Evaluation of quantitative biosensor for glucose-6-phosphate dehydrogenase activity detection

  • Pairat Pengboon,

    Roles Data curation, Formal analysis, Investigation

    Affiliation Graduate Program in Medical Technology, Faculty of Allied Health Sciences, Thammasat University, Pathumtani, Thailand

  • Areenuch Thamwarokun ,

    Contributed equally to this work with: Areenuch Thamwarokun, Khaimuk Changsri

    Roles Data curation

    Affiliation Graduate Program in Biomedical Sciences, Faculty of Allied Health Sciences, Thammasat University, Pathumtani, Thailand

  • Khaimuk Changsri ,

    Contributed equally to this work with: Areenuch Thamwarokun, Khaimuk Changsri

    Roles Conceptualization, Funding acquisition, Methodology

    Affiliation Department of Medical Technology, Faculty of Allied Health Sciences, Thammasat University, Pathumtani, Thailand

  • Chollanot Kaset,

    Roles Formal analysis, Writing – review & editing

    Affiliation Department of Medical Technology, Faculty of Allied Health Sciences, Thammasat University, Pathumtani, Thailand

  • Sirinart Chomean

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Validation, Writing – original draft, Writing – review & editing

    sirinat.c@allied.tu.ac.th

    Affiliation Department of Medical Technology, Faculty of Allied Health Sciences, Thammasat University, Pathumtani, Thailand

Abstract

Neonatal jaundice is a common and severe disease in premature infants with Glucose-6-Phosphate Dehydrogenase (G-6-PD) deficiency. The World Health Organization (WHO) has recommended screening for G-6-PD deficiency in newborns for early recognition as well as to prevent unwanted outcomes in a timely manner. The present study aimed to assess a point-of-care, careSTARTTM G6PD biosensor as a quantitative method for the diagnosis of G-6-PD deficiency. Factors influencing the evaluation of G-6-PD enzyme activity were examined in 40 adults, including ethylenediaminetetraacetic acid (EDTA) anticoagulant, hematocrit concentration, storage temperature and time. Analytic performance of the careSTARTTM G6PD biosensor was evaluated in 216 newborns and compared with fluorescent spot test (FST) and standard quantitative G-6-PD enzyme activity (SGT) assay. The results of factors affecting the G-6-PD enzyme activity showed that the activity determined from finger-prick was not statistically different from venous blood (p = 0.152). The G-6-PD value was highly dependent on the hematocrit and rose with increasing hematocrit concentration. Its activity was stable at 4°C for 3 days. Reliability analysis between the careSTARTTM G6PD biosensor and SGT assay showed a strong correlation with a Pearson’s correlation coefficient of 0.82 and perfect agreement by intraclass correlation coefficient (ICC) of 0.90. Analysis of the area under the Receiver Operating Curve (AUC) illustrated that the careSTARTTM G6PD biosensor had 100% sensitivity, 96% specificity, 73% positive predictive value (PPV), 100% negative predictive value (NPV) and 97% accuracy at 30% of residual activity. While the diagnostic ability for identifying G-6-PD deficiency had 78% sensitivity, 89% specificity, 56% positive predictive value (PPV), 96% negative predictive value (NPV) and 88% accuracy when stratified by gender. The careSTARTTM G6PD biosensor is an attractive option as a point-of-care quantitative method for G-6-PD activity detection. Quantification of G-6-PD enzyme activity in newborns is the most effective approach for the management of G-6-PD deficiency to prevent severe jaundice and acute hemolysis.

Introduction

Glucose-6-phosphate dehydrogenase (G-6-PD) deficiency affects millions of people worldwide and comprises the most common inherited blood disorder [13]. Defective G-6-PD enzymes cause the increased susceptibility of red blood cells to reactive oxygen species, leading to hemolysis [1, 4]. Its clinical manifestations vary from asymptomatic to severe hemolytic anemia depending on the residual G-6-PD enzyme activity [1, 5]. Affected individuals generally present with acute hemolytic anemia, favism, or chronic non-spherocytic hemolytic anemia [1]. The clinical symptoms tend to increase as the red blood cells undergo oxidative stress triggered by agents such as anti-malarial drugs (primaquine), mothballs, infection or the ingestion of fava beans [1, 39]. Several studies have reported that diabetes, myocardial infarction and strenuous physical exercise can stimulate hemolysis in individuals with G-6-PD deficiency [1, 7]. Notably, affected neonates frequently show severe jaundice. Lack of awareness concerning G-6-PD deficiency in newborns could cause extreme hyperbilirubinemia, bilirubin neurotoxicity, kernicterus and, eventually, mental retardation [8, 10]. Accordingly, the World Health Organization (WHO) has recommended screening for G-6-PD deficiency in newborns to promote early diagnosis and prevent unwanted outcomes in a timely manner [8, 1013]. Moreover, individuals suffering from severe G-6-PD deficiency with residual activity less than 30% and 70% should be excluded from primaquine and tafenoquine administration, respectively [14, 15]. Thus, the measurement of G-6-PD activity is a crucial step before starting malaria treatment. Although the fluorescent spot test (FST) is widely used for qualitative screening of G-6-PD deficiency, its discriminative power appears limited, requiring UV visualization and highly specific skills for interpretation [6, 8, 1619]. The reference method for G-6-PD deficiency detection is based on a quantitative G-6-PD enzymatic assay. However, it is time-consuming, laborious and requires a spectrophotometer, which may not be suitable for the field or large-scale testing [17, 18]. Quantitative point-of-care G-6-PD tests are an essential tool in low-resource settings. Recently, the point-of-care quantitative careSTARTTM G6PD biosensor was developed as an alternative to the labor-intensive standard G-6-PD enzymatic method.

