High-Resolution Echo-Planar Spectroscopic Imaging of the Human Calf

Abstract

Background

This study exploits the speed benefits of echo-planar spectroscopic imaging (EPSI) to acquire lipid spectra of skeletal muscle. The main purpose was to develop a high-resolution EPSI technique for clinical MR scanner, to visualise the bulk magnetic susceptibility (BMS) shifts of extra-myocellular lipid (EMCL) spectral lines, and to investigate the feasibility of this method for the assessment of intra-myocellular (IMCL) lipids.

Methods

The study group consisted of six healthy volunteers. A two dimensional EPSI sequence with point-resolved spectroscopy (PRESS) spatial localization was implemented on a 3T clinical MR scanner. Measurements were performed by means of 64×64 spatial matrix and nominal voxel size 3×3×15 mm3. The total net measurement time was 3 min 12 sec for non-water-suppressed (1 acquisition) and 12 min 48 sec for water-suppressed scans (4 acquisitions).

Results

Spectra of the human calf had a very good signal-to-noise ratio and linewidths sufficient to differentiate IMCL resonances from EMCL. The use of a large spatial matrix reduces inter-voxel signal contamination of the strong EMCL signals. Small voxels enabled visualisation of the methylene EMCL spectral line splitting and their BMS shifts up to 0.5 ppm relative to the correspondent IMCL line. The mean soleus muscle IMCL content of our six volunteers was 0.30±0.10 vol% (range 0.18–0.46) or 3.6±1.2 mmol/kg wet weight (range: 2.1–5.4).

Conclusion

This study demonstrates that high-spatial resolution PRESS EPSI of the muscle lipids is feasible on standard clinical scanners.

Introduction

Proton (1H) magnetic resonance spectra of lipids in skeletal muscle are dominated by two spectral lines originating from the methylene protons (-CH2-)n of extra- (EMCL) and intramyocellular (IMCL) lipids (EMCLCH2, IMCLCH2). Less intense resonances are also observable, originating from the methyl protons (CH3), methylene protons α and β to COO (-CH2-CH2-COO), α to C = C (-CH2-CH = CH-CH2-), diallylic methylene ( = CH-CH2-CH = ), and methine protons (-CH = CH-) [1], [2]. IMCL are stored in the form of small droplets in a magnetic susceptibility homogeneous environment (cytoplasm) of muscle cells. In such circumstances, the positions of IMCL spectral lines are independent in orientation of the myocytes relative to the external magnetic field B0 [2], [3]. The bulk fat (EMCL) is, in contrast, stored subcutaneously, in bone marrow and as interstitial (adipose) tissue. The anisotropic shape and geometrical arrangement of EMCL compartments result in a bulk magnetic susceptibility (BMS) shift of its spectral lines [3]. If the shape of the EMCL compartments resemble “infinitely long” cylinders, the spectral line shifts range from 0.2 ppm to −0.1 ppm between parallel and perpendicular orientations of the cylinders relative to B0 [3][6]. However, EMCL compartments in skeletal muscle do not follow a strict cylindrical shape. BMS effects could be therefore more complex. This expectation is supported by the study of Steidle et al [7]. The authors proposed a deconvolution method to compute magnetic field distribution (MFD) inside the voxel using the spectra of muscle lipids and bone marrow. Spectra of soleus muscle revealed that magnetic field deviations experienced by the EMCL compartments were in the interval from ca 0.6 to −0.2 ppm in respect of IMCLCH2. From Steidle et al study [7] it follows that EMCL spectral line shifts somewhat larger than 0.2 ppm or less than −0.1 ppm relative to IMCL can be expected.

The quantification of IMCL became important mainly because of its relation to insulin resistance in obese and non-obese adults as well as in non-diabetic offspring of type 2 diabetic subjects [8][10]. IMCLs are involved in lipid metabolism with a turnover of some hours. IMCL concentrations vary during physical exercise and also depend on training status, diet, fasting, oxygen supply, etc [4]. Fitting of IMCL resonances is often hampered by overlap with intense EMCL spectral lines. Single-voxel MRS with its large voxel size (>1 cm3) is the most sensitive to this effect [2], [6]. The voxel must be placed in the muscle regions free from large deposits of EMCL tissues. Improper voxel positioning will result in a spectrum with indistinguishable IMCL lines. To overcome this limitation, magnetic resonance spectroscopic imaging techniques (MRSI) were proposed [11][17]. The reduction of IMCL contamination was achieved by using smaller voxel sizes and by a post-acquisition selection of voxels with acceptable separation of IMCLCH2 line from EMCLCH2. Conventional STEAM, PRESS or slice selective 2D MRSI techniques reached nominal voxel sizes 5×5×10 mm3 using spatial matrices 32×32 or 36×36 [11][15]. Acquisition time was 20–35 minutes. The fundamental problem of MRSI is inter-voxel signal contamination (signal bleeding) of the strong EMCL signals from subcutaneous fat, bone marrow or larger facial layers into the adjacent regions of muscle due to limited k-space sampling. One solution to this problem is presaturation of large fat structures followed by removing the residual signals by a lipid extrapolation procedure [14], [15]. Another solution is to acquire MRSI data set with high spatial resolution [16], [17]. We previously used 2D MRSI based on a gradient echo sequence with step increments of TE [16]. Spectral matrix 128×256 led to the acquisition voxel size 1.2×0.6×15 mm3. Newcomer et al [17] used a conventional 2D MRSI sequence with decreased flip angles, short TR (146 ms), spectral matrix 64×64, and voxel size 2.5×2.5×10 mm3. The disadvantages of both high-resolution methods were poor water suppression, decreased spectral quality, and the inability to detect other than most intense EMCLCH2 and IMCLCH2 lines.

