A safe, low-cost, easy-to-use 3D camera platform to assess risk of obstructed labor due to cephalopelvic disproportion

  • Rudolph L. Gleason Jr. ,

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing

    rudy.gleason@me.gatech.edu

    Affiliations The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, United States of America, The Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology, Atlanta, GA, United States of America, Because of Kennedy, Inc., Acworth, GA, United States of America

  • Mahlet Yigeremu,

    Roles Conceptualization, Investigation, Methodology, Project administration, Supervision, Writing – review & editing

    Affiliation Department of Obstetrics & Gynecology, Addis Ababa University, Addis Ababa, Ethiopia

  • Tequam Debebe,

    Roles Conceptualization, Data curation, Formal analysis, Methodology, Supervision, Writing – review & editing

    Affiliation Department of Radiology, Addis Ababa University, Addis Ababa, Ethiopia

  • Sisay Teklu,

    Roles Conceptualization, Data curation, Project administration, Writing – review & editing

    Affiliation Department of Obstetrics & Gynecology, Addis Ababa University, Addis Ababa, Ethiopia

  • Daniel Zewdeneh,

    Roles Data curation, Formal analysis, Writing – review & editing

    Affiliation Department of Radiology, Addis Ababa University, Addis Ababa, Ethiopia

  • Michael Weiler,

    Roles Conceptualization, Data curation, Formal analysis

    Affiliation LymphaTech, Inc., Atlanta, GA, United States of America

  • Nate Frank,

    Roles Data curation, Methodology

    Affiliation LymphaTech, Inc., Atlanta, GA, United States of America

  • Lorenzo Tolentino,

    Roles Data curation, Formal analysis, Methodology, Project administration, Writing – review & editing

    Affiliation Because of Kennedy, Inc., Acworth, GA, United States of America

  • Shehab Attia,

    Roles Data curation, Methodology

    Affiliation The Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology, Atlanta, GA, United States of America

  • J. Brandon Dixon,

    Roles Conceptualization, Funding acquisition

    Affiliation The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, United States of America

  • Catherine Kwon,

    Roles Data curation

    Affiliation The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, United States of America

  • Anastassia Pokutta-Paskaleva,

    Roles Conceptualization, Funding acquisition

    Affiliation The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, United States of America

  • Katie A. Gleason

    Roles Data curation, Writing – review & editing

    Affiliation Because of Kennedy, Inc., Acworth, GA, United States of America

Abstract

Cephalopelvic disproportion (CPD)-related obstructed labor is accountable for 3–8% of the maternal deaths worldwide. The consequence of CPD-related obstructive labor in the absence of a Caesarian section (C/S) is often maternal or perinatal mortality or morbidity to the mother and/or the infant. Accurate and timely referral of at-risk mothers to health facilities where C/S is a delivery option could reduce maternal mortality in the developing world. The goal of this work was to develop and test the feasibility of a safe, low-cost, easy-to-use, portable tool, using a Microsoft Kinect 3D camera, to identify women at risk for obstructed labor due to CPD. Magnetic resonance imaging (MRI) scans, 3D camera imaging, anthropometry and clinical pelvimetry were collected and analyzed from women 18–40 years of age, at gestational age ≥36+0 weeks with previous C/S due to CPD (n = 43), previous uncomplicated vaginal deliveries (n = 96), and no previous obstetric history (n = 148) from Addis Ababa, Ethiopia. Novel and published CPD risk scores based on anthropometry, clinical pelvimetry, MRI, and Kinect measurements were compared. Significant differences were observed in most anthropometry, clinical pelvimetry, MRI and Kinect measurements between women delivering via CPD-related C/S versus those delivering vaginally. The area under the receiver-operator curve from novel CPD risk scores base on MRI-, Kinect-, and anthropometric-features outperformed novel CPD risk scores based on clinical pelvimetry and previously published indices for CPD risk calculated from these data; e.g., pelvic inlet area, height, and fetal-pelvic index. This work demonstrates the feasibility of a 3D camera-based platform for assessing CPD risk as a novel, safe, scalable approach to better predict risk of CPD in Ethiopia and warrants the need for further blinded, prospective studies to refine and validate the proposed CPD risk scores, which are required before this method can be applied clinically.

Introduction

Obstructed labor accounts for 3–8% of all maternal deaths worldwide and, in many countries, is almost as prevalent today as it was 30 years ago [13]. Cephalopelvic disproportion (CPD) is an inadequate size of the maternal pelvis, compared to the fetal head, which prevents the fetus from passing through the pelvic cavity during delivery causing obstructed labor [4]. CPD-related obstructed labor requires delivery via Caesarean section (C/S). In the developed world, even if an Obstetrician suspects that a first-time pregnancy is at risk of CPD-related obstructed labor, the clinician will often prescribe trial of labor and, if the labor fails to progress, an emergency C/S is performed. Given the nearly universal access to C/S in much of the developed world, the utility of diagnosing CPD risk has been diminished because C/S capabilities are almost always nearby.

In rural Ethiopia, and many low-resource settings around the world, it may require hours or days of travel for mothers to reach facilities with the infrastructure and expertise to perform a C/S. The consequence of CPD-related obstructive labor (in the absence of C/S) is often maternal and/or perinatal mortality or long-term morbidity to the mother and/or the infant. CPD is especially prevalent in regions like Ethiopia, where girls are small in stature, grow up malnourished, marry at a young age, and become pregnant before the pelvis is fully grown [5, 6]. Maternal and perinatal mortality in Ethiopia are among the highest in the world [7]; up to 22% of maternal deaths in Ethiopia have been attributed to CPD-related obstructed labor [79]. Further, maternal morbidity (e.g., obstetric fistulas) from CPD is widespread in Ethiopia and carries significant stigma and social exclusion [10]. In Ethiopia, 1 out of every 250 women report having an obstetric fistula from obstructed labor [11]. Taken together, there is a pressing need to identify safe, low-cost methods for timely and accurate assessment of risk of obstructed labor from CPD in low resource settings [12].

Mengert summarized well that “there are five components of CPD [related obstructed labor]: (1) size and shape of the bony pelvis, (2) size of the fetus, (3) force exerted by the uterus, (4) moldability of the head, and (5) fetal presentation and position” [13]. Over the last century, many have undertaken the aim of developing methods to accurately assess the risk of CPD, prior to labor, based on external and internal physical measurements to assess the size of the pelvis, radiological techniques to assess and compare the pelvic capacity and the fetus size, or anthropometry; the latter is based on the assumption that the size of the pelvis (and perhaps the size of the fetus) correlate, to some degree, with other more clinically tractable measurements of maternal size, such as height or foot size.

Radiological pelvimetry, using X-ray, magnetic resonance imaging (MRI), or computed tomography (CT), can provide accurate quantification of the pelvic planes and fetal size, from which a number of CPD risk scores have been proposed [1320]; however, the accuracy of radiological pelvimetry at predicting CPD-related obstructive labor is controversial and lacks testing via rigorous randomized trials [14, 21]. Further, MRI and CT are cost prohibitive in low-resource settings and X-ray poses an unwanted health risk from radiation exposure to the fetus and may represent a significant barrier to widespread scale-up in the developing world. Assessment of fetal size is routinely done via ultrasound imaging; however, ultrasound technology and expertise are also limited in most low-resource settings.

Anthropometric features such as maternal height [2246], foot length [25, 35, 42, 45, 47], shoulder diameter [35], lower limb length [42], BMI [32, 4851], maternal head circumference-to-height ratio [32], and external measurements to assess hip and pelvis dimensions [28, 34, 35, 45, 52] have been proposed to assess risk of CPD. Often, their diagnostic capabilities rely heavily on the accuracy of the measurements, which contain significant inter- and intra-user variability [53, 54]. CPD risk assessment based on anthropometry have shown low sensitivity and specificity [3235, 45, 52, 55, 56]. Further, clinical pelvimetry to assess the size of the pelvic cavity by means of the systematic vaginal palpation of specific bony landmarks has shown marginal success and requires skilled personnel and training [57, 58].

With the recent explosion in 3D camera technology, driven in large part by video gaming and virtual reality applications, the resolution, ease-of-use, and widespread availability of 3D cameras, including smartphone-based 3D cameras, are expected to increase dramatically in the years ahead, while cost is expected to continue to fall. The Microsoft Kinect 3D camera, together with innovative software, can generate a 3D point-cloud of the objects in the field of view and uses skeletal mapping algorithms to approximate the location of key joints. Given the required cost, infrastructure, and expertise required for MRI pelvimetry and the inter- and intra-user variability of anthropometry, we hypothesize that a 3D camera-based platform for assessing CPD risk could provide a novel, scalable approach to better predict risk of CPD in the rural developing world. The goal of this paper is to develop and test the feasibility of using a low-cost, easy-to-use, portable technology using a Microsoft Kinect 3D camera to quantify risk of CPD-related obstructed labor.

