A metal artifact reduction algorithm in CT using multiple prior images by recursive active contour segmentation

  • Haewon Nam,

    Affiliation Department of Liberal Arts and Science, Hongik University, Sejong, South Korea

  • Jongduk Baek

    jongdukbaek@yonsei.ac.kr

    ¤ Current address: School of Integrated Technology and Yonsei Institute of Convergence Technology, Yonsei University, 162-1, Songdo-dong, Yeonsu-gu, Incheon, South Korea

    Affiliations Yonsei Institute of Convergence Technology, Yonsei University, Incheon, South Korea, School of Integrated Technology, Yonsei University, Incheon, South Korea

Abstract

We propose a novel metal artifact reduction (MAR) algorithm for CT images that completes a corrupted sinogram along the metal trace region. When metal implants are located inside a field of view, they create a barrier to the transmitted X-ray beam due to the high attenuation of metals, which significantly degrades the image quality. To fill in the metal trace region efficiently, the proposed algorithm uses multiple prior images with residual error compensation in sinogram space. Multiple prior images are generated by applying a recursive active contour (RAC) segmentation algorithm to the pre-corrected image acquired by MAR with linear interpolation, where the number of prior image is controlled by RAC depending on the object complexity. A sinogram basis is then acquired by forward projection of the prior images. The metal trace region of the original sinogram is replaced by the linearly combined sinogram of the prior images. Then, the additional correction in the metal trace region is performed to compensate the residual errors occurred by non-ideal data acquisition condition. The performance of the proposed MAR algorithm is compared with MAR with linear interpolation and the normalized MAR algorithm using simulated and experimental data. The results show that the proposed algorithm outperforms other MAR algorithms, especially when the object is complex with multiple bone objects.

Introduction

Metal artifacts are one of the most common problems in computed tomography (CT). Various metal implants in the human body such as dental fillings [1], orthopedic implants [2], hip prostheses [3], and implanted marker bins [4], generate dark and bright streaks owing to the high attenuation of metals, which degrade the image quality and diagnostic value of CT images. To reduce metal artifacts, several metal artifact reduction (MAR) algorithms have been developed [5]. These are classified into two groups: projection completion methods and iterative methods.

Projection completion methods treat projections through the metal trace region as missing data and estimate them using neighboring measured data by linear [6], higher-order [79], wavelet [10, 11], prior knowledge [12, 13], and diffusion [14, 15] approximations. Interpolation-based MAR algorithms are simple, but produce additional artifacts when large metal implants are present. To prevent this, MAR algorithms utilizing a prior image have been suggested [1620]. In these algorithms, the metal trace region is replaced by the forward projection of the prior image obtained by applying thresholding on the pre-corrected CT images [1619] with the assumption that pixel values do not vary substantially within materials. Instead of using simple thresholding, Zhang et al. [20] used a TV minimization regularized algebraic reconstruction technique (ART) to find the prior image. In general, MAR algorithms with prior images demonstrate better performance in artifact reduction, but acquiring a high quality prior image is essential.

Iterative methods attempt to iteratively reduce the mismatch between measured data and forward projection of the object [2127]. By excluding data in the metal trace region of the sinogram, MAR with iterative methods becomes an ill-posed problem that is stabilized by using a priori knowledge of the image as a regularization term [2124]. Wang et al. [21] proposed two iterative deblurring algorithms that used an expectation maximization (EM) formula and ART adapted for MAR. Williamson et al. [22] proposed statistical iterative reconstruction using prior information about the geometry of metal objects and the detector response model. Choi et al. [23] and Zhang et al. [24] proposed iterative MAR based on constrained optimization with total variation (TV) and a quadratic smoothness function as a regularizer, respectively. Verburg and Seco [25] used a regularized iterative method only applied on a high-Z implants traced sinogram. Lemmens et al. [26] used maximum a posteriori (MAP) reconstruction with thresholding-based multimodal prior, and Nasirudin et al. [27] used material decomposed images of spectral CT as priors, and reduced the metal artifacts by regularized maximum likelihood approach. Compared to the projection completion methods, MAR with iterative methods are robust to noise because they are controlled by a regularization term. However, the high computational cost is a practical obstacle.

