A novel ultra-light suction device for mechanical characterization of skin

  • Bettina Müller ,

    Contributed equally to this work with: Bettina Müller, Julia Elrod

    Roles Data curation, Formal analysis, Investigation, Methodology, Software, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliation Institute for Mechanical Systems, Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland

  • Julia Elrod ,

    Contributed equally to this work with: Bettina Müller, Julia Elrod

    Roles Data curation, Investigation, Methodology, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliation Department of Surgery, University Children’s Hospital Zurich, Zurich, Switzerland

  • Marco Pensalfini,

    Roles Data curation, Methodology, Software, Writing – original draft, Writing – review & editing

    Affiliation Institute for Mechanical Systems, Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland

  • Raoul Hopf,

    Roles Data curation, Software, Writing – original draft, Writing – review & editing

    Affiliation Institute for Mechanical Systems, Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland

  • Oliver Distler,

    Roles Conceptualization, Writing – original draft, Writing – review & editing

    Affiliation Department of Rheumatology, University Hospital Zurich, Zurich, Switzerland

  • Clemens Schiestl,

    Roles Conceptualization, Resources, Writing – original draft, Writing – review & editing

    Affiliation Department of Surgery, University Children’s Hospital Zurich, Zurich, Switzerland

  • Edoardo Mazza

    Roles Conceptualization, Formal analysis, Methodology, Project administration, Resources, Supervision, Writing – original draft, Writing – review & editing

    mazza@imes.mavt.ethz.ch

    Affiliations Institute for Mechanical Systems, Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland, Empa, Swiss Federal Laboratories for Materials Science and Technology, Dübendorf, Switzerland

Abstract

Suction experiments have been extensively applied for skin characterization. In these tests the deformation behavior of superficial tissue layers determines the elevation of the skin surface observed when a predefined negative (suction) pressure history is applied. The ability of such measurements to differentiate between skin conditions is limited by the variability of the elevation response observed in repeated experiments. The scatter was shown to be associated with the force exerted by the observer when holding the instrument against the skin. We have developed a novel suction device and a measurement procedure aiming at a tighter control of mechanical boundary conditions during the experiments. The new device weighs only 3.5 g and thus allows to minimize the force applied on the skin during the test. In this way, it is possible to reliably characterize the mechanical response of skin, also in case of low values of suction pressure and deformation. The influence of the contact force is analyzed through experiments on skin and synthetic materials, and rationalized based on corresponding finite element calculations. A comparative study, involving measurements on four body locations in two subjects by three observers, showed the good performance of the new procedure, specific advantages, and limitations with respect to the Cutometer®, i.e. the suction device most widely applied for skin characterization. As a byproduct of the present investigation, a correction procedure is proposed for the Cutometer measurements, which allows to partially compensate for the influence of the contact force. The characteristics of the new suction method are discussed in view of future applications for diagnostic purposes.

1. Introduction

The mechanical behavior of the skin primarily depends on the connective tissue structures which are present in the epidermis, dermis—consisting of the more superficial papillary dermis and the collagen-rich reticular dermis—and the subcutaneous tissues[1]. The skin, particularly the epidermis and the stratum corneum, functions as a barrier protecting the body from potentially harmful external influence. The characterization of the epidermis is often associated with the development of therapies for pressure ulcers or of cosmetic products[2]. The biomechanical properties of the skin are dominated by the dermis, which consists mostly of collagen (77% of dry weight), elastin (4%), and ground substances[3].

Several skin diseases are associated with structural tissue changes, which lead to abnormal mechanical behavior of the skin. An example is systemic sclerosis (SSc), which is a heterogeneous autoimmune inflammatory disorder characterized by thickening of connective tissues in skin. A progressive fibrotic process and enhanced deposition of collagen results in the disease, leading to significant stiffening of the skin[4,5]. The non-invasive diagnosis of SSc includes a subjective palpation method, so that an objective procedure for characterization of skin biomechanical properties is expected to improve the assessment of disease progression. Another example of potential clinical use of skin biomechanical analysis is for monitoring of healing scars, in particular for cases of large wounds covered by skin grafts. Over the time angiogenesis, continued wound contraction, and eventually connective tissue remodeling might result in the formation of extensive scars. Many ancillary treatment modalities to improve scar quality have been developed, yet the question whether a therapy is indicated for individual patients remains and poses a substantial challenge. This motivates the need for methods to objectively assess scar progression as well as the properties of mature scars[68].

In vivo methods for mechanical characterization of superficial skin layers might be considered for such diagnostic applications. They can be grouped in five major classes, including in-situ tensile tests, torsional test, indentation, ballistometry (using the impact of a mass on the skin surface), and tissue elevation methods[9]. The approach most widely used for determining mechanical characteristics of healthy and diseased skin is the suction method[1,2,1013]. In suction experiments, a negative pressure is applied on the skin surface through an aperture of the suction device. In response to this stimulus, skin tissue is drawn into the probe cavity. The elevation of the skin tissue is quantified and tissue stiffness determined from the ratio of negative pressure and skin displacement. Besides the stiffness, other parameters associated with the viscoelastic and viscoplastic behavior of skin can be determined by controlling the time history of pressure applied in suction experiments. Additionally, different suction openings can be used to observe the mechanical properties of the most superficial layers (2 mm probe aperture) or deeper skin tissues (4 mm– 8 mm) [14,15]. The largest number of skin suction investigations reported in literature used the Cutometer® MPA 580 (Courage + Khazaka electronic GmbH, Germany) [16] which was applied to study skin aging[17,18], epidermal hydration[19], systemic sclerosis[13,14,20], and for elasticity measurements in scars[6,21]. Control of pressure increase and decrease rates with the Cutometer can be exploited to provide information on skin stiffness, recoverable and dissipative deformation, and viscoelasticity of the skin[1,22]. For measurements on the same skin location, results variability is associated with the compressive force exerted to push the instrument against the skin to ensure contact tightness, as well as possible relative movements between patient and observer during suction. Bonaparte et al. [23] reported the force applied by the observer on the probe, and therefore on the skin, as the most important limitation of suction measurements with the Cutometer. Increasing force on the probe resulted in reduction of pliability and elasticity values. Even minor changes in contact force led to significantly different outcomes, affecting interpretation of patient data. Not only inter-rater variability suffers from varying contact forces, but also comparison between studies is difficult. Accordingly, recent investigations introduced specific protocols to minimize the effect of these factors. Weickenmeier et al. [24] and Pensalfini et al. [25] introduced a custom-modified headrest for rigid fixation of participant’s head to prevent variations in probe placement on the face, control of contact pressure and relative movement between probe and measured location. Probably due to limited ability to discriminate between different skin conditions, the Cutometer MPA 580 has never actually become part of standardized applications in clinics.

Based on our experience with suction experiments for soft organs characterization[2430] we developed a novel suction device and measurement procedure specifically designed to overcome the above limitations for measurements on skin. We call the new device “Nimble”. Based on the same principle as the Cutometer MPA 580, the new device was optimized in terms of weight (down to few grams), in order to avoid the effect of the contact force as well as drastically reduce the influence of patient movement during the measurement. The specific configuration evaluated here was selected such that skin deformation remains relatively low, which was hypothesized to improve the ability of this method to distinguish between skin conditions. The paper presents the new technique; it compares its performance with that of the Cutometer when applied on synthetic materials and skin. The influence of contact force is studied through measurements on skin and synthetic materials and corresponding finite element calculations. A procedure is proposed to compensate for this effect in Cutometer measurements. Finally, three observers performed suction experiments on four different body locations in two volunteers. Quantitative comparison of the two suction methods includes their ability to distinguish between skin locations as well as the intra- and inter-observer variability. The new suction procedure is shown to offer significant advantages.