The drawbacks of conventional [11][15] and high-resolution MRSI techniques with short TR [16], [17] can be overcome by echo-planar spectroscopic imaging (EPSI) [18][23]. With the EPSI approach, a gradient echo train simultaneously encodes one spatial and one spectral dimension. It enables the acquisition of large spatial matrices with long TR in a short measurement time.

The focus of this study was to develop and implement a high-resolution 2D PRESS EPSI technique on a clinical MR scanner, to visualise the bulk magnetic susceptibility (BMS) shifts of EMCL spectral lines, and to investigate the feasibility of this method for the assessment of IMCL lipids.

Methods

Ethics Statement

The study was approved by The Regional Ethical Review Board in Uppsala and written consent was obtained from all participants.

Subjects

Six male volunteers participated in this study. The subjects were healthy, non-obese, passive in free time physical activity and with sedentary occupations. Median age and body mass index were 30.5 years (range: 22–61), and 23.3±1.5 kg/m2 (range: 21–25.5), respectively.

Data Acquisition

All the experiments were performed by using a 3 T scanner (Achieva, Philips Healthcare, Best, Netherlands). The trapezoidal echo-planar readout gradient train was implemented into a manufacturer point-resolved spectroscopy (PRESS) localization module (Fig. 1). It enabled utilization all possibilities of the manufacturer spectroscopic imaging package including water/fat and outer volume suppression modules, excitation pulse shapes, etc. The PRESS localization technique was chosen because this method effectively reduces risk of signal contamination from outside the VOI, e.g. from the other leg (in our case). However, PRESS excitation (Fig. 1) was effective only for the slice selection (Fig. 2), because our intention was to perform spectroscopic imaging in the whole slice (calf). A whole-body coil was used for excitation, and the standard, manufacturer provided flexible two-element coil (20 cm diameter) placed in front of and behind the calf served as the receiver. The subject lay in a supine position with the most extended part of the calf in the center of the receiver coil. 2D PRESS EPSI measurement was performed in the transversal plane by means of 64×64 spatial matrix (Fig. 2); FOV 192 mm; nominal voxel size 3×3×15 mm3 (0.135 cm3); acquisition bandwidth 128 kHz; TR 1500 ms, and TE 38 ms (minimum TE). One and four signal averages were used for non-water and water suppressed scans, respectively. Typical size of the excitation volume (PRESS box) was 160×160 mm in the transversal plane (Fig. 2). Thickness of the PRESS box was 15 mm in feet-head direction. Chemical shift displacement between PRESS box of trimethyl ammonium containing compounds (TMA) at ca 3.2 ppm and methylene (-CH2-)n PRESS box (∼1.5 ppm) was 44.9, 16.4, and 4.2 mm in anterior-posterior, left-right, and feet-head directions, respectively (Fig. 2). 90o RF-pulse with broader bandwidth (BW) produces lower chemical shift artefact in the left-right direction than 180o RF-pulse with narrow BW in the anterior-posterior direction (Fig. 2). Radiofrequency of the PRESS pulses was centered to ∼1.5 ppm in water-suppressed scans, i.e. there was no chemical shift displacement between the water and methylene (-CH2-)n PRESS boxes. The period of readout gradients was 1.6108 ms and readout gradient train consisted of 128 positive and 128 negative lobes. Trapezoidal readout gradient waveforms consisted of linear ramps of 0.1566 ms duration, a constant plateau of 0.4922 ms duration and an amplitude 15.66 mT/m. The duration of the linear ramps was limited by the slew rate of 100 mT/m/s fixed by manufacturer. Conventional phase-encoding was performed in the anterior-posterior direction (Fig. 1). Magnetic field homogeneity in the whole slice was improved by iterative first order shimming (Fig. 2). Typical linewidth of the water line was between ∼15 and ∼25 Hz. Water suppression was performed by applying two bandwidth selective RF pulses and spoiler gradients. Suppression was achieved by adjusting the tip angle of the second RF pulse such that the longitudinal magnetization of the water signal was minimal at the time of the first PRESS excitation RF pulse. Water suppression bandwidth was 140 Hz. An acquisition of two interleaved gradient echo trains was performed in order to increase spectral bandwidth [21]. The beginning of the second echo train was shifted by a half period (0.8054 ms) of the readout gradients. Combination of both echo trains resulted in the effective inter-echo time 0.8054 ms corresponding to a spectral bandwidth of 9.72 ppm. The total net measurement time of our sequence with two interleaves of gradient echo train was 3 min 12 sec for non-water-suppressed (1 acquisition) and 12 min 48 sec for water-suppressed scans (4 acquisitions).

thumbnail
Figure 1. 2D PRESS EPSI sequence.

