with cubic spline interpolation for reslicing. Only voxels within the MRSI volume of interest were included in the NMF analysis, as only these voxels had a complete set of MRI features. A total of 24 MRI features was finally obtained, making up the rows of the input matrix X.

The UZ Leuven patient cohort consisted of 14 high-grade glioma patients. Written informed consent was obtained from every patient before participation. MRI acquisition was performed on a 3T Philips Achieva scanner (Best, The Netherlands), using a body coil for transmission and an 8-channel head coil for signal reception. A detailed description of the UZ Leuven acquisition protocol and image processing methods can be found in the Appendix 2. A total of 29 MRI features was obtained from the multi-parametric scanning sequence. NMF analyses were restricted to one axial image slice, as the MR spectroscopic data were only acquired in 2D. Rigid coregistration with reslicing was applied as for the UZ Ghent dataset.

NMF and initialization methods

NMF methods.

We consider 3 different NMF problem formulations, published before in the literature, namely: single-level NMF [19], Convex NMF [20] and Hierarchical NMF (hNMF) [12]. Single-level NMF, called shortly NMF in the sequel, formulated as in Eq (2), solves for W and H in one step. Convex NMF imposes additional constraints on the source vectors in W and hence considers a more restricted single-level NMF problem. Hierarchical NMF, denoted as hNMF in the sequel, considers a multi-level NMF approach by splitting the initial problem hierarchically over several levels as a sequence of NMF problems in order to solve for W and H. Each NMF problem can be solved using a variety of optimisation methods. In this paper we consider 3 different algorithms for solving the single-level NMF problem: accelerated hierarchical alternating least squares NMF (aHALS NMF), gradient descent NMF (GD NMF) and projected gradient NMF (PG NMF). Convex NMF is solved using multiplicative update rules. For hNMF, we consider a 2-level approach, using aHALS NMF at each level. The proposed NMF methodologies are discussed in more detail below.

aHALS NMF.

aHALS NMF [21] is a member of the family of alternating least squares (ALS) NMF methods. These methods rely on the observation that finding the optimal factor matrix H when W is fixed, and finding the optimal W when H is fixed, are convex problems, as opposed to the original non-convex NMF problem as defined in Eq (2). In the current study we chose to use aHALS because of its computational efficiency and fast convergence compared to other ALS methods [21].

GD NMF.

Gradient-descent NMF (GD NMF) methods are based on non-linear least squares optimization. As these methods aim at directly minimizing the NMF cost function, they show faster convergence compared to standard NMF algorithms [22]. A transformation of variables is used to convert the constrained optimization problem in Eq (2) into an unconstrained problem, by squaring the entries of the factor matrices. The Gauss-Newton algorithm with dogleg trust region is used to solve the resulting non-linear least-squares problem [23]. The Gauss-Newton algorithm linearizes Eq (2) using Taylor’s expansion. The conjugate gradients method, which is particularly suitable for large-scale problems, can be used to calculate the gradient step from the resulting linear least squares problem at each iteration.

PG NMF.

Lin et al. proposed an efficient implementation for solving the NMF problem by alternating non-negative least squares using projected gradients (PG NMF) [24]. W and H are alternatingly updated by taking steps along the negative gradient direction. Negative elements occurring throughout the iterative procedure are brought back to the non-negative orthant by setting them to zero. The “Armijo rule along the projection arc” strategy is used to determine the gradient step size.

Convex NMF.

Convex NMF [20] imposes the constraint that the source vectors (i.e. the columns of W) must lie within the column space of X. As such, each source is a weighted sum of the data points. An auxiliary matrix A is introduced to define these weights: (3) Multiplicative update rules are defined for A and H. Without additional constraints, Convex NMF results in a sparse H matrix.

hNMF.

Hierarchical NMF (hNMF) has been introduced in previous MRI studies on brain tumor characterization [12, 25]. It consists of a 2-level approach, assigning tissue types which are most similar to the same source after a first level of rank-2 NMF. Tissue-specific sources are obtained after the second level of NMF. The sources are then recombined in a final step to calculate the tissue abundances using non-negative least squares. hNMF has shown improved differentiation and segmentation of the pathologic tissue types in brain tumors compared to single-level aHALS NMF. As hNMF consists of 2 levels of NMF, the initialization methods are applied at both NMF stages.

Initialization methods

Successive projection algorithm (SPA).

To better understand why SPA is suitable for finding the NMF sources, it is worthwhile to consider the NMF problem from a geometrical perspective. Let’s consider a non-negative input matrix X of rank 2, containing n data points in three-dimensional space (i.e. ). The fact that X has rank 2 implies that all its data points belong to a two-dimensional subspace, and they are all located in the non-negative orthant. To solve the NMF problem, we need to find 2 sources which will be located in the same two-dimensional subspace as the data points. Let’s further assume that for each source, there is at least one data point in X that purely contains that source (i.e. the pure-pixel assumption). In that case, it can be seen that all data points are confined within a convex cone, whose edges intersect with the pure data points. Any data point within this cone can be obtained as a weighted sum of the extreme vectors of the cone. From this geometrical point of view, NMF comes down to finding the vertices of a convex cone spanning all the data points.

