A Micro CT Study in Patients with Breast Microcalcifications Using a Mathematical Algorithm to Assess 3D Structure

  • David Kenkel ,

    david.kenkel@usz.ch

    Affiliation Department of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland

  • Zsuzsanna Varga,

    Affiliation Department of Pathology and Molecular Pathology, University Hospital Zurich, Zurich, Switzerland

  • Heike Heuer,

    Affiliation Department of Gynecology, University Hospital Zurich, Zurich, Switzerland

  • Konstantin J. Dedes,

    Affiliation Department of Gynecology, University Hospital Zurich, Zurich, Switzerland

  • Nicole Berger,

    Affiliation Department of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland

  • Lukas Filli,

    Affiliation Department of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland

  • Andreas Boss

    Affiliation Department of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland

Abstract

Purpose

The aim of this study was to evaluate the relevance of the three-dimensional (3D) structure of breast microcalcifications (MC) as a predictor of malignancy using highly resolved micro-computed tomography (micro-CT) datasets of biopsy samples.

Material and Methods

The study included 28 women with suspicious MC in their mammogram undergoing vacuum-assisted biopsy. Directly after the intervention, the specimens were scanned in a micro-CT with an isometric spatial resolution of 9 μm. Datasets were analysed regarding the number, volume and morphology of suspicious non-monomorphic MC (fl—fine linear, fp—fine pleomorphic, ch—coarse heterogeneous) and the structure model index (SMI). Histological evaluation was performed according to the B-classification: normal tissue or benign (group A: B1, B2), unclear malignant potential or suspicious of malignancy (group B: B3, B4) and malignant lesions (group C: B5).

Results

In all groups, suspicious non-monomorphic MC were found: group A exhibited fp MC in 38.5% of samples, no fl/ch; group B: fl 14.3%, fp 28.6%, ch 14.3%; group C always had at least one type of suspicious non-monomorphic MC (fl (57.1%) or fp (57.1%)) in each sample. The different histologic groups showed a similar mean SMI (benign: 2.97 ± 0.31, malignant: 3.02 ± 0.10, unclear: 2.90 ± 0.28). Between the three groups, no significant differences were found regarding number, volume or SMI value of MC.

Conclusion

3D structure based on the SMI of MC analysed with highest spatial resolution is not significantly associated with the B-classification of breast lesions. Thus, magnification views of MC may be omitted in the analysis of MC detected in mammograms.

Introduction

Breast cancer is the most frequent malignancy in women, with yearly 1.38 million new breast cancer cases worldwide representing 10.9% of all new diagnosed cancer [1]. The World Health Organization (WHO) predicts that the importance of breast cancer is likely to increase in the next decades in developing and developed countries due to the increased incidence of breast cancer caused by increased life expectancy [2]. These circumstances urge an improvement in early diagnosis, firstly to improve the outcome of patients and secondly to reduce the healthcare costs related to breast cancer therapy. In Canada, for instance, the average lifetime cost per breast cancer case was higher at higher stages ($36,340 for stage IV) compared to lower stages ($23,275 for stage I) [3].

Mammography screening is the current gold standard for the early detection of breast cancer. The evaluation of mammograms is standardized according to the American College of Radiology (ACR) Breast Imaging Reporting and Data System (BIRADS) classification. One of the key findings for the detection of suspicious breast lesions and an important sign of malignancy is microcalcifications (MC) [4, 5]. Typically, the shape of the MC in breast lesions is used to classify the finding as benign or malignant [69]. According to the BIRADS classification, there are three categories of MC including typically benign, MC of intermediate concern and calcification with a high probability of malignancy. Benign calcifications are mostly larger, calcifications of intermediate concern are coarse heterogeneous and typical malignant MC are fine pleomorphic and fine linear [10].

Mammography is the only screening method that has been proved to be efficient and cost-effective [1113]. The early diagnosis of breast cancer causes a decrease in the mortality associated with the disease [14]. A meta-analysis including 11 randomized trials showed that the relative risk reduction of women taking part in the screening compared to a control group not undergoing mammography screening is 20% (95% CI 0.73–0.89) [15].

The aim of this study is to evaluate whether mathematical modelling of the three-dimensional (3D) structure of MC may serve as an objective predictor of malignancy as proved by minimal invasive breast biopsy histological diagnostics.