2. Material and methods

2.1 Measurement devices

2.1.1 Cutometer.

In previous studies, the Cutometer was shown to be applicable to various body locations, leading to good reliability, safety, and significance of measurement outcomes[21]. In the Cutometer probe a negative pressure is created, drawing the skin into a defined cavity through a circular opening (diameters ranging from 2 to 8 mm). A 6 mm probe opening was selected in order to perform a measurement which includes deeper tissue and not only the most superficial skin layers, as typically achieved with a 2 mm probe opening. In addition, this size has often been applied in previous suction studies on scar tissue [31,32], and this is among the planned future applications of the new device. Inside the probe, skin elevation is determined by a non-contact optical measurement system consisting of a light source and a light receptor, determining the light intensity decrease associated with skin protrusion. Applied pressure history and corresponding elevation curves constitute the main measurement outcome, from which several parameters are extracted. In particular, the parameter R0 (mm) was considered here, which is a measure of the maximum tissue elevation in response to a prescribed negative pressure ramp.

2.1.2 Nimble.

The novel instrument is named “Nimble” and is based on the design of an aspiration device previously applied in clinical studies for the characterization of the uterine cervix[28,33,34]. As described in [35], the new device was minimized in terms of components and features, so to reduce its weight as far as possible. A dedicated set-up and protocol for applications on human skin was realized for the present investigations.

Fig 1 shows a schematic drawing of the Nimble. The main part is the aspiration probe, which is connected to two pressure sensors, PS1 and PS2, via silicone tubes. Between the pressure sensors and the aspiration probe, filters are integrated to avoid contamination associated with airflow. The silicone tubes T1 and T2 are connected to short metal tubes within the measurement probe (Fig 2C). T1 is glued at a distance of h = 1 mm from the lower front side (Fig 2B). The pressure tube T1 is connected to the peristaltic pump, which generates a progressive negative pressure within the aspiration probe. As a consequence, skin tissue is sucked into the aspiration probe opening until it reaches a distance h. As soon as the tissue touches and seals the tube T1, a difference is detected between the pressure values reported by the pressure sensors PS1 and PS2. Thus, the measurement is automatically stopped and the corresponding closing pressure, pcl (PS2), is recorded. The valve between Filter 1 and the peristaltic pump releases the pressure to ensure safe detachment of the probe after the measurement. The probe opening diameter was selected here as d = 6 mm to match the corresponding Cutometer characteristic.

thumbnail
Fig 1. Schematic illustration of the Nimble with all components.

The aspiration probe is connected (tubes T1 and T2) to pressure sensors PS1 and PS2 via filters 1 and 2. T1 is connected to a peristaltic pump, which generates the progressive negative pressure in the cavity. The valve releases the pressure at the end of the measurement.

https://doi.org/10.1371/journal.pone.0201440.g001

thumbnail
Fig 2. Aspiration probe.

(A) Schematic of aspiration probe: The position of the vertical tube T1 defines the elevation height h. (B) The tissue is drawn into the cavity until it closes T1. (C) The dimensions of the aspiration probe are indicated.

https://doi.org/10.1371/journal.pone.0201440.g002

The Nimble probe is, with Ø 10 mm x 10 mm and a total weight of 3.5 g, smaller and lighter than all existing suction devices, including the lightweight DermaLab USB (Cortex Technologies) and the LASTIC [36]. In particular, when compared with the Cutometer (120.8 mm x 26 mm, with a weight of 80 g) the volume and weight of the Nimble are more than one order of magnitude smaller.

Different from the Cutometer, which measures the elevation obtained for a given suction pressure, the Nimble determines the pressure required for a given elevation h, i.e. it is a displacement-controlled experiment, which inherently increases the safety of the measurement procedure. The specific value of h can be selected for each application. The typical skin elevation obtained in corresponding Cutometer measurements ranges between 1.1 and 2 mm [25,37]. For the present study, a value of h = 1 mm was selected, thus leading to rather small deformations of the skin in Nimble measurements.

The Nimble is held by the observer through light and highly flexible silicon tubes and is approached to the skin surface while the peristaltic pump is running. In this way, contact is established without significant forces being exerted on the skin. Due to the lightweight of the probe, the Nimble follows possible body movements during the measurements, minimizing possible detrimental influences on the measurement. To facilitate the creation of tight contact between the aspiration probe and the skin surface at the beginning of the measurement, a thin Vaseline layer is applied on the outer edge of the Nimble. No Vaseline was required for the Cutometer due to the large surface area in contact with skin.

2.2 Suction measurement protocol

2.2.1 Cutometer.

The so-called “Mode 2” of the Cutometer software was used for this study. A ramp load with 15 mbar/s was applied up to a maximum suction pressure of pmax = 400 mbar. Unloading was with the same rate (-15 mbar/s). The Elevation-Time and the Pressure-Elevation curves were extracted for each measurement. Fig 3 shows an Elevation-Time curve, and the maximum elevation of the tissue is the parameter called “R0”, often used in literature[21]. Fig 3 also includes a Pressure-Elevation curve. We used this curve to extract a parameter directly comparable to the closing pressure measured with the Nimble: this is the suction pressure needed for a tissue elevation of 1 mm and we call it . From both, and R0 we determined corresponding stiffness values (mbar/mm) of the tissue, as and respectively.

thumbnail
Fig 3. Output curve of Cutometer measurements.

Representative Elevation-Time curve (left) and Pressure-Elevation curve (right) of a Mode 2 Cutometer measurement. The maximum elevation is indicated with R0 and the loading and unloading curves are displayed. From the loading part of the Pressure-Elevation curve the closing pressure at 1 mm elevation is extracted for comparison with the Nimble parameter pclnimble. T0 and TR0 indicate the start of the experiment and the timepoint when R0 is reached respectively. These graphs were generated with a maximum suction pressure of 400 mbar.

https://doi.org/10.1371/journal.pone.0201440.g003

2.2.2 Nimble.

The pressure-loading ramp can be defined through the setting of the peristaltic pump. A rate of 15 mbar/s was selected to match the one of the Cutometer. The measurement parameter was recorded for every measurement. Fig 4 demonstrates a Pressure-Time curve of a Nimble measurement. The pressure progression measured by the PS2, see Fig 1, and PS1 are indicated. Δp defines the threshold difference (5 mbar) applied to detect the closure of the tube T1. When Δp is reached, the closing pressure is retrieved from the reading of PS1. The stiffness parameter k (mbar/mm) for the Nimble measurements is calculated as .

thumbnail
Fig 4. Output curve of Nimble measurements.