For now, we assume that the pure-pixel assumption holds, meaning that some data points in X contain purely one source. We further assume that the sum of the abundances is not greater than 1 in any data point, which is called the sum-to-one constraint. For such input data, the NMF problem is said to be near-separable [26]. Under the assumptions of near-separable NMF, the data points of X will span a convex hull in m-dimensional space, and the sources correspond to the vertices of this hull. SPA fits well with this geometrical interpretation of NMF, as it aims at finding the vertices [16]. SPA works as follows: in a first step, the data point p1 with the highest l2-norm is selected, as it will correspond to a vertex of the convex hull. Next, all data points are projected onto the orthogonal complement of p1. The data point p2 with the highest l2-norm in this projected subspace will be another vertex. The next vertex p3 is found as the data point with the highest l2-norm after projection onto the orthogonal complement of p1 and p2, and so on. The columns of W0 will be formed by p1, p2, …, and pr. Algorithm 1 describes the SPA procedure.

Algorithm 1: SPA

Input: , rank r

Output: Set of vectors {p1, p2, …, pr} ∈ X

 1. Initialize matrix S = X

 2. for j = 1: r

 3. find index i* such that

 4. set pj = X(:, i*), sj = S(:, i*)

 5. update S

 6. end

SPA does not directly provide an initialization for the abundance matrix, H0. Based on W0, we used non-negative least squares fitting to obtain H0. For Convex NMF, an initialization of the auxiliary matrix A0 is required. Solving Eq (3) for A, we find: (4) Since A must be non-negative, we obtain A0 as proposed in [20]: (5) with A+ being the positive part of A, 〈A〉 = ∑m,n|An,k|/∥An,k0 and ∥An,k0 is the number of nonzero elements in A. E is a matrix of ones with the same dimensions as A. A0 is similarly obtained from H0 for the other initialization methods.

Repetitive random initialization (rRandom).

Initialization obtained from SPA is compared with rRandom initialization, where the elements of W0 and H0 are set to uniformly distributed non-negative values between 0 and 1 at each run. We ran NMF 30 times with random initialization to obtain reproducible results. For rRandom, the final NMF result is selected as the random run with the lowest residual error as defined in Eq (2).

Non-negative double singular value decomposition (NNDSVD).

NNDSVD is based on 2 levels of SVD. First, the best rank-r approximation of the input data matrix X is computed based on truncated SVD: (6) with ui and vi being the ith left and right singular vectors of X. As we require a fully non-negative initialization, the positive component of Ci, is withheld, being the nearest non-negative rank-2 approximation of Ci. It was shown that the dominant singular triplet of can easily be obtained by decomposing the singular vectors ui and vi into their positive and negative components. W0 and H0 are then initialized based on the dominant singular triplet of each . A detailed description of the NNDSVD algorithm can be found in [1]. Several studies have considered NNDSVD for NMF initialization [9, 27]. It must be noted that NNDSVD in its original form introduces a significant number of zero elements in W0 and H0. This can be problematic for some NMF algorithms, i.e. elements initialized at zero will remain zero in the final solution. This is for instance the case for the original multiplicative update NMF algorithm proposed by Lee and Seung [19], and also for the GD NMF algorithm that we are considering. To overcome this kind of convergence problem, we set all zero elements in W0 and H0 to a small positive value, i.e. to 1% of the mean value of the positive elements of W0 and H0, respectively.

Fuzzy c-means clustering (FCM).

FCM aims at partitioning the input data into c fuzzy clusters. The data points have a degree of belonging to each of the clusters centers, as in fuzzy logic, rather than belonging completely to just one cluster. The degree of membership of a data point i belonging to cluster j is defined by a weighting value wij, which depends on a distance metric between the data point and the cluster centroid (e.g. Euclidean distance). FCM is an iterative algorithm, in which the cluster centroids and the weighting values are iteratively updated until convergence [28]. FCM has been applied as an initialization method for NMF in several studies [7, 8], with the cluster centroids being assigned to the columns of W0 and the membership weights per cluster serving as the rows of H0. It has to be noted that FCM in itself is non-deterministic, meaning that the final output will depend on the initialization. Therefore, we applied the same approach as with random initialization: run NMF with randomly initialized FCM 30 times and select the run with the lowest residual Frobenius norm as defined in Eq (2) as the final solution.

Validation

Tissue segmentation masks are obtained by applying k-means clustering to the voxel-wise abundance values of H. As it is assumed that each source and its associated abundances correspond to one tissue type, we initialized the cluster centroids by setting one abundance value to 1 and all others to 0. The NMF segmentations are compared with manual segmentation performed by an experienced neuroradiologist using MRIcron [29]. The overlap between segmentations was quantified using the Dice-score: (7) where Atissue,NMF is the area segmented by NMF and Atissue,radiol the area segmented by the radiologist for the same tissue type. The Dice-score is one of the most commonly used metrics to quantitatively evaluate segmentation performance in brain tumors [14, 30]. It normalizes the number of true positives to the average size of the area segmented by both methods. Dice-scores were calculated for 3 tissue classes, as defined in the BraTS challenge [30]: active tumor, the tumor core (i.e. active tumor and necrosis) and the whole tumor (tumor core and edema). Segmentation results were also assessed for direct SPA output, to compare performance of using SPA as an endmember extraction tool to using it as an initialization method. We will denote the NMF results obtained with the different initialization methods as NMFSPA, NMFNNDSVD, NMFFCM and NMFrRandom, respectively.