Materials and methods

Participant enrollment and baseline data

This study was approved by the Institutional Review Board Committees at Addis Ababa University, College of Health Sciences (Protocol number: 054/15/gyn, approved 2/5/2016) and the Georgia Institute of Technology (Protocol number: H15314, approved 2/24/16). Women 18 to 40 years of age at gestational age 36 weeks and 0 days (denoted 36+0) or above with vertex presentation and singleton pregnancy that fit into one of the follow groups were recruited between April and November 2016 from Tikur Anbessa (Black Lion) Specialized Referral Hospital in Addis Ababa, Ethiopia: Group 1: High risk, (n = 43): Women with a previous Caesarian section due to CPD and no previous vaginal delivery; Group 2: Low risk (n = 96): Women with at least one previous uncomplicated vaginal delivery and no previous Caesarian sections; Group 3: Unknown risk, (n = 148): Primigravida women. Women in labor, with non-vertex presentation, with Diabetes mellitus, preeclampsia or other hypertensive disease, women that had a previous C/S for reason other than CPD, women that did not intend to try labor (i.e., had a planned C/S), and women with contraindication to MRI were excluded. Written informed consent was obtained from the participant prior to enrollment. Based on both obstetric history and outcome of the current pregnancy, we categorized subjects into two groups: CPD or VD. The CPD group included all women with at least one CPD-related C/S; namely, all women from Group 1 and all women that delivered via CPD-related C/S from Group 2 and Group 3. The VD group included all women that delivered vaginally, with no current or previous C/S; namely, vaginal deliveries from Group 2 and Group 3. We excluded non-CPD-related C/S from Groups 2 and Group 3.

Participant information

Age, gestational age, systolic and diastolic blood pressure, and weight were recorded from the hospital card of each participant. The number of previous live births and how each previous live birth was delivered (vaginally/instrumentally or C/S) were recorded. For subjects that have had a previous C/S, the Obstetrician and Gynecologist reviewed the hospital card describing the previous deliveries to assess the reason for the previous C/S and, if necessary, confirmed the reason with the participant.

Anthropometry

Body height was measured with a stadiometer (Infant/Child/Adult ShorrBoard®, Weights and Measures, LLC, USA). Head circumference was measured as the circumference at the level of the mid-forehead and 1.5 cm above the ear [32]. Bisacromial (shoulder) width measured as the distance between the lateral extremities of the acromion processes [35]. Shoulder height was measured from the top of the acromion process to the floor. Waist circumference was measured at the navel, at the level below the lowest rib, and at the narrowest diameter of the back. Waist height was measured as the distance from the naval to the floor. Hip circumference was measured at the widest part of the hips and hip height was measured from widest part of the hips to the floor. Foot length was measured from back of heel to the tip of the big toe [35]. External conjugate was measured with an anthropometer (Sammons Preston, Large) as the distance from the point below the lowest lumbar vertebra to the upper edge of the pubic symphysis [45]. All measurements, except for height and external conjugate, were made with a 200 cm non-stretching anthropometry tape measure. All measurements were made by nurses trained by the research team and under the direct observation of the research team. All measurements were recorded to the nearest millimeter. Ratios and differences between the several anthropometric features were calculated.

Clinical pelvimetry

Clinical pelvimetry was performed by an Obstetrician and Gynecologist (authors MY and ST) following published measurement definitions [5860]. The diagonal conjugate was scored as ‘reachable’ or ‘unreachable’. The pelvic side walls were scored as ‘divergent’, ‘straight’, or ‘convergent’. The sacrospinus ligament was scored as accommodating two fingers (‘yes’ or ‘no’). The ischial spines were scored as ‘prominent’ or ‘not prominent’. The subpubic angle was scored as accommodating two fingers (‘yes’ or ‘no’). The intertuberous ligament was scored as accommodating the knuckles of a fist (‘yes’ or ‘no’). The fetal head station (the position of the fetus' head in relation to the distance from the ischial spines) was scored as -3, -2, -1, 0, 1, 2, or 3 cm from the ischial spine; negative numbers indicate that the head is further inside than the ischial spines and positive numbers show that the head is below the level of the ischial spines.

Radiological pelvimetry by MRI

Radiological pelvimetry was performed following published methods [18, 6165]. MRI was performed using a 1.5-Telsa MRI (Achieva Philips) at Tikur Anbessa Specialized Referral Hospital while the patient was in a supine position, at a 30-degree angle, with their head up. A fast-spin echo (FSE) proton density-weighted pulse sequence was used, which is the most commonly used fetal imaging modality of MRI [66]. A T2-Turbo Spin Echo sequence was used with 1200 msec repetition time, 80 msec echo time, 80–102 cm field of view, 276-292-by-316-332 matrix, 1–2 averages, echo number 1, 5 mm section thickness, and 6.0–6.6 mm spacing between slices.

The pelvic inlet (superior strait) is bounded posteriorly by the promontory and alae of the sacrum, laterally by the linea terminalis, and anteriorly by the horizontal rami of the pubic bones and pubis symphysis. The pelvic inlet anterior-posterior diameter (APD), also called the obstetric conjugate, was measured from the anterior cortical surface of the sacrovertebral angel (promontory) to the closest point on the convex posterior-superior aspect of the pubis symphysis. The diagonal conjugate was measured from the lower border of the pubic symphysis to the sacral promontory. The pelvic inlet transverse diameter (TD) was measured as the greatest linear width of the linea terminalis on either side at the level of the sacral promontory. The pelvic inlet oblique diameters were measured from one of the sacroiliac synchondroses to the iliopectineal eminence on the opposite side of the pelvis and are designated right and left, according to whether they originate at the right or left sacroiliac synchondrosis.

The midpelvis is at the level of the ischial spines and is the plane of smallest pelvic dimensions. The mid pelvis APD was measured from the posterior-inferior aspect of the pubis symphysis to the anterior cortical surface of the sacrococcygeal junction. The mid pelvis TD (also called the interspinous diameter) was measured as the distance between the inschial spines. The pelvic outlet APD was measured from the lower margin of the symphysis pubis to the tip of the sacrum. The pelvic outlet TD (also known as the intertuberous diameter) was measured between the inner edges of the ischial tuberosities. The pelvic outlet posterior sagittal diameter (PSD) extends from the tip of the sacrum to a right-angle intersection with a line between the ischial tuberosities. The vertical Michaelis sacral rhomboid dimension was measured as the distance between the spinus process tip of the fifth lumbar vertebra to fifth sacral vertebra; the transverse Michaelis sacral rhomboid dimension was measured as the distance between the superior posterior iliac spines. The femoral head separation was measured as the distance between the medial edge of the right and left femoral heads. The intertrochanteric distance was measured as the distance between the lesser trochanter of the left and right femur.

The fetal head biparietal diameter (BPD) was measured as the transverse diameter on an axial plane that traverses the thalami, and cavum septum pellucidum, which extends from one parietal boss to the other. The occipitofrontal diameter (OFD) was measured as the anterior-posterior diameter from just above the root of the nose to the most prominent portion of the occipital bone. The occipitomental diameter (OMD) was measured from the chin to the most prominent portion of the occiput. The suboccipitobregmatic diameter (SOBD) was measured from the middle of the large fontanel to the undersurface of the occipital bone just where it joins the neck. The fetal abdomen APD and TD were measured from a transverse section through the upper abdomen that demonstrate the fetal stomach, umbilical vein, and portal sinus.

Pelvimetry and fetal head measurements were obtained by using midsagittal, transverse, and coronal oblique planes with the scan range covering the entire placenta to below lesser trochanters and oblique section through pubic symphysis and sacral promontory. The pelvic inlet, mid-pelvis, and pelvic outlet APD, the pelvic PSD, and the vertical sacral rhomboid dimension were measured from the midsagittal plane. The pelvic inlet oblique diameters and inter-femoral trochanteric distance were measured in the oblique coronal scan. The pelvic inlet, mid-pelvis, and pelvic outlet TD’s, the transverse sacral rhomboid dimension, and the inter-femoral head distance were measured in the transverse plane. The fetal measurements were made on the planes showing the correct reference structures. Fetal head BPD and OFD were measured in the trans-thalamic axial section of the fetal head. Fetal head OMD and SOBD were measured in the midsagittal section. Fetal abdomen APD and TD were measured from the transverse section of the fetal abdomen at the level of the liver where the intrahepatic umbilical vein and stomach is visible. All measurements were made by a trained radiologist. Measurements were taken using RadiAnt DICOM Viewer version 2.2 and recorded to the nearest millimeter.

The pelvic inlet, mid-pelvis, and pelvic outlet circumferences, areas, and pelvic capacities were estimated as C = π(TD + APD)/2, A = π(TD + APD), and V = πd3/6, respectively, where d is the smaller of the TD and APD in the respective plane. The fetal head C and A were also quantified at the level of the BPD and OFD and fetal abdomen C and A were calculated; the fetal head volume was estimated as V = π(BPD)(SOBD)(OFD + OMD)/12. Ratios between the specific diameters, circumferences, areas, and volumes of the fetus, with those of the pelvic inlet and mid-pelvis were calculated. The fetal-pelvic index (FPI) was calculated by adding the two highest circumference differences between the fetal head or abdomen circumference minus the inlet or mid-pelvis circumference [19]. The index of Abitbol et al., was calculated as the difference between the fetal BPD and the smaller of the inlet APD or mid-pelvis TD [16]. Following Friedman and Taylor and Sporri et al., the difference between the fetal head volume and the inlet, the mid-pelvis, and the smaller of the inlet and mid-pelvis capacity were calculated [17, 18].

Kinect scanning

A Kinect v2 sensor (Model #1520, Microsoft, Inc.) with Kinect for Windows SDK 2.0 software was employed to collect clips of the participants in the anterior, posterior, and left and right lateral views, while in a standing position with their legs positioned ~0.5 meters apart and their arms raised at shoulder height in the lateral direction (Fig 1A). The Kinect software algorithms determine the approximate locations of body parts and approximates the location of 24 key joints from the 3D point-cloud. From this skeletal map, custom algorithms to calculate shoulder width, shoulder height, hip width, hip height, hip-to-shoulder length, head height, face height, neck height, shoulder-to-head length, right and left femur length, and right and left leg length from the anterior scan were developed.

thumbnail
Fig 1. Illustrative kinect image and 3D reconstruction.