The careSTARTTM G6PD biosensor is an electrochemical biosensor that measures the electron transfer from the change of nicotinamide adenine dinucleotide phosphate (NADPH) into its reduced form by using G-6-PD enzymes. It is proposed for the quantitative measurement of G-6-PD enzyme activity in whole blood. Therefore, the present study aimed to examine the performance of the careSTARTTM G6PD biosensor compared with the gold standard enzymatic assay. Moreover, the most widely used FST method was evaluated. Measurement interference effects were also investigated, including ethylenediaminetetraacetic acid (EDTA), hematocrit concentration and enzymatic stability.

Materials and methods

Sample collection

Ethical approval for this study was obtained from the Third Ethics Committee of Thammasat University, Pathum Thani Province, Thailand (COA No. 150/2561). Drawing blood from a newborn can be difficult, potentially harmful and may yield limited sampling blood volume. Therefore, the factors affecting the G-6-PD enzyme activity measurement were evaluated in 40 adults. Prior to enrolment, all volunteers provided informed written consent to participation in the study as well as publication of the results. The volunteers included 20 males and 20 females ranging in age between 18 and 24 years old. The blood samples were collected from both finger-prick (5 μL) and venous puncture (3 mL in K2EDTA vacutainer tubes), after which the samples were analyzed by standard quantitative G-6-PD enzymatic assay (OSMMR2000-D G-6-PD kit, R&D Diagnostics, Ltd., Greece)

To evaluate the performance of the careSTARTTM G6PD biosensor (WELLS BIO, INC., Korea), 216 neonatal blood samples ranging in age between 1 to 35 years old were collected and kept in a microtainer tube with K2EDTA anticoagulant. The ethics committee provided a waiver of informed consent due to the neonatal blood samples used in this study being leftover samples following routine analysis. All data samples were de-identified before access. Complete blood count (CBC) (DxH 800 hematology analyzer, Beckman Coulter, Inc., USA) and fluorescent spot test (FST) (SQMMR500, R&D Diagnostics, Ltd., Greece) were assessed to screen for G-6-PD deficiency. G-6-PD enzyme activity was measured by the careSTARTTM G6PD biosensor compared with standard quantitative G-6-PD enzymatic assay (SGT).

All blood samples were carried out a single time following standard operating procedures under controlled time and temperature.

Assessment of factors influencing G-6-PD enzyme activity

Capillary blood samples were immediately assessed for G-6-PD enzyme activity by the standard SGT. Meanwhile, venous blood was aliquoted into 3 parts for assessment of the factors that affect G-6-PD enzyme activity, including anticoagulant (EDTA), hematocrit (Hct) concentration and enzymatic stability.

The effect of EDTA anticoagulant collected from venous blood was investigated by measuring G-6-PD enzyme activity and comparing with the sample from finger-prick.

Due to the activity of G-6-PD enzyme depending on the storage time and temperature, EDTA blood was aliquoted into a sealed, dark 1.5 mL microcentrifuge tube and kept at room temperature (25°C) and 4°C for 3 days. G-6-PD enzyme activity was measured at 1, 2, 6, 12, 24, 48 and 72 hours.

The proportion of red blood cells to total blood volume (Hct) varies substantially depending on race, age and sex, which are commonly used to determine the presence of anemia. In this study, the Hct levels of each of the 40 samples were adjusted to 30%, 40%, 50% and 60% by using their own plasma. Each Hct level was assessed for G-6-PD enzyme activity and compared to each other.

The study of factors influencing G-6-PD enzyme activity was performed using the SGT assay according to the manufacturer’s instructions. Briefly, five microliters of blood were mixed with the Elution Buffer (75 μL) in a U-bottom microtiter plate for 10 min at room temperature (25±2°C). The eluted (15 μL) blood was transferred to 75 μL of the Reagent Mixture contained in a new flat-bottom microtiter plate and mixed thoroughly. Then, 100 μL of Color Reagent Mixture was added to each well and G-6-PD activity reaction (kinetic mode) was read with an ELISA reader (800TS Microplate Reader, BioTek, USA) at 550 nm for 15 min at 1 min intervals. After the final reading was taken, the plate was read at wavelength 405 nm to get the hemoglobin (Hb) content of each sample. G-6-PD enzyme activity was calculated by comparing the rates of a blood sample to the rate of normal control with known G-6-PD activity. The reported G-6-PD activity was normalized to Hb and the results expressed directly into international units per gram of hemoglobin (IU/gHb).