Results

Fig 1 shows an example of a coregistered image set of a glioblastoma patient. The segmentation results for active tumor and necrosis are shown for hNMF (in blue) using the different initialization methods, and are compared with the manual segmentation by the radiologist (in green). Segmentation of the active tumor region is similar for SPA, NNDSVD and rRandom, although NNDSVD and rRandom show more overlap with manual segmentation, but also some additional false positive segmentation. FCM shows a considerable oversegmentation outside of the pathologic region for active tumor. SPA obtained good segmentation correspondence for necrosis, whereas the other initialization methods show some underestimation of the necrotic area.

thumbnail
Fig 1. Coregistered set of MRI images.

Showing T1+C(A), CBV(B), ADC(C) and Lac(D). The MRSI region of interest is marked in green. Segmentation results are shown below for active tumor and necrosis, respectively, for hNMF with each type of initialization: SPA (E,F), NNDSVD (G,H), FCM (I,J) and rRandom (K,L). NMF segmentation is shown in blue, segmentation by the radiologist in green and overlap in cyan.

https://doi.org/10.1371/journal.pone.0180268.g001

Table 1 shows the mean Dice-scores on the UZ Ghent dataset for SPA, NNDSVD, FCM and rRandom initialization. Looking at the results per tissue class and per NMF method, variations in mean Dice-scores among the different initialization methods are mostly below 5%. The largest difference is found for active tumor with GD NMF and Convex NMF, where NMFFCM gives 5% lower Dice-scores than NMFSPA. The highest mean Dice-score per NMF method and per tissue class is marked in bold. SPA has the highest mean Dice-score for 12 out of 15 comparisons, NNDSVD has the highest Dice-score in only 2 cases, FCM in 3 cases and rRandom in 7 cases. Dice-scores for NMFNNDSVD are lower than or at best equal to NMFSPA for all tissue classes and for all NMF methods, but differences are never higher than 3%. The last column of Table 1 shows the segmentation results obtained directly from SPA output (W0, H0). The Dice-scores obtained from SPA were lower or at best equal to those from the NMFSPA methods for each tissue class. Statistical significance of the higher Dice-scores for NMFSPA compared to direct SPA was found in about half of the cases.

thumbnail
Table 1. Comparison of the mean Dice-scores and their standard deviation between different initialization methods for the UZ Ghent dataset.

The highest Dice-score per NMF method and per tissue class is marked in bold. * indicates statistically significantly higher Dice-scores with SPA initialization compared to direct SPA endmember extraction (right column), using a one-tailed Wilcoxon signed rank test (p < 0.05).

https://doi.org/10.1371/journal.pone.0180268.t001

Table 2 reports the mean Dice-scores on the UZ Leuven dataset for the different initialization methods. Looking at the results per NMF method and per tissue class, differences in the mean Dice-score are never higher than 5%. SPA obtains the highest mean Dice-score in 10 out of 15 comparisons. NNDSVD has the highest Dice-score in 5 cases, FCM in 7 cases and rRandom in 7 cases. Dice-scores obtained from direct SPA output were lower or at best equal to those from the NMFSPA methods for each tissue class. Statistical significance of the higher Dice-scores for NMFSPA compared to direct SPA could be shown in most cases, except for the active tumor region and for the whole tumor region with Convex NMF.

thumbnail
Table 2. Comparison of the mean Dice-scores and their standard deviation between different initialization methods for the UZ Leuven dataset.

The highest Dice-score per NMF method and per tissue class is marked in bold. * indicates statistically significantly higher Dice-scores with SPA initialization compared to direct SPA endmember extraction (right column), using a one-tailed Wilcoxon signed rank test (p < 0.05).

https://doi.org/10.1371/journal.pone.0180268.t002

Table 3 shows the mean number of iterations to convergence for the 4 single level NMF methods using the different initialization schemes. These values could not directly be reported for hNMF, as it consists of 2 levels of aHALS NMF. The same stopping criteria were applied to all NMF methods: a maximum number of iterations of 10000 and a convergence tolerance of 10−5 for the relative difference of two subsequent values of the residual error ∥XWHF. For aHALS NMF, the mean number of iterations is below 200 for all initialization methods. Convergence rate is low for Convex NMF, where all initialization methods result in a mean number of iterations close to 9000. Only for GD NMF, we see a clear difference in convergence rate among the initialization methods. NMFSPA and NMFNNDSVD have the fastest convergence, with a mean number of iterations of 218 and 175, respectively. NMFrRandom requires on average 418 iterations to reach convergence, which is about twice as many. NMFFCM has the lowest convergence rate, with a mean number of iterations of 832. PG NMF shows the highest convergence rate for all initialization methods, with only NMFrRandom requiring more than 100 iterations on average. Fig 2 illustrates the convergence behavior of the 4 single-level NMF methods when comparing initialization schemes for a particular glioblastoma patient. Initial errors are lowest for SPA and FCM. The main reduction of the residual error occurs within the first 100 iterations in all cases, except for Convex NMF with rRandom. For aHALS NMF, Convex NMF and PG NMF, all initialization methods converge to approximately the same residual error, whereas for GD NMF SPA and FCM converge to a higher final residual.

thumbnail
Table 3. Mean number of iterations to reach convergence for the different initialization methods on the UZ Ghent dataset.