(a) Illustrative image, point cloud, and skeletal mapping overlay from an anterior view with joint locations from skeletal mapping shown as red dots. Note that the color of the face was darkened to protect the identity of the subject. (b) Illustration of the resulting three-dimensional model of the torso region obtained by stitching together the surface maps from four orthogonal views (anterior, left lateral, posterior, and right lateral). The red and blue dots represent the boundary points that were matched across adjacent scans.

https://doi.org/10.1371/journal.pone.0203865.g001

Fully-automated algorithms to detection of anthropometric features from anterior and posterior 3D point clouds were developed. Each algorithm employed statistical analysis and gradient analysis on the 3D point cloud data to identify the feature. From the posterior 3D point-clouds, algorithms determined height (distance from the top of the head to the floor), shoulder height (distance from the top of the shoulder directly above the arm pit to the floor), hip height (distance from the widest part of the hip to the floor), torso length (length from v-point at the arm pit to the v-point where the legs meet the torso), hip-to-shoulder length (difference between the shoulder height and hip height), hip diameter (the widest diameter in the pelvic area), waist diameter (the smallest diameter of the torso), shoulder diameter (the distance between the left to right arm pit), and five diameters equally spaced between the hip and shoulders (denoted diameter 1, 2, 3, 4, and 5). From the anterior 3D point-clouds we determined head length (distance from the top of the head to the bottom of the chin), head diameter (widest diameter of the head), fundal height (measured from the pubic symphysis to uppermost point of the fundus), and belly height (the distance of the fundus from the plane made by the pubis and the crest of belly). All landmarks were consistent after manual verification by visually inspection of the point cloud, with the features overlaid on each scan.

In addition to assessing measurements from point-clouds from individual scans, algorithms were developed to ‘stitch’ the point-clouds of the torso together from the four orthogonal views into an accurate 3D model of that participant (Fig 1), using an iterative closest point (ICP) algorithm. The ICP algorithm is well-studied and commonly used in image processing to match features between adjacent frames in two or more images to stitch them together. The algorithm relies on two main features to successfully stitch two images, or in our case, point clouds: (i) initial transformations of the point clouds that will be stitched; and, (ii) matching surface features or reference points on the adjacent point clouds. An initial transformation was achieved by translating and rotating the left, right lateral scans and posterior scans to be in the correct anatomical direction, relative to the anterior scan. Challenges arose in matching surface features and creating reference points using scans that are adjacently orthogonal due to the lack of similar features between adjacent scans; e.g., the anterior torso scan may have a curved belly and a symmetric shape; however, the left torso lateral scan lacks the symmetry of its partner and the Kinect may not capture the full belly from the sides, hence there would not be an adequate number of references for the ICP algorithm to match the two scans. To overcome this issue, we used the geometrical shape of the edges of the scans in 2D and then 3D as reference points. The edges, such as the curvature of the anterior scan from the shoulder, along the hips and down to the thigh, matched the curvature of the left and right lateral scans along similar sides, providing adequate information for the ICP algorithm to reconstruct the shape of the subject.

Pregnancy outcomes

After delivery, maternal and perinatal mortality, birth date, method of delivery, reason for C/S, and baby weight, length, head circumference, and Apgar score were recorded. The pregnancy outcome was recorded as “Vaginal Delivery (VD)” or “Caesarean Section (C/S)” by review of the hospital card after delivery. For delivery via C/S, the hospital card was reviewed and, if necessary discussion with the attending physician was conducted, to assess the reason that C/S was prescribed. Reason for C/S were recorded in the following categories: CPD after prolonged labor, elective C/S due to previous CPD (Group 1 only), abnormal presentation, fetal distress, placental concerns, failed induction, or other.

Risk score framework

We performed binomial logistic regression to develop predictive risk scores based on either (i) anthropometry, (ii) clinical pelvimetry, (iii) MRI, or (iv) Kinect measurements. The risk score was defined as (1) where j = A,P,K,M denotes risk scores based on anthropometry, clinical pelvimetry, Kinect 3D camera, or MRI measurements, respectively, and βo and βi are model parameters. We considered two different models for ri; namely, (2) where xi are the values of the features fi (e.g., fi = height, xi = 155 cm) measured with the respective tool, and are the normal Gaussian probability distribution functions of the feature fi for women who delivered via CPD-related C/S or uncomplicated vaginal delivery, respectively, and are means and and standard deviations for women who delivered via CPD-related C/S or via vaginal delivery, respectively. To illustrate, , , and ri are shown for fi = height (Fig 2). We denote models using Eq (2)1 as r1 and those using Eq (2)2 as r2. Since the independent variables for clinical pelvimetry are categorical, r2 was not considered for clinical pelvimetry models.

thumbnail
Fig 2. Illustrative risk model, ri.

Histograms and Gaussian distribution functions of feature fi = height for all CPD cases (shaded histogram bars and solid line NCPD) and all uncomplicated vaginal delivery cases (blue histogram bars and dashed line NVD) and from Eq (2)2 is shown in the dotted line. Note that the histograms and distribution functions refer to the vertical axis to the left and ri refers to the vertical axis to the right.

https://doi.org/10.1371/journal.pone.0203865.g002

Risk score assessment

Best fit models were determined using the glmfit subroutine in MATLAB (MathWorks) for possible permutations of feature sets and parameters were calculated for each value of k for all subjects, where is the total number of features considered for inclusion in the model and k is the actual number of features included in the model. For each of the np parameter sets, for each value of k, the area under the ROC curve, denoted AUC, was calculated with the perfcurve MATLAB subroutine and the best model was defined as the feature set and model parameters with the greatest AUC, denoted AUCbest. For the pelvimetry- and anthropometry-based models, the total number of features was = 7 and 21, respectively; for these models, we calculated the best fit parameters (βo and βi) for all np possible permutations of feature sets.

For Kinect and MRI models, = 54 and 49, respectively; for these models, np was far too large to compute all possible permutations of parameter sets. Therefore, we devised a parameter elimination routine for cases when was above 30-million permutations. Features were eliminated based on the percent of occurrences that each feature appears in the top 1,000 models from the models with lower feature counts k (forward elimination) and independently from models with a higher feature count (backward elimination); parameters with the lowest occurrences in models with lower (or higher) feature count were eliminated, thereby reducing for that parameter count k and yielding a manageable np that was <10-million. Based on the forward and backward elimination routines, all evaluated models were ranked by AUC and the percentage of time that each feature appears in the top 10,000 models was calculated. Another ~10-million permutations were evaluated, wherein the features that appeared most often in the top models were fixed (i.e., not permutated) and the features that appeared less often were permutated. The models were again ranked by AUC and AUCbest was the feature set and parameters associated with the model with the highest AUC from all permutations considered.

Since all of the data were used to identify the best model, these models were over-fitted (i.e., optimistic). To evaluate the optimism, O, and to assess the predictive capabilities of the best models, we adopted the approach of Harrell et al. [67]. Briefly, nB = 2,000 non-parametric bootstrap samples were generated from the original data set. For each bootstrap sample set, the top feature sets were selected and the model parameters (βo,b and βi,b) were identified and the AUC for each bootstrap sample, denoted AUCb,boot, was calculated. The feature set and fitted model parameters (βo,b and βi,b) were then applied to the original data set and the AUC, denoted AUCb,orig, was calculated. The optimism was calculated as (3) The adjusted AUC, denoted AUCadj, which represents the expected AUC if the model was applied to a new data set (i.e., the predictive capability), was calculated as (4) For each number of features, k, O and AUCadj were calculated for the top 200 feature sets and parameters with the highest AUCbest. The most predictive model was determined as the model with the highest AUCadj.

A 2x2 contingency table and 4-point risk score (very high, high, moderate, and mild risk) table were determined for the model with the highest AUCadj. The cut-off risk score for the 2x2 contingency table and for the high-moderate cut-off was set as the value in which the true positive rate (TPR, a.k.a., sensitivity) was equal to the true negative rate (TNR, a.k.a., the specificity). The very high-high cut-off was the highest risk value where the false positive rate (FPR) was ≤10%. We define the detection rate as the percentage of CPD cases that score as very high risk. The moderate-mild cut-off was the highest risk value where the false negative rate (FNR) was ≤10%. The likelihood ratios for CPD and VD are defined as and

Sample size determination and statistical analysis

Given that use of the Kinect 3D camera is a new strategy to predict CPD risk there are no data in the literature that can accurately estimate expected standard deviations or correlations between experimental Kinect-based measures across groups. Indeed, the purpose of this study was to generate the first data on these measures to assess the feasibility of the Kinect-based approach, compared to MRI, anthropometry, and clinical pelvimetry. The sample sizes for each group were selected to (i) recruit 50 women considered as high risk for CPD and (ii) to quantify the prevalence of CPD in this Ethiopian population in women with no history of CPD and in women with no obstetric history; the latter will inform future sample size determination. The target sample size of 50 cases was based on previous studies in literature, aimed at assessing risk of CPD based on anthropometry [16]. We considered all women with a history of CPD as high-risk cases. We estimated the prevalence of CPD in primigravida women to be 3–5%; thus, a sample size of n = 148 was expected to yield between 4 and 7 CPD cases. We estimated the prevalence in women with previous uncomplicated vaginal deliveries to be 2–4%; thus, a sample size of n = 96 was expected to yield between 2 and 4 CPD cases. Together, these sample sizes were expected to yield 50 high-risk subjects and should yield an adequate estimate of the prevalence of CPD in primigravida and multigravida women with no history of CPD.