All experiments were conducted for the quality control of deficient blood (2.0 IU/g Hb with the specification of 1.0–3.0 IU/gHb) and normal blood (14.6 IU/gHb with the specification of 9.5–19.7IU/gHb in duplicate.

Performance of the careSTARTTM G6PD biosensor for G-6-PD enzyme activity detection

The diagnostic ability of the quantitative careSTARTTM G6PD biosensor was tested using 216 neonatal blood samples. According to the manufacturer’s instructions, ten microliters of blood pricked by lancet were applied to the end of the test strip, which was inserted into the analyzer. After 4 mins, the G-6-PD enzyme activity result was displayed on the analyzer’s monitor in terms of units per deciliter (IU/dL). The activity derived from the biosensor measured the G-6-PD enzymes from RBC in one deciliter, which should be divided by Hb (IU/gHb). The EDTA blood was also used for identifying G-6-PD deficiency by qualitative FST assay. Five microliters of EDTA blood was added to 100 μL of G-6-PD substrate reagent, mixed thoroughly and incubated at 25°C for 10 min. After that, the mixture (100 μL) was applied to filter paper and observed under UV light. The results acquired by the careSTARTTM G6PD biosensor and FST were compared with standard quantitative G-6-PD assay (SGT) as a reference assay.

Statistical analysis

Statistical analysis was performed using MedCal software (version 18.2.1; MedCalc, Mariakerke, Belgium) and IBM SPSS statistics (IBM SPSS Statistics for Windows, Version 22.0. Armonk, NY: IBM Corp software). Since all data were normally distributed, the parametric statistics test was used throughout the study. The mean, median, standard deviation (SD) and ranges were determined for all G-6-PD enzyme activity values with a 95% confidence interval (CI). Normal G-6-PD activity or 100% activity can be defined by the median value of hemizygous males with a G-6-PD normal allele, called adjusted male median (AMM), which is calculated according to the description of Domingo GJ et al. [20]. Any males and females who have G-6-PD activity less than to 30% of the AMM must be regarded G-6-PD deficient [2022]. Meanwhile, males with G-6-PD activity of 30% or more of the AMM can be identified as G-6-PD normal. Heterozygous females can have intermediate status with G-6-PD activity levels ranging from 30% to 80% of AMM. A G-6-PD activity value of 80% or more of the AMM is considered G-6-PD normal [2022]. Also, p values less than 0.05 indicate a statistically significant difference between the groups. A paired t-test was used to evaluate the difference of G-6-PD enzyme activity measured from finger-prick and venous blood keeping in EDTA vacutainer tube. The effect of Hct concentration and storage time was assessed by one-way ANOVA running with a post hoc test. A linear regression model was used to predict the time at which G-6-PD enzymatic activity fell below 90% of the median initial value. Pearson correlation (r) and Bland-Altman plots were performed to evaluate the relationship, the degree of agreement between careSTARTTM G6PD biosensor and the reference SGT assay. Analysis of the area under the Receiver Operating Curve (AUC) under the receiver operating characteristic (ROC) curve was performed between the careSTARTTM G6PD biosensor and a quantitative SGT assay at 30%, 70% and 80% of AMM. The performance of the careSTARTTM G6PD biosensor included sensitivity, specificity, positive likelihood ratio (+LR), negative likelihood ratio (-LR), positive predictive value (PPV), negative predictive value (NPV) and disease prevalence was tested applying standard formulas [23]. The diagnostic potential of the careSTARTTM G6PD biosensor and FST analyzed in clinical samples was determined between genders, assuming the SGT as the reference method.

Results

Study population and distribution of G-6-PD enzyme activity

Study population characteristics including mean, median and range of G-6-PD activity were stratified by gender (Table 1). G-6-PD enzyme activity measured by quantitative SGT was taken for normal distribution. Forty adult samples from males and females with an average age of 21 (21 ± 1 year) and Hct level of 42% (42 ± 3.7%) were studied. The results demonstrated that there were no significant differences in enzyme activity between females (5.8 ± 1.9 IU/gHb) and males (6.4 ± 2.1 IU/gHb) (p = 0.328). The adult adjusted male median was 6.6 IU/gHb. In newborns with an average age of 4 (4 ± 4 day), complete blood count (CBC) and G-6-PD enzyme activity were investigated. The CBC results showed that the RBC count, hemoglobin (Hb) and hematocrit (Hct) had statistically significant differences between males and females (p<0.05) (S1 Table). As shown in Table 1, no significant differences in G-6-PD enzyme activity were observed between females (7.6 ± 1.9 IU/gHb) and males (7.1 ± 3.4 IU/gHb) (p = 0.235). The AMM of newborns was 8.1 IU/gHb, representing 100% G-6-PD activity in the study population.

thumbnail
Table 1. Proposed reference values for G-6-PD enzyme activity in this study population.