Convergence tolerance was set to 10−5 and the maximum number of iterations to 10000.

https://doi.org/10.1371/journal.pone.0180268.t003

thumbnail
Fig 2. Convergence plots for aHALS NMF (A), GD NMF (B), Convex NMF (C), and PG NMF (D) with the different initialization methods.

The residual error, ∥XWHF, is shown on a log scale. For rRandom and FCM, the shown curve corresponds to the selected run with the lowest residual error.

https://doi.org/10.1371/journal.pone.0180268.g002

The software code used within the study has been made available in the manuscript’s Supporting Information files (see S1 File), along with one patient’s anonymized dataset. Interested researchers may run the code on this examplary dataset. Other patients included in the study did not consent to make their data publicly available.

Discussion

NMF performance

This study illustrates the behaviour of SPA as an initialization method for NMF in comparison to other common initialization methods for brain tumor segmentation. As noted in [1], there is no consensus on how to assess NMF performance when comparing different initialization methods. Many studies report the final residual error as a measure of performance [2, 3, 9], which has the advantage of being a generally applicable measure. Other studies use a validation which is specific to the problem at hand. Ortega-Martorell et al. calculated correlation coefficients as a quality measure of the obtained tissue sources for brain tumor characterization using single-voxel magnetic resonance spectroscopic data [13]. Wild et al., besides looking at the residual error, assessed the quality of image sources for a facial image reconstruction problem based on visual inspection [6]. We assessed quality of the segmentation results using the Dice-scores, because a lower residual error does not warrant a better parts-based solution. For instance, one can observe in Fig 2 that higher residual errors remain for Convex NMF compared to aHALS and GD NMF. This is due to the additional constraint imposed by Convex NMF that the sources have to be a weighted sum of the input data. As can be seen in Table 1, Convex NMF reaches similar Dice-scores as the other single-level NMF methods. Therefore we decided to assess the quality of the NMF result using direct validation of the problem at hand, i.e. segmentation of the pathologic tumor regions. Mean Dice-scores did not differ by more than 5% for any NMF method using the different initialization strategies. In terms of the highest Dice-score per tissue class, SPA came out best on both datasets. SPA had the highest mean Dice-score in 12 out of 15 cases for the UZ Ghent dataset and in 10 out of 15 cases for the UZ Leuven dataset. These results suggest that SPA performs slightly better than the other initialization methods in terms of segmentation quality.

For most advanced initialization methods, similar performance has been reported compared to random initialization, although residual errors generally tend to be rather higher than lower. Wild et al. discussed how most advanced methods obtain a head start in terms of residual error compared to random initialization, but this advantage is mostly lost towards the end of the iteration process [6]. In some cases, random initialization will achieve a lower error towards the end of the convergence process, which is also illustrated in [6] and in Fig 2 for GD NMF. This is explained by the fact that advanced initialization might impose some restrictions on the factorization which favor other local minima, leading to more meaningful tissue segmentations. In Fig 2C, SPA and FCM show lower initial errors but higher final residual errors compared to rRandom and NNDSVD. The initial factor matrices obtained from SPA and FCM are more representative for the data at hand, and tend to converge to a nearby local optimum for GD NMF. As such, GD NMF is directed towards a factorization result which is meaningful in terms of tissue segmentation.

Convergence and computational cost

Although a head start in terms of the initial error has been reported for most advanced initialization methods compared to random initialization, mixed results have been found in terms of the number of iterations to reach convergence [9]. As shown in Table 3, for aHALS NMF and Convex NMF we did not find considerable differences in convergence rate among the different initialization methods. For GD NMF, SPA and NNDSVD converge significantly faster than rRandom and FCM. For PG NMF, rRandom converges somewhat slower than the advanced initialization methods. Besides convergence speed, another important concern regarding computation time is the fact that random and FCM initialization are non-deterministic. Therefore, individual runs of these initialization strategies will not provide reproducible results, and some particular initializations might lead to a local optimum which is far from the global optimum. We decided to run NMF with repetitive random and FCM initialization, with a sufficient number of runs such that results were reproducible. This of course dramatically increased computation times for NMFrRandom and NMFFCM.

Concerning computational complexity, SPA is a straightforward and computationally inexpensive method, requiring a total of 6mnr operations [16]. For NNDSVD, the main computational cost comes from calculating the truncated SVD of the input data matrix X. The SVD is commonly calculated by means of some iterative algorithm, with a computational complexity of O(mnr) operations per iteration [1]. This means that the leading constant, which depends on the number of iterations, will usually be much higher for NNDSVD compared to SPA. Similarly, FCM is an iterative procedure with a complexity of O(mnr) per iteration [9], such that the leading constant will be higher than for SPA. Mean computation times of the advanced initialization methods on the UZ Ghent dataset were 0.25s for NNDSVD and 4s for FCM. SPA took on average 0.02s to initialize W0. It must be noted, however, that both NNDSVD and FCM deliver W0 and H0, whereas SPA only delivers W0. In the current study, we used non-negative least squares fitting after SPA, taking on average 8s per patient to initialize H0. Computationally more efficient methods to initialize H0 are available. Random initialization is sometimes used for H0 in combination with a more advanced initialization of W0 [2], although this partially annihilates the advantages of using a fixed and input-specific initialization. Another possibility is to initialize H0 based on least-squares fitting without non-negativity constraints, then setting all negative values to zero or a small positive value. This actually comes down to one updating step of the basic block coordinate descent ALS algorithm. Standard least-squares will produce H0 with a higher initial error, but at a lower computational cost. Due to the relatively low cost of the initialization compared to the NMF computation itself, we did not explore more efficient strategies to initialize H0.