For continuous variables, a two-sample t-test was performed and for categorical variables, χ2 analysis was performed to determine statistical significance across groups (p < 0.05); for categorical variables with more than two categories, a pairwise Marascuillo procedure was performed to compare differences in proportions across groups for each category (p < 0.05). The standard error and 95% confidence interval with the nonparametric assumption was determined for the AUC of the ROC curves and statistical significance (p<0.05) against the null hypothesis that the AUC = 0.50 was reported. Statistical analysis was performed in MATLAB 2018a (MathWorks) and IBM® SPSS® Statistics, version 24.

Results

Pregnancy outcomes

The incidence of CPD related C/S was 79%, 4%, and 4%, among women with previous CPD-related C/S, women with a successful previous vaginal delivery, and primigravida women, respectively (Fig 3). Of the 45 non-CPD related C/S (across Groups 1, 2, and 3), 21 were due to fetal distress, 8 were due to failed induction, 3 were due to abnormal presentation, 1 was due to placental issues, and the remaining 12 were due to other reasons. All women in Group 2 and Group 3 tried labor prior to C/S; however, some women in Group 1 underwent a scheduled C/S, without trial of labor. No maternal mortality was observed across all 287 subjects. Perinatal mortality occurred in three deliveries, all following vaginal deliveries in primigravida women (Group 3).

thumbnail
Fig 3. Pregnancy outcomes.

Pregnancy outcomes were categorized as CPD-related C/S, non-CPD-related C/S, or VD. Based on both obstetric history and outcome of the current pregnancy, subjects were categorized into two groups: CPD (shaded red) or VD (shaded green). Exclusions are shaded in blue.

https://doi.org/10.1371/journal.pone.0203865.g003

Measurement differences across CPD and vaginal delivery outcomes

Participant age and 1-minute APGAR scores were higher and parity lower in the CPD group compared to the VD group (Table 1). Total height, shoulder height, waist height, hip height, shoulder width, and foot length were significantly lower and head-circumference-to-height ratio and BMI were significantly higher in the CPD group compared to the VD group. Inlet APD, TD, right and left OD, circumference, cross-sectional area, and capacity, diagonal conjugate, midpelvic APD, circumference, and cross-sectional area, and outlet APD, circumference, cross-sectional area, and capacity were all significantly smaller in the CPD group compared to the VD group. No differences were observed in any fetal head or fetal abdomen measurements. BPD to smallest inlet diameter ratio, fetal head circumference- and fetal abdomen circumference to inlet circumference and to mid-pelvis circumference ratios, and fetal head- and fetal abdomen- to inlet area and mid-pelvis area ratios were all higher in the CPD compared to the VD groups. The indices proposed by Friedman and Taylor (on the pelvic inlet), Sporri et al., and FPI were significantly different in the CPD group compared to the VD group. The vertical and transverse sacral rhomboid dimension and the distance between the right and left femoral heads was significantly lower in the CPD group compared to the VD group.

thumbnail
Table 1. Anthropometry, clinical pelvimetry, MRI, and kinect measurements and novel risk scores.

Model parameters (βo and βi) are for the model specified in Eq (1), with either Eq (2)1 or (2)2. Eq (2)1 was used for the clinical pelvimetry scores since these data were categorical. Eq (2)1 outperformed Eq (2)2 for the anthropometry risk score, whereas Eq (2)2 outperformed Eq (2)1 for the MRI and Kinect risk scores. The red bold-italic font denotes features that are statistically different (p<0.05) between the CPD and VD groups. Note: CPD = cephalopelvic disproportion, VD = vaginal delivery, cm = centimeters, kg = kilograms, BMI = body mass index, APD = anterior-posterior diameter, TD = transverse diameter, AC = abdomen circumference, OD = oblique diameter, PSD = posterior sagittal diameter, BPD = biparental diameter, OFD = occipitofrontal diameter, OMD = occipitomental diameter, SOBD = suboccipitobregmatic diameter, HC = head circumference, and AC = abdomen circumference.

https://doi.org/10.1371/journal.pone.0203865.t001

Clinical pelvimetry showed a higher percentage of women in the CPD group had diagonal conjugate scored as reached, pelvic side walls scored as convergent, sacrospinous ligament scored as not accommodating two fingers, ischial spines scored as prominent, subpubic angle scored as not accommodating two fingers, and intertuberous ligament scored as not accommodating the knuckles of the fist. Kinect skeletal mapping measurements for head, face, neck, shoulder, hip, and right and left leg heights, right femur length, and shoulder-to-head length were significantly smaller in the CPD group compared to the VD group. Kinect 3D camera measurements for height, shoulder height, hip height, head length, hip-to-shoulder length, hip diameter, diameter 1, and head diameter were significantly lower in the CPD group compared to the VD group. Belly Height-to-Hip Diameter (3D), Belly Height-to-Hip Diameter (3D), Torso length-to-Hip height (3D), Height-to-Belly height (3D), and Head height (SM)-to-Belly height (3D) all differed between CPD and VD groups.

Anthropometry and CPD risk

Of all the individual anthropometric features, height alone showed the greatest predictive power (Table 2). Height alone identified 36% of the CPD cases as very high risk and 40% of the vaginal delivery cases as mild risk with likelihood ratios of 4.4 and 4.2 for CPD and VD, respectively (Table 3); however, the remaining CPD cases and vaginal deliveries, for women of height between 151–160 cm showed poor likelihood ratios. Therefore, height alone is a poor predictor of CPD risk for women of medium stature.

thumbnail
Table 2. Area under the receiver operator characteristic (ROC) curve.

The novel risk scores from Eq (1), with Eq (2)2 based on MRI and Kinect measurements, outperformed all other parameters and published CPD indices. The MRI- and Kinect-based CPD risk scores were the only scores with an AUCadj > 0.80, indicating very good predictive capability. The novel risk scores from Eq (1), with Equation (2)1 based on anthropometry were the only score with an AUCadj = 0.79, was the next best model. AUC in the range of 0.70–0.80 are generally considered to have fair predictive capabilities and those in the range of 0.60–0.70 are generally considered poor predictors. Note: AUC = area under the receiver operator characteristic curve, S.E. = standard error, C.I. = confidence interval, HC = maternal head circumference, SM = skeletal mapping, FPI = fetal pelvic index. p < 0.05 denotes statistical significance against the null hypothesis that the AUC = 0.50.

https://doi.org/10.1371/journal.pone.0203865.t002

thumbnail
Table 3. 2x2 Contingency table and 4-point risk score contingency table for select indices.

Tabulates the percentage of CPD and VD that scored in the range indicated. For example, 68% of the CPD cases had height ≦156.1 cm and 27% of VD had height in the range greater 156.1 cm and less than 160.3 cm. The 2x2 contingency table for the Kinect risk score shows that 84% of the CPD cases and 87% of the VD can be deected with a FNR of 13% and a FPR of 16%. The 4-point scale for the Kinect risk score shows that 78% of the CPD cases were scored as very high risk with a FPR ≤10% and 74% of the VD were scored at mild risk with a FNR ≤10%. Note: CPD = cephalopelvic disproportion, VD = vaginal delivery, FPI = fetal pelvic index, LRCPD = CPD likelihood ratio = TPR/FPR within the range of risk values of interest, TPR = true positive rate, FPR = false positive rate, LRVD = vaginal delivery likelihood ratio = TNR/FNR within the range of risk values of interest, TNR = true negative rate, and FNR = false negative rate. Green shaded areas denote ‘true’ predictions (i.e., TPR or TNR) and the orange shaded areas denote ‘false’ predictions (i.e., FPR or FNR).

https://doi.org/10.1371/journal.pone.0203865.t003

The anthropometry-based risk model, following Eq (1) with Eq (2)1 yielded a higher AUCbest and AUCpred, compared to Eq (2)2 (Fig 4). An 8-parameter risk model yielded the highest AUCpred = 0.793, with the corresponding AUCbest = 0.824 (Table 2 and Fig 5). Model parameters (βo and βi) for each model are provided in Table 1. The optimism, O, increased linearly with the number of features. This risk model was able to identify 53% of the CPD cases as very high risk and 47% of the vaginal delivery cases as mild risk, with likelihood ratios of 5.5 and 5.0 for CPD and VD, respectively (Table 3); the remaining CPD cases and vaginal deliveries, for the high risk and moderate risk range, showed likelihood ratios of 1.2 and 1.6 for CPD and VD, respectively.

thumbnail
Fig 4. AUCbest, AUCadj and O versus number of features (k) included in the model.

(a) AUCbest increases monotonically with increasing number of features. (b) AUCadjk curves show that as the number of features included in the model increased, the AUCadj increased to reach a maximum value, plateaued in some cases, then decreased in models with a high number of features. (c) The optimism, O, increased linearly with an increasing number of features.

https://doi.org/10.1371/journal.pone.0203865.g004

thumbnail
Fig 5. Receiver operator characteristic (ROC) curves.