Gr, Gp, and Gs are gradients in read, phase-encoding, and slice selection directions, respectively. B1 depicts excitation rf pulses.

https://doi.org/10.1371/journal.pone.0087533.g001

thumbnail
Figure 2. Planning of 2D PRESS EPSI experiment with a spatial matrix 64×64 on the background of T1-weighted transversal and coronal spin echo images.

Large thin and thick rectangles show lipid (-CH2-)n and TMA PRESS boxes, respectively. Dashed rectangle depicts shimming volume.

https://doi.org/10.1371/journal.pone.0087533.g002

Data Processing

Measured raw data were transferred to a Linux workstation for processing. Data processing software was developed in-house. Measured data were reorganized to the data sets containing either odd or even echoes. The first 128 echoes arising either from positive or negative readout gradients were processed for each element of the receiver coil separately, i.e., four (kread, kphase, kt) = (64, 64, 128) data matrices were processed. Measured matrices were zero-filled to size (64, 64, 512). Data processing began with optimized 2D Hanning filter [24] applied across the kread and kphase directions to reduce signal bleeding. Radial profile (Fig. 3a) of the filter (Fig. 3b) was computed using the weighting function [24]: w(kn) = β/2[1+ cos(2πknΔx/α)], where kn = [n – (N +1)/2]Δk; n = 1, 2, …, N; N is the number of phase-encoding steps; Δk = 1/FOV is the k-space increment; Δx = FOV/N is the nominal spatial resolution, α = 1.71, and β = 1.47. The first FFT was performed along the kt axis. The chemical shift artifacts caused by readout gradients were removed using a first-order phase correction [25], [26]. Data processing continued with 2D FFT along kread and kphase dimensions. The spatial distribution of the magnetic field ΔB(x, y) in the measured slice was computed from the position of the highest spectral line of non-water suppressed magnitude spectra [25], [26]. The voxel spectra were then corrected by shifting about ΔB/B0 along the spectral axis (Fig. 4). The magnitude spectra of the resulting four spectral matrices were then averaged. Averaging of real spectra was avoided because large quantities of voxel spectra (4×64×64) cannot be reliably phase-corrected before summation. Eddy currents corrections were unnecessary. Unsuppressed water line (not shown) as well as the most intense lipid lines did not reveal signs of eddy current distortions.

thumbnail
Figure 3. K-space filters of raw data and correspondent fat images.

(a) Profile of the usual k-space weighting functions: optimized Hanning (full line), standard Hanning (dashed line) and Hamming (dotted line). (b) Optimized 2D Hanning filter. (c) Standard 2D Hanning filter. (d) Fat image computed without application of k-space filter. (e) Fat image computed from filtered data set using an optimized 2D Hanning filter. (f) Fat image computed from filtered data set using a standard 2D Hanning filter.

https://doi.org/10.1371/journal.pone.0087533.g003

thumbnail
Figure 4. Magnitude spectra (vertical columns) of the voxel rows, depicted by gray and white line in the corresponding fat image.

Spectra represent, tibialis anterior (ta) muscle, bone marrow (mar), soleus (sol) and gastrocnemius medialis (gm) muscles. Spectra were corrected for shifts caused by magnetic field deviations. EMCLCH2 sub-peaks in soleus can be seen between 0.4 and 0.5 ppm, from IMCLCH2 line (arrows).

https://doi.org/10.1371/journal.pone.0087533.g004

Spectrum Processing

Fat images (Fig. 3) were computed by integrating EMCLCH2 and IMCLCH2 spectral lines. Irregularly shaped VOIs were drawn in the fat image (Fig. 3e) after appropriate thresholding of the fat intensities, i.e. the highest fat intensities were removed. Positions of the VOIs were then visualized by superposition with corresponding T1W images. Figures 57 show VOIs used in this study. The average magnitude spectra per voxel were computed by summation of the voxel spectra from the defined volume of interest (VOI) and division by the number of voxels. The spectra of soleus muscle were chosen for quantification. The threshold levels applied for soleus VOI selections were accepted when the EMCLCH2 spectral line amplitude was comparable to the IMCLCH2 (Fig. 5b, c). Spectra were fitted by LCModel [27] in the interval <−0.2; 8>ppm. No apodization of the FIDs to improve SNR was used in this study. Complex FIDs for LCModel were computed by inverse FFT of the magnitude spectra [16]. The fat content in volume percentage (vol%) was computed using the EMCLCH2 intensity of the voxels with a 100% fat content as an internal reference [16]. The reference fat spectrum was obtained from the voxel group indicated by grey pixels in the central part of tibial bone marrow (Fig. 6). The bone marrow fat content of 1.17 mol/kg wet weight was used to convert vol% into mmol/kg wet weight [15].

thumbnail
Figure 5. Magnitude spectra of soleus muscle (108 voxels). VOI indicate white MRSI voxels in T1-weighted image.