SPA: Initialization or direct source extraction

The current study is not the first one to combine SPA with NMF. Several studies have applied SPA as an endmember extraction algorithm for NMF in hyperspectral unmixing [16, 31]. However, to the authors’ knowledge, SPA has so far not been considered as an initialization strategy. We hypothesized that there might be some benefit to assigning the SPA sources to W0 instead of directly using them as the final columns of W, because SPA will not always directly provide the most suitable sources for the NMF problem at hand. One of the main limitations of SPA is its sensitivity to outliers. As SPA looks for the outer vertices of a geometrical hull, outliers might be wrongly assigned as endmembers. Several methods have been proposed to either make SPA less sensitive to outliers [17] or to discard the outliers beforehand [32]. Another important caveat is that SPA assumes the NMF problem to be near-separable [16], which is equivalent to the spatial representation of the input data by a convex hull. The assumptions that are intrinsic to near-separable NMF, i.e. the pure-pixel assumption and the sum-to-one constraint on the abundances, will often not hold, especially in the case of highly mixed NMF problems. But when applying SPA as an initialization method, the NMF procedure might still correct the source estimates. For the multi-parametric MRI datasets considered in this study, the pure-pixel assumption holds for most of the MRI modalities due to their high spatial resolution. However, the poor resolution of the MR spectroscopic data leads to significant partial volume effects, such that the pure-pixel assumption does not hold for these data. We can further assume that the sum-to-one constraint approximately holds, as each voxel represents a fixed tissue volume. By comparing the Dice-scores of NMFSPA to those obtained directly from SPA output matrices W0 and H0, it was shown that we can obtain better segmentation results when using SPA as an initialization method rather than as a direct source extraction tool. Mean SPA Dice-scores were lower than the Dice-scores obtained with any of the NMF methods and for all tissue classes. Statistical significance of the higher Dice-scores for NMFSPA could be shown in about half of the cases for the UZ Ghent dataset and in most cases for the UZ Leuven dataset.

Conclusion

This study has proposed SPA as an initialization method for NMF. Whereas SPA has previously been applied as a direct endmember extraction algorithm, using it as an initialization method might allow further enhancement of the sources during the NMF iterative procedure. This advantage has been shown on 2 multi-parametric MRI datasets for brain tumor segmentation. Compared to other initialization methods, SPA provides a realistic and reproducible initialization. Looking at the highest mean Dice-scores per NMF method and per tissue class, it was found that SPA performs slightly better than the other initialization methods in terms of segmentation quality. It is also competitive with the other initialization methods in terms of convergence rate. As the feasibility of using SPA initialization has been shown for multi-parametric MRI based tumor segmentation, we encourage the exploration of SPA in other NMF applications as well.

Appendix

Appendix 1: UZ Ghent dataset

Structural imaging included a 3-dimensional T1-weighted gradient-echo sequence (MPRAGE) with isotropic voxels (176 sagittal slices, field of view (FOV) read = 220mm, voxel size 0.9 × 0.9 × 0.9mm3, Repetition Time (TR) = 1550ms, Echo Time (TE) = 2.39ms, Inversion Time (TI) = 900ms, matrix size = 256 × 256, GRAPPA factor 2, flip angle = 9°) and a 3-dimensional T2-weighted inversion recovery sequence (FLAIR) with isotropic voxels (176 sagittal slices, FOV read = 250mm, voxel size 1 × 1 × 1mm3, TR = 6000ms, TE = 421ms, TI = 2100ms, matrix size = 256 × 238, GRAPPA factor 2). The MPRAGE sequence was repeated after administration of gadolinium contrast, namely following the acquisition of the Dynamic Susceptibility Contrast (DSC) perfusion-weighted imaging.

Perfusion-weighted imaging was performed by using a lipid-suppressed, T2*-weighted echo-planar imaging sequence with the following parameters: TR = 1000ms, TE = 29ms, FOV = 230 × 230mm2, slice thickness = 5.0mm, matrix size = 128 × 128, in-plane voxel size = 1.8 × 1.8mm2, 15 slices, distance factor = 30%, GRAPPA factor 2, flip angle = 90°. A series of 90 multi-section acquisitions was acquired at 1 second intervals. The first 10 acquisitions were performed before contrast agent injection to establish a pre-contrast baseline. At the tenth acquisition, a 0.1mmol/kg body weight bolus of gadobutrol (Gadovist, Bayer) was injected with a power injector (Spectris, Medrad Inc., Indianola, PA) at a rate of 4ml/s through a 18-gauge intravenous catheter, immediately followed by a 20 ml bolus of sodium chloride solution at 4ml/s. Relative Cerebral Blood Volume (CBV) maps were derived from the dynamic signal intensity curves in DSCoMAN (Duke University, Durham, NC) using the method proposed by Boxerman et al. [33].