The proposed MAR algorithm is categorized in the projection completion group. To reduce estimation errors in the metal trace region, we generate multiple prior images by applying a recursive active contour (RAC) segmentation technique to the MAR image with linear interpolation [6], where the number of prior image is controlled by the RAC depending on the object complexity. A sinogram basis is acquired by conducting forward projection of prior images. The metal trace region of the original sinogram is then replaced by the linearly combined sinogram of prior images. When the object is heterogeneous and the data acquisition condition is not ideal, residual errors occur between the original sinogram and the linearly combined sinogram. Additional correction is performed to compensate. We iteratively continue the process until the norm of the residual errors is minimized. As a result, the proposed algorithm finds good prior images and effectively reduces metal artifacts. To evaluate the performance of the proposed MAR algorithm, both numerical and physical experiments are conducted. Quantitative evaluations are performed and compared with the MAR algorithm with linear interpolation (LIN) [6] and normalized MAR (NMAR) algorithm [18].

Materials and methods

In this section, we briefly review the RAC segmentation technique and then describe the proposed MAR algorithm.

Recursive active contour segmentation

RAC segmentation is based on classical active contour segmentation proposed by Chan-Vese [28]. The active contour model uses a level set function to divide an image into two separate regions.

Let I be a given image on such that . The active contour model minimizes the following equation [28]: (1) where H is the Heaviside function and D is the distributional derivative. A function φ is the levelset function which separates an image into two subregions, and c1 and c2 are mean values of corresponding subregions; μ is a regularization parameter to control the length of the active contour and segmentation sensitivity.

Image segmentation into multiple subregions can be performed by recursively applying the active contour model. Suppose image I0 has two objects with different intensities as shown in Fig 1(a). By applying the active contour model to I0, two objects are separated together from the background by the levelset function. To further segment subregions, the background values in I0 are replaced by the minimum value of the foreground regions in I0. Then a new image I1 is acquired as shown in Fig 1(b). By continuing the segmentation process until the entire image is filled with a constant as shown in Fig 1(c), multiple subregions are segmented. Note that red lines in Fig 1 indicate a region where the values of the levelset function are zero. The levelset function has positive values in the foreground regions.

thumbnail
Fig 1. Concept of recursive active contour segmentation.

(a) I0: initial image, (b) I1: image obtained by replacing the background values of I0 with one, and (c) I2: image by replacing the background values of I1 with two.

https://doi.org/10.1371/journal.pone.0179022.g001

The main advantage of the RAC technique is that the number of subregions is determined adaptively during the segmentation process. Although we describe image segmentation using a piecewise linear object, the RAC technique is still applicable to heterogeneous objects by controlling the regularization parameter μ in Eq (1) [29]. Since the regularization parameter controls the segmentation sensitivity [29], small μ is needed for the segmentation of low contrast intensity object, which increases the number of segmented subregions. Fig 2 illustrates the segmentation results depending on the regularization parameter, where the background intensity increases gradually from zero to one toward the right direction.

thumbnail
Fig 2. Segmentation results with different regularization parameter μ.

https://doi.org/10.1371/journal.pone.0179022.g002

Proposed method

The proposed MAR algorithm consists of three steps as described in Fig 3. In the first step, the initial image is reconstructed by the Feldkamp, David and Kress (FDK) algorithm using the original sinogram p0, and metal regions are segmented by the active contour [28]. Note that the active contour technique is more effective in metal segmentation than the simple thresholding technique when multiple metals are present. Then, the metal trace region, TM, is identified by the forward projection of the segmented metals within the sinogram domain Ω. The initial corrected image, fk for k = 1, is acquired by applying linear interpolation on the metal trace region TM [6].

In the second step, fk is divided into N subregions, I1, …, IN by RAC. In our implementation, we chose the regularization parameter in Eq (1) as 0.1, and the number of subregions N is determined accordingly. The segmented subregions are used as prior images, which have the intensity as one; thus, finding the proper coefficient dj for each subregion Ij is necessary to satisfy the following equation: (2) where is the Radon transform operator and .

The optimal coefficients dj are estimated by solving the following minimization problem: (3) where d = (d1, …, dN)T, is the complement of TM, and ∂TM is the boundary of TM. The first term in Eq (3) minimizes the mismatch between the original sinogram and the linearly combined sinogram on , and the second term regulates the smoothness along the boundary with the regularization parameter α > 0. In this work, the minimization problem is solved using Nelder-Mead simplex direct search [30] with α = 0.1. The mean value of each segmented subregion is used as an initial guess for solving Eq (3). With computed coefficients dj, a new linearly combined sinogram is generated.