Representative Pressure-Time curve of a Nimble measurement for a defined elevation of 1.0 mm. Pressure measured by PS1 and PS2 are indicated. The closing pressure is recorded when Δp = 5 mbar is recognized.

https://doi.org/10.1371/journal.pone.0201440.g004

2.3 Preliminary measurements

A two-component silicone elastomer, Ecoflex 0030 (Smooth-On Inc.), was used in this study as a reference material to compare Nimble and Cutometer. The samples were prepared based on the protocol introduced in Bernardi et al. [38], using a mixing ratio of Part A (base polymer) and Part B (crosslinker) of 1:1. Young’s modulus of the elastomer was measured by indentation tests (FemtoTools AG, FT-MTA02). A vertical arrangement (S1 Fig) was used to induce a controlled contact force in every suction measurement, which was equal to the specimen weight. As shown in S1 Fig, the 10 mm elastomer sample was placed on top of the 6 mm diameter opening for the Cutometer. A radial overlap of 2 mm was considered sufficient to generate conditions comparable to those of contact with a half space elastomer sample.

The influence of contact force was further analyzed through preliminary measurements on skin. The contact force of the Nimble corresponded to its weight (3.5 g) while for the Cutometer probe measurements were performed for the probe weight (80 g) as well as for increased force (Fig 5B, grey blocks add a mass of each 20.0 g).

thumbnail
Fig 5. Influence of contact force on volar forearm.

Weight study with Nimble (A) and Cutometer (B) on human volar forearm. The grey blocks indicate added mass of each 20.0 g, corresponding to increased contact force; (C) shows results of measurements on volar forearm with the Cutometer (red) and the Nimble (blue), and corresponding FE calculations on skin (orange).

https://doi.org/10.1371/journal.pone.0201440.g005

Measurements on elastomers and skin were rationalized with corresponding Finite Element (FE) calculations. S2 Fig shows the corresponding FE meshes and boundary conditions. For the simulation of measurements on skin, the dermis, the subcutaneous tissue and the underlying muscle layer were included in the model as proposed in Weickenmeier et al. [39]. The parameters for the constitutive model of each tissue are reported in S1 Table. For simulations of the measurements on Ecoflex 0030 the model was homogeneous and the elastomer was represented using a Neo-Hookean hyperelastic formulation with the parameter C10 = 0.01167 MPa, corresponding to a Young’s modulus E = 6 * C10 = 0.07 MPa. More information on the numerical model and the implemented constitutive equations are reported in Weickenmeier et al. [39] and Rubin et al. [40].

2.4 Measurements on volunteers

Two volunteers (VO1, VO2, aged between 20 and 30 years) were recruited in summer 2017 at ETH Zurich in agreement with the ethical approval EK 2015-N-63. Signed informed consent was given by each volunteer. Measurements were carried out by three observers (O1, O2, O3) with both devices, Cutometer and Nimble. The suction measurements were performed on four body locations, namely volar forearm (VF), forehead (FH), back of the hand (BH), and lower back (LB). Body locations were selected in order to test the performance of suction measurements for different conditions and anatomical features of dermis substrates. Each location was measured in triplicates for participant 1 (female) and in sextuples for participant 2 (male). In order to evaluate the influence of the number of measurement repetitions, suction experiments in the second subject were repeated six times. Between every measurement on the same location, a waiting time of at least 45 seconds was ensured, based on prior experience on tissue recovery after suction experiments[25]. A total of 216 suction experiments were performed.

2.5 Statistical analysis

For statistical analysis, the Python library scipy.stats was used (Python Software Foundation). The analysis included descriptive statistics, inter alia, calculation of minima and maxima, means (MEAN), standard deviations (STD), and coefficient of variation CV = STD/MEAN. Pairwise correlation of repeated measurements was plotted. Statistical significance of effect of the chosen method, different positions, and multiple observers was assessed with a two-sided t-test (stats.ttest_ind) and the level of significance is indicated for every measurement (p < 0.05 “*”, p < 0.01 “**” and p < 0.001 “***”). The reliability of the devices was examined by means of the intraclass correlation coefficient (ICC), where a class 2 ICC (all observers evaluate all locations) was applied and the absolute agreement calculated. Concurrent validity was assessed between the stiffness measurements of both suction devices by means of a Pearson correlation (stats.pearsonr). The calculation of the root mean square (RMS) is used for analysis of results.

3. Results

3.1 Preliminary measurements

Measurements on the Ecoflex specimen were performed in order to evaluate the reliability of the Nimble compared to the Cutometer. The repeatability of the measurements is excellent for both devices, and the closing pressure measured with the Nimble is in perfect agreement with the prediction of corresponding FE calculations for m = 0 g contact force (S3 Fig). Larger contact forces lead to a larger pre-deformation with the elastomer protruding more into the probe S1B Fig. As confirmed by corresponding simulations (S2 Table), this pre-deformation reduces the suction pressure (in case of an elastomer) required to reach 1 mm elevation, for both devices.

The difference between Cutometer and Nimble measurements is statistically significant, with the Cutometer reading about 10% higher than Nimble. Various factors were analyzed and a series of tests were performed in order to rationalize this difference. We considered the influence of probe geometry, inaccuracies in pressure readings and the influence of friction coefficient. For the latter, a FE based parametric study was performed considering conditions ranging from frictionless to dry friction (leading for both devices to a friction coefficient of μ = 1). The results indicate that these factors can justify the observed systematic difference between Nimble and Cutometer, which will be accounted for in the following sections.

The results of the measurements on elastomer indicate that the contact force present during the measurement influences the measured tissue stiffness. Since this effect is expected to depend on the mechanical behavior of the material investigated, the influence of contact force is analyzed for measurements on skin with both Nimble and Cutometer. We performed tests on the volar forearm of the volunteer VO1, Fig 5.

Results reported in Fig 5 indicate a notable increase in closing pressure with increasing contact force. Interestingly, a linear interpolation through the Cutometer data extrapolated down to zero force matches to a great extent the measurement obtained with the Nimble. Corresponding FE calculations were performed with constitutive model parameters for skin selected to match the Nimble measurement. The prediction obtained using these parameters for increased contact forces qualitatively confirms the experimental observations, with increasing apparent stiffness for larger force values. The corresponding data obtained from measurements on Ecoflex (section 3.1) are reported in the supplementary information (S3 Fig). They show that for an elastomer, the apparent stiffness (slightly) decreases for increasing contact force. This result is in line with corresponding FE calculations and is associated with the difference in strain stiffening behavior between skin and elastomer.

The devices were further analyzed in terms of their sensitivity to patient movements (rotation in the horizontal plane and vertical motion). Results are reported in supplementary information (S4 Fig) and indicate that the effect of skin movement was negligible for the Nimble while significant influences were observed for the Cutometer.

The results of Fig 5 confirm previous data[23], indicating a notable influence of contact force on suction measurements. While improved measurement protocols can help minimizing this[25], in clinical applications, the contact force applied with a hand—held device is difficult to control. For this reason, a simple correction procedure was developed to compensate (to some extent) for this influence, as described in the next section.

3.2 Correction scheme for contact force in Cutometer measurements

The contact force between suction probe and skin surface causes an initial deformation, leading to an initial tissue protrusion into the suction cavity. This initial deformation is measured by the Cutometer and displayed as “Offset”, which is proportional to the force applied by the operator. As illustrated in Fig 6A, the Cutometer measures tissue elevation starting from the initial position. Thus, when the software indicates 1 mm protrusion, the total elevation is higher in that the Offset should be added to it. When compared to the situation of the Nimble (negligible force and pre-deformation), the actual skin deformation at 1 mm elevation is higher for the Cutometer. As an alternative, it is proposed to compare the pressure values for a total elevation of 1 mm of the Cutometer, calculated as the reported elevation plus the Offset (Fig 6C).

thumbnail
Fig 6. Representation of the correction procedure.