https://doi.org/10.1371/journal.pone.0226927.t001

Assessment of factors influencing G-6-PD enzyme activity

The effect of certain factors on G-6-PD activity measurement was assessed in 40 adults by quantitative SGT. G-6-PD enzymatic activity collected by finger-prick was compared to venous blood keeping in EDTA vacutainer tube. The results demonstrated that the mean of G-6-PD activity from venous blood (6.1 ± 1.9 IU/gHb, 95%CI: 5.5 to 6.7) was slightly higher than the finger-prick (5.6 ± 1.6 IU/gHb, 95%CI: 5.1 to 6.1) with no statistically significant difference (p = 0.257) (Fig 1A). Activity of G-6-PD enzyme rose significantly with increasing hematocrit (Hct) concentration, accounting for 4.5 ± 0.3, 6.4 ± 0.2, 7.9 ± 0.3 and 8.7 ± 0.2 IU/gHb for 30%, 40%, 50% and 60%, respectively (Fig 1B). A Tukey post hoc test revealed that G-6-PD activity showed a statistically significant difference among groups (p<0.05) (S2 Table). The results implied that hematocrit concentration is a dependent factor influencing G-6-PD activity detection.

thumbnail
Fig 1.

Effect of G-6-PD enzyme activity detection on type of blood collection (A) and hematocrit concentration (B). * indicates a significant difference among groups.

https://doi.org/10.1371/journal.pone.0226927.g001

The stability of G-6-PD enzymatic activity was evaluated after storage for 3 days at room temperature (temperature range from 25°C to 30°C) and 4°C (Fig 2). G-6-PD activity was measured immediately (0), 1, 2, 6, 12, 24, 48, and 72 hours after blood collection. For the room temperature condition, the mean of G-6-PD enzyme activity was decreased gradually over time and fell to 5.1 ± 0.2 IU/gHb (95%CI: 4.7 to 5.5) at 72 hours (a mean fractional fall of 1.1 IU/gHb) (Fig 2A). On the other hand, there was no statistically significant difference when the blood was kept at 4°C for 72 hours with mean activity of 6.1 ± 0.2 IU/gHb (95%CI: 5.7 to 6.4) and a mean fractional fall of 0.2 IU/gHb) (Fig 2B). A drop in G-6-PD activity was statistically correlated with its activity at baseline within 72 hours at room temperature and 4°C (p = 0.001 and p = 1.000) (S3 Table). The regression analysis indicated that G-6-PD activity would decrease to 5.5 IU/gHb (10% from median 6.1 IU/gHb) for 2 days at room temperature and 182 days at 4°C (Fig 2C and 2D, respectively). The results suggest that G-6-PD enzyme activity can remain stable at 4°C for 3 days.

thumbnail
Fig 2.

Variation of G-6-PD activity detection after storage for 72 hours at room temperature (A) and 4°C (B). The ●, ■, ▲, and ♦ symbols indicate outlier data with less or greater than the interquartile range (<5th or > 95th percentile). Regression analysis of G-6-PD activity detection after storage for 72 hours under room temperature (C) and 4°C (D).

https://doi.org/10.1371/journal.pone.0226927.g002

Performance of the careSTARTTM G6PD biosensor for G-6-PD enzyme activity detection

G-6-PD status was assessed in all 216 neonates by CBC, FST, careSTARTTM G6PD biosensor and quantitative SGT. The mean delay between blood collection and laboratory processing was 22.4 hours (range from 12 to 26 hours). Efficacy evaluation involving the correlation analysis, sensitivity, specificity, positive likelihood ratio (+LR), negative likelihood ratio (-LR), positive predictive value (PPV) and negative predictive value (NPV), the hemoglobin (Hb) data from CBC results were applied to normalize the G-6-PD enzyme activity. Correlation analysis for G-6-PD enzyme activity measurement between the careSTARTTM G6PD biosensor and SGT assay was very strong with Pearson’s correlation coefficient of 0.82 (95%CI: 0.78 to 0.86, p = 0.000) (Fig 3A). Intraclass correlation analysis between the two assays showed good agreement with the coefficient of 0.90 (95%CI: 0.87 to 0.92). The mean difference between the SGT and careSTARTTM G6PD biosensor was 2.9 U/g Hb with 95% limit of agreement ranging from 6.6 to -0.8 U/g Hb (Fig 3B). The results indicated that 95% of the differences between the two assays were within this range.

thumbnail
Fig 3.

Correlation analysis between the careSTARTTM G6PD biosensor and standard G-6-PD enzymatic test (SGT) analyzed by Pearson’s correlation analysis (A) and Bland-Altman plot (B).

https://doi.org/10.1371/journal.pone.0226927.g003

The classification of G-6-PD deficiency relies on the guide of G-6-PD deficiency rapid diagnostic testing to support the curative treatment of malaria [14, 22] and WHO prequalification [24]. The receiver operator curve (ROC) analysis between the careSTARTTM G6PD biosensor and SGT was considered at 30%, 70% and 80% G-6-PD activity (Fig 4). Analysis of the area under the ROC curve (AUC) was 0.982 (95%CI: 0.954 to 0.995), 0.993 (95%CI: 0.971 to 1.000) and 0.978 (95%CI: 0.948 to 0.993) in 30%, 70% and 80% activity, respectively. The comparison analysis of AUC showed a statistical difference between areas (p < 0.001) at all thresholds (Fig 4). The performance of the careSTARTTM G6PD biosensor included sensitivity, specificity, +LR, -LR, PPV, NPV and disease prevalence, as illustrated in Table 2. A suitable cut-off was observed at 30% residual activity (≤ 2.4 IU/gHb) with 100% sensitivity, 96% specificity, 28 of +LR, 73% PPV, 100% NPV, 97% accuracy and 8.8% disease prevalence.

thumbnail
Fig 4.