(a) Spectrum computed without application of k-space filter during data processing. (b) Spectrum computed from filtered data set, using an optimized 2D Hanning filter. (c) Spectrum computed from filtered data set, using a standard 2D Hanning filter.

https://doi.org/10.1371/journal.pone.0087533.g005

thumbnail
Figure 6. Magnitude spectra of bone marrow (mar, 4 voxels) and in tibialis anterior (ta, 45 voxels), deep posterior compartment (dp, 36 voxels), gastrocnemius medialis (gm, 40 voxels), and soleus (sol, 42 voxels) muscles.

Spectrum of soleus muscle is completed by LCModel fits and residue. VOIs indicate white and gray MRSI voxels in T1-weighted image.

https://doi.org/10.1371/journal.pone.0087533.g006

thumbnail
Figure 7. Volumes of interest employed for computing the spectra shown in figures 8 and 9.

VOIs V1–V5 contains four MRSI voxels (white squares).

https://doi.org/10.1371/journal.pone.0087533.g007

Results

The proposed 2D PRESS EPSI technique and data processing provided spectroscopic fat images (Fig. 3e) with acceptable spatial resolution and magnitude spectra with a very good signal-to-noise ratio (SNR) in all subjects. Figure 3a shows the profile of the usual 2D k-space weighting functions: standard Hanning, Hamming and optimized Hanning [24]. Optimized and standard 2D Hanning filters are shown in Fig. 3b and c, respectively. Figure 3d represents the fat image computed from the data set without application of k-space filter. From Fig. 3d it follows that the signal bleeding is a non-negligible problem although k-space sampling was increased to 64×64. Corresponding fat images of filtered data sets, using an optimized and standard 2D Hanning filters, are shown in Fig. 3e and f, respectively. Figures 3e, f demonstrate that k-filters reduce inter-voxel signal contamination, but decrease spatial resolution. In this study, a 2D optimized Hanning filter (Fig. 3b) was applied as a best compromise between signal bleeding suppression, spectrum quality and spatial resolution [24]. This filter increased the nominal voxel size by a factor 1.71 in both spatial directions [24]. Our “true” voxel shape could be best approximated as a cylinder with a diameter of 5.13 mm and height 15 mm (volume 0.31 cm3). From comparison of Fig. 3e and f, it is evident that standard Hanning filter produces significantly larger true voxel size than optimized Hanning. The representative spectra of a normal volunteer are shown in Fig. 5 and 6. The most pronounced lines are residual water (4.7 ppm), singlet of creatine methylene protons (3.9 ppm), TMA (3.2 ppm), total creatine (tCr) (3 ppm), broader EMCLCH2 line at ∼1.5 ppm, and IMCLCH2 at 1.3 ppm.

Figure 5 demonstrates the impact of k-space filters on the spectrum of soleus muscle. The spectrum computed without application of k-space filter (Fig. 5a) reveals increased EMCLCH2 line at ∼1.5 ppm and a little shoulder at ∼1.8 ppm. This is due to the signal bleeding from the neighboring voxels to the selected VOI. Signal bleeding was suppressed by optimized (Fig. 5b), as well as by standard Hanning filter (Fig. 5c), as follows from missing shoulder at ∼1.8 ppm and reduced EMCLCH2 line. From a comparison of Fig. 5b and c, it follows that optimized Hanning filter (Fig. 3b) sufficiently suppressed inter-voxel signal contamination because significant sharpening of k-space filter, by standard Hanning (Fig. 3c), did not lead to a further decrease in the EMCLCH2 line. In fact, EMCLCH2 intensity increased slightly (Fig. 5c) due to larger true voxel (VOI) size.

Figure 6 shows the reference fat spectrum computed from the central tibial bone marrow and muscles in the tibialis anterior, deep posterior compartment, gastrocnemius medialis, and soleus. The spectrum of the soleus muscle is completed by LCModel fits and residue. The mean soleus IMCL content of our six volunteers was 0.30±0.10 vol% (range: 0.18–0.46) or 3.6±1.2 mmol/kg wet weight (range: 2.1–5.4). The Cramér-Rao lower bounds (CRLB) of the LCModel fits of all evaluated soleus spectra were ≤2% and ≤7% for EMCLCH2 and IMCLCH2 lines, respectively. Slight different IMCL concentration 0.26±0.1 vol% (range: 0.17–0.42) reported previously [28] was caused by different chemical shift interval <0.2; 4>ppm used in LCModel data processing.