Methods

Subjects

Patients with MC classified as BIRADS 2–5 without corresponding findings in the ultrasound examinations were prospectively included in this study. Twenty-nine biopsy specimens of 28 women (age range 50–86 years, mean age 60.5 ± 8.8 years) were evaluated. Informed consent was obtained from each patient; all patients signed a written informed consent that their tissue can be used for further imaging and histological research. The Department of Pathology and Molecular Pathology, University Hospital Zurich included this study into a larger breast pathology project (KEK-2012-554). There was no need for a specific approval of this subproject by the local ethical committee, as informed consent was available in each patient and only the histological diagnosis of the minimal invasive breast biopsies were included in the study without any further analyses on the breast tissues.

Mammography

The patients were examined using a mammography system (MAMMOMAT Inspiration; Siemens AG, Healthcare Sector, Erlangen, Germany) in the two standard projections: mediolateral oblique view and craniocaudal view. All mammographies were classified according to the BIRADS classification [7]. Prior to biopsy, mammographies were discussed by an interdisciplinary board including radiologists and gynaecologists; after biopsy, histological results were discussed by an interdisciplinary board between pathologists, radiologists and gynaecologists.

Breast biopsy

Patients underwent vacuum-assisted stereotactic breast biopsy using an EnCor Enspire breast biopsy system (Bard, Tempe) with a needle size of 7 gauge or 10 gauge. Patients were rested in a prone position on the biopsy table. The biopsies were executed by a senior attending physician from the Department of Gynaecology or the Department of Radiology. Immediately after extraction, samples were fixed in formalin.

Micro-CT imaging

Each specimen was removed from the formalin-filled tube and covered with plastic wrap in order to protect the sample from drying during measurement. Imaging was performed on a micro-CT system (SkyScan 1176, Bruker microCT, Kontich, Belgium). The biopsies were positioned in the isocentre of the micro-CT bed and stabilized with adhesive tape. Images were acquired after a localizer scan using the following scan parameters: rotation step 0.5°, tube voltage 90 kV, tube current 280 μA, covered angle 360°, copper filter 0.1 mm, voxel size 9 × 9 × 9 μm3. The total scanning time for each sample was 120 min. Images were then reconstructed using a modified Feldkamp cone-beam algorithm (NRecon, SkyScan/Bruker microCT, Kontich, Belgium). The applied MicroCT Skyscan 1176 uses a Feldkamp filtered backprojection reconstruction for cone-beam CTs. In principle, the reconstruction of each single image is performed in two dimensions, however, as a volume dataset is acquired the 2D slice are reconstructed with isotropic voxel size.

Image analysis

After reconstruction, different morphological parameters were calculated using CTAn/CTVol (SkyScan/Bruker microCT) including the object volume and SMI. For assessment of the 3D structure of the MC the structure model index (SMI) was calculated. SMI is a dimensionless quantity, that is defined according to the following equation of Hildebrand et al [16]: where S is the object surface area before dilation, S’ is the change in surface area caused by dilation and V is the undilated object volume. SMI ranges from 0 to 4: a value of 0 signifying a plate-like structure, 3 a cylindrical structure and 4 a sphere. Intermediate values indicate a mixed structure. Further, for each sample the mean object volume and the diameter assuming spherical and cubic shape were calculated.

In analogy to the BIRADS classification, the MC were categorized according to the following scheme: fine linear (fl) MC defined as calcifications with a mean estimated diameter (cube) of < 200 μm and an SMI of 2.92–3.08, fine pleomorphic (fp) with a mean estimated diameter (cube) of < 250 μm and a standard deviation (SD) of the SMI of more than 0.29, coarse heterogeneous (ch) defined as calcifications with an estimated diameter > 500 μm and an SD of the SMI > 0.5 and no suspicious MC defined if none of the other definitions is accurate.

Histological evaluation

The specimens were fixed in 4% formaldehyde and processed and stained according to the standard procedures including serial sections and hematoxylin and eosin staining. Histological diagnoses were made based on the current B-classification [17]. The B1 category indicates normal tissue that might show MC, e.g. within involutional lobules. The B2 category includes benign lesions such as fibroadenomas, fibrocystic changes, ductectasia and sclerosing adenosis. B3 lesions are of uncertain malignant potential including atypical ductal hyperplasia (ADH), classical lobular neoplasia (LN), phylloides tumours and papillary lesions. The B4 category contains suspicious lesions for malignancy, but not affirmative diagnostic for malignancy. Lesions categorized as B5 lesions are definitely malignant [18, 19].