If the segmentation is performed without errors and the coefficients dj are estimated correctly, is equivalent to p0 on with ideal data acquisition condition. However, real projectin data are polychromatic and contain some amount of scatter, and thus residual errors (i.e., ) still exist on ; thus, using data in the metal trace region of is suboptimal for metal artifact reduction. In step three, the residual errors in the metal trace region are compensated by adding a linearly interpolated sinogram of δ on the metal trace region and . The new sinogram coincides with the original sinogram on since is equivalent to δ on .

Then, k is updated by k + 1, and steps two and three are repeated using the reconstructed image fk of . This process continues until the norm of the residual errors is less than tolerance τ.

In NMAR, interpolation errors in the metal trace region are minimized by using a normalized sinogram. If the object is piecewise constant and the X-ray spectrum is monochromatic, NMAR and the proposed method produce the same MAR results. When the object is heterogeneous and the X-ray spectrum is polychromatic, the denormalization step of NMAR may amplify the interpolation errors. This effect will be more significant when the metals exist on the surface of the object. In contrast, our approach is more stable since the interpolation errors are minimized by error subtraction. This effect will be described more clearly in our result section.

Experiments

Both numerical and physical experiments were conducted to evaluate the performance of the proposed MAR algorithm.

Numerical simulations.

The numerical simulations were performed with a fan beam CT geometry using Shepp-Logan, jaw and abdomen phantoms. A Shepp-Logan phantom (shown in Fig 4(a)) contains two rectangular gold inserts, a jaw phantom (shown in Fig 4(b)) contains three gold implants, and an abdomen phantom (Fig 4(c)) has two gold implants. For each phantom, polychromatic projection data were acquired with a 120 kVp tube voltage using the Siemens X-ray spectrum [31]. Attenuation coefficients of muscle, bone, and gold were obtained from the NIST X-ray Attenuation Database (ICRU-44) [32]. Each detector cell received 106 photons without the object, and Poisson noise was simulated. Each phantom consists of 512 × 512 pixels with a 0.2 × 0.2 mm2 pixel size. To obtain the reference images, noiseless projection data of tissues and bones were reconstructed using a polychromatic X-ray spectrum. The simulation parameters are summarized in Table 1.

thumbnail
Fig 4. Three numerical phantoms.

(a) Shepp-Logan phantom with two rectangular gold inserts indicated by blue squares, (b) jaw phantom with three gold implants indicated by blue circles, and (c) abdomen phantom with two gold implants indicated by blue circles.

https://doi.org/10.1371/journal.pone.0179022.g004

Physical experiments.

For the physical experiments, a tooth and polymethylmethacrylate (PMMA) phantoms were used. As shown in Fig 5(a), the tooth phantom consists of an orthodontic bracket with one silver clip at the fore tooth and two gold rings at the cheek tooth. A PMMA phantom consists of a 10 cm-diameter cylinder with five holes and a 16 cm diameter extension annulus with four holes (Fig 5(b)). A metal bar was placed at the center of the PMMA phantom, and two screws were placed at the edge of the PMMA phantom. Projection data were acquired from the bench-top cone beam CT (CBCT) system. The bench-top system includes a generator (Indico 100, CPI Communication & Medical Products Division, Georgetown, ON, Canada), a tungsten target X-ray source (Varian G-1592, Varian X-ray Product, Salt Lake City, UT, USA) with a 0.6 × 0.6 mm2 focal spot, and a 400 × 300 mm2 flat-panel detector (PaxScan 4030CB, Varian Medical Systems, Salt Lake City, UT, USA) with a tube voltage of 120 kVp and current of 5 mA. The experimental parameters are summarized in Table 2.

thumbnail
Fig 5. Two physical phantoms.

(a) Tooth phantom containing an orthodontic bracket and three metals, and (b) PMMA phantom with a metal bar and two screws.

https://doi.org/10.1371/journal.pone.0179022.g005

Performance comparison

We compared the performance of the proposed algorithm with LIN [6] and NMAR [18]. The performance of NMAR was evaluated using three different prior images: prior image by simple thresholding (NMARth), prior image by RAC segmentation of LIN (NMARsegs), and the same prior image with the proposed algorithm (NMARsege). In NMARth, the thresholding value was chosen by the attenuation coefficient of the material (i.e., air = 0, soft tissue = 0.28, and bone = 0.4). In NMARsegs and NMARsege, the coefficients of the prior images were chosen as the mean value of the corresponding subregions.