The ROC curves for fetal pelvic index (FPI), clinical pelvimetry, height, the MRI score of Mengert, and the novel anthropometry-, clinical pelvimetry-, Kinect-, and MRI-based risk scores. The area under the ROC curves, shown in Table 2, for MRI-, Kinect-, anthropometry-, clinical pelvimetry-based scores, Mengert, height, and FPI.

https://doi.org/10.1371/journal.pone.0203865.g005

Clinical pelvimetry and CPD risk

Of all the individual features from clinical pelvimetry, pelvic side walls showed the greatest predictive power, followed by ischial spines and fetal head station. A 4-parameter risk model yielded the highest AUCpred = 0.721, with the corresponding AUCbest = 0.732; model parameters (βo and βi) for each model are provided in Table 1. This risk model was able to identify only 22% of the CPD cases as very high risk and 44% of the vaginal delivery cases as mild risk.

Radiological pelvimetry and CPD risk

The inlet area (AUC = 0.760), which is proportional to the metric of Mengert [13], was the best individual or previously published metric based on radiological pelvimetry, followed by inlet circumference and diagonal conjugate. The model of Abitbol et al.[16] was a poor predictor of CPD in this subject sample, with AUC = 0.556, which was not statistically different from the AUC = 0.50, which indicates a random sampling. We found that the model of Friedman and Taylor [17] based on inlet capacity was a fair predictor, but that based on midpelvic capacity was a poor predictor. The index of Sporri et al. was also a poor predictor, since mid-pelvic capacity was typically lower than inlet capacity [18]. We found that the FPI was also a marginal predictor of CPD, with an AUC = 0.616 [19].

The MRI-based risk model, following Eq (1) with Eq (2)2 yielded a higher AUCbest and AUCpred, compared to Eq (2)1 (Fig 4). A 24-parameter risk model yielded the highest AUCpred = 0.825, with the corresponding AUCbest = 0.906; model parameters (βo and βi) for each model are provided in Table 1. The novel MRI risk model yielded the highest AUCbest and AUCadj of all risk models (Figs 4 and 5) and identified 65% of the CPD cases as very high risk and 69% of the vaginal delivery cases mild risk with high likelihood ratios of 6.7 and 8.4 for CPD and VD, respectively; the remaining CPD cases and vaginal deliveries, for the high risk and moderate risk range, showed likelihood ratios of 1.4 and 0.9 for CPD and VD, respectively.

Kinect 3D camera measurements and CPD risk

Of all the individual Kinect measurements, head height, neck height, and face height from the skeletal map, showed the greatest predictive power; each was on par with anthropometric height. The Kinect-based risk model, following Eq (1) with Eq (2)1 yielded a higher AUCbest, but Eq (2)2 yielded a higher AUCpred (Fig 4). A 17-parameter risk model yielded the highest AUCpred = 0.801, with the corresponding AUCbest = 0.871 (Fig 5); model parameters (βo and βi) for each model are provided in Table 1. The novel Kinect-based risk score identified 57% of the CPD cases as very high risk and 67% of the vaginal delivery cases as mild risk with high likelihood ratios of 6.0 and 7.4 for CPD and VD, respectively; the remaining CPD cases and vaginal deliveries, for the high risk and moderate risk range, showed likelihood ratios of 2.7 and 1.4 for CPD and VD, respectively.

Risk score evaluation and assessment

The two models considered in this study, r1 and r2, showed differing results across risk modalities. For MRI and Kinect, r2 generated a higher AUCadj, compared to r1, whereas for the anthropometry risk score, r1 was superior to r2. For each model, as the number of features included in the model (k) increase, the AUCbest also increased asymptotically to a maximum value (Fig 4A). For Kinect, the asymptotic value of AUCbest was higher for r1, compared to r2, even though r2 yielded the higher AUCadj. In contrast, as the number of features included in the model increased, the AUCadj increased to reach a maximum value, plateaued in some cases, then decreased in models with a high number of features (Fig 4B). Thus, although the AUCbest generally increases monotonically with an increasing number of features, the most predictive models will likely correspond to a model with a moderate number of features. Indeed, for these models, the optimism, O, increased linearly with an increasing number of features (Fig 4C). Generally, at lower values of k, the slope of the AUCbestk curve (Fig 4A) is greater than the slope of the optimism Ok curve (Fig 4C), so that the slope of the AUCadjk (Fig 4B) is positive. In contrast, at higher values of k, the slope of the AUCbestk curve is less than the slope of the optimism Ok curve, so that the slope of the AUCadjk is negative. The maximum AUCadj occurs at values of k in which the slope of the AUCbestk curve is equal to that of the Ok curve.

Discussion

Given that Ethiopia’s density of doctors is 1 doctor for every 40,000 people, a large majority of women who seek care do not see an Obstetrician and Gynecologist during antenatal care and delivery [68]. In most cases, care is provided by a midwife or health extension worker. While clinically trained obstetricians and gynecologists have built intuition, based on training and experience, most of the antenatal care givers in Ethiopia lack the knowledge and expertise to identify patients at risk of obstructed labor due to CPD. The goal of this paper is to develop and validate simple tools to translate a clinician’s intuition to a rural midwife or health worker in Ethiopia, by providing an automatic, safe, easy-to-use, easy-to-interpret, low-cost assessment of risk of CPD.

CPD risk assessment from anthropometry

Height has long been recognized as a risk indicator for C/S due to CPD [2224] across multiple nationalities and regions of the world [22, 2535], including sub-Saharan Africa [24, 3646]; however, the sensitivity and specificity of height as a predictor of CPD are generally low [69]. The specificity, sensitivity, and likelihood ratios reported herein are consistent with many reports in the literature [24, 27, 31, 38, 40, 44, 46, 52, 6973]. Other anthropometric features or combinations of features have also been identified as potential risk indicators for CPD, including foot length [25, 35, 42, 45, 47], shoulder diameter [35], lower limb length [42], BMI [32, 4851], maternal head circumference-to-height ratio [32], and external pelvimetry (e.g., external intercrestal, interspinous, intertrochanteric and intertuberous transverse pelvic diameters, antero-posterior external conjugate or Baudelocque diameter; and transverse and vertical diagonals of the Michaelis sacral rhomboid area) [28, 34, 35, 45, 52]. CPD risk assessment based on anthropometry have shown low sensitivity and specificity [3235, 45, 52, 55, 56]. For the current study, all height measurements, shoulder width, foot length were lower, and head circumference-to-height ratio and BMI were higher with CPD versus VD. The external pelvic dimension measured in this study were not different across groups. We did not measure the sacral rhomboid dimensions with a tape measure, but did measure these distances from our MRI measurements; the vertical and transverse dimension were lower with CPD. Nevertheless, in the current study, height alone out-performed (i.e., yielded a higher AUC value than) foot length, shoulder diameter, BMI, maternal head circumference-to-height ratio, and all external pelvimetry features as a predictor of CPD.

The proposed novel CPD risk score based on anthropometry, showed an AUCbest = 0.824 and AUCadj = 0.793; the latter was higher than any previously reported risk score that we considered herein, applied to this group of Ethiopian women. Typically, AUC in the range of 0.80 and 0.90 are considered good predictors, whereas those between 0.70 and 0.80 are considered only fair predictors. In the current study, the tape measurements were made by nurses trained by the research team and under the direct observation of the research team, to minimize measurement errors, whenever possible. In the field, however, traditional anthropometry is prone to measurement errors and shows poor inter- and intra-user variabilities, which may limit the scale up potential of this approach [54, 7476]. Examples include the following: (i) Improper positioning of the subject; e.g., the subjects’ heels and back are not flush to the length board, the subjects’ head is tilted forward or to the side or not held in the Frankfort plan, the subject’s knees are bent, the subjects’ feet are not flat, the subject is wearing shoes, or the subjects’ posture is not completely upright when measuring height. (ii) Improper positioning of the tape measure; e.g., not measuring hip or waist circumference at the widest part of the of the hip or waist or head circumference measurements are not taken at appropriate location on the forehead. (iii) Insufficient (or excess) tension in the measuring tape or including clothing under measuring tape. (iv) Errors in reading or recording of values from the tape measure. (v) Broken, poor quality, or inaccurate length boards and tape measures. (iv) Inaccurate calculations (e.g., waist-to-hip ratio), (vi) Dujardin et al. suggests that errors occur due to the readers attraction to round numbers, assignment of some standard sizes, and intentional over-recording of at-risk women [53]. Nevertheless, once validated against a larger sample size, this CPD risk assessment approach, based on anthropometry, could be packaged into an easy-to-use smartphone or computer application to provide a low-cost, easy-to-use tool to assess risk with good predictive capability.

CPD risk assessment from clinical pelvimetry

Clinical pelvimetry assesses the size of the pelvic cavity by means of the systematic vaginal palpation of specific bony landmarks in the pelvis and an estimation of the distances between them. The proposed novel CPD risk score based on pelvimetry, based on Eq (1), using Eq (2)1 with the parameters specified in Table 1, yielded an AUC below that of height alone. One limitation to scaling this approach is that this technique requires highly skilled personnel [57, 58, 77]. Given that Ethiopia’s density of doctors (0.025/1000 people) and nurse/midwifery personnel (0.252/1000 people) are amongst the lowest in the world [68], the number of qualified personnel that could be trained in this technique in Ethiopia is too low to effectively scale. Further, given that this technique is subjective and shows only fair predictive capability, scaling this technique may not be a useful investment of resources [57, 58, 77].