Statistical analysis

The mean and SD of the size and the morphological parameters were calculated. Data were tested for normal distribution using the Kolmogorov–Smirnov (KS-test). The means of the morphological parameters of the different histological groups were compared using an independent samples Kruskal–Wallis test and the chi-square test was used to compare the amount of BIRADS-MC with the histological groups, including those which were histologically proven to be normal or benign (group A: B1 and B2), of uncertain potential or suspicious (group B: B3 and B4) and malignant (group C: B5). Due to the multiple comparisons, a Bonferroni correction was performed. The data were analysed using IBM SPSS-Statistics version 22. A p-value of less than 0.05 was considered to be significant.

Results

Subjects and histology

In 29 biopsies from 28 patients 829 MC were detected resulting in a mean number of MC of 28.6 ± 34.7 per specimen. The majority of the examined patients were radiologically considered to be BIRADS 4 (n = 25), two were assessed as BIRADS 5, one lesion was classified as BIRADS 2 and one lesion as BIRADS 3. The histology showed that 8 of 27 biopsies were malignant (29.6%). Table 1 gives an overview of the patients age, the histological (B-classification) and radiological (BIRADS) evaluation. For the morphologic evaluation, two samples containing just one MC were excluded. At the interdisciplinary board these biopsies were considered to be non-representative since more than one MC was observed on the mammograms.

thumbnail
Table 1. Patient age and overview of the histologic (B-classification) and radiologic (BIRADS) evaluation.

https://doi.org/10.1371/journal.pone.0169349.t001

Morphological evaluation using micro-CT

Table 2 reports the results of the morphological parameters (number of MC, mean object volume, morphology, mean SMI) for each sample. No significant differences between benign (group A) lesions, lesions with unclear malignant potential (group B) and malignant (group C) lesions were observed regarding the number of MC (P = 0.113), object volume (P = 0.881), mean SMI (P = 0.756) or the SD of the SMI (P = 0.976). Table 3 and Fig 1 show the results of the different morphological parameters including number of MC, object volume and mean SMI for group A–C lesions. The different histologic groups exhibited a similar mean SMI (group A: 2.97 ± 0.31, group B: 2.90 ± 0.28, group C: 3.02 ± 0.1) and SD of the SMI (group A: 0.32 ± 0.19, group B: 0.32 ± 0.18, group C: 0.31 ± 0.11. Moreover, group C lesions (33.6 ± 12.8) showed a tendency to comprise fewer calcifications than group B lesions (41.4 ± 52.7) and more than group A lesions (23.2 ± 32.6), without reaching significance.

thumbnail
Fig 1. Overview of different morphological parameters including mean number, volume (μm3) and SMI of MC for all different groups (benign: including B-classification B1 and B2; unclear malignant potential: including B-classification B3 and B4; malignant lesions: including B-classification B5).

Error bars indicate the standard error of the mean (SEM).

https://doi.org/10.1371/journal.pone.0169349.g001

thumbnail
Table 2. Morphological parameters including object volume (μm3x105) and the structure model index (±SD) of every specimen.

Morphology: fl: fine linear, fp: fine pleomorphic, ch: coarse heterogeneous, ns: no suspicious MC, 1 indicating presence and 0 indicating absence of the respective type of microcalcification.

https://doi.org/10.1371/journal.pone.0169349.t002

thumbnail
Table 3. Overview of the results of different morphological parameters including mean SMI, SD SMI, mean number of calcifications, the object volume for histologically proven benign, malignant and unclear biological potential lesions.

Moreover the presence of the morphological MC types in the different groups is reported.

https://doi.org/10.1371/journal.pone.0169349.t003

Categorization of MC

Table 2 reports the presence of different types of MC for each specimen and Table 3 provides an overview for group A–C lesions. Group A lesions showed no fine linear or coarse heterogeneous MC; fine pleomorphic were found in 38.5% (5/13) of the samples and no suspicious MC were found in 61.5% (8/13). Group B lesions exhibited a more heterogeneous distribution of MC types than group A and C lesions: fine linear in 14.3% of samples (1/7), fine pleomorphic in 28.6% (2/7), coarse heterogeneous in 14.3% (1/7) and no suspicious calcification in 42.9% (3/7) of the samples. Group C lesions always showed at least one type of suspicious MC: 57.1% of the samples exhibited fine linear (4/7) and 57.1% showed fine pleomorphic MC (4/7).