For the quantitative evaluation, the mean square error (MSE) and standard deviation were used. For the simulation data, the MSE between the reference image and each MAR image was computed using the entire image except the metal region. For the experimental phantoms, the region of interest (ROI) was selected within homogeneous regions, and the standard deviation was calculated for each MAR image.

Results

Numerical simulations

Fig 6(a) shows the MAR results of the Shepp-Logan phantom. Due to significant beam hardening artifacts in the uncorrected image, high quality prior images could not be acquired by simple thresholding (shown in Fig 6(b)). As a result, the NMARth image contained severe streak artifacts. Using good prior images improved the image quality as shown in the NMARsegs image, and further improvement was observed in the NMARsege image. It is also shown that the proposed MAR image shows similar performance to the NMARsege image since the error amplification in the denormalization step of NMAR is not significant because of the negligible beam hardening effect of the soft tissues in the original sinogram.

thumbnail
Fig 6. Results of the Shepp-Logan phantom (C = 250 HU/W = 750 HU).

(a) Top row: reconstruction images by (left) Uncorrected, (middle) LIN, and (right) NMARth images. Bottom row: reconstruction images by (left) NMARsegs, (middle) NMARsege, and (right) the proposed MAR. (b) Prior images by (top) simple thresholding, (middle) RAC segmentation of the LIN image, and (bottom) RAC segmentation of the final iteration image.

https://doi.org/10.1371/journal.pone.0179022.g006

Fig 7(a) shows the MAR results of the jaw phantom and Fig 7(b) compares the prior images acquired by a simple thresholding and RAC technique. Unlike the Shepp-Logan results, the proposed MAR method shows better performance than the NMARsege image. Fig 8(a)–8(f) show the sinogram of the reference image, prior images by simple thresholding, the NMARth image, prior images by the RAC technique, the NMARsege image, and the proposed MAR image, respectively. Profiles of the sinogram images (indicated by the magenta line in Fig 8(a)) are also compared. As shown in Fig 8(g) and 8(h), the NMARth produces significant errors within the metal trace region. Using good prior images reduces the estimation errors in the metal trace region as shown in Fig 8(i). However, residual errors still exist in the profile of NMARsege as shown in Fig 8(j). Note that the original sinogram contains beam hardening effects generated from multiple bones; thus, the denormalization step of NMAR amplifies the interpolation errors on the metal trace region. The proposed MAR algorithm reduces the residual errors by the additional residual error correction step (i.e., Step 3 in Fig 3), and thus, further improvement in MAR can be achieved.

thumbnail
Fig 7. Results of the jaw phantom (C = 250 HU/W = 750 HU).

(a) Top row: reconstruction images by (left) Uncorrected, (middle) LIN, and (right) NMARth images. Bottom row: reconstruction images by (left) NMARsegs, (middle) NMARsege, and (right) the proposed MAR. (b) Prior images by (top) simple thresholding, (middle) RAC segmentation of the LIN image, and (bottom) RAC segmentation of the final iteration image.

https://doi.org/10.1371/journal.pone.0179022.g007

thumbnail
Fig 8. Sinogram profiles.

(a) reference, (b) Priorth, (c) NMARth, (d) Priorsege, (e) NMARsege, and (f) the proposed algorithm. Comparison of sinogram profiles of (g) reference, Priorth, and NMARth, and (h) enlarged version of the profiles within metal traces 1 and 2 of Fig 8(g). Comparison of sinogram profiles of (i) reference, Priorsege, NMARsege, and the proposed algorithm, and (j) enlarged version of the profiles within metal traces 1 and 2 of Fig 8(i).

https://doi.org/10.1371/journal.pone.0179022.g008

Fig 9(a) shows the MAR results of the abdomen phantom. Compared to the previous numerical phantoms, the abdomen phantom is more heterogeneous, and thus more accurate prior images would be required to reduce metal artifacts effectively. However, due to the severe noise and metal artifacts in the uncorrected image, prior images generated by simple thresholding miss details of inner structures in the abdomen phantom (shown in Fig 9(b)). As a result, the NMARth image contains severe streak artifacts. In contrast, prior images generated by the RAC technique contain more details, which improve the performance of the NMAR. While all other MAR algorithms contain residual streaks artifacts between two metals, the proposed algorithm reduces them effectively.

thumbnail
Fig 9. Results of the abdomen phantom (C = 500 HU/W = 1500 HU).