Receiver Operator Characteristic (ROC) analysis of careSTARTTM G6PD biosensor. (biosensor) with standard G-6-PD enzymatic test (SGT) at 30% (A), 70% (B) and 80% (C) cut-off. AUC means the area under the ROC curve.

https://doi.org/10.1371/journal.pone.0226927.g004

thumbnail
Table 2. Performance of the careSTARTTM G6PD biosensor for different residual G-6-PD activity.

https://doi.org/10.1371/journal.pone.0226927.t002

Assessment for the clinical diagnostic ability of the careSTARTTM G6PD biosensor was analyzed with standard SGT by gender (Table 3). Normal G-6-PD enzyme activity was identified from AMM in the study population. The cut-off value was assigned at 30% and 80% according to the above mentioned. Individuals with G-6-PD enzyme activity less than 30% of AMM (< 2.4 IU/gHb) were identified as having G-6-PD deficiency. Considering female heterozygous having G-6-PD activity at 30% to 80% of AMM (2.4–6.5 IU/gHb) was defined as G-6-PD intermediate. Females with G-6-PD activity more than 80% of AMM (> 6.5 IU/gHb) were deemed normal. Sensitivity, specificity, accuracy, PPV and NPV analyzed by the careSTARTTM G6PD biosensor were 89%, 93%, 92%, 67% and 98% in males, respectively, 64%, 83%, 80%, 43% and 92% in females, respectively, and 78%, 89%, 87%, 56% and 96% in total, respectively (Table 3). The results were considered with a commonly used qualitative FST. As shown in Table 4, the overall efficiency of FST was 18% sensitivity, 100% specificity, 88% accuracy, 100% PPV and 88% NPV. These findings suggested that the careSTARTTM G6PD biosensor possessed higher sensitivity than FST.

thumbnail
Table 3. Diagnostic performance of the careSTARTTM G6PD biosensor classified by gender.

https://doi.org/10.1371/journal.pone.0226927.t003

thumbnail
Table 4. Diagnostic performance of the FST classified by gender.