Axial diffusion-weighted images were acquired using a fast single-shot gradient-echo echo-planar imaging sequence with diffusion gradient b-values of 0, 500 and 1000s/mm2 (voxel size 2.0 × 2.0 × 3.0mm3, TR = 5400ms, TE = 80ms, FOV read = 264mm, number of averages = 3). An affine coregistration was applied to account for eddy currents. Apparent Diffusion Coefficient (ADC) maps were derived from the 3 b-values using weighted linear least squares fitting [34]. The b0 images were also added to the dataset, serving as a T2-weighted reference.

A 3D proton magnetic resonance spectroscopic imaging (MRSI) protocol with long TE was included. In the two-slice MRSI examination, a volume of interest of 80 × 80mm2 including tumour, perilesional edema and normal brain tissue was positioned on reconstructions of the 3D FLAIR sequence. A Stimulated Echo Acquisition Mode (STEAM) pulse sequence was used for 3D spatial localization. The VOI was completely enclosed within the brain and 8 outer volume suppression slabs were placed outside the volume of interest. Magnetic resonance parameters were TR = 1700ms, TE = 135ms, flip angle = 90°, FOV = 160 × 160mm2, voxel size 10 × 10 × 15mm3, acquisition bandwidth 1200Hz, number of averages = 3. Weak water suppression was used. Automatic shimming with manual fine-tuning of the B0 magnetic field was used as well as iterative semi-automatic optimization of the transmitter voltage. The SPID software [35] was used to quantify the following metabolites with AQSES-MRSI [36]: lipid (Lip), lactate (Lac), N-acetyl aspartate (NAA), glutamine+glutamate (Glx), total creatine (Cre), total choline (Cho). Maximum phase FIR filtering was applied for residual water suppression [37]. One or two voxel bands at the outer edges of the spectroscopic grid were omitted from the analyses when showing severe chemical shift displacement effects or lipid contamination.

Twelve MRI imaging features were obtained from the raw acquired data after pre-processing: T1, T1+C, FLAIR, CBV, ADC, b0, Lip, Lac, NAA, Glx, Cre, Cho. In addition, 3 × 3 and 5 × 5 smoothing windows were applied to the 6 non-MRSI features, and these averaged features were also added to the feature set to improve robustness of the segmentation results [12]. A total of 24 MRI features was finally obtained for each voxel, making up the rows of the input matrix X.

Appendix 2: UZ Leuven dataset

Structural MRI consisted of T2-weighted imaging, T1-weighted imaging with contrast enhancement and FLAIR. An axial spin echo T2-weighted MRI was acquired with the following parameters: TR/TE = 3000/80ms; slice/gap = 4/1mm; turbo factor = 10; FOV = 230 × 184mm2; acquisition matrix = 400 × 300. A T1-weighted 3D spoiled gradient echo MRI scan with contrast administration was performed with the following parameters: TR/TE/TI = 9.7/4.6/900ms; flip angle = 8°; turbo field echo factor = 180; acquisition voxel size = 0.98 × 0.98 × 1mm3; 118 contiguous partitions. An axial FLAIR MRI scan was acquired with the following parameters: TR/TE/TI = 11000/120/2800ms, slice/gap = 4/1 mm, FOV = 230 × 184mm2, acquisition matrix = 240 × 134.

Perfusion-weighted imaging was obtained using dynamic susceptibility-weighted contrast-enhanced MRI with a gradient-echo EPI sequence. A series of 60 multi-section scans was acquired at 1.35 second intervals, using the following parameters: TR/TE = 1350/30ms; section thickness/gap = 3/0mm; FOV = 200 × 200mm2; acquisition matrix = 112 × 109; EPI data were acquired during the first pass following a rapid injection of a 0.1mmol/kg body weight bolus of meglumine gadoterate (Dotarem, Guerbet, France) via a mechanical pump at a rate of 4ml/s, followed by a 20ml bolus of saline. Relative Cerebral Blood Volume (CBV) maps were derived from the dynamic signal intensity curves in DSCoMAN (Duke University, Durham, NC) using the method proposed by Boxerman et al. [33].

An extensive diffusion kurtosis imaging (DKI) protocol was included at UZ Leuven. A spin echo EPI diffusion-weighted sequence was used to acquire the DKI data. Implemented b-values were 0, 700, 1000, and 2800s/mm2, applied in respectively 1, 25, 40, and 75 uniformly distributed directions. The following parameters were used throughout the DKI acquisition sequence: TR/TE = 3200/90ms; gradient duration (δ) = 20ms; diffusion time interval (Δ) = 48.3ms; FOV = 240 × 240mm2; acquisition matrix = 96 × 96; number of signal averages = 1; section thickness/gap = 2.5/0mm; parallel imaging: SENSE with factor 2 in the antero-posterior direction. DKI parameters were estimated using a constrained weighted linear least squares (WLLS) model [34]. Mean diffusivity (MD), fractional anisotropy (FA) and mean kurtosis (MK) maps were computed from the DKI tensor as described in [38].