(a) Top row: reconstruction images by (left) Uncorrected, (middle) LIN, and (right) NMARth images. Bottom row: reconstruction images by (left) NMARsegs, (middle) NMARsege, and (right) the proposed MAR. (b) Prior images by (top) simple thresholding, (middle) RAC segmentation of the LIN image, and (bottom) RAC segmentation of the final iteration image.

https://doi.org/10.1371/journal.pone.0179022.g009

In simulations, the number of prior images was chosen by RAC automatically, and the tolerance τ was set to 10% of the first norm of the residual errors. Table 3 summarizes the number of prior image and coefficients for numerical simulations, and Table 4 summarizes the MSE values for different MAR images. For all numerical phantoms, the proposed algorithm shows the best performance in metal artifact reduction.

thumbnail
Table 3. Number of prior image N and coefficients dj in numerical simulations.

https://doi.org/10.1371/journal.pone.0179022.t003

Physical experiment

For the tooth phantom, three slices from the reconstructed three-dimensional volume were chosen. Fig 10 shows MAR results and prior images of slice 17. Notice that the NMARth image contains significant streak artifacts owing to the poor quality of the prior images. Although the streak artifacts are reduced in the NMARsegs and NMARsege images, residual streak artifacts are still observed. Since the tooth phantom contains the metals on the surface of the object, small interpolation errors in the metal trace region are significantly amplified during the denormalization step of the NMAR algorithm. In contrast, the proposed MAR algorithm effectively reduces the streak artifacts.

thumbnail
Fig 10. Results of tooth phantom with orthodontic bracket, on slice # 17 (C = -250 HU/W = 1250 HU).

(a) Top row: reconstruction images by (left) Uncorrected, (middle) LIN, and (right) NMARth images. Bottom row: reconstruction images by (left) NMARsegs, (middle) NMARsege, and (right) the proposed results. ROI 1 and ROI 2 are indicated by red lines on the uncorrected image. (b) Prior images by (top) simple thresholding, (middle) RAC segmentation of LIN image, and (bottom) RAC segmentation of final iteration image.

https://doi.org/10.1371/journal.pone.0179022.g010

Figs 11 and 12 show the MAR results and prior images of slice 114 and 155, respectively, where multiple metals are present. Compared to the other MAR algorithms, the proposed method reduces metal artifacts more effectively.

thumbnail
Fig 11. Results of tooth phantom with orthodontic bracket, on slice # 114 (C = -250 HU/W = 1250 HU).

(a) Top row: reconstruction images by (left) Uncorrected, (middle) LIN, and (right) NMARth images. Bottom row: reconstruction images by (left) NMARsegs, (middle) NMARsege, and (right) the proposed results. ROI 3 and ROI 4 are indicated by red lines on the uncorrected image. (b) Prior images by (top) simple thresholding, (middle) RAC segmentation of LIN image, and (bottom) RAC segmentation of final iteration image.

https://doi.org/10.1371/journal.pone.0179022.g011

thumbnail
Fig 12. Results of tooth phantom with orthodontic bracket, on slice # 155 (C = -250 HU/W = 1250 HU).

(a) Top row: reconstruction images by (left) Uncorrected, (middle) LIN, and (right) NMARth images. Bottom row: reconstruction images by (left) NMARsegs, (middle) NMARsege, and (right) the proposed results. ROI 5 and ROI 6 are indicated by red lines on the uncorrected image. (b) Prior images by (top) simple thresholding, (middle) RAC segmentation of LIN image, and (bottom) RAC segmentation of final iteration image.

https://doi.org/10.1371/journal.pone.0179022.g012

For the quantitative evaluation, two ROIs were chosen on the homogeneous region, which is indicated by the red lines in Figs 10(a)12(a). The standard deviation of each ROI was calculated and summarized in Table 5. For the tooth phantom, the proposed MAR algorithm has the smallest standard deviation for all three slices.

thumbnail
Table 5. Standard deviation of the ROIs for physical phantoms.

https://doi.org/10.1371/journal.pone.0179022.t005

Fig 13 shows MAR results and prior images of the PMMA phantom with three metal inserts, where the central slice from the reconstructed three-dimensional volume is selected. Notice that the prior images by simple thresholding and the RAC technique are very similar and thus, all NMAR algorithms show similar MAR results with the proposed MAR algorithm.

thumbnail
Fig 13. Results of a PMMA phantom (C = 0 HU/W = 600 HU).