(A) Cutometer measures tissue elevation from a baseline, defined by the apex of an initial deformation (response to contact force). (B) Nimble measures tissue elevation from skin surface (negligible initial deformation) due to its low weight. (C) The correction scheme accounts for this discrepancy and adds the Offset to the elevation measured by the Cutometer.

https://doi.org/10.1371/journal.pone.0201440.g006

The proposed correction accounts for the deformation induced by the contact force as if it would be a consequence of suction. In order to analyze the consequences of this simplification, corresponding FE calculations were performed, in which an initial deformation was generated through corresponding contact forces of 50 g and 100 g. The results are summarized in Table 1. The reference values of closing pressure for skin are those calculated for zero initial force. The error associated with the pre-deformation (without correction) is much larger for skin than for the elastomer (S2 Table). The proposed correction scheme provides a realistic estimation of the expected closing pressure. Note that the correction leads to an overestimation of the closing pressure for the skin while it underestimates the stiffness of the elastomer. This confirms that the effect of contact force depends on the constitutive behavior of the examined material and its strain-stiffening or softening response. A generalized correction scheme can therefore not be determined a priori. However, it seems that the proposed correction drastically reduces the effect of contact force for measurements on skin and it is therefore considered for the analysis of the human skin data reported in the next section. Note that contact force fluctuation during the measurements cannot be accounted for by this correction.

thumbnail
Table 1. Analysis of the proposed correction scheme for Cutometer measurements.

Results of FE calculations with enforced initial deformation (corresponding to contact forces of 50 g and 100 g) are compared for closing pressure values before and after correction. Calculations were performed for skin.

https://doi.org/10.1371/journal.pone.0201440.t001

3.3 Measurements on volunteers

Measurement results are summarized in Table 2 and Table 3. Mean values and standard deviation of data obtained by the three independent observers (O1, O2, and O3) on participants VO1 and VO2 at four different body locations (VF, FH, BH, and LB) are reported for three measurements (for VO1) or six measurements (VO2) at each location (complete data shown in S3 Table). The closing pressures obtained with the Nimble are generally much lower than those of the Cutometer. Note that no correction is applied for these data (see section 3.2). The elevation obtained with the Cutometer for 400 mbar suction pressure is much larger than the one applied with the Nimble (1 mm). As shown in Table 2 and Table 3, no significant improvement of variability was obtained when repeating the measurements six times (VO2) instead of three times (VO1). A systematic analysis of these data and comparison of the results obtained with the two devices is reported in the next sections.

thumbnail
Table 2. Measurements on volunteer VO1.

Reported are mean data and standard deviation of repeated measurements by three observer (O1, O2 and O3) on the subject VO1 at four body locations (VF, FH, BH and LB). Data include the Offset Δ, the maximum elevation R0 and the closing pressure of Cutometer, and of Nimble.

https://doi.org/10.1371/journal.pone.0201440.t002

thumbnail
Table 3. Measurements on volunteer VO2.

Reported are mean data and standard deviation of repeated measurements by three observer (O1, O2 and O3) on the subject VO2 at four body locations (VF, FH, BH and LB). Data include the Offset Δ, the maximum elevation R0 and the closing pressure of Cutometer, and of Nimble.

https://doi.org/10.1371/journal.pone.0201440.t003

3.3.1 Comparison of mean stiffness values.

As a first general comparison of the tissue characterization based on the two instruments, the mean stiffness values of each of the four locations in the two subjects were compared. The corresponding concurrent validity was calculated between knimble and the stiffness measured with the Cutometer (kcuto and kR0) by means of a Pearson correlation, see Table 4. In particular, the values -1 or +1 imply a linear relationship with direct and inverse correlation, whereas 0 implies no correlation.

thumbnail
Table 4. Pearson correlation coefficient.

r and p-values, p, between stiffness values evaluated with Nimble (knimble) and Cutometer (kcuto and kR0).

https://doi.org/10.1371/journal.pone.0201440.t004

Interestingly, the correlation of knimble with kcuto and kR0 is weak and negative. On the contrary, after application of the correction scheme proposed in section 3.2, a positive correlation is found between knimble and . The correction does not improve the correlation of the Nimble measurements with that based on R0, confirming that, due to the pronounced strain stiffening behavior of skin, the characterization at low values of suction pressure (knimble) provides a different information with respect to that at higher deformations (kR0).

To analyze the correlation between knimble with kcuto and further, Fig 7 shows the analysis of the linear regression between mean stiffness values measured by Cutometer over mean stiffness values measured by Nimble, based on the 1 mm closing pressure. Data points correspond to eight skin locations (four in each volunteer). According to the results of section 3.1, linear regression with linear least squares and zero intercept was analyzed for a slope of 1.1 (see dotted lines in the diagrams), indicating an expectation of 10% higher stiffness values for Cutometer measurements compared to Nimble. While the comparison with the non-corrected values indicates a much higher stiffness determined with Cutometer, after correction the results align reasonably well (Fig 7B). In addition, the correction notably reduces the scatter of the kcuto values, in particular for the forehead (Fig 7C). These results indicate that the proposed correction scheme can effectively be used to reconciliate Nimble and Cutometer.

thumbnail
Fig 7. Mean stiffness and standard deviation for each location and subject.

(A) and (B): linear regression for a slope of 1.1 between mean stiffness kcuto and knimble (A) and corrected Cutometer data (B), for eight different skin locations. (C): standard deviation of stiffness measurements for each body location separated for two subjects (VO1 and VO2).

https://doi.org/10.1371/journal.pone.0201440.g007

Mean values and standard deviations of stiffness values obtained for the different skin locations in each volunteer were analyzed to evaluate the ability of each parameter to distinguish among different skin conditions. To this end, each location has been considered as a specific specimen for each volunteer, see Fig 8. Nimble measurements indicate significant differences for 10 out of 12 inter-specimen comparisons. Corrected values differentiated 8 out of 12 and kR0 6 inter-specimen comparisons. The consistency between corrected and knimble is evident in Fig 8, further confirming the usefulness of the correction. Interestingly, the correction leads to a homogenization of the values so that the ability of this parameter to distinguish locations decreases. For the non-corrected values, the Cutometer indicates the lower back as the softest location for both volunteers, while it is among the stiffest based on knimble as well as corrected .

thumbnail
Fig 8. Mean values and standard deviation of stiffness values kR0, kcuto, and knimble.

Values are shown for each location separated by subjects. Significant difference between locations measured by each device are indicated. Significance level are indicated as **p < 0.01 and *** p < 0.001.

https://doi.org/10.1371/journal.pone.0201440.g008

As a next step, each of the four locations of the two subjects was considered as a specimen and a coefficient was calculated which reflects the ability of measurement procedure to differentiate among specimens[41], i.e. the Intraclass Correlation Coefficient (ICC) reported in Table 5. It is a class 2 ICC, in which all observers evaluate all specimens, and the absolute agreement is evaluated. In line with the data of Fig 8, the Nimble leads to the highest ICC (rVO1 = 0.88 and rVO2 = 0.94, strong reliability), followed by kR0 the corrected . The stiffness evaluation based on is less reliable and the stiffness based on R0 does not improve with the correction.

thumbnail
Table 5. Reliability of measurements with Cutometer (kR0 and kcuto, non-corrected and corrected), and Nimble (knimble).