https://doi.org/10.1371/journal.pone.0226927.t004

Discussion

The detection of G-6-PD deficiency has been promoted to identify newborns in many countries [5, 6, 10, 25]. Early diagnosis is the most effective management strategy not only for correcting the problem of jaundice in newborns, but also making patients more aware of their deficiency. Boonpeng et al. reported that newborns in Thailand with more than 4.4 mg/dL of microbilirubin had a higher risk of G-6-PD deficiency [26]. This finding created a significant impact on the health policy and clinical management of newborns for improving the quality of life and protecting the activities associated with life expectancy. Most G-6-PD deficient individuals should avoid the oxidative stress triggered by agents such as antimalarial drugs (primaquine, dapsone or tafenoquine), infection, or the ingestion of fava beans [1, 4, 7]. The FST assay is an easy, rapid and cost-effective qualitative method recommended by the International Committee for Standardization in Hematology (ICSH) for screening G-6-PD deficiency [5, 19]. However, the quantitative G-6-PD enzymatic activity assay (SGT) measured by spectrophotometric and automated UV enzymatic methods remains the reference method [5, 16, 25, 27]. The threshold of G-6-PD activity has also been used as criteria for malarial drug execution [7, 8, 1113]. The WHO recommended that patients having less than 30% G-6-PD enzymatic activity and 70% normal activity should be excluded from primaquine and tafenoquine treatment [14, 22]. The SGT method requires a source of electricity, refrigeration for reagents and substrates, information on patient hemoglobin levels, a spectrophotometer to measure the absorbance and personal expertise to perform the test. These limitations restrict the SGT for routine testing and field application, especially in endemic countries with malaria infection [18]. The careSTARTTM G6PD biosensor has been created due to the current trend of developing point-of-care diagnostics. It meets not only all of the point-of-care testing (POCT) requirements, but is also an interesting quantitative tool to identify G-6-PD deficiency in newborns. The careSTARTTM G6PD biosensor used in the present research is a new model (version 2017), which has improved reliability for results by adding the reference electrode. The performance and factors influencing G-6-PD activity measurement were elucidated in this investigation. Basic information about G-6-PD activity in the study population was analyzed by gender (Table 1). There were no significant differences in G-6-PD activity between males and females. Moreover, the mean and AMM of G-6-PD enzyme activity in newborns were significantly higher than adults (p < 0.05), corresponding with previous study [27]. The factors affecting G-6-PD activity measurement were examined in 40 adults involving the type of blood collection, Hct concentration, storage time and temperature. The results found that the venous blood preserved in EDTA anticoagulant did not interfere with G-6-PD enzyme activity measurement compared to finger blood (Fig 1A). This finding confirmed the report by Roca-Feltrer et al.[28], which solved the question of von Fricken ME’s studies [25]. Accordingly, the hematocrit in newborns has a higher concentration than adults [29, 30], corresponding to this study (42±3.7% in adults and 49±7.7% in newborns). This means that newborns should have higher G-6-PD activity than adults, which is consistent with the results in Fig 1B. The results suggest that the reference value of newborns should be differentiated from adults [31]. G-6-PD activity drops gradually as time progresses [6]. Because the acceptance decreasing rate of G-6-PD activity should be no more than 10% [6, 32], G-6-PD activity was reduced over acceptance value at room temperature for 50 hours (Fig 2A) and remained stable at 4°C for 3 days (Fig 2B). The regression analysis found that blood samples should be stored at 4°C for 3 days (Fig 2D). Similar to a previous report, the G-6-PD activity did not fall by 10% until after 9 days of storage [6]. The performance evaluation of the careSTARTTM G6PD biosensor was analyzed using 216 neonates. The results demonstrated a strong positive correlation between the careSTARTTM G6PD biosensor and SGT (Fig 3). Thus, it can be described that both methods directly determine the kinetics of G-6-PD enzymatic activity reactions [18]. Bland-Altman plot analysis showed the absolute activity values of the same samples obtained from SGT, which were slightly higher than the careSTARTTM G6PD biosensor (+2.9 IU/gHb). Contrary to the automated UV enzymatic assay, G-6-PD activity was significantly higher than that of the SGT method [16]. This results from the different detectors; the careSTARTTM G6PD biosensor measures the electrons of Fe3+ from the change of NADP+ to NADPH reaction upon G-6-PD enzyme, meaning the electrons may be lost during operation. In this study, we set the cut-off point for G-6-PD deficiency at 30%, 70% and 80% of AMM based on previous suggestions [14, 20, 22]. The current thresholds used as exclusion criteria for primaquine treatment (<30% activity) and tafenoquine (<70% activity) [5, 28]. The diagnostic ability of the developed careSTARTTM G6PD biosensor represented an ideal method compared to a reference SGT assay at all thresholds, especially at 30% cut-off (Fig 4 and Table 2). With ROC analysis, the results suggested the optimal cut-off value to provide appropriate diagnostic performance (Table 2). The indication is that individuals with G-6-PD enzyme activity less than 2.4 IU/gHb can be identified as having G-6-PD deficiency, while those with G-6-PD enzyme activity between 2.5 to 6.4 IU/gHb can be identified as having intermediate G-6-PD deficiency. As presented in this study, the main advantage for clinical use in quantitative point-of-care diagnostics utilizing the careSTARTTM G6PD biosensor is to provide a more accessible way of defining G-6-PD activity at bedside and identifying individuals with intermediate G-6-PD deficiency, especially in terms of differentiating heterozygous females. The study also illustrates the sensitivity, specificity, PPV, NPV and accuracy of the careSTARTTM G6PD biosensor and FST compared to standard SGT (Table 3 and Table 4). The diagnostic efficacy of the careSTARTTM G6PD biosensor was found to be slightly higher than reported in the past [6]. Such variance in the results may be caused by the small-sized group and different population (216 neonates in the present study versus 900 adults in the previous study). Table 4 illustrates the low sensitive achievement by FST assay at all thresholds. This result confirmed that all current qualitative tests perform poorly [6, 17, 25, 28]. However, it has been recommended by ICSH as the appropriate qualitative method for screening G-6-PD deficiency in the field [5, 19]. The careSTARTTM G6PD biosensor showed a high potential for identifying G-6-PD deficiency in newborns and partial G-6-PD deficiency in females, imposing a risk for primaquine the same as tafenoquine administration. Moreover, quantitative G-6-PD activity has more alternative, attractive methods than molecular analysis because some cases of intermediate or G-6-PD deficiency could not be identified in the mutation [1, 3, 33, 34]. More than 400 different mutations have been found in individuals with G-6-PD deficiency [1, 3, 33, 34]. Some cases contain the mutations in cis-acting regulatory sequences or in the non-coding region of the G-6-PD gene, which may interfere with its expression [16]. However, the results by the careSTARTTM G6PD biosensor should be normalized with Hb concentration, especially in hemolytic anemia patients, because false positive results may occur.

Conclusion

The careSTARTTM G6PD biosensor has a high diagnostic ability to quantify G-6-PD enzymatic activity. It meets the needs for point-of-care testing and an appropriate setting for health promotion. Particularly, the careSTARTTM G6PD biosensor is suitable for early identification of intermediate and G-6-PD deficiency in newborns, which is the most effective strategy for the prevention of severe hemolytic anemia and the initiation of corrective treatment in a timely manner.

Supporting information

S1 Table. The CBC results of 216 neonates.

Results of RBC count, Hb and Hct in males were significantly higher than females.

https://doi.org/10.1371/journal.pone.0226927.s001

(DOCX)

S2 Table. Multiple comparisons of G-6-PD activity among hematocrit (Hct) concentration in 40 adults.