Proton MRSI consisted of a 2D short echo time sequence. Acquisition parameters were as follows: TR/TE = 2000/35ms; FOV = 160 × 160mm2; maximal region of interest = 80 × 80mm2; section thickness = 10mm; reconstruction voxel size = 5 × 5mm2; receiver bandwidth = 2000Hz; parallel imaging: SENSE. Automated pre-scanning optimized the shim in order to yield a peak width consistently under 20Hz full-width half-maximum. AQSES-MRSI was used to quantify the following metabolites: lipid (Lip), lactate (Lac), N-acetyl aspartate (NAA), glutamine+glutamate (Glx), total creatine (Cre), total choline (Cho), myo-inositol (mI) and glycine (Gly). Maximum phase FIR filtering was applied for residual water suppression [37]. One or two voxel bands at the outer edges of the spectroscopic grid were omitted from the analyses when showing severe chemical shift displacement effects or lipid contamination.

Fifteen MRI features were obtained from the raw acquired data after pre-processing: T2, T1+C, FLAIR, CBV, MD, FA, MK, Lip, Lac, NAA, Glx, Cre, Cho, mI and Gly. In addition, 3 × 3 and 5 × 5 smoothing windows were applied to the 7 non-MRSI parameters, and these averaged features were also added to the feature set to improve robustness of the segmentation results [12]. A total of 29 MRI features was finally obtained for each voxel.

Supporting information

S1 File. Software code.

The software code used within the study has been made available in the file S1_File.zip, along with one patient’s anonymized dataset. Interested researchers may run the code on this examplary dataset. The code has been written in matlab. After unzipping the file, please consult the file README_Code.docx on how to run an NMF analysis and validate the segmentation result.

https://doi.org/10.1371/journal.pone.0180268.s001

(ZIP)

Acknowledgments

This work has been funded by the following projects: Research Foundation Flanders (FWO), grant number G.0869.12N; IWT IM 135005; Henri Benedictus Fellowship of the Belgian American Educational Foundation; Interuniversity Attraction Poles Program (P7/11) initiated by the Belgian Science Policy Office. European Research Council under the European Union’s Seventh Framework Programme (FP7/2007-2013), ERC Advanced Grant: BIOTENSORS (n°339804) and the EU MC ITN TRANSACT 2012 (n°316679).