Intraclass correlation coefficient reflects the ability of the parameter to differentiate among specimens in each subject (here: locations) with nVO1 = 36 and nVO2 = 72.

https://doi.org/10.1371/journal.pone.0201440.t005

3.3.2 Intraobserver and interobserver variability.

Based on the repeated measurements performed at each location by each observers (n = 3 for VO1 and n = 6 for VO2), the intraobserver variability was analyzed in terms of stiffness values. Each measurement result is reported against the successive one in the diagrams of Fig 9. Table 6 summarizes the coefficient of determination (R2) of the latter for each location. The measurements with the Nimble as well as those based on R0 show very low variability at every location (0.964 ≤ R2 ≤ 0.991 and 0.934 ≤ R2 ≤ 0.998 respectively). Even better is the correlation for corrected values (0.982 ≤ R2 ≤ 0.998), but note that points tend to cloud indicative of the increased homogeneity of these results (which leads to a lesser ability to distinguish skin conditions, see previous section). The kcuto measurements show larger variability, with an improvement associated with the correction. As expected, the excellent repeatability of measurements performed on synthetic materials (section 3.1) could not be observed in measurements on skin (CVnimble = 8.68% and CVcuto = 17.93%). These findings indicate that skin is very sensitive to interactions with the observer, and that this effect can be reduced with the Nimble due to its lightweight solution.

thumbnail
Fig 9. Intraobserver variability in terms of linear dependency.

Linear dependency of stiffness values kR0, kcuto, knimble, and based on repeated measurements at each location (VF, FH, BH and LB).

https://doi.org/10.1371/journal.pone.0201440.g009

thumbnail
Table 6. Coefficient of determination R2. R2 of intraobserver variability (in terms of linear regression) in stiffness values kR0, kcuto, knimble, and based on repeated measurements at each location (VF, FH, BH and LB).

https://doi.org/10.1371/journal.pone.0201440.t006

The error between measurements of different observers is defined as the percentage difference of each measured parameter with respect to the mean value for that location. Corresponding histogram plots are shown in Fig 10 (bin width = 10). The variance was comparable between Nimble and Cutometer measurements. The corrected error is the lowest, which is in line with the result of Fig 9 and associated with the general homogenization of these values over different locations.

thumbnail
Fig 10. Histogram plot of error between measurements of different observers (n = 3) for stiffness values kR0, kcuto, knimble, and .

The error is defined as the percentage difference of each measured parameter with respect to the mean value for that sample. Root mean square (RMS) is indicated for each parameter (RMSR0 = 19%, RMScuto = 46%, RMSnimble = 24%, = 13% and = 53%).

https://doi.org/10.1371/journal.pone.0201440.g010

3.3.3 Offset.

The results of the previous sections highlight the differences between corrected and non-corrected stiffness values obtained with the Cutometer. Accounting for the initial Offset reduces measurement variability and leads to a general agreement between knimble and . Importantly, the correlation between observers is significantly improved by the correction and the relative ranking of locations is also affected by the correction of stiffness values. This indicates a possible systematic difference in the Offset generated by different observers and at the different locations. A corresponding analysis is proposed in Fig 11 which reports average Offset values and STD for the three observer (left plot) or the four locations of two different participants (VO1 and VO2), middle and right plot. Mean values of Offset ranges between 0.2 and 0.8 mm, which is relatively large when compared to the reference distance of 1 mm of the Nimble. Observer 1 seems to exert lower forces, while O2 and O3 are similar. For both volunteers, the highest Offset was observed for the forehead. These values are significantly higher than for most other locations and this might explain the apparent high stiffness obtained with kcuto for the forehead (Fig 8). These results indicate that systematic differences of Offset values might be associated with measurements by one specific observer or at one specific body location.

thumbnail
Fig 11. Mean and standard deviation of Offset values in mm (measured by the Cutometer software).

Left: Mean Offset over all locations for each observer (O1, O2 and O3). Middle/Right: Mean Offset over all observers for each location (VF, FH, BH and LB) separated for two subjects (VO1 and VO2).

https://doi.org/10.1371/journal.pone.0201440.g011

4. Discussion

Based on extensive experience with aspiration experiments and, in particular, with the use of the Cutometer for skin characterization, the fundamental idea motivating this research was to realize a suction device allowing to minimize the contact force and the influence of patient movements during the measurements. As discussed by Bonaparte et al. [23], the contact force constitutes a relevant influential factor for skin suction experiments. Higher contact force leads to larger initial elevation levels and thus stiffer response. While previous investigations [24,25] analyze the applicability of stable stands to apply controlled contact forces, a novel light-weight suction device is introduced in this study for which such auxiliary measures are not required. The present work aimed to describe the characteristics of the Nimble but also to quantify its ability to characterize the mechanical response for different skin locations.

The quantitative assessment of the ability to differentiate among all four tested locations in two subjects was provided through the intraclass correlation coefficient, which was very large for the stiffness measured by the Nimble (knimble, ICCVO1 = 0.88, ICCVO2 = 0.94). These results confirm that the new technique might improve the reliability of skin characterization.

The mean stiffness determined with the Nimble does not correlate with the one obtained using the Cutometer. This discrepancy is attributed to the systematic effects associated with the contact force applied during Cutometer measurements. On the one hand, the data reported in Fig 5 convincingly demonstrate the effect of the contact force on the Cutometer reading. On the other hand, Fig 11 shows that the contact force (resulting in a corresponding Offset value) can differ systematically between body locations and observers.

Supported by corresponding finite element simulations, a correction procedure is proposed to compensate, to some extent, the influence of the contact force. Application of the correction procedure leads to a realignment of stiffness values obtained with the Cutometer and Nimble at 1 mm tissue elevation (knimble and kcuto). In addition, after correction, the scatter associated with Cutometer parameters was considerably reduced and the inter-observer Pearson’s coefficient improved for at levels similar to knimble. These results confirm both, the validity of the Nimble measurements, as well as the effectiveness of the correction. Since the Offset value is normally available in Cutometer measurements, it is recommended to apply the proposed correction in order to improve the reliability of skin characterization. It should be noted, however, that the reliability of the proposed correction depends on the mechanical behavior of the substrate to be characterized, as shown by the results on Ecoflex, S2 Table.

Another hypothesis analyzed in the present work was that a lower level of deformation would improve the ability of the suction experiment to distinguish between skin locations. In fact, due to the strong strain stiffening response of the dermal layers, at larger values modest changes of elevation occur despite significant increase in suction pressure, thus reducing the sensitivity of the measurements. This hypothesis was confirmed in that the kR0 parameters (corresponding to mean elevations of 1.5 mm) was less able to distinguish between skin locations as compared with knimble (1 mm), despite similar values in inter- and intra-observer variability. As a consequence, it is expected that application of lower values of maximum suction pressure (e.g. 200 mbar) would improve the results of a Cutometer based characterization.