The volumes of interest V1–V5 (Fig. 7) were chosen in the regions with more complex BMS shifts of EMCLCH2 spectral lines. The purpose of such VOI selections was to find spectra with split EMCLCH2 lines and with increased BMS shifts. The left panel in Fig. 8 contains spectra computed without application of k-space filter during data processing. Middle and right panel show spectra computed from filtered data set using an optimized and standard 2D Hanning k-space filter, respectively. Figure 9 shows spectra from the VOI V3 together with their individual single-voxel spectra V3a-V3d. Whereas the left panel depicts spectra computed without k-filtering of raw data, middle and right panel contains spectra obtained from the filtered data set using an optimized and standard 2D Hanning k-filter, respectively. Impact of k-filter is in figures 8 and 9 visualized by significant spectral shape differences between “raw” spectra in the left panels and “k-filtered” spectra in middle and right panels. Differences between shapes of the spectra in middle and right panel can be explained by a larger true voxel size of the spectra computed from data sets filtered by a standard 2D Hanning k-filter compare to optimized Hanning filter. The spectra in Fig. 8 and 9 reveal splitting of the EMCLCH2 lines and their position shifts up to 0.4–0.5 ppm from the IMCLCH2 line. Such spectra were found in all volunteers.

thumbnail
Figure 8. Magnitude spectra computed from the VOIs V1–V5 (four voxels).

Left panel represents spectra computed without application of k-space filter during data processing. Middle and right panel depict spectra computed from filtered data set using an optimized and standard 2D Hanning k-space filter, respectively. Figure 6 shows VOIs.

https://doi.org/10.1371/journal.pone.0087533.g008

thumbnail
Figure 9. Magnitude spectra computed from the volume of interest V3 (four white voxels in Fig. 7) and its individual voxel spectra V3a (left up), V3b (right up), V3c (left bottom), and V3d (right bottom).

Left panel represents spectra computed without application of k-space filter during data processing. Middle and right panel depict spectra computed from filtered data set using an optimized and standard 2D Hanning k-space filter, respectively.

https://doi.org/10.1371/journal.pone.0087533.g009

Discussion

Slice selective EPSI of the human brain was performed in the majority of previous studies [20], [22], [23], [29][32]. Maximum size of spatial matrices was 32×32 and 32×32×16 in 2D and 3D experiments, respectively. Subcutaneous fat suppression was achieved by outer volume suppression (OVS) slices and/or by a broadband inversion pulse. Acquisition times varied between ∼8 and ∼30 minutes. With the recent improvements in gradient hardware performance, it was demonstrated, that spectral quality (linewidths) of brain spectra and SNR per unit time and unit volume were similar to those obtained with conventional phase-encoded PRESS MRSI techniques [23]. Low spectral intensities of brain metabolites hindered utilizing the speed of EPSI techniques. It should be noted, that readout bandwidth (echo sampling frequency), instead of spectral bandwidth, is the determining factor for SNR of EPSI techniques [33]. This leads to unavoidable loss in SNR compared to conventional MRSI with pure phase-encoding of signals and readout in the absence of imaging gradients.

EPSI techniques are, however, useful in the applications where the penalty paid in SNR loss is an acceptable trade for the considerable acquisition time reduction. Such applications are spectroscopy of lipids in skeletal muscle, liver, heart, etc. The net measurement time of our 2D PRESS EPSI sequence with two interleaves of gradient echo train was 12 min 48 sec for water-suppressed scans (4 acquisitions). However, the spectra of the bone marrow and muscles in Fig. 6 reveal SNR 114, and ≥28, respectively. SNR was computed by LCModel. The measurement time of water-suppressed scans could therefore be reduced by factor 2 (2 acquisitions). For comparison, minimum measurement time needed for conventional 2D MRSI (TR 1500 ms, spectral matrix 64×64, 1 acquisition) is 102 min 24 sec for non-water-suppressed and the same time for water-suppressed scans.

Our module spectra have sufficient spectral resolution to distinguish TMA from tCr, IMCLCH2 from EMCLCH2, and to detect EMCLCH2 splitting. TMA and tCr spectral intensities reveal differences between muscles (Fig. 6, 8, 9) and dipolar coupling effects, in agreement with previous studies [34], [35]. Both spectral lines are highest in soleus muscle and lowest in gastrocnemius medialis and tibialis anterior. Sensitivity of our EPSI sequence was not sufficient to detect TMA and tCr lines in all muscles due to small voxel size and a short acquisition time. TMA and tCr intensities were at the level of the noise in gastrocnemius medialis and tibialis anterior muscles, as follows from the individual voxel spectra shown in Fig. 4. Summation of such voxel spectra cannot improve the detectability of TMA and tCr.