CPD risk assessment from radiological pelvimetry

Use of X-ray, and later ultrasound, MRI, and CT to measure fetal and pelvic dimensions, and assess risk of adverse pregnancy outcomes including CPD, has been an active area of research for nearly a century [1320]. When applied to the Ethiopian women in our cohort, aside from the risk scores from Mengert [13] and Friedman and Taylor [17], other published radiological pelvimetry-based risk scores proved to be poor predictors of CPD; however, our novel CPD risk score based on MRI measurements showed the highest AUCadj. MRI and CT are cost prohibitive in low-resource settings and a majority of mothers in the developing world do not have access to MRI or CT. X-ray machines are more readily available, but radiation exposure to the fetus poses an unwanted health risk and may represent a significant barrier to widespread scale-up in the developing world. Assessment of fetal size can be done via ultrasound imaging; however, ultrasound technology and expertise are also limited in most low-resource settings. Further, the accuracy of radiological pelvimetry at predicting CPD-related obstructive labor is controversial and lacks testing via rigorous randomized trials [14, 21]. In the current study, since we conducted MRI to quantify of the maternal pelvic dimensions, we also assessed fetal dimensions using MRI. Ultrasound fetal measurements were not routinely collected on study participants during their normal antenatal care; thus, to collect these data would require an additional exam, which we felt was unwarranted.

CPD risk assessment from kinect 3D camera

We developed a low cost, safe, easy-to-use, portable platform to assess risk of obstructed labor due to CPD, by employing a Kinect 3D camera. The automated Kinect-based CPD risk score showed very good predictive capabilities and performed better than all previously published metrics. Automated measurements, through the use of 3D camera technology offer the potential to significantly reduce measurement errors and the inherent inter- and intra-user variabilities in measurement, compared to tape measurements, thereby providing a more reliable assessment. Further, whereas length board and tape measurements are limited to ‘linear’ measurements of exterior landmarks, camera technology offers the possibility of producing new measurements that cannot be measured with a tape measure. Examples include the following: (i) Total volume of a certain region (e.g., the arm, leg, or fundus) can be accurately and automatically measured from 3D models of a subject. (ii) Curvatures (e.g., curvature of the hips or fundus) can be measured from 3D images. (iii) The skeletal mapping feature of the Kinect sensor and software offer the ability to obtain ‘internal’ dimensions; e.g., novel algorithms, developed by Microsoft, conduct real-time analysis to accurately estimate 25 joint locations (e.g., hip, shoulder, and knee joints). For two women with the same pelvic dimensions, one woman at a healthy weight and the other overweight, the hip and waist circumference of these two women may be very different, but the hip joint location, determine via skeletal mapping may be similar. (iv) 3D models and skeletal maps can also be helpful in measuring accurate heights, even if the subject is not positioned properly. For example, if the participants knees are bent and/or their back is not perfectly vertical, the sum of the distance, in 3-dimension, from the ankle joint to the knee joint, the knee joint to the hip, the hip to the shoulder, and the shoulder to the top of the head eliminate the errors due to bending of the subject out of the measurement plane. (v) Some measurements have been shown to be useful indicators of CPD, that require calipers (anthropometer) to measure; e.g., external conjugate and inter-trochanteric diameter[35]. These measurements can be accurately and automatically measured from 3D models of a subject, without the need for an anthropometer.

Limitations of this study

The current study has several important limitations that should be mentioned. (i) Given the small sample size of this feasibility study, the proposed risk models require additional refinement, based on a larger sample size and cross-section of the population, and a broader, blinded, prospective, validation studies before this method can be applied clinically. (ii) We evaluated two risk models (Eq 1, with r1 and r2); however, there are numerous other options that can be taken from machine learning and data science that should be evaluated in future work. (iii) This study relied on recruitment of women with previous obstetric history of CPD, some of whom did not try labor. Thus, it is possible that some of the participants that underwent elective C/S could have successfully delivered vaginally in the current pregnancy. (iv) With regard to the clinical pelvimetry, one limitation is that the examiners were not blinded to the previous obstetric history. (v) We have not assessed the participant and caregivers views, perception, and acceptability of the Kinect 3D camera. Although the Kinect camera is safe and has no unacceptable exposure limit, participant and caregivers perceived risk of using this tool may be an important barrier that should be quantified and addressed in future studies.

(vi) We did not include some parameters as variables in the risk model, because these data are potentially biased across our experimental groups. Labor is a multivariate process, influenced by multiple factors that were not included in the current risk model; factors such as maternal age, parity, obstetric history, gestational age, paternal factors, amongst others. The sample population used in this study is not a representative cross section of all pregnant mothers from the TASR hospital catchment area. For example, the high-risk group (Group 1) was older than our primigravida group (Group 3). This is not necessarily associated with CPD; rather, this is (at least partly due) to our experimental design, since any group of multigravida women will, on average, be older than a group of primigravida women from the same population.

(vii) All measurements were taken at >36 weeks of gestation; there is a need to develop tools that accurately predict CPD over the 1st, 2nd, and 3rd trimester. We believe that many of the key features in our preliminary risk score will not change significantly throughout gestation; e.g., maternal height, maternal shoulder diameter, or maternal head circumference. Therefore, a risk score that is accurate, based on measurements made at 36–42 weeks of gestation, may be modified to be accurate based on measurements made at any gestational age. Validating this risk assessment tool across a wide range of gestational ages will expand the tools utility; e.g., healthcare providers could assess risk at multiple gestational ages, which could help identify risk earlier in pregnancy and affirm risk assessment after multiple measurements. Given that it may require hours or days of travel for mothers in the rural developing world to reach facilities capable of performing a C/S, early and accurate diagnosis of risk of CPD may be crucial for mothers to plan for labor and delivery. Further, given that a majority of women in Ethiopia have 3 or fewer antenatal care visits [7], expanding the validated gestational age range across which this tool is valid may increase the number of pregnant women who access the tool.

(viii) The Kinect skeletal mapping joint locations are not an accurate representation of the actual anatomical joint locations. Thus, data on the actual joint locations should be used with caution. That said, we expect that the skeletal mapping outputs are repeatable measurements, despite not accurately representing the actual anatomical locations, that may be important features used to assess risk of CPD. (ix) The scan-acquisition method employed in this study captured 3D point clouds from four orthogonal views (anterior, posterior, left and right lateral). In many cases, this method limited our ability to stitch these four point-clouds together into an accurate 3D model due to lack of common landmarks across views. In the future, an improved acquisition protocol, with a larger number of views is recommended [78].

Opportunities and limitations to scale-up

Our long-term goal is to develop novel strategies to improve the diagnosis of risk of CPD in areas where emergency C/S is not readily available during labor. Despite a century of work towards developing such a tool, no tool to assess risk of CPD is adopted as standard clinical practice. This is likely because of two distinct reasons: (i) given the nearly universal access to emergency C/S in the developed world, the utility of such a tool in the developed world has diminished; and, (ii) previous risk scores have shown only fair predicative capabilities. Nevertheless, the pressing need remains for accurate and timely diagnosis of CPD in the rural developing world, where access to C/S may require significant travel for mothers. With the rapid advancement both in the areas of low-cost 3D camera technology and in data science, there is a unique opportunity to develop such a low cost, usable tool.

Although the primary goal of this paper was to demonstrate the feasibility of using a Kinect 3D camera to assess risk of CPD, and important secondary outcome is the development of a tape measurement-based tool, which is also a novel contribution of this work and may be an important part of the potential scale up strategy. Although research has shown that some anthropometric features correlate to risk of CPD, such measurements are not currently used to assess risk of CPD. Indeed, in Ethiopia, CPD typically goes undiagnosed. Given its ultralow cost and ease of use, the novel tape measurement-based risk score may provide and important initial screening to ‘triage’ low risk subjects and refer potentially high-risk subjects for further screening with a 3D camera.

One potential barrier to wide-spread uptake and participant acceptability is the need for participants to disrobe during the Kinect scan. For the purpose of this feasibility study, subjects’ were asked to disrobe, with the exception of tight-fitting undergarments, as clothing would interfere with the generation of representative 3D measurements that may be important in predicting CPD; e.g., the fundal geometry, waist circumference, hip circumference, etc. Thus, for this feasibility study, we sought to identify all of these potentially important parameters by scanning disrobed subjects. That said, one may imagine a risk score that is based only on measurements that could be collected while the subject is clothed; e.g., heights, shoulder dimensions, and most skeletal mapping outputs. Indeed, the Kinect scanner is designed to perform these analyses on clothed subjects. Such analysis was not included in the current study since additional measurement error may be introduced when data is collected from clothed participants but should be a focus of future studies.

Conclusions

A novel application of the Kinect 3D camera to assess risk of CPD is presented, which performed on par with the novel MRI-based score and outperformed all previously reported risk scores applied to this group of Ethiopian women. This work demonstrates the feasibility of a safe, low-cost, easy-to-use 3D camera technology to assess risk of CPD. This work demonstrates the need for further broader blinded, prospective studies to refine and validate the proposed CPD risk scores, which are required before this method can be applied clinically.

Supporting information

S1 Table. Data collection form used for ensuring the participants meet the eligibility criterion (Section 1) and recording anthropometric measurements (Section 2), clinical pelvimetry (Section 3), radiological pelvimetry (Section 4), and pregnancy outcomes (Section 5).

The columns in S2 Table correspond to the rows in this data collection form. This Table is provided in the supplemental file named “S1_Table.docx”.

https://doi.org/10.1371/journal.pone.0203865.s001

(DOCX)

S2 Table. Raw data for all participants for eligibility criteria, anthropometry, clinical pelvimetry, radiological pelvimetry, and pregnancy outcomes.