A chi-square test demonstrated that the absence of suspicious MC between groups A–C was not significant after Bonferroni correction (P = 0.028); however, the presence of fine linear MC was significantly different between groups A–C (P = 0.007). No significant differences were observed regarding the presence of fine pleomorphic (P = 0.539) and coarse heterogeneous MC (P = 0.227) between the groups.

Fig 2 shows representative 3D images from different biopsies from group A (A–C), group B (D–F) and group C lesions (G–I). Typically, group C lesions (H and I) exhibited fine linear MC more often than group A lesions, which never showed a fine linear pattern (A–C). Group B lesions showed the broadest variety of MC including all kinds of MC and also lesions with no suspicious MC. In Fig 3, it is demonstrated that the different types of MC can easily be assessed in micro-CT images due to the excellent spatial resolution.

thumbnail
Fig 2. Representative 3D images from different biopsies.

Benign (group A: A–C) lesions, unclear biological potential lesions (group B: D–F) and malignant (group C: G–I) lesions: A (sample number 1, fine pleomorphic), B (sample number 6, no suspicious MC), C (sample number 7, no suspicious MC), D (sample number 15, coarse heterogeneous), E (sample number 16, fine pleomorphic), F (sample number 20, fine linear), G (sample number 22, fine linear), H (sample number 25, fine linear) and I (sample number 24, fine pleomorphic). For more details concerning the different samples, compare sample numbers in Tables 1 and 2.

https://doi.org/10.1371/journal.pone.0169349.g002

thumbnail
Fig 3. Representative 3D images and mammograms from two different patients.

Left column: sample number 6 (fibrosis, sclerosing adenosis, normal ductal hyperplasia, intraductal papilloma), only spherical MC classified as no suspicious MC; right column: sample number 25 (DCIS high grade), fine linear MC. CC indicates craniocranialcaudal view and MLO indicates mediolateral oblique view.

https://doi.org/10.1371/journal.pone.0169349.g003

Discussion

In the present study, we examined the form and morphological characteristics of MC detected by mammography. Samples from vacuum-assisted stereotactic breast biopsy underwent micro-CT measurements with a highest spatial resolution of 9 μm; the MC were 3D-segmented with subsequent morphological analysis using an automatic algorithm providing the number of MC, volume and SMI. Moreover, MC were categorized as suspicious MC exhibiting a fine linear, fine pleomorphic or coarse heterogeneous pattern. None of the morphological parameters or any of the types of suspicious MC showed a statistical correlation to the B-classification type of the lesion. Therefore, the 3D shape of the MC does not allow benign lesions to be distinguished from lesions with unknown malignant potential or breast cancers.

In our study, 93% of the examined patients showed lesions radiologically classified as BIRADS 4 and 5; in the histological examination of the biopsy samples 29.6% were classified as B5 lesions corresponding to a positive predictive value (PPV) of suspicious MC in mammograms of approximately 30% at our institution. This estimated PPV is in accordance with other studies reporting a PPV in the order of 30–50% for MC undergoing biopsy [4, 5]; therefore, it may be assumed that a typical mammography screening cohort was evaluated in this prospective analysis.

Only one other study assessing the 3D shape of MC using micro-CT has been previously published, by Willekens et al [20]. Compared to the Willekens study, we applied a substantially higher isotropic spatial resolution of 9 μm compared to 35 μm. Further, Willekens et al excluded MC that were smaller than three micro-CT voxels, only analysing calcifications larger than 105 μm. Furthermore, a nearly three times larger sample size was included in our study, compared to the 11 samples evaluated in the Willekens study. Some of our findings are in line with the data of Willekens et al: Patients with histologically proven benign lesions had fewer MC than those with malignant lesions; this finding was, however, not significant in our study. Likewise, our study demonstrated that SMI values do not allow distinction between malignant and benign lesions.