Broad EMCLCH2 and narrow IMCLCH2 lines dominate in typical spectra of m. soleus (Fig. 5, 6). EMCLCH2 originates from extracellular fatty strands located between muscle fibers. It was suggested that these strands can be modeled by “infinite long” cylinders with the long axis at an angle θ relative to the direction of the static magnetic field B0 [5], [6]. According to this model, the EMCLCH2 shift ranges from 0.2 ppm to −0.1 ppm relative to IMCLCH2 corresponding to a change in orientation 0o ≤ θ ≤90o [5], [6]. EMCLCH2 position and linewidth are determined by the dominant angle θ and by a degree of dispersion in orientations of fatty strands within the volume of interest, respectively [36]. The model of long cylinders agrees well with the spectra obtained from the large volumes of interest (Fig. 5, 6). However, this model is unable to explain spectra originating from small VOIs shown in Fig. 8 and 9. EMCLCH2 signals are split into more well resolved peaks. Moreover, EMCLCH2 shifts are increased up to 0.4–0.5 ppm relative to IMCLCH2. These results are in agreement with the theoretical simulation and experimental evidence published by Steidle et al [7]. The authors found that EMCL compartments in soleus muscle experienced magnetic field deviations in the interval from ca 0.6 to −0.2 ppm in respect of IMCLCH2. It should be also noted that EMCLCH2 spectral line splitting and shifts up to ∼0.3 ppm relative to IMCLCH2 can be seen in the spectra of Hwang et al [12]. However, the authors did not point out or discuss these spectral features.

It should be emphasized that EMCLCH2 spectral line splits and their shifts up to 0.4–0.5 ppm relative to IMCLCH2 cannot be explained by signal bleeding from regions with strong lipid signals, in combination with B0 shifts in the voxels. High effectiveness of signal bleeding suppression using Hanning filters was demonstrated theoretically as well as experimentally in the previous studies [24], [37]. In addition, the significance of signal bleeding artefacts decreases with increasing size of acquisition matrix. In our case was signal bleeding negligible due to combination of large acquisition matrix 64×64 and application of the Hanning k-filter. Figures 3e, f, and spectra in Fig. 4, 5, 8, 9 demonstrate high effectiveness of the Hanning k-filters in signal bleeding suppression. Figure 4 shows BMS shifts of the EMCLCH2 lines in more detail. This phenomenon is best seen in the voxel spectra of the soleus muscle. While TMA, tCr, and IMCLCH2 intensities are in the correct positions (in line), EMCLCH2 peaks at the same voxels show splitting, dispersion and shifts in vertical (ppm) direction. Some EMCLCH2 sub-peaks can be seen between 0.4 and 0.5 ppm from the correspondent IMCLCH2 line (arrows in Fig. 4). This finding reveals that BMS effects are more complex than previously demonstrated. Spectra measured from the large VOIs (>1 cm3) reveal EMCLCH2 shifts up to 0.2 ppm. In such large volumes dominates long plate EMCL structures which resemble “infinite long” cylinders. However, very small and isolated EMCL clusters could exist in the lean muscle. These spatial structures do not resemble infinite long cylinders. We suppose that another kind of BMS effects could occur on the discontinuities or interfaces of such fat clusters.

Quantification of IMCL content was restricted to m. soleus because this muscle group was studied in many previous studies, and because of its center position between the flexible receiver coils, i.e. in the most homogeneous region of the receiver coils sensitivity. Mean IMCL concentration of our six normal volunteers was, 0.30±0.10 vol% or 3.6±1.2 mmol/kg, wet weight. This result agrees well with published values [12], [15], [16]. Despite good agreement, our result has to be taken with some caution because spectral intensities were not corrected for spatial sensitivity of the receiver coils. Measurements of coils sensitivities are challenging. In our case, this is because the mutual position of the flexible coils is different in each measurement session and the water/fat intensity ratio is different in each voxel. This problem could be circumvented by using a quadrature receiver knee coil. Such coil was not available in our laboratory at the time of this study.

To our knowledge, this is the first application of the EPSI technique in quantification of muscle lipids. PRESS localization was used only for slice selection. We were not forced to select PRESS box inside the slice because of negligible signal bleeding from subcutaneous fat and bone marrow. It enabled selection the bone marrow spectrum as the internal concentration reference. Combination of EPSI and PRESS VOI selection is relatively unusual. Only three research groups used the PRESS EPSI approach previously [38][41]. PRESS excitation improves VOI definition, reduces risk of signal contamination from outside the VOI, enables omitting inversion fat suppression pre-pulse and reduces the number of OVS slabs or even avoids them in spectroscopy of the brain [39], [41], prostate [40], heart [42], etc. More complicated data processing of EPSI techniques is in our application counterbalanced by high spatial resolution and by significantly shorter measurement time compared to conventional MRSI. The high spatial resolution enables noncontiguous and irregularly shaped VOIs (Fig. 5, 6). It was possible to improve the reliability of the muscle lipid estimations by avoiding the influence of relaxation effects because the fat (bone marrow) spectral line was used as the internal concentration reference.

Conclusion

The EPSI sequence with PRESS volume of interest localization was developed and implemented in a 3T scanner. The use of a large spectral matrix 64×64, together with the optimized Hanning k-filter significantly reduced inter-voxel signal contamination of the strong EMCL signals from subcutaneous fat, bone marrow or layers of adipose tissues. The small voxels enabled visualisation of EMCL methylene (-CH2-)n spectral lines splitting and their BMS shifts up to 0.5 ppm relative to correspondent IMCL methylene line. This study demonstrates that high-spatial resolution PRESS EPSI of the muscle lipids is feasible on clinical scanners.