The G-6-PD enzyme activity at certain Hct was subtracted with another getting the different of G-6-PD activity. The mean difference of G-6-PD activity measures the absolute difference between the mean values in two groups.

https://doi.org/10.1371/journal.pone.0226927.s002

(DOCX)

S3 Table. Multiple comparisons of G-6-PD enzyme storage stability at room temperature and 4°C for 72 hours.

The mean difference measures the absolute difference between the mean value in two groups.

https://doi.org/10.1371/journal.pone.0226927.s003

(DOCX)

Acknowledgments

We are grateful to thank the Thammasat University Hospital for supplying neonates blood samples.

References

  1. 1. Cappellini MD, Fiorelli G. Glucose-6-phosphate dehydrogenase deficiency. Lancet. 2008;371(9606):64–74. pmid:18177777
  2. 2. Minucci A, Moradkhani K, Hwang MJ, Zuppi C, Giardina B, Capoluongo E. Glucose-6-phosphate dehydrogenase (G6PD) mutations database: review of the "old" and update of the new mutations. Blood Cells Mol Dis. 2012;48(3):154–65. pmid:22293322
  3. 3. Monteiro WM, Val FF, Siqueira AM, Franca GP, Sampaio VS, Melo GC, et al. G6PD deficiency in Latin America: systematic review on prevalence and variants. Mem Inst Oswaldo Cruz. 2014;109(5):553–68. pmid:25141282
  4. 4. Schrier SL. Human erythrocyte G6PD deficiency: pathophysiology, prevalence, diagnosis, and management. Compr Ther. 1980;6(7):41–7. pmid:7471680
  5. 5. Abdul-Ghani R, Mahdy MAK, Saif-Ali R, Alkubati SA, Alqubaty AR, Al-Mikhlafy AA, et al. Glucose-6-phosphate dehydrogenase deficiency among Yemeni children residing in malaria-endemic areas of Hodeidah governorate and evaluation of a rapid diagnostic test for its detection. Malaria J. 2016;15.
  6. 6. Alam MS, Kibria MG, Jahan N, Thriemer K, Hossain MS, Douglas NM, et al. Field evaluation of quantitative point of care diagnostics to measure glucose-6-phosphate dehydrogenase activity. PLoS One. 2018;13(11):e0206331. pmid:30388146
  7. 7. Arese P, Gallo V, Pantaleo A, Turrini F. Life and Death of Glucose-6-Phosphate Dehydrogenase (G6PD) Deficient Erythrocytes—Role of Redox Stress and Band 3 Modifications. Transfus Med Hemother. 2012;39(5):328–34. pmid:23801924
  8. 8. Frank JE. Diagnosis and management of G6PD deficiency. Am Fam Physician. 2005;72(7):1277–82. pmid:16225031
  9. 9. Jalloh A, Tantular IS, Pusarawati S, Kawilarang AP, Kerong H, Lin K, et al. Rapid epidemiologic assessment of glucose-6-phosphate dehydrogenase deficiency in malaria-endemic areas in Southeast Asia using a novel diagnostic kit. Tropical Medicine & International Health. 2004;9(5):615–23.
  10. 10. Fu C, Luo S, Li Q, Xie B, Yang Q, Geng G, et al. Newborn screening of glucose-6-phosphate dehydrogenase deficiency in Guangxi, China: determination of optimal cutoff value to identify heterozygous female neonates. Sci Rep. 2018;8(1):833. pmid:29339739
  11. 11. Uthman OA, Graves PM, Saunders R, Gelband H, Richardson M, Garner P. Safety of primaquine given to people with G6PD deficiency: systematic review of prospective studies. Malar J. 2017;16(1):346. pmid:28830424
  12. 12. Uthman OA, Saunders R, Sinclair D, Graves P, Gelband H, Clarke A, et al. Safety of 8-aminoquinolines given to people with G6PD deficiency: protocol for systematic review of prospective studies. BMJ Open. 2014;4(5):e004664. pmid:24833685
  13. 13. Lee SM, Geetha D. Dapsone induced hemolysis in a patient with ANCA associated glomerulonephritis and normal G6PD level and implications for clinical practice: case report and review of the literature. Springerplus. 2015;4:29.
  14. 14. WHO. Testing for G6PD deficiency for safe use of primaquine in radical cure of P. vivax and P. ovale (Policy brief). Geneva: World Health Organization. 2016:28.
  15. 15. Lacerda MVG, Llanos-Cuentas A, Krudsood S, Lon C, Saunders DL, Mohammed R, et al. Single-Dose Tafenoquine to Prevent Relapse of Plasmodium vivax Malaria. New Engl J Med. 2019;380(3):215–28. pmid:30650322
  16. 16. Anantasomboon P, Chanda M, Jugnam-ang W, Witoonpanich P, Cheepsunthorn P, Nuchprayoon I, et al. Evaluating the performance of automated UV enzymatic assay for screening of glucose 6-phosphate dehydrogenase deficiency. Int J Lab Hematol. 2019;41(2):192–9. pmid:30383322