The category of B3 lesions comprises a heterogeneous group of different histological lesion types including atypical hyperplasia (AH) with ADH and classical LN including atypical lobular hyperplasia (ALH) and classical lobular carcinoma in situ, flat epithelial atypia (FEA), phylloides tumour, papillary lesions and radial scars. Often, B3 lesions undergoing resection are later upgraded to B5 lesions because of a small invasive focus in the removed tissue [19, 21]. The B3 and B4 lesions included in our study showed a wide variety of different types of MC with one type of suspicious MC. However, two preliminary aspects were discovered for which a high resolved depiction of MC might provide additional helpful information. Statistically significant differences between groups A and C were detected regarding the presence of fine linear MC. Fine linear MC were only present in groups B and C; therefore, the subtype of fine linear MC seems to be a potential predictor of malignancy or lesions of unknown malignant potential. Moreover, each malignant lesion showed at least one type of suspicious MC, whereas benign lesions in the majority of cases did not show suspicious MC, demonstrating that the absence of suspicious MC might be a weak predictor for the benignancy of a lesion. In our study these differences were not significant; this could be attributed to a sample size that was too low to detect such rather weak correlations.

Several approaches have been undertaken to improve the assessment of MC including X-ray magnification views [4, 6, 2229] and computer-aided detection with partially successful results: the contrast-limited adaptive histogram equalization (CLAHE) technique for example showed promising results for the computer-aided detection in digital mammograms with a sensitivity of 92.54% and a specificity of 92.50% [30]. However, assessment of the real 3D structure of MC remains challenging in 2D projection views. A technique offering the possibility of obtaining a 3D view is breast tomosynthesis [31]. If tomosynthesis could reach a resolution high enough to distinguish fine linear MC from fine pleomorphic MC, an increase of the PPV of X-ray breast imaging with a reduction of over-diagnosis and avoidance of unnecessary biopsies might become feasible. Although tomosynthesis might be a promising tool in the future it has to be taken into account that, in general, CT images suffer from noise and artifacts, especially in low dose mode. Therefore, previously published special low dose imaging algorithms should be implemented [3236].

Our study also has limitations: (a) a relatively small sample size was evaluated with heterogeneous findings regarding histology. A larger cohort might provide a deeper insight into whether the fine linear MC subtype exhibits a predictive value for the presence of malignancy. It would be desirable to have a database like the Mammographic Image Analysis Society (MIAS) database of digital mammograms. This database contains 322 digitizeddigitised films. To the best of our knowledge there is no public micro-CT database of breast MC. However, we believe that a database would be desirable for the assessment of MC since it would allow the evaluation of a higher sample size. (b) Further, we only assessed the micro-CT images alone. Consecutively, other predictors of malignancy like the distribution of calcification, which is just visible on the mammograms, were not taken into account. (c) We only assessed the Structural Model Index SMI. However, it has to be mentioned that there are many other descriptors for the 3D structure, e.g. fractal dimension [37]. However, the evaluation of other assessment measures was beyond the scope of this study.

In conclusion, we showed that the shape (based on the SMI) of MC assessed even with highest spatial resolution using micro-CT is not significantly associated with the B-classification of the underlying lesion, especially regarding the presence of malignancy. Taken together, it seems to be justified to omit magnification views for the assessment of MC.

Author Contributions

  1. Conceptualization: DK ZV HH KJD NB LF AB.
  2. Data curation: DK.
  3. Formal analysis: DK.
  4. Investigation: DK ZV HH KJD AB.
  5. Methodology: DK ZV HH KJD NB LF AB.
  6. Project administration: DK ZV AB.
  7. Resources: DK ZV AB.
  8. Supervision: DK ZV HH KJD NB LF AB.
  9. Validation: DK ZV HH KJD NB LF AB.
  10. Visualization: DK ZV HH KJD NB LF AB.
  11. Writing – original draft: DK AB.
  12. Writing – review & editing: DK ZV HH KJD NB LF AB.