(a) Top row: reconstruction images by (left) Uncorrected, (middle) LIN, and (right) NMARth images. Bottom row: reconstruction images by (left) NMARsegs, (middle) NMARsege, and (right) the proposed results. ROI 1- ROI 4 are indicated by red lines on the uncorrected image. (b) Prior images. Left: prior by simple thresholding, middle: prior by RAC segmentation of the LIN image, right: prior by RAC segmentation of the final iteration image.

https://doi.org/10.1371/journal.pone.0179022.g013

For the quantitative evaluation, four ROIs were selected on the homogeneous region, which are indicated by red lines in Fig 13(a). The standard deviation of each ROI was calculated and summarized in Table 5. Notice that the three NMAR images and the proposed MAR image show similar standard deviation owing to the simple structure of the PMMA phantom. Table 6 summarizes the number of iterations for each phantom. More iterations are required when the image quality of LIN is poor. Table 7 summarizes the number of prior image and coefficients for each experimental phantom.

thumbnail
Table 7. Number of prior image N and coefficients dj in physical phantoms.

https://doi.org/10.1371/journal.pone.0179022.t007

Discussion and conclusion

We presented a new MAR algorithm using multiple prior images with subsequential residual error correction. Traditional interpolation-based MAR algorithms replace the metal trace region with neighboring data depending on the distance, which introduces streak artifacts owing to interpolation errors in the metal trace region. The proposed method minimizes interpolation errors by utilizing multiple prior images generated by the RAC technique and additional residual error correction method in sinogram space. The performance of the proposed method was compared with LIN and NMAR with different prior images. For the homogeneous object, NMAR and the proposed method show similar performance in metal artifact reduction. However, when the object is complex with multiple bone objects, the proposed method shows better MAR results. For a complex object, we also showed that the performance of the NMAR algorithm can be improved significantly using prior images generated by the RAC technique.

In NMAR, interpolation errors in the metal trace region were reduced by performing the interpolation on the normalized sinogram. However, using poor quality prior images can degrade the performance of NMAR since the interpolation errors are magnified by the denormalization step of the NMAR algorithm. This effect is more significant when the metals lie on the surface of the objects. Using good prior images improves the performance of the NMAR algorithm significantly. As shown in the results of NMARsegs, RAC segmentation improves NMAR compared to NMARth, but the residual errors caused by the denormalization step still exist, especially for complex objects containing multiple bone objects. In contrast, the proposed algorithm does not amplify interpolation errors since the metal trace region is filled in by the summation of linearly combined sinograms of prior images and subsequential residual error correction.

The proposed method iteratively reduces interpolation errors. Each iteration consists of prior image generation, sinogram basis generation, and residual error correction. Each iteration takes approximately 15 s for a two-dimensional image in MATLAB (MathWorks, Natick, MA) on a work station with an Intel XEON E5-2630V2 processor at 2.6GHz. The number of iterations depends on the object’s complexity and the quality of the LIN image. Since the proposed algorithm can be applied independently for each slice, the computation time for three-dimensional volume data would be similar to the two-dimensional case by parallel computing using a graphics processing unit (GPU). The computation time can be more accelerated by C/C++ implementation, which is essential for real clinical applications.

In this work, we treated the projection data in the metal trace region as missing or completely unreliable. However, when the density of metals is not high enough, the metal trace region may contain available information of the object, and thus, further improvement of the MAR can be achieved by frequency split MAR (FSMAR) [19]. Comparison of the image quality improvement with FSMAR with NMAR images and proposed MAR images is a subject of future research. In conclusion, we presented a new MAR algorithm using multiple prior images with additional residual error correction. The results show that the proposed algorithm outperforms other MAR algorithms, especially when the object is complex with multiple bone objects.

Supporting information

S1 Fig. 2D Shepp Logan phantom.

Data related to Fig 6.

https://doi.org/10.1371/journal.pone.0179022.s001

(MAT)

S4 Fig. 2D abdomen phantom.

Data related to Fig 9.

https://doi.org/10.1371/journal.pone.0179022.s004

(MAT)

S5 Fig. 3D orthodontic phantom.