The Cutometer is a sophisticated measuring system providing recording of Pressure - Elevation curves from each measurement. These data can be analyzed to extract information on the elastic and dissipative behavior of superficial tissues and characterize the time and history dependence of the mechanical response. This set of parameters represents a far richer outcome compared to the single parameter provided by the Nimble. As indicated in reference [35], additional information on the viscoelastic and viscoplastic properties of skin might be obtained with the Nimble through measurements at different suction rates (pcl for slow and fast suction) or through cyclic application of suction processes (evolution of pcl in consecutive measurements). The absence of a continuous reading of elevation associated with the imposed pressure rate remains, however, a limitation of the new device. This was the price to pay for an ultralight solution.

Despite this limitation, the Nimble might provide significant advantages for clinical applications. The results of the present study indicate that the level of intra- and inter-rater reliability of the Nimble is comparable to the one of the Cutometer. On the other hand, the ICC of the Nimble indicates a better ability to distinguish between locations (knimble: ICCVO1 = 0.88 and ICCVO2 = 0.94, kR0: ICCVO1 = 0.64 and ICCVO2 = 0.87, corrected : ICCVO1 = 0.58 and ICCVO2 = 0.83). While this improvement might provide advantages in terms of sensitivity and specificity of the procedure, there are other specific features of the Nimble, which could be important for its use in clinical applications: (i) the working principle of the Nimble is inherently safer than the Cutometer and other suction devices prescribing the maximum negative pressure applied on the tissue. In fact, risk of skin rupture is associated with excessive deformation and the displacement controlled measurement of the Nimble allows to directly limit the imposed deformation. The present study applied a maximum displacement of 1 mm but preliminary measurements performed in preparation of clinical studies on scar tissue indicated that the same level of reliability and repeatability can be achieved with a Nimble version limiting the maximum displacement to 0.5 mm, which is appropriate for less compliant tissues [6,42]. While the maximum negative pressure of the Cutometer can be adjusted for every measurement, it is not possible to predict the level of tissue displacement associated with a specific pressure level, and thus to ensure a priory that a certain level of deformation will not be exceeded. The pressure level limiting the displacement to 0.5 mm for a compliant tissue would lead to non-measurable (too small) deformations at a stiff skin location. (ii) The Nimble can be used as a disposable device: the suction probe, silicone tubes and air filters are inexpensive and can be replaced for every measurement. In this way, any cross contamination can be excluded. Other suction devices, such as the Cutometer, have to be disinfected after each application, but cannot be sterilized, thus preventing their use on subjects for which the presence of multi resistant bacteria cannot be excluded. (iii) The lightweight of the Nimble makes it less sensitive to patient’s movements during the measurement, which is an important advantage for clinical applications. As reported in supplementary information S4 Fig a comparison of the devices was performed in terms of their sensitivity to movements. The parameters measured by the Cutometer showed significant differences compared with static measurements, whereas the effect for the Nimble was insignificant.

The low variability and high interobserver correlation reported in section 3.3.2 indicate that reliable and objective readings might be obtained independently of the body location or observer. This might facilitate the application of the Nimble for a wide range of pathologies and patients, including pediatric patients. In line with these encouraging results, ethical permission was recently obtained for the realization of two explorative clinical studies: one on the early detection of skin modification associated with scleroderma, and the other for the assessment of scar maturation after skin transplants or spontaneously healed wounds on children affected by scars from large burn or scald wounds.

5. Conclusions

A novel technique is introduced for suction experiments on skin. The advantages of realizing an ultralight suction probe are demonstrated through a systematic comparison with the Cutometer, i.e. the suction device most widely applied for skin characterization. As an inherent uncertainty associated with a hand held device, the influence of the contact force is analyzed in detail and a correction procedure is proposed to reduce its effect on measurement outcomes. Cutometer and Nimble measurements at different body locations were consistent after correction confirming both, the reliability of the new device and the effectiveness of the correction procedure. As an important byproduct of the present study, the correction is recommended for future applications of the Cutometer. Reduced scatter and tissue interrogation at lower level of deformation led to a better ability of the Nimble to distinguish between the different skin locations, leading to a high value of the intraclass correlation coefficient and high inter-observer correlation.

As a consequence of the design simplifications leading to the ultralight suction probe, no continuous measurement of tissue elevation is provided by the Nimble. While this limits the possibility of characterizing the viscoelastic and viscoplastic properties of skin, the single parameter obtained with the Nimble test might be adequate for clinical applications. Its safety, simplicity, and reliability motivates future studies to quantify the diagnostic relevance of the Nimble measurement as well as its usefulness in the management of scar treatment in large wounds.

Supporting information

S1 Fig. Vertical measurement set-up.

(A) Schematic of the measurement set-up: a holder keeps the Cutometer/Nimble probe in place and the specimen is placed on top of the probe. (B) The specimen weight leads to an initial protrusion (Δ) of the tissue, penetrating the Cutometer/Nimble probe.

https://doi.org/10.1371/journal.pone.0201440.s001

(TIF)

S2 Fig. Finite element meshes for simulation of the Nimble and the Cutometer aspiration.

The instruments are considered as rigid bodies. The skin tissue consists of dermis (t = 1.2mm), subcutaneous tissue (t = 1.5 mm) and underlying muscle tissue (t = 7.3 mm); the Rubin-Bodner constitutive model[40] was implemented for each layer. The same model was used for simulation of measurements on elastomer and all layers had same properties in this case (Neo-Hookean hyperelastic, C10 = 0.01167 MPa). The contact interaction between rigid body and skin uses a friction coefficient which was varied in a range μ = 0.0 ÷ 1.0 in a parametric study. For this analysis, the initial contact force (and thus initial protrusion) was zero for all calculations. Quadrilateral axisymmetric elements were used and the mesh size was optimized to ensure adequate discretization of the regions characterized by large stress and strain gradients.

https://doi.org/10.1371/journal.pone.0201440.s002

(TIF)

S3 Fig. Suction measurements on elastomer.

Mean and standard deviation of closing pressure measured by Nimble (lightgrey) and Cutometer (grey). Significant difference is indicated with p < 0.001 between the devices for m1 and p < 0.01 for m2. Measurements were performed on elastomer (Ecoflex 0030)–specimen weight m1 = 1.5 g and m2 = 50 g.

https://doi.org/10.1371/journal.pone.0201440.s003

(TIF)

S4 Fig. Influence of patient movement on suction measurements.

Nimble (blue) and Cutometer (red) were tested for patient movement. Data shown are mean and standard deviations of n = 3 repetitions of measurements on human volar forearm in static condition (no movement), rotation (orbital shaker, horizontal circular motion with approximately 1.5 cm/sec), and vertical movement (vertical stage, up-down motion with approximately 5 cm/sec). Rotational and vertical movements led to significantly different closing pressure results in Cutometer measurements. Significance level p < 0.05 (*) and p < 0.001 (***).

https://doi.org/10.1371/journal.pone.0201440.s004

(TIF)

S1 Table. Rubin-Bodner[40] constitutive model parameters for each soft tissue considered in the FE model.

The values are based on those presented in [39] and adapted in order to match the present measurements on skin.

https://doi.org/10.1371/journal.pone.0201440.s005

(TIF)

S2 Table. Analysis of the proposed correction scheme for Cutometer measurements and calculations on elastomer model.