Author Contributions

Conceived and designed the experiments: JW MB HA. Performed the experiments: JW FO-N MB. Analyzed the data: JW FO-N HA. Wrote the paper: JW HA. Designed the software used in analysis: JW. Idea of muscle lipids quantification: HA.

References

  1. 1. Ren J, Dimitrov I, Sherry AD, Malloy CR (2008) Composition of adipose tissue and marrow fat in humans by 1H NMR at 7 Tesla. J Lipid Res 49: 2055–2062.
  2. 2. Szczepaniak LS, Babcock EE, Schick F, Dobbins RL, Garg A, et al. (1999) Measurement of intracellular triglyceride stores by 1H spectroscopy: validation in vivo. Am J Physiol 276: E977–E989.
  3. 3. Chu SCK, Xu Y, Balshi JA, Springer CS (1990) Bulk magnetic susceptibility shifts in NMR studies of compartmentalized samples: use of paramagnetic reagents. Magn Reson Med 13: 239–262.
  4. 4. Boesch C, Machann J, Vermathen P, Schick F (2006) Role of proton MR for the study of muscle lipid metabolism. NMR Biomed 19: 968–988.
  5. 5. Szczepaniak LS, Dobbins RL, Stein DT, McGarry JD (2002) Bulk magnetic susceptibility effects on the assessment of intra- and extramyocellular lipids in vivo. Magn Reson Med 47: 607–610.
  6. 6. Boesch C, Slotboom J, Hoppeler H, Kreis R (1997) In vivo determination of intra-myocellular lipids in human muscle by means of localized 1H-MR-Spectroscopy. Magn Reson Med 37: 484–493.
  7. 7. Steidle G, Machann J, Claussen CD, Schick F (2002) Separation of intra- and extramyocellular lipid signals in proton MR spectra by determination of their magnetic field distribution. J Magn Reson 154: 228–235.
  8. 8. Krssak M, Petersen KF, Dresner A, DiPietro L, Vogel SM, et al. (1999) Intramyocellular lipid concentrations are correlated with insulin sensitivity in humans: a 1H-NMR spectroscopy study. Diabetologia 42: 113–116.
  9. 9. Sinha R, Dufour S, Petersen KF, LeBon V, Enoksson S, et al. (2002) Assessment of skeletal muscle triglyceride content by 1H nuclear magnetic resonance spectroscopy in lean and obese adolescent: relationships to insulin sensitivity, total body fat, and central adiposity. Diabetes 51: 1022–1027.
  10. 10. Jacob S, Machann J, Rett K, Brechtel K, Volk A, et al. (1999) Association of increased intramyocellular lipid content with insulin resistance in lean non-diabetic offspring of type 2 diabetic subjects. Diabetes 48: 1113–1119.
  11. 11. Hu J, Willcott MR, Moore GJ (1997) Two-dimensional proton chemical-shift imaging of human muscle metabolites. J Magn Reson 126: 187–192.
  12. 12. Hwang JH, Pan JW, Heydari S, Hetherington HP, Stein DT (2001) Regional differences on intramyocellular lipids in humans observed by in vivo 1H-MR spectroscopic imaging. J Appl Physiol 90: 1267–1274.
  13. 13. Larson-Meyer DE, Newcomer BR, Hunter GR (2002) Influence of endurance running and recovery diet on intramyocellular lipid content in women: a 1H NMR study. Am J Physiol Endocrinol Metab 282: E95–E106.
  14. 14. Vermathen P, Kreis R, Boesch C (2004) Distribution of intra-myocellular lipids in human calf muscles determined by MR spectroscopic imaging. Magn Reson Med 51: 253–262.
  15. 15. Vermathen P, Saillen P, Boss A, Zehnder M, Boesch C (2012) Skeletal muscle 1H MRSI before and after prolonged excercise. I. Muscle specific depletion of intramyocellular lipids. Magn Reson Med 68: 1357–1367.
  16. 16. Weis J, Johansson L, Courivaud F, Karlsson FA, Ahlström H (2007) Quantification of intramyocellular lipids in obese subjects using spectroscopic imaging with high spatial resolution. Magn Reson Med 57: 22–28.
  17. 17. Newcomer BR, Lawrence JC, Buchthal S, den Hollander JA (2007) High-resolution chemical shift imaging for the assessment of intramuscular lipids. Magn Reson Med 57: 853–858.
  18. 18. Mansfield P (1984) Spatial mapping of the chemical shift in NMR. Magn Reson Med 1: 370–386.
  19. 19. Matsui S, Sekihara K, Kohno H (1985) High-speed spatially resolved high-resolution NMR spectroscopy. J Am Chem Soc 107: 2817–2818.
  20. 20. Posse S, Tedeschi G, Risinger R, Ogg R, Bihan DL (1995) High speed 1H spectroscopic imaging in human brain by echo planar spatial-spectral encoding. Magn Reson Med 33: 34–40.