Data related to Figs 1012.

https://doi.org/10.1371/journal.pone.0179022.s005

(MAT)

Acknowledgments

This research was supported by the MSIP (Ministry of Science, ICT and Future Planning), Korea, under the “IT Consilience Creative Program” (IITP-2015-R0346-15-1008) supervised by the IITP (Institute for Information & Communications Technology Promotion) and Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT & Future Planning (2015R1C1A2A01054731, 2015R1C1A101052268, 2017M2A2A6A01019663).

Author Contributions

  1. Conceptualization: JB HN.
  2. Data curation: JB HN.
  3. Formal analysis: JB HN.
  4. Funding acquisition: JB HN.
  5. Investigation: JB HN.
  6. Methodology: JB HN.
  7. Resources: JB.
  8. Software: HN.
  9. Supervision: JB.
  10. Validation: HN.
  11. Visualization: HN.
  12. Writing – original draft: JB HN.
  13. Writing – review & editing: JB.

References

  1. 1. Tohnak S, Mehnert AJH, Mahoney M, Crozier S. Dental CT metal artefact reduction based on sequential substitution.Dentomaxillofac Radiol., 2011; 40: 184–190. pmid:21346086
  2. 2. Liu PT, Pavlicek WP, Peter MB, Spangehl MJ, Roberts CC, Paden RG. Metal artifact reduction image reconstruction algorithm for CT of implanted orthopedic devices: a work in progress. Phys. Med. Biol. 2013; 58: 8401–8418. pmid:19142632
  3. 3. Robertson DD, Yuan J, Wang G, Vannier MW. Total hip prosthesis metal-artifact suppression using iterative deblurring reconstruction. J. Comput. Assist. Tomogr. 1997; 21: 293–298. pmid:9071303
  4. 4. Handsfield LL, Yue NJ, Zhou J, Chen T, Goyal S. Determination of optimal fiducial marker across image-guided radiation therapy (IGRT) modalities: visibility and artifact analysis of gold, carbon, and polymer fiducial markers. J. of Appl. Clin. Med. Phys. 2012; 13: 181–189. pmid:22955665
  5. 5. Abdoli M, Dierckx RA, Zaidi H. Metal artifact reduction strategies for improved attenuation correction in hybrid PET/CT imaging. Med. Phys. 2012; 39: 3343–3360. pmid:22755716
  6. 6. Kalender WA, Hebel R, Ebersberger J. Reduction of CT artifacts caused by metallic implants. Radiology. 1987; 164: 576–577. pmid:3602406
  7. 7. Glover GH. Pelc NJ, An algorithm for the reduction of metal clip artifacts in CT reconstructions. Med. Phys. 1981;8(6): 799–807. pmid:7322078
  8. 8. Mahnken AH, Raupach R, Wildberger JE, Jung B, Heussen N, Flohr TG, et al. A new algorithm for metal artifact reduction in computed tomography: In vitro and in vivo evaluation after total hip replacement. Invest. Radiol. 2003; 38(12): 769–775. pmid:14627894
  9. 9. Yazdia M, Gingras L, Beaulieu L. An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: Experimental and clinical studies.Int. J. Radiation Oncol. Biol. Phys.2005; 62(4): 1224–1231. pmid:15927413
  10. 10. Zhao SY, Roberston DD, Wang G, Whiting B, Bae KT. X-ray CT metal artifact reduction using wavelets: arm application for imaging total hip prostheses. IEEE Trans. Med. Imag. 2000; 19: 1238–1247. pmid:11212372
  11. 11. Mehranian A, Ay MR, Rahmim A, Zaidi H. X-ray CT metal artifact reduction using wavelet domain L0 sparse regularization. IEEE Trans. Med. Imag. 2013; 32: 1707–1722. pmid:23744669
  12. 12. Veldkamp WJH, Joemai RMS, van der Molen AJ, Geleijns J. Development and validation of segmentation and interpolation techniques in sinograms for metal artifact suppression in CT. Med. Phys. 2010; 37(2): 620–628. pmid:20229871
  13. 13. Yu HY, Zeng K, Bharkhada DK, Wang G, Madsen MT, Saba O, et al. A segmentation-based method for metal artifact reduction. Acad. Radiol. 2007; 14: 495–504. pmid:17368220