Results of FE calculations with enforced initial deformation (corresponding to contact forces of 50 g and 100 g) are compared for closing pressure values before and after corrections.

https://doi.org/10.1371/journal.pone.0201440.s006

(TIF)

S3 Table. Measurements on volunteers.

Reported are data of repeated measurements by three observers (O1, O2 and O3) on two subjects (VO1 and VO2) at four body locations (VF, FH, BH and LB). Data include the Offset Δ, the maximum elevation R0 and the closing pressure of Cutometer, and of Nimble.

https://doi.org/10.1371/journal.pone.0201440.s007

(TIF)

Acknowledgments

This project is supported by grants from “Helmut Horten Stiftung”.

This work was conducted as part of the SKINTEGRITY flagship project of University Medicine Zurich (http://www.hochschulmedizin.uzh.ch/en/projekte/skintegrity.html).

References

  1. 1. Piérard GE, Piérard S, Delvenne P, Piérard-Franchimont C. In vivo evaluation of the skin tensile strength by the suction method: pilot study coping with hysteresis and creep extension. ISRN Dermatol [Internet]. 2013 Dec 2;2013. Available from: http://downloads.hindawi.com/journals/isrn.dermatology/2013/841217.pdf
  2. 2. Hendriks FM, Brokken D, Oomens CWJ, Bader DL, Baaijens FPT. The relative contributions of different skin layers to the mechanical behavior of human skin in vivo using suction experiments. Med Eng Phys [Internet]. 2006 Apr 1;28(3):259–66. Available from: http://www.sciencedirect.com/science/article/pii/S1350453305001451 pmid:16099191
  3. 3. Hussain SH, Limthongkul B, Humphreys TR. The Biomechanical Properties of the Skin. Dermatologic Surg [Internet]. 2013 Feb 1;39(2):193–203. Available from: http://content.wkhealth.com/linkback/openurl?sid=WKPTLP:landingpage&an=00042728-201302000-00004
  4. 4. Sedky MM, Fawzy SM, Baki NA El, Eishi NH El, Bohy AEMM El. Systemic sclerosis: an ultrasonographic study of skin and subcutaneous tissue in relation to clinical findings. Ski Res Technol [Internet]. 2013 Feb;19(1):e78–84. Available from: http://doi.wiley.com/10.1111/j.1600-0846.2012.00612.x
  5. 5. Roufosse F., Bellefon LM de. Systemic Sclerosis (Scleroderma). Acta Clin Belg [Internet]. 2007;62(5). Available from: http://www.tandfonline.com/action/journalInformation?journalCode=yacb20 pmid:18229466
  6. 6. Bloemen MCT, van Leeuwen MCE, van Vucht NE, van Zuijlen PPM, Middelkoop E. Dermal Substitution in Acute Burns and Reconstructive Surgery: A 12-Year Follow-Up. Plast Reconstr Surg [Internet]. 2010 May;125(5):1450–9. Available from: http://content.wkhealth.com/linkback/openurl?sid=WKPTLP:landingpage&an=00006534-201005000-00017 pmid:20440164
  7. 7. Junker JPE, Philip J, Kiwanuka E, Hackl F, Caterson EJ, Eriksson E. Assessing quality of healing in skin: Review of available methods and devices. Wound Repair Regen [Internet]. 2014 May 1;22(S1):2–10. Available from: http://doi.wiley.com/10.1111/wrr.12162
  8. 8. Herndon DN. Total Burn Care. 4th ed. 2012. 783 p.
  9. 9. Piérard GE. EEMCO guidance to the in vivo assessment of tensile functional properties of the skin. Skin Pharmacol Physiol [Internet]. 1999 Dec 2;12(6):352–62. Available from: http://www.karger.com/Article/Fulltext/29897
  10. 10. Iivarinen JT, Korhonen RK, Julkunen P, Jurvelin JS. Experimental and computational analysis of soft tissue mechanical response under negative pressure in forearm. Ski Res Technol [Internet]. 2013 Feb 1;19(1):e356–65. Available from: http://doi.wiley.com/10.1111/j.1600-0846.2012.00652.x
  11. 11. Anthonissen M, Daly D, Fieuws S, Massagé P, Van Brussel M, Vranckx J, et al. Measurement of elasticity and transepidermal water loss rate of burn scars with the Dermalab. Burns. 2013;39(3):420–8. pmid:23000371
  12. 12. Gankande TU, Duke JM, Danielsen PL, DeJong HM, Wood FM, Wallace HJ. Reliability of scar assessments performed with an integrated skin testing device–The DermaLab Combo. Burns. 2014;40(8):1521–9. pmid:24630817
  13. 13. Dobrev HP. In vivo study of skin mechanical properties in patients with systemic sclerosis. J Am Acad Dermatol [Internet]. 1999 Dec 2;40(3):436–42. Available from: http://www.sciencedirect.com/science/article/pii/S0190962299704949 pmid:10071315
  14. 14. Dobrev H. In vivo study of skin mechanical properties in Raynaud’s phenomenon. Ski Res Technol [Internet]. 2007 Dec 2;13(1):91–4. Available from: http://onlinelibrary.wiley.com/doi/10.1111/j.1600-0846.2007.00197.x/full
  15. 15. Pierard GE, Nikkels-Tassoudji N, Pierard-Franchimont C. Influence of the test area on the mechanical properties of skin. Dermatology [Internet]. 1995 Dec 2;191(1):9–15. Available from: pmid:8589499
  16. 16. Courage + Khazaka electronic GmbH, “Courage + Khazaka electronic GmbH”, 09 May 2012 [Internet]. Available from: http://www.courage-khazaka.de/index.php/en/all-downloads/downloads-en/file/164-anleitcuto580e
  17. 17. Ryu HS, Joo YH, Kim SO, Park KC, Youn SW. Influence of age and regional differences on skin elasticity as measured by the Cutometer. Ski Res Technol [Internet]. 2008 Aug 1;14(3):354–8. Available from: http://doi.wiley.com/10.1111/j.1600-0846.2008.00302.x
  18. 18. Luebberding S, Krueger N, Kerscher M. Mechanical properties of human skin in vivo: a comparative evaluation in 300 men and women. Ski Res Technol [Internet]. 2014 Dec 2;20(2):127–35. Available from: http://onlinelibrary.wiley.com/doi/10.1111/srt.12094/full
  19. 19. Dobrev H. Use of Cutometer to assess epidermal hydration. Ski Res Technol [Internet]. 2000 Nov 1;6(4):239–44. Available from: http://doi.wiley.com/10.1034/j.1600-0846.2000.006004239.x
  20. 20. Enomoto DN., Mekkes JR, Bossuyt PM., Hoekzema R, Bos JD. Quantification of cutaneous sclerosis with a skin elasticity meter in patients with generalized scleroderma. J Am Acad Dermatol [Internet]. 1996 Sep 1;35(3):381–7. Available from: http://www.sciencedirect.com/science/article/pii/S0190962296906015
  21. 21. Draaijers LJ, Botman YAM, Tempelman FRH, Kreis RW, Middelkoop E, Van Zuijlen PPM. Skin elasticity meter or subjective evaluation in scars: A reliability assessment. Burns. 2004;30(2):109–14. pmid:15019116