References

  1. 1. Boutsidis C, Gallopoulos E. SVD based initialization: A head start for nonnegative matrix factorization. Pattern Recognition. 2008;41(4):1350–1362.
  2. 2. Langville AN, Meyer CD, Albright R, Cox J, Duling D. Initializations for the nonnegative matrix factorization. In: Proceedings of the twelfth ACM SIGKDD international conference on knowledge discovery and data mining. Citeseer; 2006. p. 23–26.
  3. 3. Qiao H. New SVD based initialization strategy for Non-negative Matrix Factorization. Pattern Recognition Letters. 2014;63:71–77.
  4. 4. Zhao L, Zhuang G, Xu X. Facial expression recognition based on PCA and NMF. In: 7th World Congress on Intelligent Control and Automation, 2008. WCICA 2008. IEEE; 2008. p. 6826–6829.
  5. 5. Yu S, Zhang Y, Liu W, Zhao N, Xiao X, Yin G. A novel initialization method for nonnegative matrix factorization and its application in component recognition with three-dimensional fluorescence spectra. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. 2012;86:315–319.
  6. 6. Wild S, Curry J, Dougherty A. Improving non-negative matrix factorizations through structured initialization. Pattern Recognition. 2004;37(11):2217–2232.
  7. 7. Zheng Z, Yang J, Zhu Y. Initialization enhancer for non-negative matrix factorization. Engineering Applications of Artificial Intelligence. 2007;20(1):101–110.
  8. 8. Rezaei M, Boostani R, Rezaei M. An efficient initialization method for nonnegative matrix factorization. Journal of Applied Sciences. 2011;11(2):354–359.
  9. 9. Casalino G, Del Buono N, Mencar C. Subtractive clustering for seeding non-negative matrix factorizations. Information Sciences. 2014;257:369–387.
  10. 10. Niegowski M, Zivanovic M. ECG-EMG separation by using enhanced non-negative matrix factorization. In: 36th Annual International Conference of the IEEE on Engineering in Medicine and Biology Society (EMBC). IEEE; 2014. p. 4212–4215.
  11. 11. Laudadio T, Croitor Sava A, Sima D, Wright A, Heerschap A, Van Huffel S. Hierarchical non-negative matrix factorization applied to in vivo 3T MRSI prostate data for automatic detection and visualization of tumours. In: Proc. of the European Society for Magnetic Resonance in Medicine and Biology Annual meeting 2013; 2013. p. 474–475.
  12. 12. Sauwen N, Sima DM, Van Cauter S, Veraart J, Leemans A, Maes F, et al. Hierarchical non-negative matrix factorization to characterize brain tumor heterogeneity using multi-parametric MRI. NMR in Biomedicine. 2015;28(12):1599–1624. pmid:26458729
  13. 13. Ortega-Martorell S, Lisboa PJ, Vellido A, Julià-Sapé M, Arús C. Non-negative matrix factorisation methods for the spectral decomposition of MRS data from human brain tumours. BMC Bioinformatics. 2012;13(1):38. pmid:22401579
  14. 14. Bauer S, Wiest R, Nolte LP, Reyes M. A survey of MRI-based medical image analysis for brain tumor studies. Physics in Medicine and Biology. 2013;58(13):R97. pmid:23743802
  15. 15. Araújo MCU, Saldanha TCB, Galvão RKH, Yoneyama T, Chame HC, Visani V. The successive projections algorithm for variable selection in spectroscopic multicomponent analysis. Chemometrics and Intelligent Laboratory Systems. 2001;57(2):65–73.
  16. 16. Gillis N. Successive nonnegative projection algorithm for robust nonnegative blind source separation. SIAM Journal on Imaging Sciences. 2014;7(2):1420–1450.
  17. 17. Zhang J, Rivard B, Rogge D. The successive projection algorithm (SPA), an algorithm with a spatial constraint for the automatic search of endmembers in hyperspectral data. Sensors. 2008;8(2):1321–1342. pmid:27879768
  18. 18. Maes F, Collignon A, Vandermeulen D, Marchal G, Suetens P. Multimodality image registration by maximization of mutual information. IEEE Transactions on Medical Imaging. 1997;16(2):187–198. pmid:9101328
  19. 19. Lee DD, Seung HS. Learning the parts of objects by non-negative matrix factorization. Nature. 1999;401(6755):788–791. pmid:10548103
  20. 20. Ding C, Li T, Jordan M. Convex and semi-nonnegative matrix factorizations. IEEE Transactions on Pattern Analysis and Machine Intelligence. 2010;32(1):45–55. pmid:19926898
  21. 21. Gillis N, Glineur F. Accelerated multiplicative updates and hierarchical ALS algorithms for nonnegative matrix factorization. Neural Computation. 2012;24(4):1085–1105. pmid:22168561
  22. 22. Wang YX, Zhang YJ. Nonnegative matrix factorization: A comprehensive review. IEEE Transactions on Knowledge and Data Engineering. 2013;25(6):1336–1353.
  23. 23. Nocedal J, Wright SJ. Numerical Optimization, 2nd Edition. Springer; 2006.
  24. 24. Lin CJ. Projected gradient methods for nonnegative matrix factorization. Neural computation. 2007;19(10):2756–2779. pmid:17716011
  25. 25. Li Y, Sima DM, Cauter SV, Croitor Sava AR, Himmelreich U, Pi Y, et al. Hierarchical non-negative matrix factorization (hNMF): a tissue pattern differentiation method for glioblastoma multiforme diagnosis using MRSI. NMR in Biomedicine. 2013;26(3):307–319. pmid:22972709
  26. 26. Arora S, Ge R, Kannan R, Moitra A. Computing a nonnegative matrix factorization–provably. In: Proceedings of the forty-fourth annual ACM symposium on Theory of computing. ACM; 2012. p. 145–162.
  27. 27. Heinrich KE, Berry MW, Homayouni R. Gene tree labeling using nonnegative matrix factorization on biomedical literature. Computational intelligence and neuroscience. 2008;2008:2.
  28. 28. Nock R, Nielsen F. On weighting clustering. IEEE Transactions on Pattern Analysis and Machine Intelligence. 2006;28(8):1223–1235. pmid:16886859
  29. 29. Rorden C, Karnath HO, Bonilha L. Improving lesion-symptom mapping. Journal of Cognitive Neuroscience. 2007;19(7):1081–1088. pmid:17583985
  30. 30. Menze B, Reyes M, Jakab A, Bauer S, Prastawa M, Van Leemput K, et al. The Multimodal Brain Tumor Image Segmentation Benchmark (BRATS). IEEE Transactions on Medical Imaging. 2015;34(10):1993–2024. pmid:25494501
  31. 31. Mizutani T. Robustness Analysis of Preconditioned Successive Projection Algorithm for General Form of Separable NMF Problem. arXiv preprint arXiv:150608387. 2015;
  32. 32. Gillis N, Vavasis S. Fast and robust recursive algorithms for separable nonnegative matrix factorization. Pattern Analysis and Machine Intelligence, IEEE Transactions on. 2014;36(4):698–714.
  33. 33. Boxerman J, Schmainda K, Weisskoff R. Relative cerebral blood volume maps corrected for contrast agent extravasation significantly correlate with glioma tumor grade, whereas uncorrected maps do not. American Journal of Neuroradiology. 2006;27(4):859–867. pmid:16611779
  34. 34. Veraart J, Sijbers J, Sunaert S, Leemans A, Jeurissen B. Weighted linear least squares estimation of diffusion MRI parameters: strengths, limitations, and pitfalls. Neuroimage. 2013;81:335–346. pmid:23684865
  35. 35. Poullet JB. Quantification and classification of magnetic resonance spectroscopic data for brain tumor diagnosis. Ph.D. dissertation, Leuven, Belgium; 2008.
  36. 36. Croitor Sava AR, Sima DM, Poullet JB, Wright AJ, Heerschap A, Van Huffel S. Exploiting spatial information to estimate metabolite levels in two-dimensional MRSI of heterogeneous brain lesions. NMR in Biomedicine. 2011;24(7):824–835. pmid:21834006
  37. 37. Poullet JB, Sima DM, Van Huffel S, Van Hecke P. Frequency-selective quantitation of short-echo time 1H magnetic resonance spectra. Journal of Magnetic Resonance. 2007;186(2):293–304. pmid:17433741
  38. 38. Poot DH, den Dekker AJ, Achten E, Verhoye M, Sijbers J. Optimal experimental design for diffusion kurtosis imaging. IEEE Transactions on Medical Imaging. 2010;29(3):819–829. pmid:20199917