The columns in S1 Table correspond to the rows in this data collection form. This Table is provided in the supplemental file named “S2_Table.csv”.

https://doi.org/10.1371/journal.pone.0203865.s002

(CSV)

Acknowledgments

We gratefully acknowledge the support from Grand Challenges Canada (Grant Number 0748–03, RLG), as part of the Saving Lives at Birth Partnership between the United States Agency for International Development (USAID), Norwegian Agency for Development Cooperation (Norad), Bill & Melinda Gates Foundation, Grand Challenges Canada, UKaid, and the Korea International Cooperation Agency (KOICA). We also gratefully acknowledge the support from US State Department, Bureau of Educational and Cultural Affairs, Fulbright Scholars Program (Grant ID: 48153580, RLG)

References

  1. 1. Neilson JP, Lavender T, Quenby S, Wray S. Obstructed labour. Br Med Bull. 2003;67:191–204. pmid:14711764.
  2. 2. WHO. Make every mother and child count. In: The world health report 2005. Geneva, Switzerland: 2005.
  3. 3. Say L, Chou D, Gemmill A, Tunçalp Ö, Moller A-B, Daniels J, et al. Global causes of maternal death: a WHO systematic analysis. The Lancet Global health. 2014;2(6):e323–33. pmid:25103301
  4. 4. El-Hamamy E, Arulkumaran S. Poor progress of labour. Current Obstetrics & Gynaecology. 2005;15(1):1–8. 5231815714520718052related:5FpQSqwkm0gJ.
  5. 5. Wall LL. Obstetric vesicovaginal fistula as an international public-health problem. Lancet. 2006;368(9542):1201–9. pmid:17011947.
  6. 6. Konje JC, Ladipo OA. Nutrition and obstructed labor. Am J Clin Nutr. 2000;72(1):291S–7S. pmid:000088005800013.
  7. 7. Warren C. Care seeking for maternal health: challenges remain for poor women. Ethiopian Journal of Health Development. 2010;24(1). 1124977457875432220related:HLP8nAm5nA8J.
  8. 8. Victora CG, Requejo JH, Barros AJD, Berman P, Bhutta Z, Boerma T, et al. Countdown to 2015: a decade of tracking progress for maternal, newborn, and child survival. Lancet. 2016;387(10032):2049–59. pmid:26477328.
  9. 9. Tessema GA, Laurence CO, Melaku YA, Misganaw A, Woldie SA, Hiruye A, et al. Trends and causes of maternal mortality in Ethiopia during 1990–2013: findings from the Global Burden of Diseases study 2013. BMC Public Health. 2017;17(1):160. pmid:28152987; PubMed Central PMCID: PMCPMC5290608.
  10. 10. Kelly J. Ethiopia: an epidemiological study of vesico-vaginal fistula in Addis Ababa. World Health Stat Q. 1995;48(1):15–7. pmid:7571703.
  11. 11. Ethiopia Demographic and Health Survey 2016: Key Indicators Report. Addis Ababa, Ethiopia & Rockville, Maryland, USA: Ethiopian Central Statistical Agency & ICF International, 2016 Oct. Report No.
  12. 12. Dolea C, AbouZahr C. Global burden of obstructed labour in the year 2000. Evidence and information for policy (EIP). 2003. 5630576299444479567related:TyYf8U_TI04J.
  13. 13. Mengert WF. Estimation of pelvic capacity. Journal of the American Medical Association. 1948;138(3):169–74. pmid:18877482.
  14. 14. Caldwell WE, Moloy HC. Anatomical variations in the female pelvis and their effect in labor with a suggested classification. Am J Obstet Gynecol. 1933;26:479–505. pmid:000202334000068.
  15. 15. Ball RP. Roentgen pelvimetry and fetal cephalometry. Surgery Gynecology & Obstetrics. 1936;62:798–810. pmid:000188312200106.
  16. 16. Abitbol MM, Taylor UB, Castillo I, Rochelson BL. The cephalopelvic disproportion index. Combined fetal sonography and x-ray pelvimetry for early detection of cephalopelvic disproportion. The Journal of reproductive medicine. 1991;36(5):369–73. pmid:2061886.
  17. 17. Friedman EA, Taylor MB. A modified nomographic aid for x-ray cephalopelvimetry. Am J Obstet Gynecol. 1969;105(7):1110–5. pmid:5352589.
  18. 18. Spörri S, Thoeny HC, Raio L, Lachat R, Vock P, Schneider H. MR imaging pelvimetry: a useful adjunct in the treatment of women at risk for dystocia? AJR Am J Roentgenol. 2002;179(1):137–44. pmid:12076922.
  19. 19. Morgan MA, Thurnau GR, Fishburne JI. The fetal-pelvic index as an indicator of fetal-pelvic disproportion: a preliminary report. Am J Obstet Gynecol. 1986;155(3):608–13. pmid:3529966.
  20. 20. Morgan MA, Thurnau GR. Efficacy of the fetal-pelvic index in nulliparous women at high risk for fetal-pelvic disproportion. Am J Obstet Gynecol. 1992;166(3):810–4. pmid:1550146.
  21. 21. Rozenberg P. [Is there a role for X-ray pelvimetry in the twenty-first century?]. Gynécologie, obstétrique & fertilité. 2007;35(1):6–12. pmid:17188014.
  22. 22. Thomson AM. Maternal stature and reproductive efficiency. The Eugenics review. 1959;51(3):157–62. pmid:21260778; PubMed Central PMCID: PMCPMC2973796.
  23. 23. BAIRD D. The Influence of Social and Economic Factors on Stillbirths and Neonatal Deaths. The Journal of obstetrics and gynaecology of the British Empire. 1945;52(4):339–66. A1945YE05400004.
  24. 24. Everette VJ. The relationship between maternal height and cephalopelvic disproportion in Dar es Salaam. East Afr Med J. 1975;52(5):251–6. pmid:1157729.
  25. 25. Kennedy JL, Greenwald E. Correlation of shoe size and obstetric outcome: an anthropometric study. Am J Obstet Gynecol. 1981;140(4):466–7. pmid:7246666.
  26. 26. Chen HY, Chen YP, Lee LS, Huang SC. Pelvimetry of Chinese Females with Special Reference to Pelvic Type and Maternal Height. Int Surg. 1982;67(1):57–62. pmid:7096006.
  27. 27. Kappel B, Eriksen G, Hansen KB, Hvidman L, Kragolsen B, Nielsen JB, et al. Short Stature in Scandinavian Women—an Obstetrical Risk Factor. Acta Obstet Gynecol Scand. 1987;66(2):153–8. pmid:3618140.
  28. 28. Hanzal E, Kainz C, Hoffmann G, Deutinger J. An analysis of the prediction of cephalopelvic disproportion. Arch Gynecol Obstet. 1993;253(4):161–6. pmid:8161249.
  29. 29. Hin LY, Khairuddin Y, Ng KB. The predictive value of extremes of birth weight, mother&apos;s height and ethnic origin on the likelihood of emergency caesarean section. Asia Oceania J Obstet Gynaecol. 1994;20(4):389–94. pmid:7832671.
  30. 30. McGuinness BJ, Trivedi AN. Maternal height as a risk factor for caesarean section due to failure to progress in labour. Australian & New Zealand Journal of Obstetrics & Gynaecology. 1999;39(2):152–4. pmid:10755767.
  31. 31. Merchant KM, Villar J, Kestler E. Maternal height and newborn size relative to risk of intrapartum caesarean delivery and perinatal distress. Br J Obstet Gynaecol. 2001;108(7):689–96. pmid:11467692.
  32. 32. Connolly G, Naidoo C, Conroy RM, Byrne P, McKenna P. A new predictor of cephalopelvic disproportion? J Obstet Gynaecol. 2003;23(1):27–9. pmid:12623477.
  33. 33. Khunpradit S, Patumanond J, Tawichasri C. Risk indicators for cesarean section due to cephalopelvic disproportion in Lamphun hospital. J Med Assoc Thai. 2005;88 Suppl 2:S63–8. pmid:17722319.
  34. 34. Bansal S, Guleria K, Agarwal N. Evaluation of sacral rhomboid dimensions to predict contracted pelvis: a pilot study of Indian primigravidae. J Obstet Gynaecol India. 2011;61(5):523–7. pmid:23024521; PubMed Central PMCID: PMCPMC3257337.
  35. 35. Benjamin SJ, Daniel AB, Kamath A, Ramkumar V. Anthropometric measurements as predictors of cephalopelvic disproportion. Acta Obstet Gynecol Scand. 2012;91(1):122–7. 15319068463571714286related:7sAUjZpDmNQJ. pmid:21895610
  36. 36. Antenatal screening for fetopelvic dystocias. A cost-effectiveness approach to the choice of simple indicators for use by auxiliary personnel. The Kasongo Project Team. The Journal of tropical medicine and hygiene. 1984;87(4):173–83. pmid:6441000.
  37. 37. Adadevoh SWK, Hobbs C, Elkins TE. The Relation of the True Conjugate to Maternal Height and Obstetric Performance in Ghanaians. Int J Gynaecol Obstet. 1989;28(3):243–51. pmid:2564354.