  22. 22. Dobrev HP. A study of human skin mechanical properties by means of Cutometer. Folia Med [Internet]. 2002 Dec 2;44(3):5–10. Available from: https://www.researchgate.net/profile/Hristo_Dobrev/publication/10905168_A_study_of_human_skin_mechanical_properties_by_means_of_Cutometer/links/0deec51dc51505de89000000.pdf
  23. 23. Bonaparte JP, Ellis D, Chung J. The effect of probe to skin contact force on Cutometer MPA 580 measurements. J Med Eng Technol [Internet]. 2013 Dec 2;37(3):208–12. Available from: http://www.tandfonline.com/doi/abs/10.3109/03091902.2013.779325 pmid:23631522
  24. 24. Weickenmeier J, Jabareen M, Mazza E. Suction based mechanical characterization of superficial facial soft tissues. J Biomech [Internet]. 2015 Dec 2;48(16):4279–86. Available from: http://www.sciencedirect.com/science/article/pii/S0021929015005874 pmid:26584965
  25. 25. Pensalfini M, Weickenmeier J, Rominger M, Santoprete R, Distler O, Mazza E. Location-specific mechanical response and morphology of facial soft tissues. J Mech Behav Biomed Mater [Internet]. 2018 Feb 1;78:108–15. Available from: https://www.sciencedirect.com/science/article/pii/S1751616117304496 pmid:29149656
  26. 26. Mazza E, Nava A, Hahnloser D, Jochum W, Bajka M. The mechanical response of human liver and its relation to histology: an in vivo study. Med Image Anal [Internet]. 2007 Dec 7;11(6):663–72. Available from: http://www.sciencedirect.com/science/article/pii/S1361841507000667 pmid:17719834
  27. 27. Barbarino GG, Jabareen M, Mazza E. Experimental and numerical study on the mechanical behavior of the superficial layers of the face. Ski Res Technol [Internet]. 2011 Dec 2;17(4):434–44. Available from: http://onlinelibrary.wiley.com/doi/10.1111/j.1600-0846.2011.00515.x/full
  28. 28. Badir S, Bajka M, Mazza E. A novel procedure for the mechanical characterization of the uterine cervix during pregnancy. J Mech Behav Biomed Mater [Internet]. 2013 Dec 2;27:143–53. Available from: http://www.sciencedirect.com/science/article/pii/S1751616112003323 pmid:23274486
  29. 29. Hollenstein M, Bugnard G, Joos R, Kropf S, Villiger P, Mazza E. Towards laparoscopic tissue aspiration. Med Image Anal [Internet]. 2013 Dec 1;17(8):1037–45. Available from: http://www.ncbi.nlm.nih.gov/pubmed/23876854 pmid:23876854
  30. 30. Mazza E, Parra-Saavedra M, Bajka M, Gratacos E, Nicolaides K, Deprest J. In vivo assessment of the biomechanical properties of the uterine cervix in pregnancy. Prenat Diagn [Internet]. 2014 Jan 1;34(1):33–41. Available from: http://doi.wiley.com/10.1002/pd.4260 pmid:24155152
  31. 31. Draaijers LJ, Botman YAM, Tempelman FRH, Kreis RW, Middelkoop E, Van Zuijlen PPM. Skin elasticity meter or subjective evaluation in scars: A reliability assessment. Burns [Internet]. 2004;30(2):109–14. Available from: http://www.sciencedirect.com/science/article/pii/S030541790300247X pmid:15019116
  32. 32. Nedelec B, Correa JA, Rachelska G, Armour A, LaSalle L. Quantitative Measurement of Hypertrophic Scar: Intrarater Reliability, Sensitivity, and Specificity. J Burn Care Res [Internet]. 2008 May 1;29(3):489–500. Available from: https://academic.oup.com/jbcr/article/29/3/489-500/4598356 pmid:18388577
  33. 33. Mazza E, Nava A, Bauer M, Winter R, Bajka M, Holzapfel GA. Mechanical properties of the human uterine cervix: an in vivo study. Med Image Anal [Internet]. 2006 Dec 7;10(2):125–36. Available from: http://www.sciencedirect.com/science/article/pii/S1361841505000691 pmid:16143559
  34. 34. Badir S, Mazza E, Zimmermann R, Bajka M. Cervical softening occurs early in pregnancy: characterization of cervical stiffness in 100 healthy women using the aspiration technique. Prenat Diagn [Internet]. 2013 Dec 7;33(8):737–41. Available from: http://onlinelibrary.wiley.com/doi/10.1002/pd.4116/abstract pmid:23553612
  35. 35. Mazza E, Delgado F. Aspiration Device and Method for Determining Viscoelastic Properties of Biological Tissues and Synthetic Materials, EP16197195.7. 2016.
  36. 36. Schiavone P Promayon E, Payan Y. LASTIC: a Light Apiration device for in vivo Soft TIssue Characterization. Available from: http://www-timc.imag.fr/Emmanuel.Promayon/
  37. 37. Langton AK, Graham HK, McConnell JC, Sherratt MJ, Griffiths CEM, Watson REB. Organization of the dermal matrix impacts the biomechanical properties of skin. Br J Dermatol [Internet]. 2017 Sep;177(3):818–27. Available from: http://doi.wiley.com/10.1111/bjd.15353 pmid:28132410
  38. 38. Bernardi L, Hopf R, Ferrari A, Ehret AE, Mazza E. On the large strain deformation behavior of silicone-based elastomers for biomedical applications. Polym Test [Internet]. 2017;58:189–98. Available from: http://www.sciencedirect.com/science/article/pii/S0142941816312089
  39. 39. Weickenmeier J, Jabareen M. Elastic–viscoplastic modeling of soft biological tissues using a mixed finite element formulation based on the relative deformation gradient. Int j numer method biomed eng [Internet]. 2014 Dec 2;30(11):1238–62. Available from: http://onlinelibrary.wiley.com/doi/10.1002/cnm.2654/full pmid:24817477
  40. 40. Rubin MB, Bodner SR. A three-dimensional nonlinear model for dissipative response of soft tissue. Int J Solids Struct [Internet]. 2002 Dec 2;39(19):5081–99. Available from: http://www.sciencedirect.com/science/article/pii/S0020768302002378
  41. 41. Streiner DL, Norman GR, Cairney J. Health Measurement Scales: A Practical Guide to Their Development and Use—David L. Streiner, Geoffrey R. Norman, John Cairney—Google Books [Internet]. Fifth edit. Oxford University Press; 2015. 399 p. Available from: https://books.google.de/books?hl=de&lr=&id=JH3OBAAAQBAJ&oi=fnd&pg=PP1&dq=streiner+2015+reliability&ots=tj_Bwl6G8h&sig=q6K0z2YUz6HDFIOOlw849VB9Awk#v=onepage&q=streiner 2015 reliability&f=false
  42. 42. Jaspers MEH, Brouwer KM, van Trier AJM, Groot ML, Middelkoop E, van Zuijlen PPM. Effectiveness of Autologous Fat Grafting in Adherent Scars. Plast Reconstr Surg [Internet]. 2017 Jan;139(1):212–9. Available from: http://insights.ovid.com/crossref?an=00006534-201701000-00039 pmid:27632398