References

  1. 1. Ferlay JSH, Bray F, Forman D, Mathers C, Parkin DM. Estimates of worldwide burden of cancer in 2008: GLOBOCAN 2008. Int J Cancer. 2010:2893–917.
  2. 2. Available from: http://www.who.int/cancer/detection/breastcancer/en.
  3. 3. Will BP, Berthelot JM, Le Petit C, Tomiak EM, Verma S, Evans WK. Estimates of the lifetime costs of breast cancer treatment in Canada. European Journal of Cancer. 2000;36(6):724–35. http://dx.doi.org/10.1016/S0959-8049(99)00340-8. pmid:10762744
  4. 4. Fallenberg EM, Dimitrijevic L, Diekmann F, Diekmann S, Kettritz U, Poellinger A, et al. Impact of magnification views on the characterization of microcalcifications in digital mammography. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2014;186(3):274–80. pmid:23999780
  5. 5. Weigel S, Decker T, Korsching E, Hungermann D, Böcker W, Heindel W. Calcifications in Digital Mammographic Screening: Improvement of Early Detection of Invasive Breast Cancers? Radiology. 2010;255(3):738–45. pmid:20501713
  6. 6. Gruner JH, Berndt G, Borchert B, Farber A, Gmelin E, Ratmann R et al. DIMA enlargement mammography in microcalcifications: a prospective study with ROC analysis. Eur Radiol. 2001:284–91. pmid:11218029
  7. 7. Elizabeth S. Burnside M, Fowler J, Salkowski R, Rubin L, Gale A et al. Use of Microcalcification Descriptors in BI-RADS 4th Edition to Stratify Risk of Malignancy. Radiology. 2007;242:388–95. pmid:17255409
  8. 8. Priscilla J, Slanetz M, Giardino AA, Oyama T, Koerner FC, Kopans B et al. Mammographic Appearance of Ductal Carcinoma In Situ Does Not Reliably Predict Histologic Subtype. The Breast Journal. 2001;7:417–21. pmid:11843854
  9. 9. Rominger M, Wisgickl C, Timmesfeld N. Breast Microcalcifications as Type Descriptors to Stratify risk of Malignancy: a Systematic Review and Meta-Analysis of 10665 Cases with Special Focus on Round/Punctate Microcalcifications. Fortschr Röntgenstr. 2012;184(12):1144–52. Epub 24.08.2012.
  10. 10. Muttarak M KP, Sukhamwang N. Breast calcifications: which are malignant? Singapore Med J. 2009:907–14. pmid:19787181
  11. 11. Sankatsing VDV, Heijnsdijk EAM, van Luijt PA, van Ravesteyn NT, Fracheboud J, de Koning HJ. Cost-effectiveness of digital mammography screening before the age of 50 in The Netherlands. International Journal of Cancer. 2015;137(8):1990–9. pmid:25895135
  12. 12. Zelle SG, Vidaurre T, Abugattas JE, Manrique JE, Sarria G, Jeronimo J, et al. Cost-Effectiveness Analysis of Breast Cancer Control Interventions in Peru. PLoS ONE. 2013;8(12):e82575. pmid:24349314
  13. 13. Feig S. Cost-Effectiveness of Mammography, MRI, and Ultrasonography for Breast Cancer Screening. Radiologic Clinics of North America. 2010;48(5):879–91. http://dx.doi.org/10.1016/j.rcl.2010.06.002. pmid:20868891
  14. 14. Murray PJ, Wivell G, Denton E. Breast cancer screening and diagnosis in the 21st century within the UK. Post Reproductive Health: The Journal of The British Menopause Society. 2015.
  15. 15. Marmot MG, Altman DG, Cameron DA, Dewar JA, Thompson SG, Wilcox M. The benefits and harms of breast cancer screening: an independent review. Br J Cancer. 2013;108(11):2205–40. pmid:23744281
  16. 16. Hildebrand P. Quantification of Bone Microarchitecture with the Structure Model Index. Computer Methods in Biomechanics and Biomedical Engineering. 1997;1(1):15–23. pmid:11264794
  17. 17. Rageth CJ, O’Flynn EA, Comstock C, Kurtz C, Kubik R, Madjar H, et al. First International Consensus Conference on lesions of uncertain malignant potential in the breast (B3 lesions). Breast Cancer Research and Treatment. 2016;159(2):203–13. pmid:27522516
  18. 18. AWMF. Mammakarzinom der Frau; Diagnostik, Therapie und Nachsorge. S3 Leitlinie 2012.
  19. 19. Ellis IO, Humphreys S, Michell M, Pinder SE, Wells CA, Zakhour HD. Best Practice No 179: Guidelines for breast needle core biopsy handling and reporting in breast screening assessment. Journal of Clinical Pathology. 2004;57(9):897–902. pmid:15333647
  20. 20. Willekens I, Casteele EVd, Buls N, Temmermans F, Jansen B, Deklerck R, et al. High-resolution 3D micro-CT imaging of breast microcalcifications: a preliminary analysis. BMC Cancer. 2014.