  38. 38. Sokal D, Sawadogo L, Adjibade A. Short stature and cephalopelvic disproportion in Burkina Faso, West Africa. Operations Research Team. Int J Gynaecol Obstet. 1991;35(4):347–50. pmid:1682184.
  39. 39. van Roosmalen J, Brand R. Maternal height and the outcome of labor in rural Tanzania. Int J Gynaecol Obstet. 1992;37(3):169–77. pmid:1351003.
  40. 40. Kwawukume EY, Ghosh TS, Wilson JB. Maternal height as a predictor of vaginal delivery. Int J Gynaecol Obstet. 1993;41(1):27–30. pmid:8098291.
  41. 41. Tsu VD. Antenatal screening: its use in assessing obstetric risk factors in Zimbabwe. J Epidemiol Community Health. 1994;48(3):297–305. pmid:8051531; PubMed Central PMCID: PMCPMC1059963.
  42. 42. Van Bogaert LJ. The relation between height, foot length, pelvic adequacy and mode of delivery. Eur J Obstet Gynecol Reprod Biol. 1999;82(2):195–9. pmid:10206415.
  43. 43. Ould El Joud D, Bouvier-Colle MH, Group M. Dystocia: a study of its frequency and risk factors in seven cities of west Africa. Int J Gynaecol Obstet. 2001;74(2):171–8. pmid:11502297
  44. 44. Brabin L, Verhoeff F, Brabin BJ. Maternal height, birthweight and cephalo pelvic disproportion in urban Nigeria and rural Malawi. Acta Obstet Gynecol Scand. 2002;81(6):502–7. pmid:12047302.
  45. 45. Rozenholc AT, Ako SN, Leke RJ, Boulvain M. The diagnostic accuracy of external pelvimetry and maternal height to predict dystocia in nulliparous women: a study in Cameroon. BJOG: An International Journal of Obstetrics & Gynaecology. 2007;114(5):630–5. pmid:17439570.
  46. 46. Aitken IW, Walls B. Maternal Height and Cephalopelvic Disproportion in Sierra-Leone. Trop Doct. 1986;16(3):132–4. pmid:3765098.
  47. 47. Frame S, Moore J, Peters A, Hall D. Maternal height and shoe size as predictors of pelvic disproportion: an assessment. Br J Obstet Gynaecol. 1985;92(12):1239–45. pmid:4084467.
  48. 48. Witter FR, Caulfield LE, Stoltzfus RJ. Influence of Maternal Anthropometric Status and Birth-Weight on the Risk of Cesarean Delivery. Obstet Gynecol. 1995;85(6):947–51. pmid:7770265.
  49. 49. Jensen H, Agger AO, Rasmussen KL. The influence of prepregnancy body mass index on labor complications. Acta Obstet Gynecol Scand. 1999;78(9):799–802. pmid:10535344.
  50. 50. Young TK, Woodmansee B. Factors that are associated with cesarean delivery in a large private practice: The importance of prepregnancy body mass index and weight gain. Am J Obstet Gynecol. 2002;187(2):312–8. pmid:12193918.
  51. 51. Chen G, Uryasev S, Young TK. On prediction of the cesarean delivery risk in a large private practice. Am J Obstet Gynecol. 2004;191(2):616–24. pmid:15343249.
  52. 52. Liselele HB, Boulvain M, Tshibangu KC, Meuris S. Maternal height and external pelvimetry to predict cephalopelvic disproportion in nulliparous African women: a cohort study. BJOG: An International Journal of Obstetrics & Gynaecology. 2000;107(8):947–52. pmid:10955423.
  53. 53. Dujardin B, Clarysse G, Mentens H, De Schampheleire I, Kulker R. How accurate is maternal height measurement in Africa? Int J Gynaecol Obstet. 1993;41(2):139–45. pmid:8099029.
  54. 54. Ayele B, Aemere A, Gebre T, Tadesse Z, Stoller NE, See CW, et al. Reliability of measurements performed by community-drawn anthropometrists from rural Ethiopia. PloS one. 2012;7(1):e30345. pmid:22291939; PubMed Central PMCID: PMCPMC3265464.
  55. 55. Møller B, Lindmark G. Short stature: an obstetric risk factor? A comparison of two villages in Tanzania. Acta Obstet Gynecol Scand. 1997;76(5):394–7. pmid:9197438.
  56. 56. Shepard MJ, Saftlas AF, Leo-Summers L, Bracken MB. Maternal anthropometric factors and risk of primary cesarean delivery. Am J Public Health. 1998;88(10):1534–8. pmid:9772858; PubMed Central PMCID: PMCPMC1508477.
  57. 57. Adinma JI, Agbai AO, Anolue FC. Relevance of clinical pelvimetry to obstetric practice in developing countries. West Afr J Med. 1997;16(1):40–3. pmid:9133823.
  58. 58. Maharaj D. Assessing cephalopelvic disproportion: back to the basics. Obstet Gynecol Surv. 2010;65(6):387–95. pmid:20633305.
  59. 59. Larimore WL, Petrie K. Management of labor. In: Ratcliffe SD, Baxley EG, Byrd JE, Sakornbut EL, editors. Family practice obstetrics, 2nd edition. Philadelphia: Hanley & Belfus; 2001. p. 359–441.
  60. 60. Baskett TF, Calder AA, Arulkumaran S. Munro Kerr’s Operative Obstetrics. Twelfth Edition: Saunders Elsevier; 2014.
  61. 61. Stark DD, McCarthy SM, Filly RA, Parer JT, Hricak H, Callen PW. Pelvimetry by Magnetic-Resonance Imaging. AJR Am J Roentgenol. 1985;144(5):947–50. pmid:3872578.
  62. 62. Bull HC. Pelvimetry in obstetrics. Postgrad Med J. 1949;25(285):310–8. pmid:18135155; PubMed Central PMCID: PMCPMC2529979.
  63. 63. Parkar AP, Olsen ØE, Gjelland K, Kiserud T, Rosendahl K. Common fetal measurements: A comparison between ultrasound and magnetic resonance imaging. Acta Radiol. 2010;51(1):85–91. A0779EF6-3388-4259-A680-B5B332190887. pmid:20088642
  64. 64. Keller TM, Rake A, Michel SCA, Seifert B, Efe G, Treiber K, et al. Obstetric MR Pelvimetry: Reference Values and Evaluation of Inter- and Intraobserver Error and Intraindividual Variability1. Radiology. 2003;227(1):37–43. CD1874A1-A2A6-41B6-BF40-FB97029A3EEA. pmid:12601187
  65. 65. Zaretsky MV, Alexander JM, McIntire DD, Hatab MR, Twickler DM, Leveno KJ. Magnetic resonance imaging pelvimetry and the prediction of labor dystocia. Obstet Gynecol. 2005;106(5 Pt 1):919–26. pmid:16260507.
  66. 66. Saleem SN. Fetal MRI: An approach to practice: A review. Journal of Advanced Research. 2014;5(5):507–23. pmid:25685519
  67. 67. Harrell FE, Lee KL, Mark DB. Multivariable prognostic models: issues in developing models, evaluating assumptions and adequacy, and measuring and reducing errors. Stat Med. 1996;15(4):361–87. pmid:8668867.
  68. 68. World Health Organization's Global Health Workforce Statistics, OECD, supplemented by country data.: The World Bank; 2017. Available from: https://data.worldbank.org/indicator/SH.MED.NUMW.P3?view=map&year=2010.
  69. 69. Dujardin B, Van Cutsem R, Lambrechts T. The value of maternal height as a risk factor of dystocia: a meta-analysis. Trop Med Int Health. 1996;1(4):510–21. pmid:8765460
  70. 70. Scott RT, Strickland DM, Hankins GDV, Gilstrap LC. Maternal Height and Weight-Gain During Pregnancy as Risk-Factors for Cesarean-Section. Mil Med. 1989;154(7):365–7. A1989AE91700009. pmid:2503778
  71. 71. Tsu VD. Maternal height and age: risk factors for cephalopelvic disproportion in Zimbabwe. Int J Epidemiol. 1992;21(5):941–6. pmid:1468857.
  72. 72. Burgess HA. Anthropometric measures as a predictor of cephalopelvic disproportion. Trop Doct. 1997;27(3):135–8. PubMed pmid:9227004.
  73. 73. Sheiner E, Levy A, Katz M, Mazor M. Short stature—an independent risk factor for Cesarean delivery. Eur J Obstet Gynecol Reprod Biol; Jun 012005. p. 175–8. pmid:15925047
  74. 74. Ulijaszek SJ, Kerr DA. Anthropometric measurement error and the assessment of nutritional status. The British journal of nutrition. 1999;82(3):165–77. pmid:10655963.
  75. 75. Johnson TS, Engstrom JL, Warda JA, Kabat M, Peters B. Reliability of length measurements in full-term neonates. Journal of obstetric, gynecologic, and neonatal nursing: JOGNN / NAACOG. 1998;27(3):270–6. pmid:9620819.
  76. 76. Sicotte M, Ledoux M, Zunzunegui M-V, Ag Aboubacrine S, Nguyen V-K, group A. Reliability of anthropometric measures in a longitudinal cohort of patients initiating ART in West Africa. BMC medical research methodology. 2010;10:102. pmid:20969785; PubMed Central PMCID: PMCPMC2988008.
  77. 77. Anderson B. Echocardiography: The Normal Examination of Echocardiographic Measurements. 1 ed: Wiley-Blackwell; 2002 Nov 08. 251 p.
  78. 78. Cui Y, Chang W, Nöll T, Stricker D. KinectAvatar: Fully Automatic Body Capture Using a Single Kinect. 7729. Berlin, Heidelberg: Springer, Berlin, Heidelberg; 2012. p. 133–47.