  21. 21. Ericsson A, Weis J, Sperber GO, Hemmingsson A (1995) Measurements of magnetic field variations in the human brain using a 3D-FT multiple gradient echo technique. Magn Reson Med 33: 171–177.
  22. 22. Ebel A, Soher BJ, Maudsley AA (2001) Assessment of 3D proton MR echo-planar spectroscopic imaging using automated spectral analysis. Magn Reson Med 46: 1072–1078.
  23. 23. Otazo R, Mueller B, Ugurbil K, Wald L, Posse S (2006) Signal-to-noise ratio and spectral linewidth I mprovements between 1.5 and 7 Tesla in proton echo-planar spectroscopic imaging. Magn Reson Med 56: 1200–1210.
  24. 24. Pohmann R, von Kienlin M (2001) Accurate phosphorus metabolite images of the human heart by 3D acquisition-weighted CSI. Magn Reson Med 45: 817–826.
  25. 25. Weis J, Ericsson A, Hemmingsson A (1997) 1H-spectroscopic imaging with read gradient during acquisition in inhomogeneous fields: analysis, measurement strategy and data processing. MAGMA 5: 201–212.
  26. 26. Weis J, Ericsson A, Hemmingsson A (1999) Chemical shift artifact-free microscopy: spectroscopic microimaging of the human skin. Magn Reson Med 41: 904–908.
  27. 27. Provencher SW (1993) Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med 30: 672–679.
  28. 28. Weis J, Bruvold M, Ortiz-Nieto F, Ahlström H (2012) High-Resolution Echo-Planar Spectroscopic Imaging of Human Calf. In Proceedings of the 19th Annual Meeting of ISMRM, Melbourne (abstract 1625).
  29. 29. Guimaraes AR, Baker JR, Jenkins BG, Lee PL, Weisskoff RM, et al. (1999) Echo planar chemical shift imaging. Magn Reson Med 41: 877–882.
  30. 30. Mathiesen HK, Tscherning T, Sorensen PS, Larsson HBW, Rostrup E, et al. (2005) Multi-slice echo-planar spectroscopic MR imaging provides both global and local metabolite measures in multiple sclerosis. Magn Reson Med 53: 750–759.
  31. 31. Maudsley AA, Darkazanli A, Alger JR, Hall LO, Schuff N, et al. (2006) Comprehensive processing, display and analysis for in vivo MR spectroscopic imaging. NMR Biomed 19: 492–503.
  32. 32. Posse S, Otazo R, Caprihan A, Bustillo J, Chen H, et al. (2007) Proton echo-planar spectroscopic imaging of J-coupled resonances in human brain at 3 and 4 Tesla. Magn Reson Med 58: 236–244.
  33. 33. Mulkern RV, Panych LP (2001) Echo planar spectroscopic imaging. Concepts Magn Reson 13: 213–237.
  34. 34. Vermathen P, Boesch C, Kreis R (2003) Mapping fiber orientation in human muscle by proton MR spectroscopic imaging. Magn Reson 49: 424–432.
  35. 35. Hu J, Xia Y, Feng W, Xuan Y, Shen Y, et al. (2005) Orientational dependence of trimethyl ammonium signal in human muscles by 1H magnetic resonance spectroscopic imaging. Magn Reson Imaging 23: 97–104.
  36. 36. Khuu A, Ren J, Dimitrov I, Woessner D, Murdoch J, et al. (2009) Orientation of lipid strands in the extracellular compartment of muscle: effect on quantitation of intramyocellular lipids. Magn Reson Med 61: 16–21.
  37. 37. Scheenen TWJ, Klomp DWJ, Röll SA, Fütterer JJ, Barentsz JO, et al. (2004) Fast acquisition-weighted three-dimensional proton MR spectroscopic imaging of the human prostate. Magn Reson Med 52: 80–88.
  38. 38. Lipnick S, Verma G, Ramadan S, Furuyama J, Thomas MA (2010) Echo planar correlated spectroscopic imaging: Implementation and pilot evaluation in human calf in vivo. Magn Reson Med 64: 947–956.
  39. 39. Cunningham CH, Vigneron DB, Chen AP, Xu D, Nelson SJ, et al. (2005) Design of flyback echo-planar readout gradients for magnetic resonance spectroscopic imaging. Magn Reson Med 54: 1286–1289.
  40. 40. Chen AP, Cunningham CH, Ozturk-Isik E, Xu D, Hurd RE, et al. (2007) High-speed 3T MR spectroscopic imaging of prostate with flyback echo-planar encoding. J Magn Reson Imaging 25: 1288–1292.
  41. 41. Furuyama JK, Wilson NE, Thomas MA (2012) Spectroscopic imaging using concentrically circular echo-planar trajectories in vivo. Magn Reson Med 67: 1515–1522.
  42. 42. Weis J, Bruvold M, Ortiz-Nieto F, Ahlström H (2012) Assessment of myocardial triglycerides by 2D PRESS echo-planar spectroscopic imaging. In Proceedings of the 19th Annual Meeting of ISMRM, Melbourne (abstract 1125).