  17. 17. Fan YH, Lazenbery L, Foster E, Duelm F, Grant E Jr. Improved quantitative method for G6PD deficiency detection. J Clin Lab Anal. 2007;21(2):107–13. pmid:17385678
  18. 18. Ley B, Bancone G, von Seidlein L, Thriemer K, Richards JS, Domingo GJ, et al. Methods for the field evaluation of quantitative G6PD diagnostics: a review. Malar J. 2017;16(1):361. pmid:28893237
  19. 19. Nadarajan V, Shanmugam H, Sthaneshwar P, Jayaranee S, Sultan KS, Ang C, et al. Modification to reporting of qualitative fluorescent spot test results improves detection of glucose-6-phosphate dehydrogenase (G6PD)-deficient heterozygote female newborns. Int J Lab Hematol. 2011;33(5):463–70. pmid:21501392
  20. 20. Domingo GJ, Satyagraha AW, Anvikar A, Baird K, Bancone G, Bansil P, et al. G6PD testing in support of treatment and elimination of malaria: recommendations for evaluation of G6PD tests. Malar J. 2013;12:391. pmid:24188096
  21. 21. Pal S, Bansil P, Bancone G, Hrutkay S, Kahn M, Gornsawun G, et al. Evaluation of a Novel Quantitative Test for Glucose-6-Phosphate Dehydrogenase Deficiency: Bringing Quantitative Testing for Glucose-6-Phosphate Dehydrogenase Deficiency Closer to the Patient. Am J Trop Med Hyg. 2019;100(1):213–21. pmid:30350771
  22. 22. WHO. Guide to G6PD deficiency rapid diagnostic testing to support P. vivax radical cure. Geneva: World Health Organization. 2018:34.
  23. 23. Simundic AM. Measures of Diagnostic Accuracy: Basic Definitions. EJIFCC. 2009;19(4):203–11. pmid:27683318
  24. 24. WHO. Technical Specifications Series for submission to WHO prequalification–diagnostic assessment: in vitro diagnostics medical devices to identify Glucose-6-phosphate dehydrogenase (G6PD) activity. Geneva: World Health Organization. 2016:20.
  25. 25. von Fricken ME, Weppelmann TA, Eaton WT, Masse R, de Rochars MVEB, Okech BA. Performance of the CareStart Glucose-6-Phosphate Dehydrogenase (G6PD) Rapid Diagnostic Test in Gressier, Haiti. Am J Trop Med Hyg. 2014;91(1):77–80. pmid:24778197
  26. 26. Boonpeng P. BB, Traivaree C, Changsri K. Molecular Characterization of G6PD Gene Mutation in Neonatal Jaundice in Phramongkutklao Hospital. RTA Med J. 2018;71(3):8.
  27. 27. Kim S, Nguon C, Guillard B, Duong S, Chy S, Sum S, et al. Performance of the CareStart (TM) G6PD Deficiency Screening Test, a Point-of-Care Diagnostic for Primaquine Therapy Screening. Plos One. 2011;6(12).
  28. 28. Roca-Feltrer A, Khim N, Kim S, Chy S, Canier L, Kerleguer A, et al. Field trial evaluation of the performances of point-of-care tests for screening G6PD deficiency in Cambodia. PLoS One. 2014;9(12):e116143. pmid:25541721
  29. 29. Lindblade KA, Mwololo K, van Eijk AM, Peterson E, Odhiambo F, Williamson J, et al. Evaluation of the WHO Haemoglobin Colour Scale for diagnosis of anaemia in children and pregnant women as used by primary health care nurses and community health workers in western Kenya. Trop Med Int Health. 2006;11(11):1679–87. pmid:17054747
  30. 30. Marn H, Critchley JA. Accuracy of the WHO Haemoglobin Colour Scale for the diagnosis of anaemia in primary health-care settings in low-income countries: a systematic review and meta-analysis. Lancet Glob Health. 2016;4(4):E251–E65. pmid:26905314
  31. 31. Li MJ, Yang WC, Tai S, Hsu CL, Shao PL, Peng CT. The Reference Levels of Glucose-6-Phosphate Dehydrogenase Enzyme Activity for Infants in Taiwan. Blood. 2017;130.
  32. 32. Tang J, Zhou X, Liu X, Ning L, Zhou W, He Y. External quality assessment program for detection of glucose-6-phosphate dehydrogenase deficiency in the Guangxi region. Exp Ther Med. 2017;14(3):2021–4. pmid:28962119
  33. 33. Yan TZ, Cai R, Mo OH, Zhu DL, Ouyang H, Huang LH, et al. Incidence and complete molecular characterization of glucose-6-phosphate dehydrogenase deficiency in the Guangxi Zhuang autonomous region of southern China: description of four novel mutations. Haematol-Hematol J. 2006;91(10):1321–8.
  34. 34. Xu W, Wang J, Hua X, Du C. Detection of point mutations in exon 2 of the G6PD gene in Chinese G6PD variants. Chin Med Sci J. 1994;9(1):20–3. pmid:8086629