  21. 21. Grunert JH, Borchert B, Farber A, Gmelin E. Magnified Radiographs as Quality Control for Detection of Microcalcifications in Histopathology of Breast Biopsies. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2000:68–72.
  22. 22. Moritz JD, Mertens C, Westerhof JP, Oestmann JW. Role of high magnification specimen radiography in surgical and core biopsies of the breast. The British journal of radiology. 2000:1170–7. pmid:11144794
  23. 23. Grunert JH, Farber A, Gmelin E. Detection of micro calcifications in breast specimen by 4-fold DIMA direct magnification radiography compared to 1.5-fold conventional magnification radiography. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 1999:302–6. pmid:10598166
  24. 24. Diekmann F, Bick U, Rogalla P, Blohmer JU, Winzer KJ, Hamm B. Comparing the visualization of microcalcifications with direct magnification in digital full-field mammography vs. film-screen mammography. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2002:297–300. pmid:11885006
  25. 25. Geller BM, Vacek PM, Skelly J, Harvey SC. The use of additional imaging increased specificity and decreased sensitivity in screening mammography. Journal of clinical epidemiology. 2005;58(9):942–50. pmid:16085198
  26. 26. Law J. Breast dose from magnification films in mammography. The British journal of radiology. 2005;78(933):816–20. pmid:16110103
  27. 27. Nakayama R, Watanabe R, Namba K, Takeda K, Yamamoto K, Katsuragawa S, et al. Computer-aided diagnosis scheme for identifying histological classification of clustered microcalcifications by use of follow-up magnification mammograms. Academic radiology. 2006;13(10):1219–28. pmid:16979071
  28. 28. Grunert JH, Khalifa R, Gmelin E. [Computer-aided segmentation, form analysis and classification of 2975 breast microcalcifications using 7-fold microfocus magnification mammography]. RoFo: Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2004;176(12):1759–65. pmid:15573286
  29. 29. Hizukuri A, Nakayama R, Nakako N, Kawanaka H, Takase H, Yamamoto K, et al. Computerized segmentation method for individual calcifications within clustered microcalcifications while maintaining their shapes on magnification mammograms. Journal of digital imaging. 2012;25(3):377–86. pmid:21989574
  30. 30. Zhang Y, Wu X, Lu S, Wang H, Phillips P, Wang S. Smart detection on abnormal breasts in digital mammography based on contrast-limited adaptive histogram equalization and chaotic adaptive real-coded biogeography-based optimization. SIMULATION. 2016.
  31. 31. Morel JC, Iqbal A, Wasan RK, Peacock C, Evans DR, Rahim R, et al. The accuracy of digital breast tomosynthesis compared with coned compression magnification mammography in the assessment of abnormalities found on mammography. Clinical radiology. 2014;69(11):1112–6. pmid:25100302
  32. 32. Chen Y, Shi L, Feng Q, Yang J, Shu H, Luo L, et al. Artifact Suppressed Dictionary Learning for Low-Dose CT Image Processing. IEEE Transactions on Medical Imaging. 2014;33(12):2271–92. pmid:25029378
  33. 33. Chen Y, Chen W, Yin X, Ye X, Bao X, Luo L, et al. Improving low-dose abdominal CT images by Weighted Intensity Averaging over Large-scale Neighborhoods. European Journal of Radiology. 2011;80(2):e42–e9. http://dx.doi.org/10.1016/j.ejrad.2010.07.003. pmid:20709478
  34. 34. Yang C, Xindao Y, Luyao S, Huazhong S, Limin L, Jean-Louis C, et al. Improving abdomen tumor low-dose CT images using a fast dictionary learning based processing. Physics in Medicine and Biology. 2013;58(16):5803. pmid:23917704
  35. 35. Yang C, Zhou Y, Yining H, Guanyu Y, Yongcheng Z, Yinsheng L, et al. Thoracic low-dose CT image processing using an artifact suppressed large-scale nonlocal means. Physics in Medicine and Biology. 2012;57(9):2667. pmid:22504130
  36. 36. Chen Y, Gao D, Nie C, Luo L, Chen W, Yin X, et al. Bayesian statistical reconstruction for low-dose X-ray computed tomography using an adaptive-weighting nonlocal prior. Computerized Medical Imaging and Graphics. 2009;33(7):495–500. http://dx.doi.org/10.1016/j.compmedimag.2008.12.007. pmid:19515533
  37. 37. Sang Hee N, Jun Young C, editors. A method of image enhancement and fractal dimension for detection of microcalcifications in mammogram. Engineering in Medicine and Biology Society, 1998 Proceedings of the 20th Annual International Conference of the IEEE; 1998 29 Oct-1 Nov 1998.