  14. 14. Zhang Y, Zhang L, Zhu RX, Chambers M, Dong L. Reducing metal artifacts in cone-beam CT by tracking and eliminating metal shadows in raw projection data. Med. Phys. 2006; 33: 2288–2288.
  15. 15. Zhang Y, Zhang L, Zhu XR, Lee AK, Chambers M, Dong L. Reducing metal artifacts in cone-beam CT images by preprocessing projection data. Int. J. Radiat. Oncol., Biol., Phys.2007; 67: 924–932. pmid:17161556
  16. 16. Prell D, Kyriakou Y, Beister M, Kalender WA. A novel forward projection based metal artifact reduction method for flat detector computed tomography. Phys. Med. Biol. 2009; 54: 6575–6591. pmid:19826202
  17. 17. Watzke O, Kalender WA. A pragmatic approach to metal artifact reduction in CT: Merging of metal artifact reduction in CT. Med. Phys. 2004; 28: 475–490. pmid:15014974
  18. 18. Meyer E, Raupach R, Lell M, Schmidt B, Kachelrieβ M. Normalized metal artifact reduction (NMAR) in computed tomography. Med. Phys. 2010; 37: 5482–5493. pmid:21089784
  19. 19. Meyer E, Raupach R, Lell M, Schmidt B, Kachelrieβ M. Frequency split metal artifact reduction (FSMAR) in computed tomography. Med. Phys. 2012; 39: 1904–1916. pmid:22482612
  20. 20. Zhang Y, Yan H, Jia X, Yang J, Jiang SB, Mou X. A hybrid metal artifact reduction algorithm for x-ray CT. Med. Phys. 2013; 40: 041910. pmid:23556904
  21. 21. Wang G, Snyder DL, O’Sullivan JA, Vannier MW. Iterative deblurring for CT metal artifact reduction. IEEE Trans. Med. Img. 1996; 15: 657–664. pmid:18215947
  22. 22. Williamson JF, Whiting BR, Benac J, Murphy RJ, Blaine GJ, O’Sullivan JA, et al. Prospects for quantitative computed tomography imaging in the presence of foreign metal bodies using statistical image reconstruction. Med. Phys. 2002; 29: 2404–2418. pmid:12408315
  23. 23. Choi J, Kim KS, Kim MW, Seong W, Ye JC. Sparsity driven metal part reconstruction for artifact removal in dental CT. J. X-Ray Sci. Technol. 2011; 19: 457–475. pmid:25214380
  24. 24. Zhang X, Wang J, Xing L. Metal artifact reduction in x-ray computed tomography (CT) by constrained optimization. Med. Phys. 2011; 38: 701–711. pmid:21452707
  25. 25. Verburg JM, Seco J. CT metal artifact reduction method correcting for beam hardening and missing projections. Phys. Med. Biol. 2012; 57: 2803–2818. pmid:22510753
  26. 26. Lemmens C, Faul D, Nuyts J. Suppression of metal artifacts in CT using a reconstruction procedure that combines MAP and projection completion. IEEE Trans. Med. Imaging. 2009; 28: 250–260. pmid:19188112
  27. 27. Nasirudin RA, Mei K, Panchev P, Fehringer A, Pfeiffer F, Rummeny EJ, et al. (2015) Reduction of Metal Artifact in Single Photon-Counting Computed Tomography by Spectral-Driven Iterative Reconstruction Technique. PLoS ONE 10(5): e0124831. pmid:26098956
  28. 28. Chan TF, Vese LA. Active Contours Without Edges. IEEE Trans. on Image Processing. 2001; 10: 266–277.
  29. 29. Lai KF, Chin RT. On regularization, formulation and initialization of the active contour models (snakes). Asian Conf. on Comp. Vis. 1993; 10: 542–545.
  30. 30. Lagarias JC, Reeds JA, Wright MH, Wright PE. Convergence properties of the Nelder-Mead simplex method in low dimensions. SIAM J. of Optim. 1998; 9: 112–147.
  31. 31. Boone JM, Fewell TR, Jennings RJ. Simulation of X-ray Spectra. Available form: http://www.oem-products.siemens.com/x-ray-spectra-simulation
  32. 32. Hubbell JH, Seltzer SM. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy- Absorption Coefficients from 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Sub- stances of Dosimetric Interest. Tech. Rep., National Institute of Standards and Technology (Available: http://physics.nist.gov/xaamdi), Gaithersburg, MD. (2004).